DEMOLIZER #47 ONE-GALLON POINT OF GENERATION SHARPS CONTAINER
K982781 · Thermal Waste Technology, Inc. · MMK · Sep 30, 1998 · General Hospital
Device Facts
Record ID
K982781
Device Name
DEMOLIZER #47 ONE-GALLON POINT OF GENERATION SHARPS CONTAINER
Applicant
Thermal Waste Technology, Inc.
Product Code
MMK · General Hospital
Decision Date
Sep 30, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The sharps container is intended for the collection of disposable sharps in both clinical and nonclinical healthcare settings, such as laboratories, dentists' offices, doctors' offices, veterinarian offices, corporate clinics, nursing homes, and other types of healthcare centers.
Device Story
The Demolizer #47 is a single-use, non-sterile, one-gallon metal sharps container. It is constructed of tin-plated steel with a friction-fit lid and black foam plastisol sealant. The device is used at the point of generation in clinical and non-clinical settings (e.g., doctors' offices, labs, nursing homes) to collect contaminated sharps, including hypodermic needles, syringes, scalpel blades, lancets, and broken glass. The container features a vertical inlet opening and a built-in vent port. Once closed, the lid is resistant to manual opening, providing a secure containment solution for hazardous medical waste. It is operated by healthcare staff or clinicians to safely dispose of sharps, reducing the risk of accidental injury during waste handling.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Material: Tin-plated steel; lap welded. Construction: Three-piece, one-gallon drum. Dimensions: 6-5/8" wide x 7-7/16" high; 2-3/8" inlet. Sealant: Black foam plastisol. Features: Friction-fit lid, 1 mm vent port. Sterilization: Non-sterile. Connectivity: None.
Indications for Use
Indicated for the disposal of contaminated medical sharps, including needles, syringes, slides, and ampoules, in clinical and laboratory healthcare settings.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
1.0 Gallon Leaktight Locking Top Translucent Red Sharps Disposal Container (K925816)
{0}------------------------------------------------
SEP 3 0 1998
### ATTACHMENT 9
### (As required by 21 C.F.R. §§ 807.87(h), 807.92)
- Manufacturer 1.
Contact Person
Thermal Waste Technologies, Inc. 19 Stony Hill Road Bethel, CT 06801 (203) 778-2210
Bill Ransone Vice President Sales & Marketing (203) 778-1039
- 3. Device
2.
- Demolizer® #47 One-Gallon Point of Trade or Proprietary Name a. Generation Sharps Container Sharps Container Common Name b. Classification Name Accessory to Hypodermic Needle C. II Classification d.
- Intended Use: The sharps container is intended for the collection of disposable sharps 4. in both clinical and nonclinical healthcare settings, such as laboratories, dentists' offices, doctors' offices, veterinarian offices, corporate clinics, nursing homes, and other types of healthcare centers.
#### క్ Predicate Devices
The Demolizer # 47 One-Gallon Point of Generation Sharps Container is substantially equivalent to the 1.0 Gallon Leaktight Locking Top Translucent Red Sharps Disposal Container marketed by Post Medical, Inc. (#K925816), and to the 5-Quart Round Sharps Container marketed by Sage Products, Inc. (#K943659).
#### 6.. General Description
The Demolizer # 47 One-Gallon Point of Generation Sharps Container is a single use, non-sterile, disposable, point of generation sharps container, constructed out of tin-plated steel. The device has a friction fit lid and black foam plastisol sealant for container closure. It is designed to hold contaminated disposable sharps, such as various-sized hypodermic, intravenous syringes or other medical needles, scalpel blades, disposable knives and lancets, and broken glass such as slides, slip covers and ampoules. The device is not a secondary container.
{1}------------------------------------------------
### 7. Detailed Description
The sharps container is a three-piece construction, one-gallon metal drum with a friction fit lid and sealant used to close the container. The minimum nominal body wall thickness is .0094 inches, and the minimum nominal top/bottom thickness is .0097 inches. The container is opaque and measures 6-5/8 inches wide by 7-7/16 inches in height, with a 2-3/8 inch sharps inlet opening in the center of the top of the container. The lid is 2-3/4 inches in diameter, with a thickness of .0088 inches. Lining the perimeter of the posterior side of the lid is a black foam plastisol which acts as an additional safety seal; this foam ensures a closed container. A built-in vent port, approximately 1 mm in diameter, is located on the top of the sharps container.
#### 8. Summary of Technological Characteristics
The Demolizer # 47 One-Gallon Point of Generation Sharps Container differs slightly from its predicate devices, in that the Demolizer sharps container is constructed of tinplated steel and lap welded. The two predicate devices are constructed of plastic. Further, unlike the predicate devices, once the Demolizer sharps container is closed, it is resistant to manual opening.
The Demolizer sharps container is similar to the two predicate devices, in that all are single use, disposable, nonsterile, opaque containers of roughly the same capacity, and all feature vertical inlet openings for the disposal of sharps.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle.
# MAR 2 7 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Bill Ransone Vice President Sales & Marketing Thermal Waste Technologies, Incorporated 19 Stony Hill Road Bethel, Connecticut 06801
Re: K982781
Trade/Device Name: Demolizer #47 One-Gallon Point of Generation Sharps Container Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Codes: MMK Dated: August 26, 1998 Received: August 31, 1998
Dear Mr. Ransone:
This letter corrects our substantially equivalent letter of September 30. 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Ransone
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 http://www.fda.gov/cdrh/organiz.html#OC for OC organization structure). Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Syute Y. Mcheioms
Chiu S. Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# ATTACHMENT 11
### INDICATIONS FOR USE STATEMENT
## (REVISION)
Page 1 of 1
3
510(k) Number: K982781 Device Name: Demolizer #47 One-Gallon Point of Generation Sharps Container Indications for Use: This device is intended for the disposal of various-sized contaminated medical sharps (including needles, syringes, slides, and ampoules) in clinical and laboratory healthcare settings.
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division Infection Control,
and General Hospital Devices
| 510(k) Number | K982781 |
|---------------|---------|
|---------------|---------|
| Prescription Use | |
|---------------------------|--|
| (Per 21 C.F.R. § 801.109) | |
OR
| Over the Counter Use | |
|----------------------|--|
|----------------------|--|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.