INTELLIVUE CL SPO2 POD AND CL NBP POD, PHILIPS INTELLIVUE PATIENT MONITORS MP5, MP5T, MP5SC
Applicant
Philips Medizin Systeme Boeblingen GmbH
Product Code
DRG · Cardiovascular
Decision Date
Nov 26, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
IntelliVue CL SpO2 Pod: The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for acquisition or monitoring of physiological patient parameters SpO2 and pulse rate wirelessly. The intended use of the IntelliVue CL SpO2 Pod when used together with a patient monitor or with a telemetry system transceiver is for monitoring, and recording of, and to generate alarms for, arterial oxygen saturation and pulse rate of adult and pediatric patients. The IntelliVue CL SpO2 Pod is also intended for local attended monitoring of, and to generate alarms for, arterial oxygen saturation and pulse rate of adult and pediatric patients. The IntelliVue CL SpO2 Pod is also intended for acquisition of arterial oxygen saturation and pulse rate data of adult and pediatric patients for a clinical information management system. The IntelliVue CL SpO2 Pod is intended for use by health care professionals inside hospitals. It is not intended for home use. It is not a therapeutic device. Local Attended monitoring refers to situations where clinical staff is close to the patient such that acoustic alarming of the IntelliVue CL SpO2 Pod will be noticed. Example: In-hospital transport of a patient by hospital staff to a procedure room. IntelliVue CL NBP Pod: The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for acquisition or monitoring of physiological patient parameters non-invasive blood pressure and pulse rate wirelessly. The intended use of the IntelliVue CL NBP Pod when used together with a patient monitor or with a telemetry system transceiver is for monitoring and recording of, and to generate alarms for, systolic, diastolic, and mean pressure, and to measure pulse rate of adult and pediatric patients. The IntelliVue CL NBP Pod is also intended for local attended monitoring of, and to generate alarms for, systolic, diastolic, and mean pressure, and to measure pulse rate of adult and pediatric patients. The IntelliVue CL NBP Pod is also intended for acquisition of systolic, diastolic, and mean pressure and pulse rate data of adult and pediatric patients for a clinical information management system. The IntelliVue CL NBP Pod is intended for use by health care professionals inside hospitals. It is not intended for home use. It is not a therapeutic device. Local Attended monitoring refers to situations where clinical staff is close to the patient such that acoustic alarming of the Intelli Vue CL NBP Pod will be noticed. Example: In-hospital transport of a patient by hospital staff to a procedure room. IntelliVue Patient Monitors MP5, MP5T, and MP5SC: The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment. The MP5, MP5SC and MP5T monitors are also intended for use during patient transport inside the hospital environment; only the MPS monitor is for use during patient transport outside of the hospital environment. The MP5, MP5SC and MP5T when used with the TRx4841A/TRx4851A IntelliVue Telemetry System Transceiver or with the IntelliVue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment. The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMIEC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment. The SSC Sepsis Protocol in the ProtocolWatch clinical decision support tool. is intended for use with adult patients only. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (I to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
Device Story
Battery-powered, wrist-worn IntelliVue CL SpO2 and NBP Pods provide cableless monitoring of SpO2, pulse, and non-invasive blood pressure; transmit data via 2.4 GHz SRR to IntelliVue Patient Monitors (MP5, MP5T, MP5SC) or Information Centers. SpO2 Pod uses Fourier Artifact Suppression Technology (FAST) based on light absorption; NBP Pod uses oscillometric method. Modifications enable local attended monitoring (audible alarms on pods during transport) and pulse tone modulation on SpO2 Pod. Monitors feature 8.4" TFT LCDs, CPU, and modular measurement sets; store trend/event data. Healthcare professionals use devices in hospitals for real-time patient status, alarm notification, and clinical decision support. Output includes numerics, waveforms, and alarms, aiding clinical assessment and patient management during transport or bedside care.
Clinical Evidence
Clinical validation performed per ANSI/AAMI SP10 and ISO 81060-2 standards for NBP accuracy using dual-observer mercury sphygmomanometer comparison. Ambulatory monitoring method (ISO 81060-2:2013) also validated. Clinical usability testing confirmed acceptance of local alarming and pulse tone modulation. Bench testing validated measurement ranges and hose durability. All tests met pass/fail criteria.
Technological Characteristics
Battery-powered, wrist-worn pods; 8.4" TFT LCD monitors. Wireless 2.4 GHz ISM band SRR connectivity. SpO2: Fourier Artifact Suppression Technology (FAST). NBP: Oscillometric. Materials: Reused from predicate hoses (biocompatible). Standards: IEC 62304, IEC 60601-1-8, ISO 9919, ANSI/AAMI SP10, ISO 81060-2.
Indications for Use
Indicated for adult, pediatric, and neonatal patients (monitors) or adult/pediatric (pods) in hospital environments requiring physiological monitoring (SpO2, pulse, NBP, ECG, etc.). Contraindications: Not for home use; ST segment monitoring not for neonates/pediatrics; IPI not for patients outside 1-12 years; PPV only for sedated, mechanically ventilated adults free from arrhythmia.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
#### 510(k) Summarv
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
The submitter of this Premarket Notification is: l. Markus Stacha Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen. Germany Fax: ++49 7031 463-2442 Tel: ++49 7031 463-2840, e-mail: markus.stacha@philips.com
This summary was prepared on October 25, 2013.
- 2. The names of the devices are:
Trade names: IntelliVue CL SpO2 Pod and CL NBP Pod Common name: Telemetry Transceiver
Trade names: Philips IntelliVue Patient Monitors MP5, MP5T, MP5SC Common name: Multiparameter Patient Monitors
| Device Panel | Classification | ProCode | Description |
|---------------------------|----------------|---------|---------------------------------------------------------------------|
| Cardiovascular<br>Devices | §870.1025. II | DSI | Detector and Alarm, Arrhythmia |
| | §870.1100, II | DSJ | Alarm. Blood-Pressure |
| | §870.1110, II | DSK | Computer. Blood-Pressure |
| | §870.1120, II | DXQ | Cuff. Blood-Pressure |
| | §870.1130. II | DXN | System. Measurement, Blood-Pressure.<br>Non-Invasive |
| | §870.1435. II | DXG | Computer. Diagnostic, Pre-Programmed.<br>Single-Function |
| | §870.2030, II | DRT | Monitor, Cardiac (incl. Cardiotachometer<br>& Rate Alarm) |
| | §870.2060, II | DRQ | Amplifier and Signal Conditioner,<br>Transducer Signal |
| | §870.2700, II | DQA | Oximeter |
| | §870.2900, I | DSA | Cable, Transducer and Electrode, incl.<br>Patient Connector |
| | §870.2910, II | DRG | Transmitters and Receivers. Physiological<br>Signal, Radiofrequency |
| | - | MSX | System. Network and Communication,<br>Physiological Monitors |
Classification of the IntelliVue CL SpO2 Pod and CL NBP Pod
{1}------------------------------------------------
# PHILIPS
| | | | Classification of the IntelliVue Patient Monitors MP5. MP5T. and MP5SC: |
|--|--|--|-------------------------------------------------------------------------|
|--|--|--|-------------------------------------------------------------------------|
| Device Panel | Classification | ProCode | Description |
|---------------------------|----------------|---------|------------------------------------------------------------------------------|
| | §870.1025. II | DSI | Detector and alarm, arrhythmia |
| | §870.1025. II | MLD | Monitor, ST Segment with Alarm |
| | §870.1025. II | MHX | Monitor, Physiological. Patient (with<br>arrhythmia detection or alarms) |
| | §870.1100, II | DSJ | Alarm, Blood Pressure |
| | §870.1110, II | DSK | Computer, Blood Pressure |
| | §870.1130, II | DXN | System. Measurement, Blood-<br>Pressure, Non-Invasive |
| | §870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function |
| | §870.1915. II | KRB | Probe, Thermodilution |
| | §870.2060. II | DRQ | Amplifier and Signal Conditioner,<br>Transducer Signal |
| | §870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) |
| | §870.2340. II | DPS | Electrocardiograph |
| | §870.2340. II | MLC | Monitor, ST Segment |
| | §870.2350, II | DRW | Electrocardiograph, Lead Switching<br>Adapter |
| | §870.2370, II | KRC | Tester, Electrode, Surface,<br>Electrocardiograph |
| | §870.2700. II | DQA | Oximeter |
| | §870.2600. I | DRJ | System, Signal Isolation |
| | §870.2770. II | DSB | Plethysmograph, Impedance |
| | §870.2800, II | DSH | Recorder, Magnetic tape, Medical |
| | §870.2810, I | DSF | Recorder, Paper Chart |
| | §870.2850. II | DRS | Extravascular Blood Pressure<br>Transducer |
| | §870.2900. I | DSA | Cable, Transducer and Electrode, incl.<br>Patient Connector |
| | | MSX | System, Network and Communication,<br>Physiological Monitors |
| | §870.2910, II | DRG | Transmitters and Receivers,<br>Physiological Signal, Radiofrequency |
| Anesthesiology<br>Devices | §868.1400. II | CCK | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase |
| | §868.1500, II | CBQ | Analyzer, Gas, Enflurane, Gaseous-<br>Phase (Anesthetic Concentration) |
| | §868.1500, II | NHO | Analyzer, Gas, Desflurane, Gaseous-<br>Phase (Anesthetic Concentration) |
| | §868.1500, II | NHP | Analyzer, Gas, Sevoflurane, Gaseous-<br>Phase (Anesthetic Concentration) |
| | §868.1500, II | NHQ | Analyzer, Gas, Isoflurane, Gaseous-<br>Phase (Anesthetic Concentration) |
| Device Panel | Classification | ProCode | Description |
| | §868.1620, II | CBS | Analyzer, Gas, Halothane, Gaseous-<br>Phase (Anesthetic Concentration) |
| | §868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | §868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-<br>Phase |
| | §868.1880, II | BZC | Data calculator Pulmonary-function |
| | §868.2375, II | BZQ | Monitor, Breathing Frequency |
| | §868.2480, II | LKD | Monitor, Carbon Dioxide, Cutaneous |
| | §868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia |
| | §880.2910, II | FLL | Thermometer, Electronic, Clinical |
Page 2 of 9
·
:
{2}------------------------------------------------
3. The modified Philips IntelliVue CL SpO2 Pod and CL NBP Pod. software Rev. B.02, are substantially equivalent to the previously cleared IntelliVue CL SpO2 Pod and CL NBP Pod marketed pursuant to K101600 and K111905. With respect to the local attended monitoring feature the modified IntelliVue CL SpO2 Pod and CL NBP Pod are substantially equivalent to the Philips IntelliVue X2 Multi-Measurement Module software Rev. J.07 marketed pursuant to K122223. K122439. K120366, K113441, K113657. K110622. K102562, K083517, K082633, K072070, and K071426.
The new Philips air hose Mobile CL Air Hose-Bayonet Connector is substantially equivalent to the previously cleared Philips air hoses Mobile CL Extension Air Hose marketed pursuant to K101600 and M1598B marketed pursuant to K052707.
The modified Philips IntelliVue Patient Monitors MP5, MP5T, and MP5SC, software Rev. J.20. are substantially equivalent to the previously cleared IntelliVue Patient Monitors MP5, MP5T, and MP5SC marketed pursuant to K122439. K122223, K120366, K113657. K113441、K110622, K102562, K091395. K083517, K083228, K082633. K081793. K071426, K063725, and K062392.
#### 4. Description of the devices
The IntelliVue CL SpO2 Pod is a small, battery powered, wrist worn pulse oximeter device for cableless monitoring of patients. It contains Philips FAST-SpO2 (Fourier Artifact Suppression Technology) to provide reliable saturation values under various artifact conditions including motion and low perfusion.
It uses a measurement principle that is based on the specific absorption characteristics of oxyhemoglobin and deoxyhemoglobin and the pulsating arteriolar vascular bed at the measurement site. It provides continuous operating mode and intermittent operating mode with configurable measurement intervals. The integrated monochrome LCD display shows measured values, measurement signal quality, battery state, and RF signal strength. It communicates measurement values and other information to an IntelliVue Patient Monitor or an IntelliVue Information Center via a telemetry device. It also acquires SpO2 and pulse rate
Page 3 of 9
{3}------------------------------------------------
data for a clinical information management system. It has three hardkeys for basic operation and navigation. It supports specialized Philips reusable and disposable SpO2 sensors.
The IntelliVue CL NBP Pod is a small. battery powered, non-invasive blood pressure and pulse rate measurement device for cableless monitoring of patients. It uses oscillometric method for measuring NBP. It produces numerics for systolic. diastolic and mean blood pressure values and pulse rate. The integrated monochrome LCD Display shows measured values, battery state, and RF signal strength. It communicates measurement values and other information to an IntelliVue Patient Monitor or an IntelliVue Information Center via a telemetry device. It also acquires systolic, diastolic, and mean pressure and pulse rate data for a clinical information management system. It has three hardkeys for basic operation and navigation. It supports specialized Philips reusable and disposable NBP cuffs and air hoses. Both measuring devices the CL SpO2 Pod and the CL NBP Pod can also be controlled from an assigned IntelliVue Patient Monitor, an IntelliVue Information Center via a telementy device, or from the IntelliVue GuardianSoftware via the hospital IT infrastructure. They communicate with the compatible devices using a wireless Short Range Radio (SRR) interface in the 2.4 GHz ISM frequency band.
The modification to the IntelliVue CL SpO2 Pod and CL NBP Pod enables local physiological alarm indication on these devices for local attended monitoring without a host patient monitor, e.g. during in-hospital transport attended by a care giver. This notifies clinical personnel about possibly threatening situations for a patient.
The modification to the IntelliVue CL SpO2 Pod introduces pulse tone and pulse tone modulation locally at the IntelliVue CL SpO2 Pod to allow clinical personnel to hear deviations of the patients pulse frequency and saturation levels also during local attended monitoring (without a patient monitor).
The IntelliVue CL NBP Pod has additionally been modified in order to enhance accuracy of diastole and Mean Arterial Pressure (MAP) measurements in order to enhance the margin to the limits specified in the recognized consensus standards ANSI/AAMI SP10 and ISO 81060-2.
To supplement the range of measurement accessories, the CL NBP Pod has also been validated for use with a series of additional, legally marketed Philips cuffs. To connect the additional Philips cuffs a new adapter hose, the Mobile CL Air Hose-Bayonet Connector (model No .: 989803187431), has been setup and added to the list of accessories. As the legally marketed Philips Mobile CL Extension Air Hose, the new hose is explicitly intended for use with the IntelliVue CL NBP Pod.
The IntelliVue Patient Monitors MP5, MP5T, and MP5SC consist of 8.4" TFT LCD display unit including built-in central processing unit (CPU) and multiple physiological measurements. All monitor models share the same architecture of CPU units and the same software is executed on each monitor. They are robust, portable, lightweight, compact in size and modular in design, with interfaces to dedicated external measurement devices. The MPS. MP5T and MP5SC monitors have varying measurement sets and network capabilities. All
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are black and appear to be slightly distressed or textured, giving them a worn look. The word is presented in a straightforward, horizontal arrangement.
models can be used with adult, pediatric and neonatal patients in a hospital and transport environment. They store data in trend, event, and calculation databases. Tabular trends (vital signs) can be watched on display and documented on a printer.
The MP5 monitor provides a comprehensive set of basic physiological measurements: NBP, SpO2, and optionally ECG, invasive blood pressure, predictive temperature, standard temperature and CO2. Through networking it provides information integration, documentation and information access.
The MP5T monitor is intended for use together with a telemetry device. It has no ECG measurement of its own but does have NBP and optionally SpO2 and predictive temperature. When the telemetry device is directly connected to the MP5T. the measurements from the MPST are transmitted with those from the telemetry device (ECG and optionally SpO2) to the Information Center.
The MP5SC monitor is customized for use as a multi-patient spot check monitor. It has SpO2 and NBP and optionally predictive temperature or Microstream CO2. A telemetry device can be connected via short range radio (SRR) to the MP5SC to provide an ECG measurement.
When used together with the IntelliVue CL SpO2 Pod and CL NBP Pod. the MP5. MPST, and MPSSC monitors receive via Short Range Radio (SRR) interface measurement values. The measurement data is displayed and recorded on the patient monitors. The patient monitors also receive technical and physiological alarms generated in the CL SpO2 and NBP Pods for their audible and visual indication on the monitors. In addition, the patient monitors can control some functions of the CL SpO2 and NBP Pods via SRR link.
The common software of the IntelliVue Patient Monitors MP5. MP5T. and MP5SC has slightly been modified to allow the operator to start and to end local attended monitoring of the IntelliVue CL SpO2 Pod and CL NBP Pod.
#### 5. Intended Use
IntelliVue CL SpO2 Pod:
The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for acquisition or monitoring of physiological patient parameters SpO2 and pulse rate wirelessly,
The intended use of the IntelliVue CL SpO2 Pod when used together with a patient monitor or with a telemetry system transceiver is for monitoring, and recording of, and to generate alarms for, arterial oxygen saturation and pulse rate of adult and pediatric patients. The IntelliVue CL SpO2 Pod is also intended for local attended monitoring of, and to generate alarms for, arterial oxygen saturation and pulse rate of adult and pediatric patients.
The IntelliVue CL SpO2 Pod is also intended for acquisition of arterial oxygen saturation and pulse rate data of adult and pediatric patients for a clinical information management system.
The IntelliVue CL. SpOs Pod is intended for use by health care professionals inside hospitals. It is not intended for home use. It is not a therapeutic device.
Local Attended monitoring refers to situations where clinical staff is close to the patient such that acoustic alarming of the IntelliVue CL SpO2 Pod will be noticed. Example: In-hospital transport of a patient by hospital staff to a procedure room.
{5}------------------------------------------------
#### IntelliVue CL NBP Pod:
The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for acquisition or monitoring of physiological patient parameters non-invasive blood pressure and pulse rate wirelessly.
The intended use of the IntelliVue CL NBP Pod when used together with a patient monitor or with a telemetry system transceiver is for monitoring and recording of, and to generate alarms for, systolic, diastolic, and mean pressure, and to measure pulse rate of adult and pediatric patients.
The IntelliVue CL NBP Pod is also intended for local attended monitoring of, and to generate alarms for, systolic, diastolic, and mean pressure, and to measure pulse rate of adult and pediatric patients.
The IntelliVue CL NBP Pod is also intended for acquisition of systolic, diastolic, and mean pressure and pulse rate data of adult and pediatric patients for a clinical information management system.
The IntelliVue CL NBP Pod is intended for use by health care professionals inside hospitals. It is not intended for home use. It is not a therapeutic device.
Local Attended monitoring refers to situations where clinical staff is close to the patient such that acoustic alarming of the Intelli Vue CL NBP Pod will be noticed. Example: In-hospital transport of a patient by hospital staff to a procedure room.
#### IntelliVue Patient Monitors MP5. MP5T, and MP5SC:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The MP5, MP5SC and MP5T monitors are also intended for use during patient transport inside the hospital environment; only the MPS monitor is for use during patient transport outside of the hospital environment. The MP5, MP5SC and MP5T when used with the TRx4841A/TRx4851A IntelliVue Telemetry System Transceiver or with the IntelliVue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMIEC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are black and appear to be slightly distressed or textured, giving them a worn look. The word is horizontally oriented and fills most of the frame, making it the central focus of the image. The background is plain and white.
The SSC Sepsis Protocol in the ProtocolWatch clinical decision support tool. is intended for use with adult patients only.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (I to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
#### 6. Technological Characteristics
The modifications to the IntelliVue CL SpO2 Pod and CL NBP Pod and the Patient Monitors MP5, MP5T. and MP5SC do not affect their technological characteristics. Design, materials. energy source, portability, user interface. radio technology, measurement principle, and performance specifications of the devices remain all unchanged as they are.
This new adapter hose Mobile CL Air Hose-Bayonet Connector is based on the design and components from two legally marketed Philips air hoses: Mobile CL Extension Air Hose. model No.: 989803163131 (tubing and plastic male connector) and air hose model No.: M1598B (metal bayonet connector). The new Mobile CL Air Hose-Bayonet Connector has the same dimensions and environmental specifications as the legally marketed Mobile CL Extension Air Hose. Biocompatibility is ensured through the reuse of components from the predicate hoses, which are made of the same materials.
#### 7. Summary of V&V activities
Non-Clinical Performance Tests (bench testing):
The measurement ranges of systolic, diastolic and mean arterial pressure of the modified IntelliVue CL NBP Pod were validated by means of comparison with the legally marketed predicate IntelliVue CL NBP Pod by way of a series of several hundred paired measurements on a blood pressure simulator with blood pressure values covering the specified measurement range.
The comparative testing has demonstrated that the modified IntelliVue CL NBP Pod has produced over the specified measurement range systolic and mean arterial pressure measurement results equivalent to those of the legally marketed predicate IntelliVue CL NBP Pod.
Performance of the new hose Mobile CL Air Hose - Bayonet Connector with respect to durability, cleaning and disinfection was validated according to the same methods and pass/fail criteria as the predicate hose. The performed tests demonstrate that the new hose is resistant against the specified disinfectants, does meet its reliability and performance specifications and its performance is equivalent to that of the predicate hose.
#### - Clinical Performance Tests
The overall efficacy of the modified IntelliVue CL NBP Pod in combination with its specialized legally marketed accessories and with the new Mobile CL Air Hose-Bayonet
{7}------------------------------------------------
Connector together with additional legally marketed Philips cuffs was clinically validated according to the ANSI/AAMI SP10 and ANSI/AAMI/ISO 81060-2 standards.
The modified IntelliVue CL NBP Pod met all safety and efficiency requirements and demonstrated that the accuracy requirements (measured by means of mean error and standard deviations, using the same Arm Sequential Method - Dual Observer - in Comparison to a Mercury Reference Sphygmomanometer) for the number of subjects. number of paired measurements, age distribution, gender distribution, arm circumference distribution, and blood pressure distribution requested in both standards were met.
An additional clinical validation was done following requirements of section 5.2.6 Ambulatory monitoring method of ISO 81060-2:2013 standard. The modified IntelliVue CL NBP Pod met all safety and efficiency requirements and demonstrated that the accuracy requirements (measured by means of mean error and standard deviations, using opposite arm simultaneous method - Dual Observer - in comparison to a Mercury Reference Sphygmomanometer) requested in this section of the standard were met.
- Clinical Usability Testing to evaluate acceptance (ease of use) and clinical usefulness of the implementation of local alarming in the Intelli Vue CL SpO2 and CL NBP Pod and pulse tone modulation in the CL SpO2 Pod. All defined pass criteria have been met.
- Testing according to the following recognized consensus standards: ।
- IEC 62304:2006 (Software life cycle processes) .
- . IEC 60601-1-8:2003 (Alarms)
- . ISO 9919:2005 (Pulse Oximeter)
- ANSI/AAMI SP10:2002/(R)2008 & ANSI/AAMI SP10:2002/A1:2003/(R)2008 & . ANSI/AAMI SP10:2002/A2:2006/(R)2008 and ANSI/AAMI/ISO 81060-2:2009 (Non-invasive sphygmomanometers)
All applicable requirements have been met.
- .' Testing as identified in the Hazard Analysis. All specified pass/fail criteria have been met. The test results have confirmed the effectiveness of implemented design risk mitigation measures.
- Function testing on the modified IntelliVue CL SpO2 Pod, CL NBP Pod and Patient -Monitors MP5. MP5T. MP5SC. All specified criteria have been met. The test results have confirmed that all modified devices have functioned safe, effective and according to the specifications and Instructions for Use in the intended environment of use.
- Regression testing of the related. unchanged software parts of the IntelliVue CL. SpO2 Pod. CL NBP Pod and Patient Monitors MP5, MP5T. All specified criteria have been met. The test results have confirmed that all modified devices have functioned safe. effective and according to the specifications and Instructions for Use in the intended environment of use.
{8}------------------------------------------------
# PHILIPS
#### 8. Conclusion
Verification and validation testing activities were conducted to establish the performance. safety, functionality, usability, effectiveness, and reliability characteristics of the modified devices. V&V testing included clinical and bench performance tests, clinical usability. function, and regression tests. All clinical and non-clinical tests were successfully completed.
The results demonstrate that the modified Philips IntelliVue CL SpO2 Pod. CL NBP Pod and Patient Monitors MP5. MP5SC are as safe, as effective and perform as well as the predicate devices.
The modified devices are substantially equivalent in intended use and fundamental technological characteristics to the appropriate predicate devices. The devices introduce no new questions concerning the safety or efficacy and are, therefore. substantially equivalent to the predicate devices.
{9}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 26, 2013
Philips Medizin Systeme Boeblingen Gmbh Markus Stacha Hewlett-packard Str. 2 Boeblingen, Bw, D-71034 GM
K131913 Re:
> Trade/Device Name: Intellivue Cl SPO2 Pod and CL NBP Pod, Philips Intellivue Patient · Monitors MP5, MP5T, MP5SC Regulation Number: 21 CFR 870.2700 Regulation Name: Telemetry Transceiver, Multiparameter Patient Monitors Regulatory Class: Class II Product Code: DRG, DXJ, MSX, DSI, DSK, DSI, DXQ, DSA, DXG, DRT, DRQ. DQA, DXN Dated: October 25, 2013 Received: October 31, 2013
·Dear Markus Stacha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or anv Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (2) CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Oweកូរខ្លួននូវ
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
#### Device Name:
- Philips IntelliVue CL SpO2 Pod -
- Philips IntelliVue CL NBP Pod -
- Philips IntelliVue Patient Monitors MP5, MP5T, MP5SC -
#### Indications for Use:
#### IntelliVue CL SpO2 Pod:
The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for acquisition or monitoring of physiological patient parameters SpO2 and pulse rate wirelessly.
The intended use of the IntelliVue CL SpO3 Pod when used together with a patient monitor or with a telemetry system transceiver is for monitoring, and recording of, and to generate alarms for, arterial oxygen saturation and pulse rate of adult and pediatric patients.
The IntelliVue CL SpO2 Pod is also intended for local attended monitoring of, and to generate alarms for, arterial oxygen saturation and pulse rate of adult and pediatric patients.
The IntelliVue CL SpO2 Pod is also intended for acquisition of arterial oxygen saturation and pulse rate data of adult and pediatric patients for a clinical information management system.
The Intelli Vue CL SpO2 Pod is intended for use by health care professionals inside hospitals. It is not intended for home use. It is not a therapeutic device.
Local attended monitoring refers to situations where clinical staff is close to the patient such that acoustic alarming of the Intelli Vue CL SpO2 Pod will be noticed. Example: In-hospital transport of a patient by hospital staff to a procedure room.
Continued on next pages
Prescription Use Yes (Part 21 CFR 801 Subpart D)
Over-The-Counter Use No AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by Owen
P. Faris-S
Date: 2013.11.26
14:16:31-05'00'
{12}------------------------------------------------
. K131913 page 2 of 3
Indications for Use (continued):
#### IntelliVue CL NBP Pod:
The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for acquisition or monitoring of physiological patient parameters non-invasive blood pressure and pulse rate wirelessly.
The intended use of the IntelliVue CL NBP Pod when used together with a patient monitor or with a telemetry system transceiver is for monitoring and recording of, and to generate alarms for, systolic, diastolic, and mean pressure, and to measure pulse rate of adult and pediatric patients.
The IntelliVue CL NBP Pod is also intended for local attended monitoring of, and to generate alarms for, systolic, diastolic, and mean pressure, and to measure pulse rate of adult and pediatric patients.
The IntelliVue CL NBP Pod is also intended for acquisition of svstolic, diastolic, and mean pressure and pulse rate data of adult and pediatric patients for a clinical information management system.
The IntelliVue CL NBP Pod is intended for use by health care professionals inside hospitals. It is not intended for home use. It is not a therapeutic device.
Local attended monitoring refers to situations where clinical staff is close to the patient such that acoustic alarming of the Intelli Vue CL NBP Pod will be noticed. Example: In-hospital transport of a patient by hospital staff to a procedure room.
#### IntelliVue Patient Monitors MP5, MP5T, and MP5SC:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The MP5. MP5SC and MP5T monitors are also intended for use during patient transport inside the hospital environment; only the MPS monitor is for use during patient transport outside of the hospital environment. The MP5, MP5SC and MP5T when used with the TRx4841 A/TRx4851 A IntelliVue Telemetry System Transceiver or with the Intelli Vue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
Continued on next page
{13}------------------------------------------------
#### Indications for Use (continued):
The ECG measurement is intended to be used for diagnostic recording of rhvthm and detailed morphology of complex cardiac complexes (according to AAMIEC II).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment.
The SSC Sepsis Protocol in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.
The Integrated Pulmonary index (IPI) is intended for use with adult and pediatric (I to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.