K132087 · Philips Health Care · JAK · Jul 24, 2013 · Radiology
Device Facts
Record ID
K132087
Device Name
PERCUNAV SYSTEM
Applicant
Philips Health Care
Product Code
JAK · Radiology
Decision Date
Jul 24, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance, (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Rotational Fluoroscopy, Endoscopy and other imaging systems. CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various combinations, such as CT with MR, MR with ultrasound, etc. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account patient movement. This is intended for treatment planning and guidance for clinical, interventional, and/or diagnostic procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device. The device is intended to be used in interventional and diagnostic procedures in a clinical setting. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another.
Device Story
PercuNav is a stereotaxic image-guided surgery system; provides diagnostic/interventional guidance. System inputs include pre-procedural and intra-procedural imaging (CT, MR, US, PET, SPECT, Rotational Fluoroscopy). Ultrasound Tracker accessory uses electromagnetic (EM) tracking to monitor ultrasound transducer position in real-time within a varying EM field. Software fuses multi-modality images; displays simulated insertion tool (needle, guidewire, probe) path on monitor relative to target organs; accounts for patient movement. Used in clinical settings by physicians. Output assists in locating areas of interest; guides instrumentation to targets; enables image-free proximity tracking between devices. Benefits include improved accuracy for biopsies, ablations, and other interventional procedures. Modification focuses on reusability of Ultrasound Tracker; redesigned for cleaning/disinfection per AAMI/ISO standards.
Clinical Evidence
Bench testing only. No clinical data. Performance validated via surgical simulation and non-clinical accuracy testing. Surgical simulation used phantom-based models to determine Euclidian System Error (Target Registration Error). Non-clinical accuracy testing evaluated image registration and fusion accuracy. Results confirmed system meets all functional requirements and performs as intended.
Technological Characteristics
Electromagnetic (EM) tracking system. Ultrasound Tracker accessory redesigned for reusability; compliant with AAMI TIR12:2010, AAMI TIR30:2011, ISO 17664:2004, and ANSI ST81:2004/(R)2010 for reprocessing. System performs spatial mapping (registration) between image space and physical space. Connectivity includes multi-modality image fusion (CT, MR, US, PET, etc.).
Indications for Use
Indicated for patients undergoing interventional and diagnostic procedures requiring image-guided navigation, including soft tissue biopsies/ablations, bone biopsies/ablations, nerve blocks, pain management, drainage, hydrodissections, bladder stimulation, fiducial placement, tumor resection, and sinus/intranasal/transphenoidal procedures.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the handwritten text "K132087" on a white background. To the right of the number is a black box with the text "510(k) Page Number" written in white. The text is likely part of a document or label, possibly related to a medical device or regulatory submission.
# 510(k) Summary of Safety and Effectiveness for Update to PercuNav
This summary of safety and effectiveness is provided as part of the Premarket Notification in compliance with 21 CFR. Part 807.92.
#### I. Manufacturer/Owner
Manufacturer's Name Philips Healthcare. 49 Spadina Avenue, #310 Toronto, Ontario Canada M5V 2J1 Tel: (416) 603-8349 Fax: (416) 603-8354
JUL 2 4 2013
#### II. Contact Person
Dave Szczupakowski Senior Manager, Quality & Regulatory (717) 667-5016 Email: dave.szczupakowski@philips.com
#### III. Propriety Name/Classification Name
PercuNav Computer assisted, image-guided surgery system Classification Name: Computed Tomography X-ray System Class II as described in 21 CFR 892.1750 JAK
Product Code:
Propriety Name:
Common Name:
#### IV. Date Prepared
April 01, 2013
#### V. Device Description
The PercuNav system provides image guided intervention and diagnostic information which guides interventional instrumentation to targets that have been defined by the physician. The target can be indicated either pre-procedurally or intra-procedurally using images or relative to an indicated position on the patient. As a diagnostic system, it combines pre-procedural and intra-procedural imaging to assist in locating areas of interest detected on one set of images on the other. The system provides fusion between different modalities. Different imaging modalities such as CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various
{1}------------------------------------------------
combinations, for example CT with MR, CT with Ultrasound, PET/CT with ultrasound, MR with ultrasound, etc.
#### VI. Purpose of Submission
Philips Healthcare is submitting this Special 510(k) to address modifications to the Ultrasound Tracker accessory. When used together with the PercuNav system, the Ultrasound Tracker device provides electromagnetic (EM) tracking of an ultrasound transducer in a varying EM field, and displays the transducer's real-time position in the PerculVav system. The Ultrasound Tracker attaches to a mounting plate or equivalent mating surface on the transducer or transducer bracket.
The Philips PercuNav Ultrasound Tracker is designed to be used with the PercuNav image-guided intervention system for diagnostic or interventional procedures using ultrasound. The Ultrasound Tracker consists of a connector, a cable, and trackedsensor housing with a release tab that attaches to the needle-guidance system.
Modifications were made to the PercuNav Ultrasound tracker accessory which focuses mainly on durability of the product. The currently released PercuNav accessory is provided as single use devices and sterile. This has been challenged by many customers who feel that these trackers should be reusable and also do not need to be sterile as the only come into contact with the intact skin of the patient (i.e. noncritical medical device as per Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling issued on May 2, 2011).
The modifications described in this Special 510(k) have not altered the basic fundamental technology of the predicate PercuNav Ultrasound tracker K121498. The modified Ultrasound tracker has the same technological characteristics as the legally marketed predicate device with exception that this new tracker is provided non sterile and intended undergo cleaning and disinfection reprocessing at the hospital facility.
#### Description of Device Modification VII.
Philips Healthcare is submitting this Special 510(k) to address modifications to the predicate Ultrasound Tracker [K121498]. The proposed Ultrasound Tracker was redesigned to withstand cleaning and disinfection associated with reuse in accordance with the following standards:
- AAMI TIR12:2010 Designing, testing, and labeling reusable medical . devices for reprocessing in health care facilities: A guide for medical device manufacturers.
- . AAMI TIR30:2011 - A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices.
- . ISO 17664:2004 - Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical dévices.
{2}------------------------------------------------
- ANSI ST81:2004/(R) 2010 Sterilization of medical devices Information . to be provided by the manufacturer for the processing of resterilizable medical devices.
## VIII. Indications for Use
PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Rotational Fluoroscopy, Endoscopy, and other imaging systems. CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various combinations, such as CT with MR, MR with ultrasound, etc. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account patient movement. This is intended for treatment planning and guidance for clinical, interventional, and/or diagnostic procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device.
The device is intended to be used in interventional and diagnostic procedures in a clinical setting. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another.
Example procedures include, but are not limited to:
- Image fusion for diagnostic clinical examinations and procedures .
- . Soft tissue biopsies (liver, lung, kidney, breast, pancreas, bladder, adrenal glands, lymph node, mesentery, etc.)
- . Soft tissue ablation (liver, kidney, breast, pancreas, lung, etc.)
- . Bone ablations
- . Bone biopsies
- Nerve Blocks & Pain Management .
- . Drainage placements
- Hydrodissections •
- . Bladder Stimulation
- Fiducial placements
- . Tumor resections
- . Sinus procedures
- . Intranasal procedures
- Transphenoidal procedures .
#### IX. Substantial Equivalence
The technological characteristics of the proposed Philips PercuNav Ultrasound Tracker is equivalent to the cleared predicated device with the exception that this new tracker is provided non-sterile and intended undergo cleaning and disinfection reprocessing at the hospital facility.
{3}------------------------------------------------
This update to the Philips PercuNav Ultrasound Tracker is substantially equivalent to the FDA cleared frameless stereotaxic system, predicate PercuNav K121498. As required by risk analysis, all verification and validation activities were performed by designated individual(s) and the results demonstrated substantial equivalence.
The device labeling contains instructions for use which is substantially equivalent to the predicate device but has been revised in accordance with the FDA Draft guidance Document Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling issued on May 2, 2011 to ensure safety and effectiveness of the reprocessing instructions provided to the user.
#### X. Non Clinical Performance Data
Testing for the Philips PercuNav Ultrasound Tracker was performed to ensure that functional requirements have been met, and that core functions execute as expected. Safety and Effectiveness has been established through bench testing and non-clinical performance testing inducing Surgical Simulation and the Clinical Accuracy simulation tests. Results from performance testing conducted demonstrate that the Philips PercuNav Ultrasound Tracker is safe and effective for use with the PercuNav System [K121498].
Product requirements are decomposed into test cases and variations, consisting of conditions and expected results. Each case set is associated with a specific requirement and is grouped according to functional characteristics. The Verification and Validation plan identifies all variations, including input and expected results (outputs), test setup conditions (environmental needs), requirements, variations tested, and procedural steps taken to execute the test case. The Verification and Validation results summary demonstrate that the device satisfies all performance and functional requirements.
### Surgical simulation
A real-world simulation of a guided surgical procedure to determine the accuracy of the PercuNav system using a control measurement gauge. It is also used to determine if the system works as intended in a simulated surgical environment. The system is validated using a phantom based simulation that is designed to determine the error in the system by determining a value of "Euclidian System Error". The Euclidian System Error is the "Target Registration Error" (TRE) commonly used for accuracy assessment in image guided surgery. All instrumentation and accessories provided with the PercuNav system are subjected to Surgical Simulation Test Case Protocol.
### Non Clinical Accuracy
The Non Clinical Accuracy test is a simulation test intended to evaluate Image Registration and Image Fusion accuracy of the Philips PercuNav Ultrasound Tracker in order to demonstrate safety and effectiveness of the system.
- a) Registration Test
The system performs spatial mapping from image space to physical space through a process called Registration. This allows the physician to correlate
{4}------------------------------------------------
image datasets with one another, as well as to the patient. The purpose of this test is to verify image registration
b) Image Fusion Accuracy Test
Image fusion provides the advantage of combining high resolution crosssectional imaging modalities with live ultrasound imaging. Image fusion displays the pre-procedure images, reformatted in real time, to continuously match live ultrasound imaging. An electromagnetic tracker affixed to the ultrasound transducer tracks the position and orientation of the ultrasound scan plane. The PercuNav software uses this information to fuse the preprocedure images and the live ultrasound into one image. The purpose of this test is to qualify each registration method that is available to the operator.
System testing to date shows that all accuracy and performance requirements for PercuNav system use with the Philips PercuNav Ultrasound Tracker have been met and the system is safe and effective for its intended use. The completed tests were conducted to ensure that the changes to the proposed device did not introduce any new issues of safety or effectiveness from our legally marketed device - K121498.
#### XI. Conclusion
Verification and Validation testing activities were required to establish the performance, functionality, and reliability characteristics of the modified Philips PercuNav Ultrasound Tracker. Testing involved system level tests, performance tests, and safety testing. Results of the verification and validation activities for the Ultrasound Tracker confirms that the device performed as intended, is safe and effective, and is substantially equivalent to the currently marketed device [K121498].
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a circular logo for the Department of Health & Human Services (HHS) in the USA. The logo features the HHS symbol, which consists of three stylized waves or lines, positioned to the right of the text. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 24, 2013
Philips Healthcare % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street N W BUFFALO MN 55313
Re: K132087
Trade/Device Name: PercuNav System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK, IYO, LLZ Dated: July 1, 2013 Received: July 12, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Job
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDcyices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
## 510(k) Number (if known): K132087
Device Name: PercuNav
Indications for Use:
PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance, (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Rotational Fluoroscopy, Endoscopy and other imaging systems. CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various combinations, such as CT with MR, MR with ultrasound, etc. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account patient movement. This is intended for treatment planning and guidance for clinical, interventional, and/or diagnostic procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device.
The device is intended to be used in interventional and diagnostic procedures in a clinical setting. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Page 1 of _ 2 _
{8}------------------------------------------------
## Indications for Use
Example procedures include, but are not limited to:
- Image fusion for diagnostic clinical examinations and procedures ●
- Soft tissue biopsies (liver, lung, kidney, breast, pancreas, bladder, adrenal . glands, lymph node, mesentery, etc.)
- Soft tissue ablation (liver, kidney, breast, pancreas lung, etc.) .
- Bone ablations .
- Bone biopsies
- Nerve Blocks & Pain Management .
- Drainage placements .
- Hydrodissections .
- Bladder Stimulation �
- Fiducial placements .
- Tumor resections
- Sinus procedures
- Intranasal procedures
- Transphenoidal procedures
Prescription Use v (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Smh.7)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) K132087
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.