K123587 · Heine Optotechnik GmbH & Co. KG · HLJ · Mar 22, 2013 · Ophthalmic
Device Facts
Record ID
K123587
Device Name
HEINE MINI 3000 LED OPHTHALMOSCOPE
Applicant
Heine Optotechnik GmbH & Co. KG
Product Code
HLJ · Ophthalmic
Decision Date
Mar 22, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 886.1570
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The HEINE mini 3000® LED Ophthalmoscope is a battery powered hand-held device for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
Device Story
The HEINE mini 3000® LED Ophthalmoscope is a handheld, battery-powered monocular device used by medical professionals to examine the eye's media and retina. It consists of an instrument head and a battery handle (powered by AA batteries or a rechargeable option). The device provides illumination via a 3W LED lamp and viewing optics with adjustable diopters (+20D to -20D) and multiple apertures. The examiner holds the device at a specific distance from the patient's eye to visualize the fundus. The device includes a red-free filter to enhance contrast. It is a non-invasive diagnostic tool used in clinical settings to assist in ocular examinations. No software or automated analysis is involved.
Clinical Evidence
Bench testing only. Testing included optical radiation hazard assessment per ISO 10942 and fundamental requirements/test methods per ISO 15004-2. No clinical data was required or provided.
Technological Characteristics
Handheld monocular ophthalmoscope; 3W LED illumination; 2.5V supply voltage (AA batteries or rechargeable); diopter range +20D to -20D; multiple apertures (small, large, semicircle, medium with reticle); red-free filter; self-extinguishing materials; non-sterile; no software.
Indications for Use
Indicated for medical professionals to examine the media (cornea, aqueous, lens, vitreous) and retina of the eye in patients.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the word "HEINE" in all capital letters. The font is bold and sans-serif. The word is horizontally oriented and centered.
HEINE Optotechnik GmbH & Co. KG · Kientalstr. 7 · 82211 Herrsching · Germany Tel. +49(0)8152/38-0 · Fax +49(0)8152/38-202 · E-Mail: info@heine.com · www.heine.com
Image /page/0/Picture/2 description: The image is a logo featuring a black triangle with a white border. Inside the triangle, there is a stylized white design that resembles a lightning bolt intertwined with a snake or cord. Below the design, the word "HEINE" is written in white, with a horizontal line extending from the base of the "E" to the right, underlining the name. The overall design is simple and geometric, with a focus on the contrast between the black background and the white elements.
MAH 2 2 201
ﺎ ﻟﻠ
K123587
## 510(k) Summary of Safety and Effectiveness
In accordance with the requirements of the Safe Medical Device Act, HEINE Optotechnik GmbH & Co. KG herewith submits a Summary of Safety and Effectiveness.
| Submitter Information: | HEINE Optotechnik GmbH & Co. KG |
|------------------------|---------------------------------|
| | Kientalstr. 7 |
| | 82211 Herrsching |
| | Germany |
| Registration Number: | 1000379039 |
| Owner/Operator Number: | 9003020 |
Mr. Jörg Rönnau Official Correspondent: Director Regulatory Affairs HEINE Optotechnik GmbH & Co. KG Phone: +49 8152 38 0
Benoit St. Jean US Agent (Contact): HEINE USA, Itd. 10 Innovation Way Dover, NH 03820 USA Phone: +1 603 7427217 E-mail: Bstjean@heine-na.com
## Date Prepared:
Common Name:
September 3rd, 2012
| Device(s) Identification: | |
|---------------------------|--|
| Device Trade Name: | |
HEINE mini3000® LED Ophthalmoscope Ophthalmoscope
## Classification of the device:
| Device Classification Name: | O |
|-----------------------------|---|
| Product Code: | H |
| Device Classification No.: | P |
| Panel: | O |
| Regulatory Status: | C |
Ophthalmoscope HLJ Part 886.1570 Ophthalmic Devices (86) Class II
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "HEINE" in bold, black letters. The letters are all capitalized and evenly spaced. The font appears to be sans-serif.
Image /page/1/Picture/1 description: The image shows a black triangle with a white logo inside. The logo features a stylized snake wrapped around a rod, a symbol often associated with medicine or healthcare. The word "HEINE" is written in white letters within the curves of the snake. The logo is simple and recognizable, likely representing a company or organization in the medical field.
ので、その他の出来ないという。 この日、 この時間を使いたとき、 出来、 も
## Device Description:
The HEINE mini 3000® LED Ophthalmoscope is a battery powered hand-held device to provide illumination and viewing optics in order to examine the retina of a patient's eve. It consists of an instrument head and a battery handle that can be screwed onto the instrument head. The ophthalmoscope can be either operated by replaceable AA batteries or a rechargeable option in combination with the HEINE mini NT® charger.
### Intended Use:
The HEINE mini 3000® LED Ophthalmoscope is a battery powered hand-held device for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
## Predicate Device:
| Device Trade Name: | Piccolight® E50 |
|--------------------|----------------------------------|
| Applicant: | Kirchner & Wilhelm GmbH & CO. KG |
| 510(k) No.: | K070270 |
The HEINE mini3000® LED Ophthalmoscope is considered substantial equivalent to the Piccolight® E50 Ophthalmoscope (K070270). There is no significant difference in intended use or technology.
{2}------------------------------------------------
HEIME
Image /page/2/Picture/1 description: The image shows a logo with a black triangle on the left side. Inside the triangle, there is a white stylized text that appears to be the word "HEINE". To the right of the triangle, there are some blurred or faded elements that are difficult to discern clearly.
..
| | HEINE mini3000® LED Ophthalmoscope | Piccolight® E50 | Assessment |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Intended Use | The HEINE mini3000® LED Ophthalmoscope is<br>a battery powered hand-held device for medical<br>professionals, containing illumination and<br>viewing optics intended to examine the media<br>(cornea, aqueous, lens, vitreous) and the retina<br>of the eye. | The Piccolight® E50 ophthalmoscope is<br>intended to be used to examine the cornea,<br>aqueous, lens, vitreous and retina of the eye. | same |
| Type | Monocular | Monocular | same |
| Method of operation | Used to examine the retina by an examiner in a<br>specific distance to the eye. | Used to examine the retina by an examiner in<br>a specific distance to the eye. | same |
| Illumination type | LED | Halogen filament bulb | refer to<br>justification 1 |
| Exposure parameters | Emission of a white LED | Emission of 2.5 V halogen bulb | refer to<br>justification 1 |
| Light output | 542 lux | 46 lux | refer to<br>justification 1 |
| Filter | Red free filter | none | refer to<br>justification 2 |
| Service life of<br>Illuminant | 50.000 hours | approx. 15 hours | refer to<br>justification 3 |
| Diopters | + 20D to -20D | + 20D to -20D | same |
| Lens power viewing<br>Optics | Diopter of used lens in steps:<br>-20, -15, -10, -8, -6, -4,-3, -2, -1,<br>0, 1, 2, 3, 4, 6, 8, 10, 15, 20 | Diopter of used lens in steps:<br>-20, -15, -10, -8, -6, -4,-3, -2, -1,<br>0, 1, 2, 3, 4, 6, 8, 10, 15, 20 | same |
| Light apertures | small circle D = 13,8 mm<br>large circle D = 27,3 mm<br>semicircle<br>medium circle with reticle D = 3,1 mm | Large circle D = 24,00 mm | refer to<br>justification 4 |
| Supply voltage | 2.5 V | 2.5 V | same |
| Power sources | 2 alkaline cells (size LR6/AA) /<br>HEINE mini 2Z rechargeable battery | Battery | refer to<br>justification 5 |
| Brightness controls | | | |
| Maximum<br>temperature of parts<br>of the device held by<br>the operator or<br>accessible to the<br>patient | none | Complies with IEC 60601-1 for temperatures<br>of external surfaces and controls3 | same |
| Flammability of<br>materials | Low probability. All measures have been taken<br>to use self-extinguishing materials. The system<br>is illuminated using a 3W LED lamp and all<br>materials used in the vicinity are specially<br>designed to safely operate in high temperature<br>environments. | Fiber-glass reinforced plastic Polyamide 6<br>GF30C | same |
| Cleaning and<br>Disinfection | Cleaning<br>Cleaning or disinfection by spraying or<br>immersion as well as sterilization is not allowed<br>and will damage the instrument! | Cleaning<br>In order to avoid dirt accumulation and dust<br>always store the unit in the packaging. The<br>outside of the unit can be cleaned with a<br>damp, soft and fluff-free fabric. For<br>disinfection, a fabric moistened with alcohol<br>can be used. If necessary, the pane of the<br>observation window can be cleaned with a<br>cotton plug and some alcohol. Please do not<br>press the pane! Never bring the<br>ophthalmoscope into contact with liquid and<br>make sure that no liquid penetrates the<br>housing! | same |
| | Instrument head<br>The housing can be wiped clean with a damped<br>cloth. Glass surfaces can be cleaned with a<br>cotton wool. | | |
| | Battery handle<br>The handle can be cleaned with a damp cloth<br>(e.g. alkaline or pH- neutral detergent). | | |
| | Disinfection<br>The instrument head and handle can be<br>disinfected with CIDEX® OPA by wiping it with<br>a soft cloth moistened with disinfectant for 5<br>minutes. | Since the ophthalmoscope is not desired for<br>operations, simple cleaning and/or<br>disinfection with a surface disinfectant on an<br>alcoholic base is sufficient. Do not use any<br>scrubbing cleaning agents! The unit cannot<br>be sterilized | same |
Page 3 / 6
{3}------------------------------------------------
HEINE
Image /page/3/Picture/1 description: The image shows a black triangle with a white logo inside. The logo appears to be a stylized letter "S" with a line through it. The logo is likely a company logo or brand mark. The image is simple and clean, with a focus on the logo.
Page 4 / 6
{4}------------------------------------------------
HEINE
Image /page/4/Picture/1 description: The image shows a logo with the word "HEINE" written inside of a triangle. The triangle is black, and the word "HEINE" is white. There is a white line that goes through the word "HEINE" and connects to the corners of the triangle. The logo is simple and modern.
. . . ..
| Note 1: Measurements taken 200 mm from output (direct ophthalmoscopes), large circle aperture<br>Note 2: All power sources comply with the relevant standard of IEC 60601-1 and IEC 60601-1-2.<br>Note 3: Chapter 11 of this submission contains the corresponding test report No. E256178-A18-CB-1. Please refer to clause 42 of test<br>report E256178-A18-CB-1 for further details. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Justification 1<br>The HEINE mini3000® LED Ophthalmoscope complies with requirements for group 2 instrument of ISO 15004-2:2007. |
| Justification 2<br>The Red free filter is used to enhance the contrast of the fundus image. Diagnosis is even possible without it, but may take longer. Thus<br>the Red-free filter does not influence the safety and effectiveness compared to the predicate device. |
| Justification 3<br>The minimum lifetime of the used LED in the HEINE mini3000® LED Ophthalmoscope is 50.000 hours. Assuming the use of the LED for a<br>continuous period of 10 hours at 220 working days a year, this would calculate to 22 years of use. Thus does not influence the<br>effectiveness and safety compared to the predicate device. |
| Justification 4<br>Offering several apertures with different diameters gives the examiner more flexibility to adjust the instrument to the examined eye. Thus<br>does not influence the effectiveness and safety compared to the predicate device. |
| Justification 5<br>The predicate device uses the same battery type as the HEINE mini3000® LED Ophthalmoscope. In addition, a rechargeable battery is<br>available for the HEINE mini3000® LED Ophthalmoscope. Thus does not influence the effectiveness and safety compared to the predicate<br>device. |
Page 5 / 6
{5}------------------------------------------------
# HEINE
, 2014/04/20 10:00:00 ID: 00/08/2019 10:00 ID: 00/08/2019 10:00 ID: 00/0
## Summary of Non-Clinical Performance Testing:
The HEINE mini 3000® LED Ophthalmoscope is tested according to the "Ophthalmoscope Guidance" in respect to optical radiation hazard with ophthalmoscopes (ISO 10942). " Additionally testing in accordance with applicable requirements of ISO 15004-2 "Ophthalmic instruments - Fundamental requirements and test methods" has been performed.
## Conclusion:
HEINE Optotechnik believes that the HEINE mini 3000® LED Ophthalmoscope is substantially equivalent to the currently legally marketed devices. It does not introduce new indications for use, have the same technological characteristics and do not introduce new potential hazards or safety risks.
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### March 22, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
HEINE Optotechnik GmbH & Co. KG % Mr. Alexander Schapovalov TUV America, Inc. 1775 Old Highway 8 NW, Suite 104 New Brighton, MN 55112
Re: K123587
Trade Name: HEINE mini3000® LED Ophthalmoscope Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: HLJ Dated: March 5, 2013 Received: March 7, 2013
Dear Mr. Schapovalov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
## Page 2 – Mr. Alexander Schapovalov
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/7/Picture/7 description: The image shows the text "Deborah L. Falls - S FDA". The text is in a simple, sans-serif font and is arranged in a single line. The letters "FDA" are stylized with a geometric pattern.
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510 (k) number: K123587
Device Name:HEINE mini 3000® Ophthalmoscope
Indications For Use:
The HEINE mini 3000® LED Ophthalmoscope is a battery powered hand-held device for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
Prescription Use_ X
AND/OR
Over-The-Counter Use_
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
## P(LEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dexiu-Shi
Digitally signed by Dexiu Shi -S
DN: c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
cn=Dexiu Shi -S,
0.9.2342.19200300.100.1.1=130
0157347/
Date: 2013:03:15 16:45:33
-04'00'
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices
510(k) Number_K123587
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.