K131959 · Heine Optotechnik GmbH & Co. KG · HLJ · Mar 7, 2014 · Ophthalmic
Device Facts
Record ID
K131959
Device Name
HEINE MINI 3000(R) OPHTHALMOSCOPE
Applicant
Heine Optotechnik GmbH & Co. KG
Product Code
HLJ · Ophthalmic
Decision Date
Mar 7, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 886.1570
Device Class
Class 2
Indications for Use
The HEINE mini 3000 Ophthalmoscope is a battery powered hand-held device for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
Device Story
HEINE mini 3000 Ophthalmoscope is a battery-powered, hand-held diagnostic instrument; used by medical professionals in clinical settings. Device provides illumination and viewing optics to visualize ocular media and retina. Input: manual operation by clinician; light from 2.5V halogen filament bulb. Output: visual examination of patient eye. Clinician views retina through instrument head; adjustable correction lens (+20D to -20D) allows for refractive compensation. Device aids in ocular health assessment and diagnosis of retinal conditions. Power source: replaceable AA batteries or rechargeable battery via HEINE mini NT charger.
Clinical Evidence
Bench testing only. Device tested for optical radiation safety per ISO 10942 and ISO 15004-2 standards. Results confirm compliance with Group 2 limits for weighted retinal radiance and thermal radiance. No clinical data presented.
Technological Characteristics
Hand-held ophthalmoscope; 2.5V halogen filament bulb; battery-powered (2x AA or rechargeable); adjustable lens range +20D to -20D; monocular viewing; includes red-free filter and multiple aperture sizes (small, large, semicircle, medium with reticle). Complies with IEC 60601-1 for surface temperatures and ISO 15004-2/ISO 10942 for optical radiation safety.
Indications for Use
Indicated for use by medical professionals to examine the ocular media (cornea, aqueous, lens, vitreous) and the retina of the eye.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
K1319554
MAR - 7 2014
# 510(k) Summary of Safety and Effectiveness
In accordance with the requirements of the Safe Medical Device Act, HEINE Optotechnik GmbH & Co. KG herewith submits a Summary of Safety and Effectiveness.
| Submitter Information: | HEINE Optotechnik GmbH & Co. KG<br>Kientalstr. 7<br>82211 Herrsching<br>Germany |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number: | 1000379039 |
| Owner/Operator Number: | 9003020 |
| Official Contact Person: | Mr. Manfred Bartsch-Tittmann<br>Director Regulatory Affairs<br>HEINE Optotechnik GmbH & Co. KG<br>82211 Herrsching, Germany<br>Phone: +49 8152 38 0 |
| US Agent (Contact): | Benoit St. Jean<br>HEINE USA, Itd.<br>10 Innovation Way<br>Dover, NH 03820 USA<br>Phone: +1 603 7427217<br>E-mail: bstjean@heine-na.com |
| Date Prepared: | February, 2nd 2014 |
| Device(s) Identification:<br>Device Trade Name: | HEINE mini 3000® Ophthalmoscope |
| Common Name: | Ophthalmoscope |
| Classification of the device: | |
## Device Classification Name: Product Code: Device Classification No .: Panel: Regulatory Status:
Ophthalmoscope HLJ Part 886.1570 Ophthalmic Devices (86) Class II
{1}------------------------------------------------
#### Device Description:
The HEINE mini 3000® Ophthalmoscope is a battery powered hand-held device to provide illumination and viewing optics in order to examine the media and the retina of a patient's eye. It consists of an instrument head and a battery handle that can be screwed onto the instrument head. The ophthalmoscope can be either operated by replaceable AA batteries or a rechargeable option in combination with the HEINE mini NT® charger.
#### Intended Use:
The HEINE mini 3000® Ophthalmoscope is a battery powered hand-held device for medical professionals, containing illumination and viewing optics intend to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
#### Predicate Device:
| Device Trade Name: | HEINE mini 3000® LED |
|--------------------|---------------------------------|
| Applicant: | HEINE Optotechnik GmbH & Co. KG |
| 510(k) No.: | K123587 |
The HEINE mini 3000® Ophthalmoscope is considered substantial equivalent to the HEINE mini 3000® LED Ophthalmoscope (K123587). There is no significant difference in intended use or technology.
{2}------------------------------------------------
| HEINE mini3000 Ophthalmoscope | HEINE mini3000 LED Ophthalmoscope | Assessment | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| The HEINE mini3000 Ophthalmoscope is a battery powered hand-held device for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye. | The HEINE mini3000 LED Ophthalmoscope is a battery powered hand-held device for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye. | Same | |
| Monocular | Monocular | Same | |
| Used to examine the retina by an examiner in a specific distance to the eye. | Used to examine the retina by an examiner in a specific distance to the eye. | Same | |
| Halogen filament bulb | LED | Different | |
| Emission of 2.5 V halogen bulb | Emission of a white LED | Different3 | |
| 315 lux | 542 lux | Different3 | |
| Red free filter | Red free filter | Same | |
| approx. 45 hours | unlimited | Different | |
| + 20D to -20D | + 20D to -20D | Same | |
| Diopter of used lens in steps:<br>-20, -15, -10, -8, -6, -4, -3, -2, -1,<br>0, 1, 2, 3, 4, 6, 8, 10, 15, 20 | Diopter of used lens in steps:<br>-20, -15, -10, -8, -6, -4, -3, -2, -1,<br>0, 1, 2, 3, 4, 6, 8, 10, 15, 20 | Same | |
| small circle D = 13,8 mm<br>large circle D = 27,3 mm<br>semicircle<br>medium circle with reticle D = 23,1 mm | small circle D = 13,8 mm<br>large circle D = 27,3 mm<br>semicircle<br>medium circle with reticle D = 23,1 mm | Same | |
| Used to examine the retina by an examiner in a specific distance to the eye | Used to examine the retina by an examiner in a specific distance to the eye | Same | |
| Correction lens adjustable with left / right hand | Yes | Same | |
| 2.5 V | 2.5 V | Same | |
| 2 alkaline cells (size LR6/AA) / HEINE mini 2Z rechargeable battery | 2 alkaline cells (size LR6/AA) / HEINE mini 2Z rechargeable battery | Same | |
| none | none | Same | |
| Maximum temperature of parts of the device held by the operator or accessible to the patient | Complies with IEC 60601-1 for temperatures of external surfaces and controls4 | Complies with IEC 60601-1 for temperatures of external surfaces and controls4 | Equivalent |
| Flammability of materials | Low probability. All measures have been taken to use self-extinguishing materials. The system is illuminated using a xhl lamp and all materials used in the vicinity are specially designed to safely operate in high temperature environments. | Low probability. All measures have been taken to use self-extinguishing materials. The system is illuminated using a 3W LED lamp and all materials used in the vicinity are specially designed to safely operate in high temperature environments. | Equivalent |
{3}------------------------------------------------
## Summary of Non-Clinical Performance Testing:
The HEINE mini 3000® Ophthalmoscope is tested according to the "Ophthalmoscope Guidance" in respect to optical radiation hazard with ophthalmoscopes (ISO 10942). Additionally testing in accordance with applicable requirements of ISO 15004-2 "Ophthalmic instruments - Fundamental requirements and test methods" has been performed.
### Conclusion:
HEINE Optotechnik believes that the HEINE mini 3000® Ophthalmoscope is substantially equivalent to the currently legally marketed devices. It does not introduce new indications for use, have the same technological characteristics and do not introduce new potential hazards or safety risks.
# Summary Report
The HEINE mini 3000® Ophthalmoscope is tested according to the "Ophthalmoscope Guidance" in respect to optical radiation hazard with ophthalmoscopes (DIN EN ISO 10942:2007). Additionally testing in accordance with applicable requirements of ISO DIN EN 15004-2:2007 *Ophthalmic instruments - Fundamental requirements and test methods" has been performed.
The bench test results are described in the following reports:
| Standard | Test Identification No. | Document No. |
|-------------|------------------------------------------------------------|--------------|
| ISO 15004-2 | ISO 15004-2_mini3000XHLOphth | 18a |
| | NR. L-56_13 Heine Mini 3000<br>(Risk group classification) | 18b |
| ISO 10942 | ISO 10942_mini3000XHL_Ophth | 18c |
{4}------------------------------------------------
# Bench Test HEINE mini3000® Ophthalmoscope According to EN ISO 15004-2:2007
Based on Test Report NR. L-56_13 Heine Mini 3000 Issued by Seibersdorf Laboratories Date: 11.06.2013 (11th of June 2013)
Table 3 has been used for this report because it represents the worst case of intended use of the device and the results are caiculated according to the methods of DIN EN ISO 15004-2:2007.
The instructions for use contain the maximum time at maximum intensity up to 47 seconds (worst case) as specified in the test report.
| Parameter | Test specification | Wavelength | Limit Group 2 | Test result | Verdict |
|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--------------------------|---------------------------------------------------------------------|---------|
| Weighted corneal<br>and<br>lenticular UV, ES-<br>CL, HS-CL | The corneal radiant exposure shall<br>be evaluated by averaging<br>highest localized radiation power<br>incident upon a circular area at<br>the corneal plane with a diameter of<br>1 mm (7,9 x 10-3 cm²). | 250 - 400 | 3<br>mJ cm-2 | 0,008<br>µW cm-2 | PASS |
| Unweighted<br>corneal and<br>lenticular UV,<br>EUV-CL,<br>HUV-CL | The corneal radiant exposure shall<br>be evaluated by averaging<br>highest localized radiation power<br>incident upon a circular area at<br>the corneal plane with a diameter of<br>1 mm (7,9 x 10-3 cm²). | 360 - 400 | 1 J cm-2<br>(t < 1000 s) | 0,019<br>µW cm-2 | PASS |
| Unweighted<br>corneal and<br>lenticular infrared,<br>EIR-CL | The corneal irradiance shall be<br>evaluated by averaging the highest<br>localized radiation power incident<br>upon a circular area at the<br>corneal plane with a diameter of 1<br>mm (7,9 x 10-3 cm²). | 770 - 2500 | 100<br>mW cm-2 | 1,65<br>µW cm-2 | PASS |
| Unweighted<br>anterior<br>segment visible<br>and<br>infrared, EVIR-AS,<br>HVIR-AS | The unweighted anterior segment<br>irradiance shall be evaluated by<br>averaging the highest localized<br>radiation power incident upon a<br>circular area at the corneal plane<br>with a diameter of 0,5 mm<br>(2,0 x 10-3 cm²). | 380 - 1200 | 20<br>W cm-2 | Not applicable | PASS |
| Weighted retinal<br>radiance, Li, A-R | Measurements of radiance shall be<br>the radiant power detectable<br>through a 7 mm diameter aperture at<br>the cornea and shall be<br>averaged over a right circular cone<br>field-of-view of 0,011 rad.<br>However, if the instrument is<br>designed to be used with an eye that<br>is immobilized, a field-of-view of<br>0,001 75 rad shall be used instead<br>of the 0,011 rad field-of-view. | 305 - 700 | 100<br>J cm-2 sr-1 | 11 mrad: 1216<br>mW cm-2 sr-1<br>1,75 mrad:<br>2124 mW cm-2<br>sr-1 | PASS |
Bench Test Table 3:
{5}------------------------------------------------
| Parameter | Test specification | Wavelength | Limit Group 2 | Test result | Verdict |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|----------------------|--------------------------------|---------|
| Weighted retinal<br>visible<br>and infrared<br>thermal<br>radiance, LVIR-R | In the expression for the limit value,<br>under normal intended use<br>conditions, dr, expressed in<br>millimetres, is the minimum retinal<br>image diameter of the source based<br>on the standard eye<br>(see Annex D for instructions on the<br>way to determine the value of<br>dr). If the calculated value of dr is<br>greater than 1,7 mm, the value<br>1,7 mm shall be used for dr. If the<br>calculated value of dr is less than<br>0,03 mm, the value of 0,03 mm shall<br>be used for dr.<br>Measurements of radiance shall be<br>the radiant power detectable<br>through a 7 mm diameter aperture at<br>the cornea and shall be<br>averaged over a right circular cone<br>field-of-view of 0,001 75 rad. | 380 - 1400 | 29,4¹<br>W cm⁻² sr⁻¹ | 1,75 mrad; 19,9<br>W cm⁻² sr⁻¹ | PASS |
*) = 20 mrad (worst case assumption)
## Summary:
.
The HEINE mini3000® Ophthalmoscope fulfills the requirements according to EN ISO 15004-2:2007, group 2.
The HEINE mini3000® Ophthalmoscope is safe and effective.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2014 ·
HEINE Optotechnik GmbH & Co. KG c/o Ms. Belinda Labourdette Executive Assistant 10 Innovation Way Dover, NH 03820
Re: K131959
Trade/Device Name: HEINE Mini 30000 Ophthalmoscope Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: HLJ Dated: January 28, 2014 Received: February 4, 2014
Dear Ms. Labourdette:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
#### Page 2 - Ms. Belinda Labourdette
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.goy/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
510(k) number (if known): K131959
Device Name: HEINE mini 3000® Ophthalmoscope
Indications For Use:
... . . . . ..................................................................................................................................................................
The HEINE mini 3000 Ophthalmoscope is a battery powered hand-held device for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
AND/OR Over-The-Counter Use Prescription Use × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Marsha L.
Burke Nicholas
-S
Digitally signed by Marsha L. Bu
Nicholas -S
DN: c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
0.9.2342.19200300.100.1.1-130
022, cn=Marsha L. Burke Nichol
Date: 2014.03.04 13:53:46 -05'00
Digitally signed by Marsha L. Burke Nicholas -S DN: c=US, o=U.S. Government, 0.9.2342.19200300.100.1.1=1300014 022, cn=Marsha L. Burke Nicholas -S Date: 2014.03.04 13:53:46 -05'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.