K132817 · Benvenue Medical, Inc. · NDN · Jan 24, 2014 · Orthopedic
Device Facts
Record ID
K132817
Device Name
KIVA VCF TREATMENT SYSTEM
Applicant
Benvenue Medical, Inc.
Product Code
NDN · Orthopedic
Decision Date
Jan 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Kiva® VCF Treatment System is indicated for use in the reduction and treatment of spinal fractures in the thoracic and/or lumbar spine from T6-L5. It is intended to be used in combination with the Benvenue Vertebral Augmentation Cement Kit.
Device Story
Kiva VCF Treatment System is a sterile, single-use implantable device for percutaneous reduction and treatment of spinal fractures. System includes nitinol Kiva Coil, PEEK Optima implant with barium sulfate, deployment cannula, and manual orthopedic access instruments. Procedure involves guiding nitinol coil into vertebral body via deployment cannula; PEEK implant is guided over coil to achieve height distraction of vertebral body. Once positioned, PMMA bone cement is injected into implant to stabilize fracture and minimize posterior extravasation. Used in clinical settings by physicians. Benefits include vertebral height restoration and fracture stabilization.
Clinical Evidence
Clinical evidence provided via the KAST trial, a prospective, multi-center, randomized, controlled study (n=300) comparing Kiva system to balloon kyphoplasty for osteoporotic VCF. Primary endpoint: composite of pain reduction and functional improvement without device-related serious adverse events. Secondary endpoints: vertebral height restoration, cement volume, quality of life, ambulatory status, subject satisfaction, cement extravasation, migration, and subsequent fracture. Results supported safety and effectiveness.
Technological Characteristics
Implant: PEEK Optima with barium sulfate. Deployment system: Nitinol coil, deployment cannula. Access instruments: Class I manual orthopedic needles, stylets, cannulas. Sterilization: Terminal irradiation (ANSI/AAMI/ISO 11137). Biocompatibility: ANSI/AAMI/ISO 10993 compliant. Energy source: Manual mechanical operation.
Indications for Use
Indicated for reduction and treatment of spinal fractures in thoracic and/or lumbar spine (T6-L5).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
K 132817 Page 1 of 5 .
JAN 2 4 2014
# 510(K) SUMMARY AS REQUIRED BY 21 CFR 807.92
#### 510(k) Owner:
Barbara S. Lindsay Vice President, Clinical, Regulatory and Quality Affairs BENVENUE MEDICAL, INC 3052 Bunker Hill Lane, Suite 120 Santa Clara, CA 95054 Phone: (408) 454-9300 Fax: (408) 982-9023
#### 510(k) Contact:
Cindy Domecus, R.A.C. (US & EU) Principal Domecus Consulting Services LLC Consultant to Benvenue Medical, Inc. Phone: (650) 343-4813 Fax: (650) 343-7822 domecusconsulting@comcast.net
Date Prepared: January 23, 2014
Kiva® VCF Treatment System Device Trade Name:
Common Name:
Polymethylmethacrylate (PMMA) Bone Cement
Classification Name:
Polymethylmethacrylate (PMMA) Bone Cement (21 CFR
888.3027)
Device Product Code:
NDN, LOD
### Page 1 of 5
{1}------------------------------------------------
Predicate Device(s): The Kiva® VCF Treatment System is substantially equivalent to the following devices:
| Device | Manufacturer | 510(k) |
|--------------------------------|--------------------------|---------|
| StaXxTM FX System | SpineWave, Inc | K063606 |
| Benvenue VCF Osteo Coil System | Benvenue Medical, Inc | K070293 |
| KyphX® HV-RTM Bone Cement | Kyphon, Inc. (Medtronic) | K041584 |
| KyphX® Inflatable Bone Tamp | Kyphon, Inc. (Medtronic) | K010246 |
#### Device Description:
The Kiva® VCF Treatment System is provided as a sterile, single use, implantable device which may be used in percutaneous procedures for the reduction and treatment of spinal fractures. The device consists of an implant, a deployment system for the implant, and a set of accessory access instruments. The deployment system component is a single-use, nonimplantable device that is used to properly position and deliver the implant. It consists of a nitino! Kiva Coil, which is guided through a deployment cannula into the bone via a hand operated mechanism. The Kiva Implant, made from PEEK Optima with barium sulfate for radiopacity, is guided over the nitinol Kiva Coil. As the Kiva Implant is pushed over the Kiva Coil, it reduces the fracture via height distraction of the vertebral body. Once the Implant is placed into the vertebral body. PMMA bone cement is deployed into the Kiva Implant. The Kiva Implant contains the PMMA bone cement and helps minimize posterior extravasation. A collection of manual surgical orthopedic instrumentation (Class I needles, stylets, cannulas) is used to gain access to the vertebral body at the start of the procedure, then again later for bone cement deployment.
#### Statement of Intended Use:
The Kiva® VCF Treatment System is indicated for use in the reduction and treatment of spinal fractures in the thoracic and/or lumbar spine from T6-L5. It is intended to be used in combination with the Benvenue Vertebral Augmentation Cement Kit.
{2}------------------------------------------------
#### Sterilization:
The Kiva® VCF Treatment System is terminally sterilized via irradiation according to the requirements of ANSI/AAMI/ISO 11137 to ensure a sterility assurance level (SAL) of 10°
#### Biocompatibility:
The materials comprising the Kiva® VCF Treatment System (implant, deployment system, and access instrumentation) have been evaluated and tested for biocompatibility according to applicable requirements of ANSI/AAMI/ISO 10993. Results demonstrated that the Kiva® VCF Treatment System is biocompatible.
#### Non-Clinical/Mechanical Testing:
The safety and performance of the Kiva® VCF Treatment System have been substantiated through extensive non-clinical mechanical testing. Results of the testing confirm that the Kiva® VCF Treatment System can reliably perform as intended.
#### Clinical Performance Discussion:
A clinical trial was conducted under an approved IDE. The study is titled Kiva® System as a Vertebral Augmentation Treatment -- A Safety and Effectiveness Trial (KAST trial). This was a prospective, multi-center, randomized, controlled study designed to evaluate the safety and effectiveness of the Kiva® VCF Treatment System in comparison to balloon kyphoplasty in the treatment of osteoporotic vertebral compression fractures (VCF). The primary endpoint was a composite of reduction in paintenance or improvement in function in the absence of device related serious adverse events. Secondary endpoints include vertebral height restoration, cement volume, quality of life, ambulatory status, subject satisfaction, cement extravasation, device or cement migration, and subsequent fracture. A total of 300 subjects were enrolled at 21 sites in the US, Canada, and Europe. The results demonstrate the safety and effectiveness of the Kiva® VCF Treatment System and support substantial equivalence to predicate devices.
{3}------------------------------------------------
### Substantial Equivalence:
The rationale for substantial equivalence of the Kiva® VCF Treatment System to cited predicates is summarized in the following bullet points:
- A substantial part of this device (Kiva Coil and Deployment System) is the cleared . Benvenue VCF Osteo Coil System.
- The implantable PEEK Optima Kiva Implant part of the device is substantially . equivalent to the stackable PEEK wafers of the Spine Wave StaXxTM FX device.
- The Kiva® VCF Treatment System reduces compression fracture via height . distraction as do the predicate devices.
- The Kiva® VCF Treatment System functions in concert with PMMA bone cement to . achieve fracture reduction and treatment as do the predicate devices. The following table describes and shows similarities of the indications for use among the predicate devices and the Kiva® VCF Treatment System.
{4}------------------------------------------------
| Device | Indications for Use |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| StaXx™ FX System | The StaXx™ FX System is indicated for use in the reduction<br>of spinal fractures. It is intended to be used in combination<br>with Stryker Spineplex Radiopaque Bone Cement. |
| Benvenue VCF Osteo Coil<br>System | The Benvenue VCF Osteo Coil System is indicated for the<br>treatment of pathological compression fractures of the<br>vertebral body that may result from osteoporosis, benign<br>lesions, and/or malignant lesions, by creating channels in the<br>existing spinal bone structure for the flow of PMMA bone<br>cement. |
| KyphX® HV-R™ Bone<br>Cement | KyphX® HV-R™ Bone cement is indicated for the treatment<br>of pathological fractures of the vertebral body due to<br>osteoporosis, cancer, or benign lesions using a balloon<br>kyphoplasty procedure. Cancer includes multiple myeloma<br>and metastatic lesions, including those arising from breast or<br>lung cancer or lymphoma. Benign lesions include hemagioma<br>and giant cell tumor. |
| KyphX® Inflatable Bone<br>Tamp | KyphX® Inflatable Bone Tamps are intended to be used as<br>conventional bone tamps for the reduction of fractures and/or<br>creation of a void in cancellous bone in the spine (including<br>use during balloon Kyphoplasty with KyphX® HV-R™ Bone<br>Cement), hand, tibia, radius and calcaneus. |
| Kiva® VCF Treatment<br>System | The Kiva® VCF Treatment System is indicated for use in the<br>reduction and treatment of spinal fractures in the thoracic<br>and/or lumbar spine from T6-L5. It is intended to be used in<br>combination with the Benvenue Vertebral Augmentation<br>Cement Kit. |
- The extensive non-clinical and clinical testing performed on this device provides . support that the Kiva® VCF Treatment System performs as intended and presents no different questions of safety or effectiveness.
The Kiva® VCF Treatment System device technological characteristics compared to predicate devices, indications for use compared with predicate devices, non-clinical test results, and clinical data demonstrate that the Kiva® VCF Treatment System is safe, effective, performs as intended, and is substantially equivalent to the currently marketed predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes extending upwards. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the bird-like figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
January 24, 2014
Benvenue Medical, Incorporated % Ms. Cindy Domecus, Regulatory Affairs Consultant Principal Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, California 94010
Re: K132817
Trade/Device Name: Kiva® VCF Treatment System Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, LOD Dated: January 6, 2014 Received: January 7, 2014
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
#### Page 2 - Ms. Cindy Domecus
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Lori A. Wiggins
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### BENVENUE MEDICAL, INC KIVA® VCF TREATMENT SYSTEM - TRADITIONAL 510(k) PREMARKET NOTIFICATION
### 4. INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K132817
Device Name: Kiva® VCF Treatment System
Indications for Use:
The Kiva® VCF Treatment System is indicated for use in the reduction and treatment of spinal fractures in the thoracic and/or lumbar spine from T6-L5. It is intended to be used in combination with the Benvenue Vertebral Augmentation Cement Kit.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Laurence D. Coyne -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K132817
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.