K131859 · Beijing Honkon Technologies Co., Ltd. · ONF · Oct 18, 2013 · General, Plastic Surgery
Device Facts
Record ID
K131859
Device Name
MULTIFUNCTIONAL SERIES
Applicant
Beijing Honkon Technologies Co., Ltd.
Product Code
ONF · General, Plastic Surgery
Decision Date
Oct 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Multifunctional Series device (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in hair removal, moderate inflammatory acne vulgaris, and benign cutaneous vascular lesions.
Device Story
The Multifunctional Series (models Aeslight-S3D, HONKON-S3C, HONKON-M40e+, HONKON-M80e+) are Intense Pulsed Light (IPL) systems. The device delivers pulsed light energy (530nm-1200nm) via a sapphire handpiece to target chromophores based on the principle of selective thermolysis. By limiting pulse width below the thermal relaxation time of skin, the device causes thermal damage to targets while sparing surrounding tissue. The system is intended for use by clinicians in surgical, aesthetic, and cosmetic settings. Operators select specific filter/wavelength settings based on the condition (hair removal, acne, or vascular lesions) and skin type. The output energy and pulse parameters are controlled by the user to achieve therapeutic effects. The device benefits patients by providing non-invasive treatment for dermatologic conditions.
Clinical Evidence
Bench testing only. Performance testing verified output energy, spot size, and compliance with electrical safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2) standards. No clinical data provided.
Technological Characteristics
IPL system; sapphire crystal delivery; wavelength ranges 530-1200nm; energy range 20-50J/cm²; pulse duration 1-25ms; pulse delay 0.1-40ms; max power 1250-2000W; spot sizes 8x40mm and 15x60mm. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for hair removal, moderate inflammatory acne vulgaris, and benign cutaneous vascular lesions in patients with Fitzpatrick skin types I-V.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
S3 510(k) Summary
#### Project #:M0382013B
# Section #3 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: ____
- 1. Date of Prepared: 06/07/2013
**OCT 18 2013**
2. Sponsor
Beijing Honkon Technologies Co., Ltd. No.3 Building, No.11 Yard, Kangding Street, BDA, 100176,Beijing, P.R.China
Establishment Registration Number: Not yet registered
Contact Person: Li Zhao Position: Management Representative Tel: +86 10 56370050 ext. 508 Fax: +86 10 56370067 ext. 610 Email: small@honkonlaser.com
- 3. Submission Correspondent
Diana Hong & Tarzan. Wang Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
{1}------------------------------------------------
S3 510(k) Summary
- 4. Proposed Device Identification
Proposed Device Name: Multifunctional Series Proposed Device Common Name: Intense Pulse Light (IPL)
Regulatory Information Classification Name: powered laser surgical instrument Classification: II; Product Code: ONF; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;
Intended Use Statement:
The Multifunctional Series device (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in hair removal, moderate inflammatory acne vulgaris, and benign cutaneous vascular lesions.
| Conditions | Filter setting<br>and<br>wavelength<br>scope | Skin type | | | | |
|---------------------|----------------------------------------------|-----------|----------|----------|----------|----------|
| | | I | II | III | IV | V |
| Hair<br>Removal | | 610-1200 | 610-1200 | 610-1200 | 610-1200 | 610-1200 |
| Acne | | 530-1200 | 530-1200 | 530-1200 | 530-1200 | 530-1200 |
| Vascular<br>Lesions | | 585-1200 | 585-1200 | 585-1200 | 585-1200 | 585-1200 |
- 5. Predicate Device Identification
510(k) Number: K122995 Predicate Device Name: Intense Pulsed Light (IPL) Systems Manufacturer: Beijing KES Biology Technology Co., Ltd.
510(k) Number: K093627 Predicate Device Name: IPULSELIGHT IPL SYSTEM Manufacturer: Shanghai APOLO Medical Technology Co., Ltd
- 6. Device Description
{2}------------------------------------------------
The Multifunctional Series device are intense pulsed light system which delivers intense pulsed light at a wavelength ranging from 530nm-1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores" thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin.
IPL. Systems are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penctration without using excessive energy levels and enables targeting of specific chromophores.
Based on this, The Multifunctional Series device (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic applications in hair removal, acne, and blood vessel lesions.
The proposed device includes four models as Aeslight-S3D, HONKON-S3C. HONKON-M40e+ and HONKON-M80e+.
#### 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- a) IEC 60601-1. Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988: Amendment i, 1991-11, Amendment 2, 1995.
- b) IEC 60601-1-2:2007. Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.
- c) Performance Test (output energy, spot size)
#### 8. Technological Characteristics Comparison
The proposed device has the same technological characteristics with the predicate device, such as light medium, wavelength, control method and intended use.
- 9. Substantially Equivalent Conclusion Based on the comparison and analysis as following, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
{3}------------------------------------------------
・・
| | | מערכת המועד היה היה להתנועד האור | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|--|
| | Proposed Device | Predicate Device | Predicate Device | | |
| ITEM | Multifunctional Series | Intense Pulsed Light (IPL) | IPULSELIGHT IPL SYSTEM | Remark | |
| | | Systems(K 122995) | (K093627) | | |
| Product | | | | | |
| | ONF | ONF | ONF | SE | |
| Code | | | | | |
| Regulation | 21 CFR 878,4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | ટાને | |
| No | | | | | |
| Class | ll | ll | 11 | ટાદ | |
| Intended<br>Use | The Multifunctional<br>Series<br>(inclusive<br>device<br>of<br>the<br>handpiece used to deliver<br>pulsed-light<br>energy)<br>are<br>indicated for use in surgical,<br>aesthetic<br>and<br>cosmetic<br>applications in hair removal.<br>moderate inflammatory acne<br>vulgaris.<br>and<br>benign<br>cutaneous vascular lesions. | آل المحافظ من المحافظ المحافظ المحافظ المحافظ المحافظ المحافظ المحافظ المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق<br>Intense Pulsed Light (IPL) S<br>(inclusive<br>ystems<br>of<br>the<br>handpiece used to deliver<br>energy )<br>pulsed-light<br>are<br>indicated for use in surgical,<br>aestheric<br>and<br>cosmetic<br>applications<br>in<br>permanent<br>hair reduction, treatment of<br>pigmented lesions, moderate<br>inflammatory acne vulgaris,<br>ephelides<br>(freckles),<br>and<br>vascular lesions. | IPulseLight IPL System (HS 300C,<br>HS 650) are identical with regard to<br>indications for use including<br>recommended filters to be used with<br>Fitzpatrick skin type, Both models<br>are intended for medical use in the<br>treatment of the following<br>dermatologic conditions:<br>Permanent hair reduction- long-term<br>stable reduction in number of hairs<br>re-growing after a treatment regiment<br>Treatment of:<br>- Moderate inflammatory acne<br>vulgaris<br>- Benign pigmented epidermal<br>lesions including dy schromia,<br>hyperpigmentation, melasma,<br>ephelides (freckles).<br>- Cutaneous lesions including scars<br>- Benign cutaneous vascular lesions<br>including port wine stains,<br>hemangiomas, facial, truncal and leg<br>telangiectasias, erythema of rosacea,<br>leg veins, spider anglomas and<br>venous malformations.<br>The integrated thermal cooling is<br>indicated for use in cooling the<br>epidermis at the treatment site prior<br>to, during, and after light treatment<br>in general aesthetic dermatologic and<br>plastic surgery procedures<br>· Reduce pain during light<br>treatment (via partial anesthesia from<br>cooling)<br>- Reduce discomfort during and/or<br>associated with light treatment<br>- Minimize thermal injury,<br>including thermal necrosis, to<br>non-larget skin and skin structures<br>during and/or associated with light<br>treatment, thus reducing possible<br>complications such as scabbing.<br>scarring, hyper - and/or hypo<br>pigmentation<br>- Allow the use of higher light or<br>laser fluencies for light treatments<br>(such as for hair removal and the<br>treatment of vascular or pigmented<br>lesions)<br>- Reduce potential side effects of<br>light treatments (such as for hair<br>removal and the treatment of<br>vascular or pigmented lesions) | ਟ ਦ | |
| ITEM | Proposed Device<br>Multifunctional Series<br>(Aeslight-S3D) | Predicate Device<br>Intense Pulsed Light (IPL)<br>Systems(K122995) | Predicate Device<br>IPULSELIGHT IPL SYSTEM<br>(K093627) | Remark | |
| Light source | Intense pulsed light | Intense pulsed light | Intense pulsed light | SE | |
| Wavelength | 610-1200nm,<br>585-1200nm,<br>530-1200nm | 430-1200nm, 530-1200nm,<br>640-1200nm,<br>Optional: 480-1200nm.<br>560-1200nm, 590-1200nm,<br>690-1200nm, 750 -1200nm | 420nm-1200, 510-1200nm,<br>560-1200nm, 610-1200nm,<br>640-1200nm (Standard):<br>480nm-1200, 585-1200nm,<br>690-1200nm, 755-1200nm<br>(Options) | Analysis<br>1 | |
| Deliver system | Sapphire | Sapphire | Sapphire | SE | |
| Energy Range | 20-50J/cm² | 10-60J/cm² | 10-60J/cm2 | Analysis<br>2 | |
| Pulse Delay | 0.1-40ms | 5 - 50ms | 5 - 50ms | Analysis<br>3 | |
| Pulse Duration | 1-25 ms | 1-20 ms | 2-20 ms | Analysis<br>4 | |
| Max. Power | 2000 W | 2000 W | 1200 W | SE | |
| Spot size | 8mm*40mm, 15mm*60mm | MED-210: 15mmX50mm<br>(optional: 12mmX33mm,<br>15mmX35mm )<br>MED-230:<br>A: 12mm X33mm;<br>B: 15mmX50mm (optional:<br>15mmX35mm) | 12x35mm, 15x50mm (Standard);<br>12x12mm, 12x25mm, 12x50mm,<br>15x35mm, 15x45mm (Options) | Analysis<br>5 | |
| Setting for Specified Indication for use:Hair Removal | | | | | |
| Wavelength<br>(nm) | 610-1200 | 640-1200/690-1200/750-1200 | 610-1200/640-1200/690-1200 | | |
| Energy Range<br>(J/cm2) | 20-50 | 10-44 | 10-46 | | |
| Pulse Duration<br>(ms) | 3-15 ms | 3-14 | 4-15 | Analysis<br>6 | |
| Pulse Delay<br>(ms) | 15-35 | 16-32 | 15-30 | | |
| Spot size | 8mm X 40mm; 15mm X 60mm | 12mm X 33mm; 15mm X 50mm;<br>15mm X 35mm | 12mm X 25mm; 12mm X 35mm;<br>12mm X 50mm; 15mm X 45mm | | |
| Setting for Specified Indication for use: Acne | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.