INTELLIVUE CL SP02 POD MODEL 865215, INTELLIVUE CL NBP POD MODEL 865216, INTELLIVUE PATIENT MONITORS MP5, MP2, X2 MODEL
K101600 · Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · DQA · Aug 31, 2010 · Cardiovascular
Device Facts
Record ID
K101600
Device Name
INTELLIVUE CL SP02 POD MODEL 865215, INTELLIVUE CL NBP POD MODEL 865216, INTELLIVUE PATIENT MONITORS MP5, MP2, X2 MODEL
Applicant
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
Product Code
DQA · Cardiovascular
Decision Date
Aug 31, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters SpO2 and pulse rate wirelessly. The intended use of the IntelliVue CL SpO2 Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A, is for monitoring, recording, and alarming arterial oxygen saturation and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device. The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters non-invasive blood pressure and pulse rate wirelessly. The intended use of the IntelliVue CL NBP Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A is for monitoring, recording, and alarming of systolic, diastolic, and mean pressure and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device. The IntelliVue Patient Monitors MP5, MP5T, MP2, X2 are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside of the hospital environment. The MP2, X2, and MP5 monitors are also intended for use during patient transport outside of the hospital environment. They are not intended for home use. The monitors are only for use on one patient at a time. They are not therapeutic devices. The IntelliVue Telemetry System Transceiver TRx4841A is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended to provide ambulatory and bedside monitoring, recording and alarming of ECG and SpO2 parameters of adult and pediatric patients in transport and hospital environments. It is intended to be used by trained healthcare personnel. It is not intended for home use. The device is only for use on one patient at a time. It is not for use with infant or neonatal patients.
Device Story
IntelliVue CL SpO2 and NBP Pods are battery-powered, wrist-worn, cableless telemetry transceivers. SpO2 Pod uses FAST-SpO2 (Fourier Artifact Suppression Technology) for saturation values; NBP Pod uses oscillometric method for systolic, diastolic, mean pressure, and pulse rate. Pods transmit digitized physiological data via short-range radio (SRR) to IntelliVue Patient Monitors (MP5, MP5T, MP2, X2) or Telemetry Transceiver (TRx4841A). TRx4841A relays data via WMTS (1.4 GHz) to IntelliVue Information Center. Healthcare professionals view waveforms, numerics, and alarms at bedside monitors or central station. System facilitates continuous or intermittent monitoring, aiding clinical decision-making by providing real-time physiological status and alarm notifications for hospitalized patients.
Clinical Evidence
Clinical validation included a desaturation study for the CL SpO2 Pod per ISO 9919 and a clinical study for the CL NBP Pod per ANSI/AAMI SP10. Studies confirmed accuracy of SpO2, systolic/diastolic blood pressure, and pulse rate within specified ranges. Bench testing verified measurement averaging and performance under artifact conditions.
Technological Characteristics
Battery-powered, wrist-worn telemetry transceivers. SpO2 sensing via FAST-SpO2 algorithm; NBP via oscillometric method. Connectivity via short-range radio (SRR) to monitors and WMTS (1.4 GHz) to central station. Biocompatible materials per ISO 10993. Monochrome LCD display, 3-button interface. Software-controlled wireless communication.
Indications for Use
Indicated for adult and pediatric patients (and neonates for MP5/MP5T/MP2/X2 monitors) in hospital environments for monitoring SpO2, pulse rate, non-invasive blood pressure, ECG, and respiration. Not for home use. TRx4841A not for neonatal use.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Nihon Kohden Transmitter ZS-940PA series (K043517)
Philips IntelliVue Telemetry System Transceiver TRx4841A (K081793, K041741, K040357)
{0}------------------------------------------------
## KIOL600 PI/7
## AUG 3 1 2010
#### 510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
1. The submitter of this premarket notification is: Markus Stacha Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-2840, Fax: ++49 7031 463-2442 e-mail: markus.stacha@philips.com
This summary was prepared on June 02, 2010.
2. The names of the devices are:
New devices: Trade name: Philips IntelliVue CL SpO2 Pod Common name: Telemetry Transceiver
Trade name: Philips IntelliVue CL NBP Pod Common name: Telemetry Transceiver
Modified devices:
Trade names: Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2
Common names: Multiparameter Patient Monitors
Trade name: Philips IntelliVue Telemetry System Transceiver TRx4841A
Common name: Telemetry Transceiver
Classification names are as follows:
| Device Panel | Classification | ProCode | Description |
|---------------------------|----------------|---------|------------------------------------------------------------------------|
| Cardiovascular<br>Devices | §870.1025, II | DSI | Detector and Alarm, Arrhythmia |
| | §870.1100, II | DSJ | Alarm, Blood-Pressure |
| | §870.1110, II | DSK | Computer, Blood-Pressure |
| | §870.1120, II | DXQ | Cuff, Blood-Pressure |
| | §870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive |
| | §870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function |
| | §870.2030, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) |
| | §870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer Signal |
| | §870.2700, II | DQA | Oximeter |
| | §870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector |
| | §870.2910, II | DRG | Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency |
{1}------------------------------------------------
| Device Panel | Classification | ProCode | Description |
|--------------|----------------|---------|-----------------------------------------------------------------|
| | - | MSX | System, Network and<br>Communication, Physiological<br>Monitors |
- 3. The new Philips IntelliVue CL SpO2 Pod and IntelliVue CL NBP Pod are substantially equivalent to the previously cleared Nihon Kohden Transmitter ZS-940PA series marketed pursuant to K043517, Philips IntelliVue Telemetry System Transceiver TRx4841A marketed pursuant to K081793, K041741, K040357, and Philips Multi-Measurement Server M3000A/M3001A marketed pursuant to K001333, K013427, K020531, K030973, K033715, K062268, K090360.
The new reusable NBP cuffs 989803163171, 989803163191, 989803163211 and Extension Air Hose 989803163131 for the CL NBP Pod are substantially equivalent to the Philips reusable cuffs M1573A, M1574A, M1575A marketed pursuant to K001333, Philips Extension Air Hose M1598B marketed pursuant to K031187.
The new disposable NBP cuffs 989803163181, 989803163201, 989803163221 for the CL NBP Pod are substantially equivalent to the Philips disposable cuffs M4574B, M4575B, M4577B marketed pursuant to K071885, Hewlett-Packard/Philips disposable cuffs M1876A, M1877A, M1878A marketed pursuant to K941811, K925910, К923682, К923343, К922058, К910490, К903771, К903523, К900032, K896030, K882609.
The modified reusable Sp02 sensor Mobile CL RSp02-1A Reusable SpO2 Sensor for IntelliVue CL SpO2 Pod is substantially equivalent to the Philips reusable SpO2 sensor M1192AN marketed pursuant to K033715 and Nellcor disposable Sp02 sensor MAX-A marketed pursuant K081937.
The modified disposable SpO2 sensor Mobile CL DSpO2-1A Single Patient Sp02 Sensor for IntelliVue CL Sp02 Pod is substantially equivalent to the Philips disposable SpO2 sensor M1134A marketed pursuant to K091572 Philips reusable SpO2 sensor M1192AN marketed pursuant to K033715, and Nellcor disposable SpO2 sensor MAX-A marketed pursuant to K081937.
The modified Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2 are substantially equivalent to previously cleared IntelliVue Patient Monitors MP5, MP5, MP2, X2 marketed pursuant to K062392, K063725, K071426, K072070, K081793, K082633, K083228, K083517, K091395, K100939.
The modified Philips IntelliVue Telemetry System Transceiver TRx4841A is substantially equivalent to previously cleared IntelliVue Telemetry System Transceiver TRx4841A marketed pursuant to K081793, K041741, K040357.
#### 4. Description of the devices
The new Philips IntelliVue CL SpO2 Pod is a small, battery powered, wrist worn pulse oximeter device for cableless monitoring of patients. It contains Philips FAST-Sp02 (Fourier Artifact Suppression Technology) to provide reliable saturation
{2}------------------------------------------------
values under various artifact conditions including motion and low perfusion. It provides continuous operating mode and intermittent operating mode with configurable measurement intervals. Integrated monochrome LCD display shows measured values, measurement signal quality, battery state, and RF signal strength. It has three hardkeys for basic operation and navigation. It supports specialized Philips reusable and disposable Sp02 sensors.
The new Philips IntelliVue CL NBP Pod is a small, battery powered, non-invasive blood pressure and pulse rate measurement device for cableless monitoring of patients. It uses oscillometric method for measuring NBP. It produces numerics for systolic, diastolic and mean blood pressure values and pulse rate. Integrated monochrome LCD Display shows measured values, battery state, and RF signal strength. It has three hardkeys for basic operation and navigation. It supports specialized Philips reusable and disposable NBP cuffs.
The Philips IntelliVue MP5, MP5T, MP2, X2 are robust, portable, lightweight, compact in size and modular in design patient monitors with integrated displays and multiple physiological measurements and interfaces to dedicated external measurement devices. With the built-in battery the monitors can also function during transport situations.
With the built-in physiological measurements the monitors MP5, MP2, and X2 provide the following measurement capability: Invasive Blood Pressure, Temperature, CO2, ECG/Respiration, SpO2, and NIBP. The MP5T model provides internal SpO2 and NIBP measurements.
The monitors MP5, MP5T, MP2, X2 can be interfaced to several dedicated external devices to extend the measurement capabilities. The monitors MP5, MP2 and X2 can also be connected to another IntelliVue patient monitor (MP20 to MP90), where they act as multi-measurement modules acquiring measurements for the host monitor. The network interfaces provide the monitors MP5, MP2 and X2 with networking capability via wired or wireless network connections. The monitors MP5, MP5T, MP2, and X2 can also interface via cable or wirelessly via short range radio to the IntelliVue Telemetry Transceiver TRx4841A.
As a result of the current modification, the MP5, MP5T, MP2, X2 patient monitors can be interfaced wirelessly via short range radio to the new CL NBP and SpO2 Pods.
The Philips IntelliVue Telemetry System Transceiver TRx4841A is a small, lightweight, battery powered patient-worn device for monitoring ECG and Sp02. The TRx4841A Transceiver uses transmission through a body (finger tip) method and Philips FAST-Sp02 algorithm to measure SpO2. The TRx4841A Transceiver provides ECG using standard ECG and EASI derived 12-lead monitoring, 6lead monitoring with 2 V-leads and EASI derived 12-lead ECG technology with only 5 leads.
{3}------------------------------------------------
: .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The TRx4841A Transceiver sends patient digitized ECG/SpO2 signals via cable or wirelessly (via SRR) to the IntelliVue Patient Monitors MP5, MP5T, MP2, X2 or wirelessly (via WMTS 1.4 GHz) to the Philips IntelliVue Information Center, which displays ECG waveforms, heart rate, and SpO2 values, detects and analyzes the ECG for cardiac arrhythmias and displays alarm conditions and INOP's. Displays, settings, recordings, and alarms are controlled from the IntelliVue Information Center or the Patient Monitors.
As a result of the current modification, the transceiver TRx4841A can be interfaced wirelessly via short range radio to the new CL NBP and SpO2 Pods. The transceiver TRx4841A sends patient digitalized SpO2/NBP signals from the connected CL NBP and SpO2 Pods wirelessly (via WMTS 1.4 GHz) to the Philips IntelliVue Information Center, which displays the physiological measurement values and indicates alarms.
#### 5. Intended Use/ Indications for Use
The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters SpO2 and pulse rate wirelessly. The intended use of the IntelliVue CL SpO2 Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A, is for monitoring, recording, and alarming arterial oxygen saturation and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
This is the same intended use as that of the predicate Nihon Kohden Transmitter ZS-940PA series, except for the compatible monitoring devices and number of physiological parameters: whereas the Transmitter ZS-940PA provides SpO2, pulse rate, noninvasive blood pressure, ECG, and respiration, the CL SpO2 Pod only provides SpO2 and pulse rate. The new CL SpO2 Pod is compatible with Philips monitoring devices whereas the Nihon Kohden Transmitter ZS-940PA is compatible with Nihon Kohden monitoring devices.
The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters non-invasive blood pressure and pulse rate wirelessly.
The intended use of the IntelliVue CL NBP Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A is for monitoring, recording, and alarming of systolic, diastolic, and mean pressure and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
This is the same intended use as that of the predicate Nihon Kohden Transmitter ZS-940PA series, except for the compatible monitoring devices and number of physiological parameters:
{4}------------------------------------------------
whereas the Transmitter ZS-940PA provides non-invasive blood pressure, pulse rate, SpO2, ECG, and respiration, the CL NBP Pod only provides non-invasive blood pressure and pulse rate. The new CL SpO2 Pod is compatible with Philips monitoring devices whereas the Nihon Kohden Transmitter ZS-940PA is compatible with Nihon Kohden monitoring devices.
The modified Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2 have the same intended use as the legally marketed monitors MP5, MP5T, MP2, X2.
The IntelliVue Patient Monitors MP5, MP5T, MP2, X2 are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside of the hospital environment. The MP2, X2, and MP5 monitors are also intended for use during patient transport outside of the hospital environment. They are not intended for home use. The monitors are only for use on one patient at a time. They are not therapeutic devices.
The modified IntelliVue Telemetry System Transceiver TRx4841A has the same intended use as the legally marketed Transceiver TRx4841A.
The IntelliVue Telemetry System Transceiver TRx4841A is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended to provide ambulatory and bedside monitoring, recording and alarming of ECG and SpO2 parameters of adult and pediatric patients in transport and hospital environments. It is intended to be used by trained healthcare personnel. It is not intended for home use. The device is only for use on one patient at a time. It is not for use with infant or neonatal patients.
#### -
#### 6.Technological Characteristics
The new IntelliVue CL SpO2 Pod and CL NBP Pod and their medical accessories have the same or substantial equivalent technological characteristics, such as design, material, energy source, portability, user interface, radio technology, measurement principle, as the legally marketed predicate devices The new IntelliVue CL SpO2 Pod and CL NBP Pod and their medical accessories use partly different patient skin contacting materials, than the predicates. The only relevant characteristic for the intended applications is the biocompatibility of these materials. The biocompatibility of the different materials, hence the equivalence to the predicate devices, has been proved by successful testing to the applicable biocompatibility standards of the IEC 10993-x standard family.
The modification of the IntelliVue Patient Monitors MP5, MP5T, MP2, X2 and the IntelliVue Telemetry System Transceiver TRx4841A is solely limited to the software changes necessary due to the
{5}------------------------------------------------
wireless communication with the introduced IntelliVue CL SpO2 Pod and CL NBP Pod. This software modification does not affect the technological characteristics of the modified patient monitors MP5, MP5T, MP2, X2 and the transceiver TRx4841A.
### 7.Brief discussion of non-clinical performance tests
The determination of the substantial equivalence of the performance of the new IntelliVue CL SpO2 Pod and CL NBP Pod and their measurement accessories was based, in addition to the clinical performance validation, on bench performance testing. For the CL Sp02 Pod the accuracy of the pulse rate in the specified measurement range has been validated. In addition, the verification of averaging of the calculated numerics (Sp02, pulse rate, perfusion index) has been verified. The bench performance testing has demonstrated that the new CL SpO2 Pod has performed within specifications, which have been substantial equivalent to those of the predicate device.
For the CL NBP Pod the accuracy of systolic/ diastolic and mean pressure, such as accuracy of pulse rate in the specified measurement range has been validated. The comparative testing has demonstrated that the new CL NBP Pod and its cuffs have produced the same results as the appropriate predicate devices.
#### 8.Brief discussion of clinical performance tests
The determination of substantial equivalence of the performance of the new IntelliVue CL SpO2 Pod and CL NBP Pod and their measurement accessories was based, in addition to the bench performance testing, on clinical performance validation. For the CL SpO2 Pod and its sensors the clinical desaturation study to determine the Sp02 accuracy in the specified measurement range, according to the recognized consensus standard ISO 9919, has been performed. The clinical desaturation study has demonstrated that the new CL SpO2 Pod and its sensors have worked within the specified accuracy and have met the performance requirements of the ISO 9919 standard and the recommendations of the draft guidance document 'Pulse Oximeters - Premarket Notification Submissions [510(k)s]' .
For the CL NBP Pod and its cuffs the clinical study to determine the systolic and diastolic accuracy of blood pressure in the specified measurement range, using the auscultatory reference method according to the recognized consensus standard ANSI/AAMI SP10, has been performed. The clinical study has demonstrated that the new CL NBP Pod and its cuffs have worked within the specified accuracy and have met the performance requirements of the ANSI/AAMI SP10 standard.
#### 9.Conclusion
ﻟﺔ ، ﺳﻴﺮ
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Verification and validation testing activities were conducted to establish the performance, safety, functionality, and reliability characteristics of the new and modified devices.
Page 19
{6}------------------------------------------------
Declarations of Conformity to the above listed recognized consensus standards are provided in the Annexes A.01 to A.06.
9.2 Summary Reports
The new IntelliVue CL SpO2 Pod and CL NBP Pod, the modified IntelliVue Patient Monitors MP5, MP5T, MP2, X2 and the modified IntelliVue Telemetry System Transceiver TRx4841A have been subject to a series of verification and validation (V&V) testing, as required by the applicable standards, relevant Sp02 and NIBP guidance documents, hazard analysis, and additional R&D V&V requirements, in order to , demonstrate their safe and effective use in accordance with specifications under simulated and clinical use conditions.
The V&V activities have confirmed safe, reliable, secure, and effective function within specifications of the new IntelliVue CL SpO2 Pod and CL NBP Pod, the modified Monitors MP5, MP5T, MP2, X2 and Transceiver TRx4841A, in the intended environment of use.
In addition to the testing required in the applicable recognized consensus standards and particular guidance documents, the following testing has been performed:
- · Environmental testing
- . Mechanical testing on the new IntelliVue CL SpO2 Pod and CL NBP Pod.
- Temperature and humidity testing on the new IntelliVue CL SpO2 Pod and CL NBP Pod.
All specified test requirements have been met. Please refer to Sub-Sec 9.2.1 'Environmental Testing' for details.
- . Radio Frequency (RF) wireless communication testing on the new CL SpO2 Pod and CL NBP Pod with the modified monitors MP5/MP5T, MP2/X2 and the modified transceiver TRx4841A, based on the FDA guidance document 'Radio-Frequency Wireless Technology in Medical Devices' .
All specified test requirements have been met.
The following specific items have been subject to the RF wireless communication tests:
- " Performance of wireless functions For details please refer to Sub-Sec. 9.2.2 'Radio Frequency (RF) Testing' and Table 16.9.1 'Summary of Tests on the IntelliVue CL SpO2 Pod/ NBP Pod, as required
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo consists of three parallel lines on the left side and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right side. The text is in a bold, sans-serif font.
Image /page/7/Picture/1 description: The image contains the text 'Public Health Service'. The text is displayed in a simple, sans-serif font. The words are arranged on a single line, with 'Public' and 'Health' appearing to be slightly larger than 'Service'.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Philips Medizin Systeme Boeblingen GmbH c/o Mr. Markus Stacha Sr. Regulatory Affairs Engineer Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany
AUG 31 2010
Re: K101600
Device Name: Philips IntelliVue CL SpO2 Pod and IntelliVue CL NBP Pod Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II (Two) Product Codes: DQA, DSI, DSJ, DSK, DXQ, DXN, DXG, DRT, DRQ, DSA, DRG, MSX, MHX Dated: June 2, 2010 Received: June 8, 2010
Dear Mr. Stacha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Markus Stacha
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M. A. Killeen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): KIO 1600
#### Device Names:
- Philips IntelliVue CL SpO2 Pod and IntelliVue CL NBP Pod ・
- Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2 -
- -Philips IntelliVue Telemetry System Transceiver TRx4841A
#### Indications for Use:
#### IntelliVue CL Sp02 Pod:
The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters SpO2 and pulse rate wirelessly. The intended use of the IntelliVue CL SpO2 Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A, is for monitoring, recording, and alarming arterial oxygen saturation and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
#### IntelliVue CL NBP Pod:
。 . ·····
> The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters non-invasive blood pressure and pulse rate wirelessly.
The intended use of the IntelliVue CL NBP Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A is for monitoring, recording, and alarming of systolic, diastolic, and mean pressure and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
continued on next page
| Prescription<br>(Part 21 CFR 801 Subpart D) | Yes |
|---------------------------------------------|--------|
| | AND/OR |
| Over-The-Counter<br>(21 CFR 807 Subpart C) | No |
| Use | Use |
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE
IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Willieme
f Cardiovascular Devices
510(k) Number
Page 11
{10}------------------------------------------------
#### Indications for Use (continued) :
#### Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2:
The IntelliVue Patient Momitors MP5, MP5T, MP2, X2 are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside of the hospital environment. The MP2, X2, and MP5 monitors are also intended for use during patient transport outside of the hospital environment.
They are not intended for home use. The monitors are only for use on one patient at a time. They are not therapeutic devices.
#### Philips IntelliVue Telemetry System Transceiver TRx4841A:
The IntelliVue Telemetry System Transceiver TRx4841A is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
Intended to provide ambulatory and bedside monitoring, recording and alarming of ECG and SpO2 parameters of adult and pediatric patients in transport and hospital environments.
It is intended to be used by trained healthcare personnel. It is not intended for home use. The device is only for use on one patient at a time. It is not for use with infant or neonatal patients.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.