K121498 · Philips Health Care · JAK · Dec 14, 2012 · Radiology
Device Facts
Record ID
K121498
Device Name
PERCUNAV
Applicant
Philips Health Care
Product Code
JAK · Radiology
Decision Date
Dec 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Rotational Fluoroscopy, Endoscopy, and other imaging systems. CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various combinations, such as CT with MR, MR with ultrasound, etc. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account patient movement. This is intended for treatment planning and guidance for clinical, interventional, and/or diagnostic procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device. The device is intended to be used in interventional and diagnostic procedures in a clinical setting. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another. Example procedures include, but are not limited to: Image fusion for diagnostic clinical examinations and procedures. Soft tissue biopsies (liver, lung, kidney, breast, pancreas, bladder, adrenal glands, lymph node, mesentery, etc.). Soft tissue ablation (liver, kidney, breast, pancreas, lung, etc.). Bone ablations. Bone biopsies. Nerve Blocks & Pain Management. Drainage placements. Hydrodissections. Bladder Stimulation. Fiducial placements. Tumor resections. Sinus procedures. Intranasal procedures. Transphenoidal procedures.
Device Story
PercuNav is a computer-assisted, image-guided surgery system; functions as an accessory to imaging modalities (CT, MR, US, PET, SPECT, Fluoroscopy). Inputs include archived image data (PACS, CD) and live imaging; system performs spatial mapping (registration) between image spaces or image-to-physical space. Operates via Electromagnetic Measurement System (EMMS) with field generator, control unit, and tool connection unit. Displays tracked instruments (biopsy needles, probes) relative to target anatomy or other devices on iU22 Ultrasound monitor. Used in ORs, procedure rooms, and scanner suites by physicians to facilitate minimally invasive interventions. Provides real-time navigation and path projection; accounts for patient movement. Benefits include improved accuracy in targeting lesions or anatomical structures during interventional procedures.
Clinical Evidence
No clinical data. Safety and effectiveness established through bench testing and surgical simulation (phantom-based). Accuracy assessed via Target Registration Error (TRE), defined as the discrepancy between system-reported location and ground truth (e.g., CT confirmation scan). Testing confirmed all functional requirements and accuracy targets were met.
Technological Characteristics
Stereotaxic accessory; consists of Electromagnetic Measurement System (EMMS), field generator, system unit, and tool connection unit. Integrates with iU22 Diagnostic Ultrasound System via keyboard, touch-screen, and display. Software-based image fusion and navigation; supports image-free mode. Connectivity via standard imaging interfaces (PACS, CD).
Indications for Use
Indicated for patients undergoing interventional or diagnostic procedures requiring image-guided navigation or fusion, including soft tissue biopsies/ablations, bone procedures, nerve blocks, drainage, and various surgical interventions (e.g., sinus, transphenoidal). Used in clinical settings for treatment planning and instrument guidance.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K12498
DEC 1 4 2012
2
# 510(k) Summary of Safety and Effectiveness for Update to PercuNav
### I. Manufacturer/Owner
Manufacturer's Name Philips Healthcare. 49 Spadina Avenue, #310 Toronto, Ontario Canada M5V 2J1 Tel: (416) 603-8349 Fax: (416) 603-8354
#### II. Contact Person
Tara MacKinnon Senior Manager, Quality and Regulatory Philips Healthcare 49 Spadina Avenue, Suite 310 Toronto, Ontario, Canada M5V 2J1 (416) 603-8349 Ext. 231 Email: tara.mackinnon@philips.com Dave Szczupakowski Senior Manager, Quality & Regulatory (717) 667-5016 Email: dave.szczupakowski@philips.com
#### III. Propriety Name/Classification Name
Propriety Name: Common Name: Classification Name:
Product Code:
### IV. Date Prepared
November 13, 2012
#### V. Device Description
The PercuNav image fusion and navigation feature for the iU22 Diagnostic Ultrasound System is supplied as an on-cart product that is packaged and installed much like a peripheral on the iU22 Diagnostic Ultrasound, similar to a third party VCR which can be placed on the ultrasound cart.
PercuNav computer assisted technology provides image guided intervention and diagnostic information which guides interventional instrumentation to targets that has
PercuNav Computer assisted, image-guided surgery system Computed Tomography X-ray System Class II as described in 21 CFR 892.1750 JAK
{1}------------------------------------------------
been defined by the physician. The target can be indicated either pre-procedurally or intra-procedurally using images or relative to an indicated position on the patient. As a diagnostic system, it combines pre-procedural and intra-procedural imaging to assist in locating areas of interest detected on one set of images on the other. The system provides fusion between different modalities. Different imaging modalities such as CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various combinations, for example CT with MR. CT with Ultrasound, PET/CT with ultrasound, MR with ultrasound, etc.
When used as a navigation aid, it also transforms two and three-dimensional patient images (scan sets), derived from for example, Computed Tomography (CT), Magnetic Resonance Imaging (MR), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Ultrasound (US), Rotational Fluoroscopy, etc. into dynamic representations on which a medical instrument can be navigated. The system performs spatial mapping from one image space to another image space or from image space to physical space ("registration") allowing the physician to correlate scan sets with each other and to the patient. The system facilitates minimally invasive interventional procedures. Like other commercially available image guided surgery systems, the PercuNav also offers computer assisted image-free navigation using the same instrumentation. In image-free mode the proximity of an interventional device is displayed relative to another device.
Images used by the PercuNav can include archived image data from a CD, PACS, etc., and live images from an imaging device such as ultrasound, etc. The system can also be used without the use of image data (image-free mode) when appropriate for the procedure being performed.
The PercuNav system consists of an Electromagnetic Measurement System (EMMS) (including a Field Generator and System Control Unit), a System Unit, Field generator stand, PercuNav software, Tool Connection Unit (TCU), and various instrumentation devices. The PercuNav System Unit and the PercuNav software interact with the keyboard, touch-screens, trackball, and visual display of the iU22 Diagnostic Ultrasound System to display and interact with images and data on the system monitor(s).
Targeted use areas for PercuNav include hospital operating rooms, outpatient surgery centers, ultrasound, CT and other scanner suites, and procedure rooms.
### VI. Purpose of Submission
The purpose of this premarket notification is to seek clearance for an update to the currently approved Indications for Use cleared for the PercuNav predicate device. The PercuNav System, originally named "Abaris" (K053610, cleared April 16, 2006) is currently marketed by Philips Healthcare. The manufacturer Traxtal Inc. was acquired by Philips Electronics and is now Philips Healthcare, a division of Philips Electronics Ltd.
{2}------------------------------------------------
# VII. Description of Device Modification
This submission provides new indications that are substantially equivalent to the named predicate device indications statement. The proposed indication for use is essentially identical to the currently cleared indications, with the exception of reference to example procedures where the PercuNav System can be used. As required by risk analysis, all verification and validation activities were performed by designated individual(s) and the results demonstrated substantial equivalence to the predicate device.
## VIII. Indications for Use
PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Rotational Fluoroscopy, Endoscopy, and other imaging systems. CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various combinations, such as CT with MR, MR with ultrasound, etc. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account patient movement. This is intended for treatment planning and guidance for clinical. interventional, and/or diagnostic procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device.
The device is intended to be used in interventional and diagnostic procedures in a clinical setting. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another.
Example procedures include, but are not limited to:
- Image fusion for diagnostic clinical examinations and procedures .
- Soft tissue biopsies (liver, lung, kidney, breast, pancreas, bladder, adrenal glands, . lymph node, mesentery, etc.)
- . Soft tissue ablation (liver, kidney, breast, pancreas, lung, etc.)
- Bone ablations .
- . Bone biopsies
- . Nerve Blocks & Pain Management
- . Drainage placements
- Hydrodissections .
- . Bladder Stimulation
- Fiducial placements .
- . Tumor resections
- . Sinus procedures
- Intranasal procedures .
- . Transphenoidal procedures
{3}------------------------------------------------
### IX. Substantial Equivalence
The technological characteristics of the proposed PercuNav device are similar to the cleared predicated device with exception of revisions to the system software which has improved image fusion and navigation workflow and added features which enhanced physician usability. It is important to note that all updates made to the PercuNav software have not impacted safety or effectiveness. This update to the PercuNav system is substantially equivalent to the FDA cleared frameless stereotaxic system, predicate PercuNav (a.k.a. Abaris) K053610. As required by risk analysis, all verification and validation activities were performed by designated individual(s) and the results demonstrated substantial equivalence.
The device labeling contains instructions for use which is substantially equivalent to the predicate device. It includes indications for use, cautions, contraindications, warnings, and planning guidance. This information assures safe and effective use of the device.
#### X. Testing and Performance Data
Testing for the PercuNav system was performed to ensure that functional requirements have been met, and that core functions execute as expected. Safety and Effectiveness of the PercuNav system has been established through bench testing and non-clinical performance data also known as Surgical Simulation Testing. Results from System Testing and Surgical Simulation testing demonstrate that the PercuNav system is safe and effective for its intended use.
## System Testing
Product requirements are decomposed into test cases and variations, consisting of conditions and expected results. Each case set is associated with a specific requirement and is grouped according to functional characteristics. The Software Validation Test Plan (SVTP) identifies all variations, including input and expected results (outputs), test setup conditions (environmental needs), requirements, variations tested, and procedural steps taken to execute the test case. The test case results summary demonstrate that the device satisfies all performance and functional requirements.
### Surgical simulation
A real-world simulation of a guided surgical procedure to determine the accuracy of the PercuNav system using a control measurement gauge. It is also used to determine if the system works as intended in a simulated surgical environment. The system is validated using a phantom based simulation that is designed to determine the error in the system by determining a value of "Euclidian System Error". The Euclidian System Error is the "Target Registration Error" (TRE) commonly used for accuracy assessment in image guided surgery.
TRE is a well established method of determining accuracy of image guided procedures, and has been use in a number of studies, particularly rigorous engineering evaluations (Maurer, Maciunas and Fitzpatrick 1998). TRE is defined as the error discrepancy between PercuNav and ground truth position as estimated for
{4}------------------------------------------------
example from a CT confirmation scan. TRE defines the error of misrepresentation of the tracked device location reported by the system and its real location and is the superposition of registration errors, motion artifacts, position sensor error etc. In the field of image guided intervention. TRE is considered by most researchers as the most clinically relevant measure of error since it is a direct measure of the full system accuracy.
The surgical simulation protocol and expected results (e.g. accuracy) is detailed in the Surgical Simulation Test Case Protocol. All instrumentation and accessories provided with the PercuNav system are subjected to the Surgical Simulation Test Case Protocol. The results of the surgical simulation are documented and referred to in a Verification and Validation Report for the PercuNav system.
System testing to date shows that all accuracy targets for the PercuNav system have been met and the system is safe and effective for its intended use. The completed tests were conducted to ensure that the changes to the proposed device did not introduce any new issues of safety or effectiveness from our legally marketed predicate device - K053610.
#### XI. Conclusion
The proposed indication for use is essentially identical to the currently cleared indications, with the exception of reference to example procedures where the PercuNav System, formerly known as Abaris, can be used. Safety and effectiveness of the example procedures have been established through clinical experience and preclinical testing including system design bench testing. Results of the verification and validation activities for the PercuNav system confirm that the system performed as intended, is safe and effective, and is substantially equivalent to the currently marketed predicate device [K053610].
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three flowing lines extending from its body, representing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 14, 2012
Philips Healthcare % Ms. Tara MacKinnon Senior Manager, Quality and Regualtory 49 Spandina Avenue, Suite 310 Toronto, Ontario, M5V 2J1 CANADA
Re: K121498
Trade/Device Name: PercuNav Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: November 12, 2012 Received: November 14, 2012
Dear Ms. MacKinnon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2-Ms. MacKinnon
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sean M. Boyd
Janine M. Morris Director. Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
for
Enclosure
{7}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K121498
## Device Name: PercuNav
## Indications for Use:
PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Rotational Fluoroscopy, Endoscopy, and other imaging systems. CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various combinations, such as CT with MR, MR with ultrasound, etc. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account patient movement. This is intended for treatment planning and guidance for clinical, interventional, and/or diagnostic procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device.
The device is intended to be used in interventional and diagnostic procedures in a clinical setting. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another. Example procedures include, but are not limited to:
- Image fusion for diagnostic clinical examinations and procedures .
- Soft tissue biopsies (liver, lung, kidney, breast, pancreas, bladder, adrenal glands, lymph node, . mesentery, etc.)
- Soft tissue ablation (liver, kidney, breast, pancreas,, lung, etc.) .
- Bone ablation .
- Bone biopsies .
- . Nerve Blocks & Pain Management
- . Drainage placements
- Hydrodissections .
- . Bladder Stimulation
- Fiducial placements .
- Tumor resections ●
- Sinus procedures .
- Intranasal procedures ●
- Transphenoidal procedures .
Prescription Use X (Part 21 CFR 801 Subpart D)
510(K)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
Page 1 of 1
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
| . " "<br>ﺗ<br>1<br>X<br>1 | |
|---------------------------|--|
| 11 | |
(Division Sign Off)
Division of Radiological Hea
Office of In Vitro Diagnostics and Rac
{8}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K121498
## Device Name: PercuNay
## Indications for Use:
PercuNav is a stereotaxic accessory for Computed Tomography (CT), Magnetic Resonance (MR), Ultrasound (US), Positron Emission Tomography (PET), Single Photon Emission Computed Tomography (SPECT), Rotational Fluoroscopy, Endoscopy, and other imaging systems. CT, Ultrasound, PET, MR, and Rotational Fluoroscopy may be fused in various combinations, such as CT with MR, MR with ultrasound, etc. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle, guidewire or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument taking into account patient movement. This is intended for treatment planning and guidance for clinical, interventional, and/or diagnostic procedures. The device also supports an image-free mode in which the proximity of the interventional device is displayed relative to another device.
The device is intended to be used in interventional and diagnostic procedures in a clinical setting. The device is also intended for use in clinical interventions to determine the proximity of one device relative to another. Example procedures include, but are not limited to:
- Image fusion for diagnostic clinical examinations and procedures .
- Soft tissue biopsies (liver, lung, kidney, breast, pancreas, bladder, adrenal glands, lymph node, . mesentery, etc.)
- Soft tissue ablation (liver, kidney, breast, pancreas,, lung, etc.) .
- Bone ablation .
- Bone biopsies ●
- Nerve Blocks & Pain Management ●
- t Drainage placements
- Hydrodissections .
- Bladder Stimulation �
- Fiducial placements .
- Tumor resections .
- Sinus procedures
- Intranasal procedures
- Transphenoidal procedures
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Sean M. Boyd
(Division Sign Off)
Division of Radiological Heatth
the of in Vitro Diagnostics and Radiological H
I(K) K121498
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.