K132671 · Stryker Corporation · JCX · Dec 20, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132671
Device Name
STRYKER NEPTUNE 2 WASTE MANAGMENT SYSTEM
Applicant
Stryker Corporation
Product Code
JCX · General, Plastic Surgery
Decision Date
Dec 20, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Indications for Use
The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor's offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices.
Device Story
Neptune 2 Waste Management System is a mobile, AC-powered surgical fluid waste collection and smoke evacuation device. System consists of a 'Rover' (mobile collection unit) and a 'Docker' (stationary docking station). Rover collects surgical waste fluids and evacuates smoke during procedures; Docker provides automated cleaning and waste offloading via connection to hospital water, sewer, and electricity. Operated by clinical staff in ORs, surgical centers, and clinics. Device uses suction to collect fluids into canisters and filters smoke via ULPA filtration. Output includes visual status on an LCD screen. System benefits include efficient waste management and smoke removal, improving surgical site visibility and reducing health hazards. Contraindicated for direct connection to chest tubes or closed wound drainage systems.
Clinical Evidence
Bench testing only. Testing verified and validated device performance, including electrically powered suction safety, electrical safety, vacuum cycling, software, smoke evacuation, and filtration. Human factors analysis and usability testing performed on user interface, labeling, and training materials. Biocompatibility testing confirmed compliance with ISO 10993-1.
Technological Characteristics
Mobile AC-powered rover and stationary docking station. Features ULPA filtration for smoke, float-style shut-off valves for overfill protection, and dry-break hydraulic couplings. Dimensions: 19 x 23 x 70 inches. Connectivity: Standalone system with facility water/sewer/electrical integration. Software-controlled vacuum regulation and user interface. Biocompatible materials per ISO 10993-1. Class I, Type CF applied part.
Indications for Use
Indicated for use in operating rooms, pathology labs, surgical centers, and doctor's offices to collect/dispose of surgical fluid waste and collect smoke from electrocautery or laser devices. Contraindicated for direct connection to chest tubes or closed wound drainage systems.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
Stryker® Neptune Waste Management System (K012991)
Valleylab OptiMumm Smoke Evacuator System (K980915)
Submission Summary (Full Text)
{0}------------------------------------------------
KV32671
DEC 2 0 2013
stryker
Instruments
### 510(k) Summary
| 510(k) Owner: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI 49001<br>(p) 269-389-3549<br>(f) 269-389-5412 |
|-------------------|-------------------------------------------------------------------------------------------------------------|
| Contact Person: | Michelle Jump |
| Registration No.: | 1811755 |
| Trade Name: | Stryker® Neptune 2 Waste Management System |
| Common Name: | Portable suction and smoke evacuation device |
| Classification Name and Regulation Number | Primary Product Code |
|-------------------------------------------|----------------------|
|-------------------------------------------|----------------------|
| Product Code | Device | Regulation Number | Class |
|--------------|----------------------------------------------------|-------------------|-------|
| JCX | apparatus, suction, ward use, portable, ac-powered | 21 CFR 872.4780 | II |
| | Secondary Product Code |
|--|------------------------|
|--|------------------------|
| Product Code | Device | Regulation Number | Class |
|--------------|------------------------------|-------------------|-------|
| FYD | apparatus, exhaust, surgical | 21 CFR 878.5070 | II |
| Predicate Devices: | Stryker® Neptune Waste Management System (K012991)<br>Dornoch Transposal Ultra System (K081047).<br>Valleylab OptiMumm Smoke Evacuator System (K980915) |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
4100 E. Milham Avenue
Kalamazoo. MI 49001
t: 269 323 7700 1: 269 389 5412
www.sstryker.com
{1}------------------------------------------------
| Device<br>Description: | The Neptune 2 Waste Management System is a fluid waste management system<br>with smoke evacuation. It collects surgical waste fluid within a closed suction<br>system, then disposes of it through a docking station. The system is comprised of a<br>rover and a docker (also referred to as a docking station).<br><br>The Neptune 2 Rover (Model Number: 0702-001-000) is a mobile, AC-powered<br>surgical fluid waste collection device. It is intended to collect waste fluids and<br>evacuate smoke during surgical procedures.<br><br>The Neptune 2 Docking Station (Model Number: 0702-014-000) is a stationary<br>device for automated rover cleaning and waste offloading. It provides a fixed<br>connection to the hospital's water supply, sewer, and electricity.<br>The following Stryker accessories are for use with the Neptune 2 Waste<br>Management System: | | | | | | | | | | | | | | | | | | | | |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| | Model Number Product Name 0700-001-026 Neptune Docking Detergent 0702-020-000 Neptune 2 4-Port Manifold 0702-020-001 Neptune 2 Specimen Collection 4-Port 0702-025-000 Neptune 2 Single-Port Manifold 0702-034-000 Fluid Suction HEPA Filter 0702-040-000 Smoke Evacuator ULPA Filter 0702-045-023 Smoke (Evacuator) Tubing, 3/8 inch x 10 feet 0700-026-000 Smoke (Evacuator) Tubing, 7/8 inch x 10 feet 0702-045-027 PenAdapt® | | | | | | | | | | | | | | | | | | | | |
| Indications<br>for Use: | The Neptune 2 Waste Management System is intended to be used in the operating<br>room, pathology, surgical centers, and doctor's offices to collect and dispose of<br>surgical fluid waste as well as collect smoke generated from electrocautery or laser<br>devices. | | | | | | | | | | | | | | | | | | | | |
| Contraindications | The Neptune 2 Waste Management System is contraindicated against:<br>• Connection directly to chest tubes.<br>• Connection to closed wound drainage systems. | | | | | | | | | | | | | | | | | | | | |
| Testing | The Neptune 2 Waste Management System meets the specification and<br>performance characteristics as identified in Stryker's internal design control<br>procedures. Bench testing was performed to demonstrate the performance of the<br>device. Testing was completed to verify and validate the device as outlined in the<br>device description. Testing included electrically powered suction safety, electrical<br>safety, vacuum cycling, software, smoke evacuation and filtration. Human factors<br>analysis and usability testing was performed on the user interface, labeling, and<br>training materials. | | | | | | | | | | | | | | | | | | | | |
| Biocompatibility | Biocompatibility testing of the Neptune 2 Waste Management System confirmed that<br>the device meets the applicable requirements of the FDA Blue Book Memorandum<br>G95-1 entitled Use of International Standards ISO-10993 Biological Evaluation of | | | | | | | | | | | | | | | | | | | | |
·
{2}------------------------------------------------
| | Medical Devices Part -1: Evaluation and Testing and are biocompatible. |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence<br>(SE) Rationale: | The Neptune 2 Waste Management System, when compared to its predicates, has an equivalent intended use, protocols for use, mode of operation, technological characteristics and performance specifications.<br>(See attached Table 1: Device Comparison Matrix) |
| Safety and<br>Effectiveness: | Based upon the comparison to the predicate devices, the Stryker Neptune 2 Waste Management System is substantially equivalent to the predicate devices. |
| Submitted by: | Michelle Jump<br>Associate Regulatory Affairs Analyst |
| | <div> <br/> Signature </div> |
| Date Submitted: | December 20, 2013 |
{3}------------------------------------------------
4100 E. Milham Avenue
Kalamazoo, Mi 49001
1: 269 323 7700 f: 269 389 5412
www.siryker.com
stryker
Instruments
Stryker Neptune 2
Covidien (Valleylab)
Dornoch Transposal
Stryker Neptune 1
# able 1: Device Comparison Matri
| Model Number | Clearance | Classification | Regulation | Product Code | Indications for Use |
|---------------------------|-----------|----------------|------------------------------------|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Rover: 0700-010-00 | 510k | II | 21 CFR 878.5070 | FYD | The Neptune Waste Management system is intended to be used in the Operating Room, Surgical Centers and Doctor's Offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices. |
| Duo: DU100<br>Quad: QD100 | K012991 | II | 21 CFR 878.4780 | JCX | The Dornoch Transposal Ultra System is self-powered suction / vacuum pump intended to collect and dispose of liquid waste within Hospital Operating Rooms, Pathology Labs, Surgical Outpatient Centers, and Doctor's Offices. |
| OptiMumm: 945 102 097 | K081047 | II | 21 CFR 878.5070 | FYD | The Indications for Use of the Valleylab OptiMumm Smoke Evacuator system are for the removal of smoke and incidental fluids produced during electrosurgery and/or laser surgery. The removal of smoke from the surgical site improves visibility and reduces potential health hazards associated with surgical smoke. |
| Rover: 0702-010-000 | K980915 | II | 21 CFR 878.4780<br>21 CFR 878.5070 | JCX<br>FYD | The Neptune Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor's offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices. |
| | Pending | II | | | |
{4}------------------------------------------------
| lest toose. Connection to closed ound drainage systems.<br>The Neptune 2 Waste Management System is contraindicated stated is the states • Connection directly so<br>hest tubes. | | 19 x 23 x 70" | swiveling hospita<br>grade caster | Rubber bumpers | urfaces of the rover vith a soft, lint-free cloth noistened with a non-<br>nstructions. Caution against using solvents, lubricants, lubricants, other chemicals,<br>brasive, hospital lisinfectant prepared ccording to the<br>Vipe the externa<br>ncluding glutaraldehyde c similar chemical<br>nanufacturer'<br>cleaners | No | と | Yes |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|--------------------------------------------------------------|-----------------------------------|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|----------------------------|----------------------------------------|
| None | | | | . :: , ' | .5 - | . " | | |
| None | | 24" x 24" x 55" or 23" x 33" x 33" x 55" depending on mode | swiveling caster | lastic bumper | | No | epending on mode<br>2 or 4 | Yes |
| None | | 18 x 25 x 51" | swiveling hospit<br>grade casters | Rubber bumper | Wipe the surfaces of the Restitio and the criation and the station and the station and the stations and the count of the count of the world coultions codes of cides to on sim<br>cleansers. | Yes | T | Yes |
| ntraindicat | ion<br>stem Configura | Size | Casters | Bumpers | aning instructi | red<br>lover is battery-power luring docking | of canister | Volume measuremen vith digital display |
{5}------------------------------------------------
| Offloading and canister<br>cleaning accessory | Neptune 1<br>Docking Station<br>(0700-005-000) | Ultra Evac<br>Docking Station<br>UL-EV100 | Neptune 2<br>Docking Station<br>(0702-014-000) |
|-----------------------------------------------------|------------------------------------------------|---------------------------------------------------------------------------------------------|------------------------------------------------|
| Rover empty weight | 230 lbs | 195 lbs or 225 lbs<br>(depending on model) | 300 lbs |
| Electrical isolation type | Class I, Type CF Applied<br>Part | Information not available | Class I, Type CF Applied<br>Part |
| Enclosure protection | IPXO Ordinary Equipment | Information not available | IPXO Ordinary Equipment |
| 120V single phase<br>electric power<br>requirements | 16 A | 4.5 A | 12 A |
| Number of uses before<br>docking | Limited by fluid capacity | Limited to 2 or 4<br>procedures<br>(depending on model) | Limited by fluid capacity |
| User display | LCD screen | LCD screen | LCD screen |
| User interaction to dock | Push rover into docker | Push Rover adjacent to<br>Docker and Insert<br>docker's gas-pump-style<br>nozzle into rover | Push rover into docker |
| Waste/water connection<br>to rover | Dry-break hydraulic<br>couplings | Information not available | Dry-break hydraulic<br>couplings |
| Rover-docker coupling<br>size | .875" body diameter | .625" OD | 1.325" body diameter |
| Rover-docker coupling<br>orientation | Horizontal | Approximately 25° from<br>horizontal | Vertical |
| Water inlet control | Solenoid valve | Solenoid valve | Solenoid valve |
| Backflow prevention<br>valve | Internal | External | Internal |
| Weight | 90 lbs | 100 lbs | 95 lbs |
| Waste/water connection<br>to facility | Quick disconnect fittings | Information not available | Quick disconnect fittings |
{6}------------------------------------------------
### Neptune docking
etergent (0700-00)
xed rubber pa 0 - 480 mm-H 23″ x 23″ x 16 igital displa
(mm-Hg) nfrared Sprinkle НЕРА 3 A Yes Yes Yes Yes Yes ನ . ---ﺎﺭﻳﺲ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ … .. 、『 ్ ప్ర ﺃ ﺍﻟ : : . . ﻳﻮ . ・ ﻨﻪ : : .............................................................................................................................................................................. 。 : ・・・・
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ . . . . . .. . :: : . ormation not availab tra Enzyme (UL-EZ20
ectrical contac 4 - 700 mm-Hj 0" x 16" x 28 igital display
(mm-Hg) Sprinkler HEPA 15 A Yes Yes Yes Yes No ਨ ters with adjustab
extensions Neptune docking
tergent (0700-00)
4 - 483 mm-H Analog gauge
mm-Hg, in-Hg, in-Hg 5" x 20" x 22 Sprinkle nfrarec HEPA Yes 3 A Yes Yes Yes (દડ I nister cleaning meth ectromagnet for
olding rover to docke cuum limit adjustab multaneous levels cuum control ranj ptional wall suctio Rinses canister with
facility water cker dimension cuum Syster 20V single phas
ectric power
equirements offloads waste to
anitary sewer Canister cleaning
Jetergent uction filtration
fficiency anister vacuun ocker support over-Docker
ommunication ndicator uction orts
### ction 005
{7}------------------------------------------------
| Canister vacuum level<br>regulation | Collection Canisters | Mechanical regulator | Electronic regulator | Electronic regulator | | 20 hrs | . ' •<br>" | 25 hrs | 80 hrs |
|--------------------------------------------|----------------------------------------|--------------------------------------------------------------|---------------------------------------------------------------|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|--------|
| Rover canister capacity | 20 L | 33 L or 52 L<br>(depending on model) | 24 L | Yes | | | ત્વ જિડ | Yes | |
| Canister graduation label | Yes | Yes | Yes | T | ﺮ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍ | . | I | రా | |
| Electronic canister<br>overfill protection | Yes | Yes | Yes | 7/8" | : | 11<br>- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - | 3/8", 7/8″ | 3/8", 7/8" | |
| Mechanical canister<br>overfill protection | Float-style shut-off valve<br>in rover | Hydrophobic filter<br>located in disposable<br>canister lid. | Float-style shut-off valve<br>in rover | Yes | | - | Yes | Yes | |
| Volume display accuracy | $\pm 175$ mL | $\pm 150$ mL | $\pm 150$ mL, $\pm 50$ mL<br>(20 L canister, 4 L<br>canister) | ED | | ్ :: :<br>- 1 - | formation not availa | EO | |
| Volume display<br>resolution | 50 mL | 10 mL | 50 mL, 20 mL<br>(20 L canister, 4 L<br>canister) | SO 10993-1 | . . | ﺎ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ | 50 10993 | 20 10993- | |
| Lighted canister contents | No | Yes | No | | | | | | |
| Smoke Evacuator | | | | fertica | | 45 degree | ્ર<br>ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ | forizonta | |
| Filtration efficiency | HEPA | ULPA | ULPA | 375" | | x .320" and 1x .37 | --<br>: : : : : : : : : : : : : : | 375" | |
| Maximum flow rate | 33 CFM | 20 CFM | 24 CFM | Single patient us | | <br>ingle patient us | | Single patient | |
| Minimum flow rate | 0 CFM | 3 CFM | 21 CFM | 1-port o 4-port | | 3-port | ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ<br> | 1-port of 4-port | |
| Maximum Suction Range<br>(100% power) | 61 mm-Hg | 99 mm-Hg | 67 mm-Hg | Yes | | Yes | :-<br>. . | Yes | |
| Minimum suction range<br>(20% power) | 0 mm-Hg | 12 mm-Hg | 34 mm-Hg | 4.7 oz | | 3.1 oz | | 1.6 oz | |
| Pre-filter | Single-use disposable | Single-use disposable | Multi-use disposable | on-sterile | | lon-steril | :<br><br>:. | Non-steril | |
| Filter life indicator | Yes | Yes | Yes | | | | | | |
| Variable speed settings | Infinite<br>(continuous rotary dial) | 5 | 9 | Yes | | Yes | | Yes | |
| ectric motor-drive | | formation not availa | | ectric motor-driv | | | | | |
| て | | ਟ | | t | | | | | |
: :
:
{8}------------------------------------------------
{9}------------------------------------------------
| Travel | Maximum height from floor | Environmental Conditions | | | |
|---------------------------|---------------------------|-----------------------------------|----------------------------------------------|-------------------------------|---------------------------|
| | | Operation Temp | Operation Humidity | Operation Barometric Pressure | Storage / Transportation |
| 29 in | 89 in | 10 - 40°C | 30 - 75% | 700 - 1060 hPa | -20 - 40°C |
| Information not available | Information not available | 10 - 40°C | 30 - 75% | 700 - 1060 hPa | -20 - 40°C |
| 32 in | 102 in | Information not available | Information not available | Information not available | Information not available |
| | | 10 - 40°C | 30 - 75% | 700 - 1060 hPa | -20 - 40°C |
| | | Storage / Transportation Humidity | Storage / Transportation Barometric Pressure | | |
| | | 10 - 75% | 500 - 1060 hPa | | |
| | | 10 - 75% | 500 - 1060 hPa | | |
| | | Information not available | Information not available | | |
| | | 10 - 75% | 500 - 1060 hPa | | |
. ﺑ
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized human figure with outstretched arms, representing care and protection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure. The logo is printed in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2013
Stryker Corporation Michelle Jump, RAC, M.S. Regulatory Affairs Associate Analyst 4100 E. Milham Ave, KALAMAZOO MI 49001
Re: K132671
Trade/Device Name: Neptune 2 Waste Management System Regulation Number: 21 CFR 872.4780 Regulation Name: Apparatus, suction, ward use, portable, ac-powered Regulatory Class: II Product Code: JCX, FYD Dated: November 14, 2013 Received: November 15, 2013
Dear Ms. Jump:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
Page 2 - Ms. Jump
.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely yours,
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/11/Picture/7 description: The image contains a signature and some text. The text reads "Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID FOR". The signature is illegible.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, . Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
510(K) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _ Stryker® Neptune 2
### Indications for Use
ﺮ ﺍﻟﻤﻌﺎﺩﻟﺔ ﻣﻮﺍﺩ ﺍﻟﻤﻮﺿﻮﻋﺎﺕ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﻠﺘﻤﻴﺔ -
The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor's offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
and/or
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth F. Glaverie -S 2013.12.17 18:13:45 -05'0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.