K132225 · Devon Medical, Inc. · OMP · May 15, 2014 · General, Plastic Surgery
Device Facts
Record ID
K132225
Device Name
EXTRICARE
Applicant
Devon Medical, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
May 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The extriCARE® 3600 Negative Pressure Wound Therapy System is indicated for wound management via the application of negative pressure to the wound by the removal of wound exudate, infectious materials, and tissue debris from the wound bed. The extriCARE® 3600 Negative Pressure Wound Therapy System is indicated for the following wound types: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafis.
Device Story
System comprises vacuum pump, canister, tubing, and NPWT bandage or foam kits; provides negative pressure to wound bed to remove exudate, infectious materials, and tissue debris. Pump features digitally programmable continuous or intermittent pressure modes. Used in clinical settings; operated by healthcare professionals. Foam kits include polyurethane foam, transparent film drape, paper ruler, and suction bell with drainage tube. Device promotes healing by maintaining negative pressure environment. Healthcare providers monitor wound status and adjust pump settings based on clinical assessment.
Clinical Evidence
Bench testing only. Performance testing included evaluation of sensors, hook, canister, clamp, air filter, and simulated wound performance. Biocompatibility testing conducted per ISO-10993, including in-vitro cytotoxicity, skin irritation, and skin sensitization, with all tests passed. Electrical safety and EMC testing performed in accordance with IEC 60601-1:2005, IEC 60601-1-2:2007, and AAMI ES 60601-1:2005.
Indicated for wound management in patients with chronic, acute, traumatic, subacute, or dehisced wounds, partial-thickness burns, ulcers (diabetic or pressure), flaps, and grafts.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
# 510K Summary
page 1 of 3
c
Cara Cara Cara Ca
# 510(K) SUMMARY
#### Submitter:
Devon Medical Inc.
# Contact Person:
Ruth Wu, COO 1100 First Avenue, Suite 202 King of Prussia, PA 19406 866.446.0092 Phone: 484.636.0211 Fax:
# Common Classification & Proprietary Names:
Negative Pressure Wound Therapy System Common Names: Proprietary Names: extriCARE 3600 extriCARE Foam Kit
#### Date Prepared:
January 16th, 2013
#### Classification
The classification name, 21 CFR Part and Paragraph number, product code and classification of the extriCARE 3600:
| FDA | Classification | Product<br>Code | Class |
|--------------------|-------------------------|-----------------|-------|
| 21 CFR<br>878.4780 | Powered Suction<br>Pump | OMP | II |
#### Predicate Devices:
The extriCARE 3600 Negative Pressure Wound Therapy System and extriCARE Foam Kit is substantially equivalent to the following:
| Predicate Device | Manufacturer | 510(k)# |
|----------------------------|--------------------------------|---------|
| extriCARE 2400 | Devon Medical | K110078 |
| V.A.C.® Freedom™ | Kinetic Concepts, Inc. | K032310 |
| Wound Pro Apex | Accuro Medical Products<br>LLC | K100823 |
| NPWT Foam<br>Dressing Kits | Smith & Nephew | K082211 |
#### Device Description
The system consists of a vacuum pump, canister, tubing, NPWT bandage or NPWT Foam Kits. In operation, the device is attached via the tubing to a
{1}------------------------------------------------
# 510K Summary
Negative Pressure Wound bandage or foam kits. Maximum pressure and continuous mode or intermittent mode selection are digitally programmable. The extriCARE 3600 pump can be sold alone or as a part of the extriCARE 3600 system with extriCARE bandages or foam kits. The extriCARE bandages are an all-in-one wound dressing; with no wound packing required which is cleared in the predicate K110078. The foam kit consists of Polyurethane foam, transparent film drape, paper ruler, Suction Bell with connecting (drainage) tube and clamp, which is included in this submission for review.
# Intended Use:
The extriCARE 3600 Negative Pressure Wound Therapy System is indicated for wound management via the application of negative pressure to the wound by the removal of wound exudate, infectious materials, and tissue debris from the wound bed. The extriCARE 3600 Negative Pressure Wound Therapy System is indicated for the following wound types: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.
# Technological Characteristics:
The manufacturer believes that the technological characteristic of the extriCARE 3600 are substantially similar to those of the predicate devices. The extriCARE 3600 has very similar components to its predicate devices and very similar principles of operation. The device consists of an electrically generated source of vacuum and a canister. Like the predicates, vacuum, is applied for a specified period of time and intensity, according to the physician's prescription.
Both extriCARE bandages and foam dressing kits are alternatively used together with the extriCARE 3600 pumps. The extriCARE bandages are the same as stated in the previously cleared submission K110078.
The extriCARE foam kits use similar components and similar principles of operation to the predicate device. Like the predicate, the foam is placed into the wound and attached to a vacuum device, which may provide negative pressure to the wound site which may promote healing by removing wound exudates.
# Performance Testing
To ensure the performance of extriCARE® 3600 device meet the design input when conjunction with NPWT bandage and foam kits, a series of bench tests were conducted to ensure device's functions, which include the sensors, hook, canister, clamp, air filter and a simulated wound performance test.
{2}------------------------------------------------
# 510K Summary
# EMC Testing
To verify that the device design met it's functional and performance requirements, representative samples of the device underwent function and mechanical testing, EMC testing in accordance with IEC 60601-1:2005, IEC 60601-1-2:2007 and electrical safety testing in accordance with AAMI ES 60601-1:2005 standards.
K13225 page 3 of 3
# Biocompatibility
The extriCARE 3600 system consists of the pump, bandages or alternative foam kits. While the pump has no direct body contact when used as indicated, the bandage and foam kits do have direct body contact. Per ISO-10993 requirements, the bandage samples and foam kits samples were tested for the following items and all tests were successfully passed.
- In-vitro Cytotoxicity .
- Skin irritation .
- Skin Sensitization .
# Statement of Substantial Equivalence
The extriCARE 3600 system is substantially equivalent in technology, function, operating parameters, and intended use to predicate devices that are currently commercially available and in distribution, and does not raise any new questions of safety or effectiveness.
### Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Devon Medical Products, believes that the extriCARE 3600 system (pump, bandage, foam, and accessories) is safe and effective and substantially equivalent to the predicate devices as described herein.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the border of the circle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
May 15, 2014
Devin Medical, Inc. Ruth Wu, Chief Operating Officer 1100 First Avenue, Suite 202 King of Prussia, Pennsylvania 19406
Re: K132225
Trade/Device Name: extriCARE® 3600 Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: February 11, 2014 Received: February 14, 2014
Dear Ms. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dr has master and regulations administered by other Federal agencies. You must or any it all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
# Page 2 - Ruth Wu, Chief Operating Officer
.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita Ashar, MD, MBA, FACS Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
cc: DMC 510(k) Staff Division D.O.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known) K132225
#### Device Name
extriCARE 3600 Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
The extriCARE® 3600 Negative Pressure Wound Therapy System is indicated for wound management via the application of negative pressure to the wound by the removal of wound exudate, infectious materials, and tissue debris from the wound bed. The extriCARE® 3600 Negative Pressure Wound Therapy System is indicated for the following wound types: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafis.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# ound Da
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the The burden time for this Colloon of information to recess, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.