K131708 · Bradford Wright Companies, Inc. · MMK · Mar 10, 2014 · General Hospital
Device Facts
Record ID
K131708
Device Name
BWCSH-1 SHARPS CONTAINER
Applicant
Bradford Wright Companies, Inc.
Product Code
MMK · General Hospital
Decision Date
Mar 10, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The BWCSH-1 sharps container has a one-gallon fill line capacity that is intended to be used for the safe disposal of hazardous sharps. The BWCSH-1 sharps container is metallic canister fitted with thermal plastic top and has a orange-red biohazard label adhered to the containers with fill line printed on the front. The BWCSH-1 device is for single use, disposable and is not a secondary container. The BWCSH-1 sharps container dimensions are 10.5 inches in width at the top, 4.75 inches deep and 8.25 inches in height. BWCSH-1 sharps container weighs 1.075 pounds when empty.
Device Story
BWCSH-1 is a single-use, disposable sharps container; 1-gallon capacity; used in healthcare facilities for safe disposal of hazardous sharps. Device consists of a tinplate steel canister base with a molded polypropylene/polystyrene rotating top and lid. Features include a locking, tamper-resistant lid; translucent components for visual content level monitoring; and a fill line to prevent overfilling. Operation involves one-handed insertion of sharps through the rotating top; lid locks permanently once full. Device provides puncture, impact, and shatter resistance; maintains stability in upright position. Benefits include reduced risk of sharps-related injury to healthcare personnel through secure containment and visual fill-level monitoring.
Clinical Evidence
Bench testing only. Testing included puncture resistance (ASTM F2132-01), leak-proof testing (29 CFR 1910.1030), impact/shatter resistance (ISO 23907:2012), and stability (29 CFR 1910.1030). All tests passed, demonstrating performance equivalent to predicate devices.
Technological Characteristics
Materials: Tinplate steel (canister), molded polypropylene (chute), molded polystyrene (rim/lid). Dimensions: 10.5" W x 4.75" D x 8.25" H. Features: Rotating top, locking tamper-resistant lid, translucent components for visual inspection. Standards: ASTM F2132-01 (puncture resistance), ISO 23907:2012 (sharps injury protection), 29 CFR 1910.1030 (leak/stability). Non-sterile, single-use, manual operation.
Indications for Use
Indicated for healthcare professionals for the safe disposal of hazardous sharps in healthcare facilities.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Demolizer #47 1 Gal. Point of Generation Sharps Container (K982781)
{0}------------------------------------------------
K 131708 Traditional 510(k) Submission
# 510(k) SUMMARY
# MAR(0 2014
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92
Preparation Date: February 24, 2014
# ADMINISTRATIVE INFORMATION
Submitted by: Bradford Wright Companies, Inc. Organization Number: 338046 . PIN: MD6068894-956733 SBD135362 4092 Orchard Way Lake Oswego, OR 97035-1871 Phone/Fax: (503) 635-6908 Email: info@bradfordwright.com
Contact Person: Mike DeMello President Email: miked@bradfordwright.com
## DEVICE IDENTIFICATION
| 510(k) Number: | K131708 |
|-----------------------|----------------------------------------------|
| Device Name: | BWCSH-1 |
| Common Name: | Sharps Container |
| Classification Name: | MMK |
| | Accessory to hypodermic single lumen needles |
| CFR Reference: | 21CFR 880.5570- Class II |
| Classification Panel: | General Hospital |
This submission is the first and only existing FDA filing by this manufacturer. We have never marketed the device in the U.S.
{1}------------------------------------------------
## Indications for Use:
The BWCSH-1 sharps container has a one-gallon fill line capacity that is intended to be used for the safe disposal of hazardous sharps. The BWCSH-1 sharps container is metallic canister fitted with thermal plastic top and has a orange-red biohazard label adhered to the containers with fill line printed on the front. The BWCSH-1 device is for single use, disposable and is not a secondary container. The BWCSH-1 sharps container dimensions are 10.5 inches in width at the top, 4.75 inches deep and 8.25 inches in height. BWCSH-1 sharps container weighs 1.075 pounds when empty.
#### Predicate Device:
| 510(k) Number: | K982781 |
|-----------------------|-----------------------------------------------------------|
| Trade Name of Device: | Demolizer #47 1 Gal. Point of Generation Sharps Container |
| Common Name: | Sharps Container |
| Classification Name: | MMK |
| | Accessory to hypodermic single lumen needles |
| | CFR Reference: |
| Classification Panel: | General Hospital |
Legally Marketed Equivalent Devices:
| Product | Company | 510(k) # |
|----------------------------------------------------------------------|-------------------------------------|----------|
| Demolizer<br>#47 1 Gallon Point of<br>Generation Sharps<br>Container | Thermal Waste<br>Technologies, Inc. | K982781 |
| B-D GuardianTM<br>Nestable Sharps<br>Collectors | Becton Dickinson & Co. | K 943575 |
{2}------------------------------------------------
# DEVICE DESCRIPTIVE INFORMATION
## Product Description:
The BWCSH-1 sharps container has a one-gallon fill line capacity that is intended to be used for the safe disposal of hazardous sharps. The BWCSH-1 sharps container is stable, puncture resistant, impact and shatter resistant, and leak-proof on the sides and bottom. The BWCSH-1 container is fitted with a rotating top, allowing safe one-handed disposal of used sharps. The container has a locking tamper resistant plastic lid that once snapped in place, cannot be manually removed. Labels are orange-red with black text and a black biohazard symbol printed on a white background. Labels are adhered to the containers at the time of manufacture with the fill line printed. The BWCSH-1 device is for single use, disposable and is not a secondary container.
The BWCSH-1 sharps container three main components are nested together to reduce storage and shipping requirements. The rotating top is made with uncolored translucent material. The translucent nature of the product allows for a visual determination of content level. The bottom of the rotating top's internal chute matches the fill line to physically prevent over filling.
The BWCSH-1 device underwent and passed performance testing that consisted of confirming its key safety and reliability aspects. The following four types of performance tests were performed:
- . Puncture Resistance Testing - According to ASTM F2132-01(2008);
- Leak-Proof Testing According to 29 CFR 1910.1030(d)(2)(viii)(c); .
- . Impact and Shatter Resistance -; According to ISO 23907:2012 and
- Stability According to 29 CFR 1910.1030(d)(4)(iii)(A)(2)(ii). .
| Product<br>Description | Access<br>opening<br>size | Access<br>opening and<br>closure | Overall<br>Size<br>L x W x H<br>(inches) | Weight<br>(pounds) | Capacity at<br>full line |
|------------------------|---------------------------|----------------------------------|------------------------------------------|--------------------|--------------------------|
|------------------------|---------------------------|----------------------------------|------------------------------------------|--------------------|--------------------------|
## General Specification Table
{3}------------------------------------------------
Bradford Wright Companies, Inc.
| Component | Material | Depth<br>(inches) | Width<br>(inches) | Height<br>(inches) |
|---------------------|----------------------|-------------------|-------------------|--------------------|
| Canister at Top | Tinplate Steel | 4" | 10" | 8" |
| Canister Bottom | Tinplate Steel | 3" | 9" | N.A. |
| Sharps Chute closed | Molded Polypropylene | 4 ¾" | 10 ½" | 2 ¼" |
| Sharps Chute alone | Molded Polypropylene | 3 ¾" | 9 ½" | 2" |
| Top Rim | Molded Polystyrene | 4 ¾" | 10 ½" | ¼" |
| Lid Plug | Molded Polystyrene | 3 ¾" | 9 ¾" | ¾" |
| Sharps Container | Metal & Plastic | 4 ¾" | 10 ½" | 8 ¼" |
Table showing sharps container component dimensions:
BWCSH-1 Sharps Container
Revision 4 Page 7 of 40
{4}------------------------------------------------
# PREDICATE DEVICES:
(K982781) Demolizer # 47 One-Gallon Point of Generation Sharps Container by Thermal Waste Technologies, Inc. and (K943575) B-D Guardian TM Nestable Sharps Collectors by Becton Dickinson & Co. are substantially equivalent to the applicant's BWCSH-1 Sharps Container.
| Element of<br>Comparison | Subject Device | Predicate Device<br>#1 | Predicate Device<br>#2 |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Bradford Wright<br>Companies, Inc. | Thermal Waste<br>Technologies, Inc. | Becton Dickinson<br>& Co. |
| Device Name | BWCSH-1 Sharps<br>Container | Demolizer # 47<br>One-Gallon Point<br>of Generation<br>Sharps Container | B-D Guardian™<br>Nestable Sharps<br>Collectors |
| FDA 510(k) Number | K131708 | K982781 | K943575 |
| Intended Use(s) | BWCSH-1 sharps<br>container is<br>intended to be used<br>for the safe disposal<br>of hazardous<br>sharps. | Intended for the<br>disposal of various-<br>sized contaminated<br>medical sharps<br>(including needles,<br>syringes, slides,<br>and ampoules) in<br>clinical and<br>laboratory<br>healthcare settings. | B-D sharps<br>containers are<br>intended to be used<br>for the safe disposal<br>of hazardous sharps |
| Target Population | Healthcare<br>professional | Healthcare<br>professional | Healthcare<br>professional |
| Where Used | Healthcare<br>Facilities | Intended for<br>clinical, nonclinical<br>healthcare settings<br>such as<br>laboratories,<br>dentists' office,<br>doctor's office,<br>corporate clinics<br>and nursing homes. | Healthcare<br>Facilities |
| Material | Polypropylene,<br>Polystyrene &<br>Tinplate steel | Tinplate steel | Polypropylene |
| Sharps Access | Sharps inserted<br>through the top | Sharps inserted<br>through the top | Sharps inserted<br>through the top |
| Sharps Closure | Lid is locked in<br>place for removal | Closure feature is<br>closed then locked | Flaps are closed<br>and locked in place |
| | | in place for<br>removal | for removal |
| Impact Resistance | Yes | Not tested | Yes |
| Puncture Resistance | Yes | Yes | Yes |
| Leak Proof | Yes | Yes | Yes |
| Non-sterile | Yes | Yes | Yes |
| Single Use | Yes | Yes | Yes |
| Capable of<br>maintaining a stable,<br>upright position | Yes | Not tested | Yes |
| No features to bend,<br>break, or shear<br>needle. | No Feature Present | No Feature Present | No Feature Present |
| Reusable Sharps<br>Containers | Labeling is "Single<br>Use | Labeling is "Single<br>Use | Labeling is "Single<br>Use |
| Overfill Indication | Labeled "Do Not<br>Fill Above This<br>Line" | Labeled "Do Not<br>Fill Above This<br>Line" | "Do Not Overfill"<br>or "Fill to this<br>Level Only" is<br>labeled or<br>embossed on the<br>container at the<br>location of the<br>overfill. Labeling<br>includes a "Fill<br>Line" |
| Clarity | Each Container has<br>a minimum of one<br>translucent<br>component, either<br>base or top | Temporary/Perman<br>ent Opening for<br>visual inspection. | Each Collector has<br>a minimum of one<br>translucent<br>component, either<br>base or top |
| Construction | Injection Molded<br>Lid and Rotating<br>top and 2 piece<br>seamed &glued<br>tinplate base | 3 piece seamed<br>&glued tinplate | Injection Molded<br>Containers, Lids<br>and Closure |
| Dimensions | 10.5 inches by 4.75<br>inches by 8.25<br>inches | 6.5 inches in<br>diameter by 8<br>inches | 12 inches by 13.5<br>inches by 6 inches |
| Fill Capacity | One US gallon | One US gallon | 7.6 Liters (2.01 US<br>gallons) |
# Comparison table with predicate device:
BWCSH-1 Sharps Container
{5}------------------------------------------------
:
.
{6}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE DISCUSSION of Similarities and Differences:
The BWCSH-1 Sharps Container is similar to both the Demolizer # 47 One-Gallon Point of Generation Sharps Container and the B-D Guardian™ Nestable Sharps Collectors in:
- . Intended use
- . Target Population
- . Materials
- . Design
- t Performance testing
#### Intended use comparison:
The intended use of the new BWCSH-1 Sharps Container is the same as both predicate devices. BWCSH-1 sharps container is intended to be used for the safe disposal of hazardous sharps.
#### Design and Material Comparison:
The design and functional characteristics of the BWCSH-1 Sharps Container and both predicate devices are similar. The BWCSH-1 Sharps Container and the B-D Guardian™ Nestable Sharps Collectors predicate device's parts are nestable, whereas the Demolizer # 47 One-Gallon Point of Generation Sharps Container is not. When assembled all form a single unit. BWCSH-1 Sharps Container and the B-D Guardian™ Nestable Sharps Collectors predicate device both have features to prevent contact between user and the contents and all are designed for a visual determination of the maximum capacity. None of the devices have features that bend, break, or shear needles. BWCSH-1 Sharps Container and both predicate devices are all designed for single use by a locking feature in the lids and closure or access door. BWCSH-1 Sharps Container's lid and sharps chute as well as the B-D Guardian 101 Nestable Sharps Collectors predicate device are constructed of polymeric materials which are injection molded. The material of construction differs from the B-D Guardian™ Nestable Sharps Collectors is the BWCSH-1 Sharps Container's lower canister is made of seamed and glued tin plated steel same as the whole Demolizer # 47 One-Gallon Point of Generation Sharps Container predicate device. BWCSH-1 Sharps Container and the predicate devices all allow for a visual determination of content level.
#### Summary of Non-Clinical Performance Bench Testing:
#### Performance Standards:
The Recognition Number 6-215 identifies ASTM F2132-01 (reapproved 2008), "Standard Specifications for puncture Resistance of Materials Used for the Collection for Discarded Medical Needles and Other Sharps."
{7}------------------------------------------------
Bradford Wright Companies, Inc.
The Recognition Number 6-293 identifies ISO 23907 First Edition 2012-09-01. "Sharps injury protection - Requirements and test methods - Sharps containers"
The performance testing demonstrates compliance with the recognized consensus standards. ASTM F2132-01 (reapproved 2008), "Standard Specifications for puncture Resistance of Materials Used for the Collection for Discarded Medical Needles and Other Sharps" and the applicable portions of ISO 23907 First Edition 2012-09-01. "Sharps injury protection - Requirements and test methods - Sharps containers"
In addition the relevant FDA guidance document, "Guidance on the Content and Format of Premarket Notification [510(k)] Submission for Sharps Collectors dated October 1993", was used to identify applicable physical and mechanical features of the BWCSH-1 Sharps Container and predicate devices. All applicable standards have been used to show that the BWCSH-1 Sharps Container is substantially equivalent to the appropriately listed predicate devices.
The performance testing summary demonstrates substantial equivalence between the BWCSH-1 Sharps Container and the predicate devices. The BWCSH-1 Sharps Container has been tested by appropriate methods with respect to the relevant standards: FDA recognized ASTM standards F2132-01 and OSHA regulations 29 CFR Part 1910:1030. No new issues of safety and effectiveness were raised with the testing performed, and the BWCSH-1 Sharps Container is considered substantially equivalent to its predicate device.
#### Performance Testing (Bench) - Product Testing
The BWCSH-1 Sharps Container incorporates equivalent collector design features and performance characteristics. The results of the product performance testing demonstrated equivalent performance to the predicate device performance and no new issues were raised.
#### Test Methods:
Puncture Resistance (Performed by and independent materials testing lab) - Passed ASTM F 2132-01 (2008) "Standard Specification for the Puncture Resistance of Materials used in containers for the Discarded Medical Needles and Other Sharps".
Leak Resistance of bottom and sides - Passed
Impact Resistance - Passed Based on ISO 23907:2012
Stability: - Passed
BWCSH-1 Sharps Container
Revision 4 Page 11 of 40
{8}------------------------------------------------
Bradford Wright Companies, Inc.
## Conclusion:
The performance testing summary demonstrates substantial equivalence between the BWCSH-1 Sharps Container and the predicate devices. BWCSH-1 Sharps Container has been tested by appropriate methods with respect to the relevant standards. FDA recognized ASTM standards F2132-01. No new issues of safety and effectiveness were raised with the testing performed and the BWCSH-1 Sharps Container is considered substantially equivalent to its predicate device.
*B-D Guardian™ Nestable Sharps Collector is a registered trademark of Becton Dickinson and Company.
BWCSH-1 Sharps Container
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or being supported by three abstract shapes. The logo is printed in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2014
Bradford Wright Companies, Inc. Mr. Mike DeMello President 4092 Orchard Way Lake Oswego, OR 97035
Re: K131708
Trade/Device Name: BWCSH-1 Sharps Container Regulation Number: 21 CFR 880.5570 Regulation Name: Sharps Container Regulatory Class: II Product Code: MMK December 31, 2013 Dated: Received: January 28, 2014
Dear Mr. DeMello:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
Page 2 - Mr. DeMello
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/10/Picture/8 description: The image contains a combination of text and graphic elements. The text includes the name "Tejashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director." There is also the acronym "DAGRID" present. The image also contains a decorative graphic element on the left side.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K131708
#### Device Name BWCSH-1 sharps container
#### Indications for Use (Describe)
The BWCSH-1 sharps container has a one-gally that is intended to be used for the safe disposal of hazardous sharps. The BWCSH-1 sharps container is metallic canister fitted with thermal plastic top and has a orange-red biohazard the containers with fill line printed on the BWCSH-1 device is for single use, disposable and is not a secondary container. BWCSH-1 sharps container dimensions are 10.5 inches in width at the top, 4.75 inches in height. BWCSH-1 sharps container weighs 1.075 pounds when empty.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY . : ...
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Elizabeth 2014.03.09 (
FORM FDA 3881 (9/13)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.