KyphX® HV-R™ bone cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a balloon kyphoplasty procedure. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesion- include hemagioma and giant cell tumor.
Device Story
Two-component PMMA bone cement system; powder component (PMMA copolymer, barium sulfate radiopacifier, benzoyl peroxide initiator) and liquid component (methylmethacrylate monomer, hydroquinone stabilizer, N,N-dimethyl-p-toluidine promoter). Components mixed manually prior to use. Used in clinical settings by physicians during balloon kyphoplasty procedures to stabilize vertebral fractures. Provides structural support to the vertebral body; benefits patients by reducing pain and restoring vertebral height. Single-use device.
Clinical Evidence
Bench testing only. Mechanical testing performed in direct comparison to Surgical Simplex® P per ISO 5833:2002. Biocompatibility testing performed per ISO 10993 and FDA Blue Book Memorandum #G95-1.
Technological Characteristics
Two-component PMMA-based bone cement. Powder: PMMA/methyl-methacrylate-styrene copolymer (68%), barium sulfate (30%), benzoyl peroxide (2%). Liquid: Methyl methacrylate monomer (99.1%), N,N-dimethyl-p-toluidine (0.9%), hydroquinone (75 ppm). Sterilization: Gamma radiation (solid components/pouch, SAL 10^-6), filtration (liquid, SAL 10^-5), ethylene oxide (outer packaging). Complies with ISO 5833:2002 and ISO 10993.
Indications for Use
Indicated for treatment of pathological vertebral body fractures due to osteoporosis, cancer (multiple myeloma, metastatic lesions from breast/lung cancer, lymphoma), or benign lesions (hemangioma, giant cell tumor) in patients undergoing balloon kyphoplasty.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Stryker Spineplex™ Radiopaque Bone Cement (K032945)
Submission Summary (Full Text)
{0}------------------------------------------------
Kyphon Inc Page 1 of 3, Amended Tab 5 KyphX® HV-RTM Bone Cement Premarket Notification [510(k)] Submission
JUL = 7 2004
K041584
## Tab 5
## Premarket Notification [510(k)] Summary
| Trade Name: | KyphX® HV-R™ Bone Cement |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Bone Cement |
| Classification /Name: | Class II<br>Bone Cement: 21 CFR, part 888.3027 |
| Device Code: | NDN |
| Manufacturer's Name:<br>Address: | Kyphon Inc.<br>1221 Crossman Avenue<br>Sunnyvale, CA 94089 |
| Corresponding Official:<br>Title:<br>Address:<br>Telephone: | Cindy Domecus<br>Vice President, Clinical Research and Regulatory Affairs<br>1221 Crossman Avenue<br>Sunnyvale, CA 94089<br>408-222-6740 |
| Predicate Device(s): | K033801: KyphX® HV-R™, Model C01A<br><br>K002652: Stryker Howmedica Surgical Simplex®<br>Radiopaque pre-packed in ACM and MixEvac II.<br><br>K032945: Stryker Spineplex™ Radiopaque Bone Cement |
| Intended Use: | KyphX® HV-R™ bone cement is indicated for the<br>treatment of pathological fractures of the vertebral body<br>due to osteoporosis, cancer, or benign lesions using a<br>balloon kyphoplasty procedure. Cancer includes multiple<br>myeloma and metastatic lesions, including those arising<br>from breast or lung cancer, or lymphoma. Benign lesion- |
Any statement regarding "substantial equivalence" made in this submission only relates to whether the intended to be interested in this submission may be lawfully market approval or reclassification, and is not intended o be interpreted as an admission of evidence in any patent proceeding, and is not infingement therein therefore should not be consider and of the scope of any patent or patent or patent or patent or patent opplication, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing.
include hemagioma and giant cell tumor.
{1}------------------------------------------------
Like the predicate devices, KyphX® HV-RTM Bone Cement Device Description: is provided as a two-component system. The powder component consists of a PMMA copolymer (polymethyl methacrylate/methyl-methacrylate-styrene-copolymer) with barium sulfate as a radiopacifier and benzoyl peroxide as an The liquid component consists initiator. of methylmethacrylate monomer, with the addition of hydroquinone as a stabilizer and N.N-dimethyl-p-toluidine as a promoter. The powder and liquid components are mixed prior to use.
> Table 5-1 compares the chemical composition of KyphX® HV-RTM Bone Cement to the Surgical Simplex® P predicate device.
| Table 5-1: Chemical Composition of KyphX® HV-R and Surgical Simplex®P | | | |
|------------------------------------------------------------------------|--|--|--|
| | | | |
| Chemical Composition | KyphX® HV-RTM | Surgical Simplex® P |
|-----------------------------------------------------------------|--------------------------------|-------------------------------------------|
| <i>Powder</i> | 20 g. packet of sterile powder | 20 g (half-dose) packet of sterile powder |
| Polymethyl methacrylate / methyl methacrylate-styrene copolymer | 68.0% w/w | 88.5% w/w |
| Barium sulfate | 30.0% w/w | 10.0% w/w |
| Benzoyl peroxide | 2.0% w/w | 1.5% w/w |
| <i>Liquid</i> | 10 ml vial of sterile liquid | 10 ml (half-dose) vial of sterile liquid |
| Methyl methacrylate (monomer) | 99.1% v/v | 97.4% v/v |
| N, N-dimethyl-p-toluidine | 0.90% v/v | 2.6% v/v |
| Hydroquinone | 75 ppm | 75 ± 15 ppm |
Mechanical Tests:
KyphX® HV-R™ Bone Cement was tested in direct comparison to the Surgical Simplex® P predicate device and verified substantially equivalent, as defined by ISO 5833:2002, "Implants for Surgery - Acrylic resin cements."
Any statement regarding "substantial equivalence" made in this submission only relates to whether the product addressed in this submission may be lawfully market approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement fitigation or proceeding before any Patent office. The present submission and statements therein the product not be construed as affecting or help to the scope of any patent or patent or patent or patent or patent or patent or patent or patent opplication, or to whether the product addressed in this submission, or its use, any parent application, of the laboration, from a a atentability perspective, from any of the other devices referenced in this filing.
{2}------------------------------------------------
- Clinical information demonstrates that the intended use of Clinical Discussion: KyphX® HV-RTM Bone Cement is substantially equivalent to the Surgical Simplex® P indication for the fixation of pathological fractures, to the KyphX® HV-R™ predicate indication for the treatment of pathological fractures of the vertebral body due to osteoporosis using a kyphoplasty procedure, and to the Spineplex™ bone cement for use in benign and metastatic lesions.
The materials used in KyphX® HV-R™ Bone Cement meet Biocompatibility: the requirements for "Implant, Tissue/Dentin/Bone, Permanent Contact" described in the FDA Blue Book Memorandum #G95-1. KyphX® HV-RTM Bone Cement was tested in direct comparison to the Surgical Simplex® P predicate device and verified substantially equivalent, as defined by ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing."
The KyphX® HV-RTM Bone Cement is sterile and non-Sterilization: pyrogenic. The solid components and pouch are sterilized with gamma radiation to a Sterility Assurance Level (SAL) of 10°. The liquid components are sterilized with filtration methods to an SAL of 10-5. The outer packaging containing the liquid component is sterilized with ethylene oxide gas. The bone cement is intended for single use only.
Substantial Equivalence: The information submitted in this pre-market notification supports a determination that KyphX® HV-RTM Bone Cement is substantially equivalent in technological characteristics and intended use to the predicates. The products have the same fundamental scientific technology their chemical composition, material properties, in performance characteristics, and biocompatibility, and also have the same intended use.
Any statement regarding "substantial equivalence" made in this submission only relates to whether the intended to be interest without may be lawfully marketed without premarket approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any petent proceeding, including paract infringement litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting or relating to the "stennssion and "statinlerity whether the product addressed in this submission, or its use, may be considered indistinct, from a patentablity perspective, from any of the other devices referenced in this filing.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three abstract human figures or forms intertwined. The symbol is rendered in black and is the focal point of the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 7 2004
Cindy Domecus Vice President, Clinical Research and Regulatory Affairs Kyphon Inc. 1221 Crossman Avenue Sunnyvale, California 94089
Re: K041584
Trade/Device Name: KyphX® HV-R™ Bone Cement, Model C01A Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: II Product Codes: NDN Dated: June 11, 2004 Received: June 14, 2004
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -- Ms. Cindy Domecus
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, PhD, MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K041584
Device Name: KyphX® HV-RTM Bone Cement
Indications for Use:
KyphX HV-R bone cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a balloon kyphoplasty procedure. Cancer includes multiple myeloma and metastatic lesions, including those atising from breast or lung cancer, or lymphoma. Benign lesions include hemagioma and giant cell tumor.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number_Ko41584
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.