K110078 · Devon Medical Products, Inc. · OMP · Feb 10, 2012 · General, Plastic Surgery
Device Facts
Record ID
K110078
Device Name
EXTRICARE
Applicant
Devon Medical Products, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Feb 10, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The extriCARE 2400 Negative Pressure Wound Therapy Pump System is a portable, battery powered pump intended to generate negative pressure or suction to remove wound exudates, infectious material, and tissue debris from the wound bed which may promote wound healing.
Device Story
Portable, battery-powered negative pressure wound therapy (NPWT) system; consists of vacuum pump, canister, and tubing. Connects to all-in-one wound dressing; no wound packing required. Operates by applying digitally programmed negative pressure to wound bed; removes exudates, infectious material, and tissue debris. Used in clinical or home settings; operated by healthcare providers or patients. Output allows clinicians to manage wound healing via controlled suction. Benefits patient by facilitating wound bed preparation and promoting healing.
Clinical Evidence
Bench testing only. Performance verified through functional and mechanical testing, including vacuum pressure calibration, intermittent mode treat time verification, flow rate performance, alarm/battery verification, and exudate removal. Bandage performance verified via peel strength, shear adhesion, and leakage tests. Biocompatibility testing conducted per ISO 10993-1 for prolonged contact (24h-30 days), including sensitization, cytotoxicity (MEM elution), and intracutaneous irritation; all tests passed.
Indicated for patients requiring negative pressure wound therapy to remove wound exudates, infectious material, and tissue debris from the wound bed to promote healing.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows a handwritten number and a fraction. The number "5110078" is written at the top of the image. Below the number, the fraction "1/3" is written in a larger font size.
# 510(K) SUMMARY
#### Submitter:
Devon Medical Products, Inc.
#### Contact Person:
Ruth Wu, COO 1100 First Avenue, Suite 202 King of Prussia, PA 19406 Phone: 866.446.0092 484.636.0211 Fax:
## Common Classification & Proprietary Names:
| Common Names: | Negative Pressure Wound Therapy System |
|-------------------|----------------------------------------|
| Proprietary Name: | extriCARE 2400 |
#### Date Prepared:
August 29, 2011
#### Classification
The classification name, 21 CFR Part and Paragraph number, product code and classification of the extriCARE 2400:
| | FDA | Classification | Product Code | Class |
|-----------------|-----|----------------------|--------------|-------|
| 21 CFR 878.4780 | | Powered Suction Pump | OMP | II |
## Predicate Devices:
The extriCARE 2400 Negative Pressure Wound Therapy System is substantially equivalent to the following.
| Predicate Device | Manufacturer | 510(k)# |
|-------------------------------|-----------------|---------|
| NPD 1000 | Kalypto Medical | K080275 |
| MobIVAC W.C. | Ohio Medical | K081185 |
| Renasys Gauze<br>Dressing Kit | Smith & Nephew | K110647 |
### Device Description
The system consists of a vacuum pump, canister, and tubing. In operation, the device is attached via the tubing to a Negative Pressure Wound bandage. Maximum pressure and mode selection are digitally programmable. The extriCARE pump can be sold alone or as a part of the extriCARE 2400 system with extriCARE bandages. The extriCARE bandages are an all-in-one wound dressing, with no wound packing required.
#### Intended Use:
{1}------------------------------------------------
# SECTION 5
The extriCARE 2400 Negative Pressure Wound Therapy System is a portable, battery powered pump intended to generate negative pressure or suction to remove wound exudates, infectious material, and tissue debris from the wound bed which may promote wound healing.
0078
## Technological Characteristics:
The manufacturer believes that the technological characteristic of the extriCARE 2400 are substantially similar to those of the predicate devices. The extriCARE 2400 has very similar components to its predicate devices and very similar principles of operation. The device consists of an electrically generated source of vacuum, a canister. Like the predicates, vacuum, is applied for a specified period of time and intensity, according to the physician's prescription.
## Performance Testing
To verify that the device design met it's functional and performance requirements, representative samples of the device underwent function and mechanical testing, EMC testing in accordance with IEC 60601-1-2 and electrical safety testing in accordance with UL 60601-1.
| Function Performance Tests |
|-----------------------------------------------------------------------|
| PQ-01 extriCARE 2400 Vacuum Pressure Calibration |
| PQ-02 Comparison on SE for Vacuum Pressure Calibration |
| PQ-03 Treat Time verification on Intermittent Mode for extriCARE 2400 |
| PQ-04 Flow Rate Performance on for extriCARE 2400 |
| PQ-05 Alarm Verification for extriCARE 2400 |
| PQ-06 Battery Verification for extriCARE 2400 |
| PQ-07 Exudate Removal test on extriCARE 2400 and extriCARE bandage |
| PQ-08 Peel strength test on extriCARE Bandage T-ZG-027 |
| PQ-09 Shear Adhesion test on extriCARE Bandage T-ZG-065 |
| PQ-10 Leakage test on extriCARE Bandage |
### Biocompatibility
The extriCARE 2400 system consists of both the pump and the bandage. While the pump has no direct body contact when used as indicated, the bandage does have direct body contact. Per ISO-10993 B-prolonged contact(24h - 30 days) requirements, the bandage samples were tested for the following items and all tests were successfully passed.
| | | Date | Protocol # | Result Summary |
|--------|---------------------------------------------------------------|--------------|--------------------------|------------------------------------------------------------------|
| Test 1 | ISO Guinea Pig<br>Maximization<br>Sensitization Test<br>(GLP) | Oct 8, 2010 | Wuxi AppTec<br>900850-23 | The test article<br>did not elicit<br>sensitization<br>response. |
| Test 2 | Cytotoxicity - MEM | Oct 29, 2010 | Nelson | Non-cytotoxic |
{2}------------------------------------------------
# SECTION 5
K110078
3/3
| | Elution without silver (ISO) | | 201003532 Rev 01 | |
|--------|-------------------------------------------------|--------------|----------------------------|---------------|
| Test 3 | Cytotoxicity – MEM<br>Elution with silver (ISO) | Oct 21, 2010 | Nelson<br>201003441 Rev 01 | Non-cytotoxic |
| Test 4 | Intracutaneous Irritation | Nov 11, 2010 | NLI# 550001 | Non-irritant |
## Statement of Substantial Equivalence
The extriCARE 2400 system is substantially equivalent in technology, function, operating parameters, and intended use to predicate devices that are currently commercially available and in distribution, and does not raise any new questions of safety or effectiveness.
### Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Devon Medical Products, believes that the extriCARE 2400, is safe and effective and substantially equivalent to the predicate devices as described herein.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is an abstract image of an eagle or bird-like figure with stylized feathers.
Devon Medical Products, Inc. % Ms. Ruth Wu. COO 1100 First Avenue, Suite 202 King of Prussia, Pennsylvania 19406
Re: K110078
Trade/Device Name: extriCARE 2400 Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: January 18, 2012 Received: January 31, 2012
Dear Ms. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB 1 0 2012
{4}------------------------------------------------
Page 2 - Ms. Ruth Woo, COO
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely y
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## SECTION 4
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _
Device Name: extriCARE 2400
The extriCARE 2400 Negative Pressure Wound Therapy Pump System is a portable, battery powered pump intended to generate negative pressure or suction to remove wound exudates, infectious material, and tissue debris from the wound bed which may promote wound healing.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kucera for MIXM
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110078
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.