K083535 · Universal Medical, Inc. · DXH · Dec 15, 2008 · Cardiovascular
Device Facts
Record ID
K083535
Device Name
HEARTRAK SMART ECAT
Applicant
Universal Medical, Inc.
Product Code
DXH · Cardiovascular
Decision Date
Dec 15, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Heartrak Smart ECAT is a wireless ambulatory, multi-channel, continuous ECG event recorder with embedded arrhythmia detection algorithms. Heartrak Smart ECAT registers symptomatic and asymptomatic cardiac events triggered by a patient manually or autotriggered by embedded arrhythmia detection algorithms. Using wireless technology, Heartrak Smart ECAT, when placed within range of a compatible RF receiver, uploads recorded ECG waveform and ECG parameter data to the receiver. When data upload is complete, data can be reviewed and analyzed at a physician's office, clinic, or monitoring center. Heartrak Smart ECAT does not deliver any energy, administer any drugs, make any diagnosis, or control a patient's life. Heartrak Smart ECAT is for prescription use only.
Device Story
Heartrak Smart ECAT is a portable, hand-held, externally applied cardiac event recorder. It continuously scans and captures ECG signals via a 3-lead patient cable. The device features embedded arrhythmia detection algorithms that auto-trigger recording for events like bradycardia, tachycardia, and atrial fibrillation; patients can also manually trigger recordings. It stores up to 30 days of ECG data. When within 10 meters of an RF-compatible receiver, the device wirelessly uploads recorded waveforms and parameters. Data is subsequently reviewed by clinicians at a physician's office, clinic, or monitoring center to assist in arrhythmia assessment. The device does not provide automated diagnosis or life-sustaining functions. It is intended for prescription use to aid in the detection of infrequent or illusive cardiac arrhythmias.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, including electrical characteristics, RF transmission performance, and verification of arrhythmia detection algorithm functionality.
Technological Characteristics
Hand-held, portable, 3-lead ECG event recorder. Dimensions: 7.4 cm x 5.3 cm x 1.8 cm; Weight: 90g. Power: Internal 3.6V Li-Ion rechargeable battery. Connectivity: RF wireless (Bluetooth 2.0 SPP profile, 10m range). Bandwidth: 0.05-30 Hz. Input impedance: 2 MOhm. CMRR: 60 dB. Features: Lead-off detection, auto-triggering for bradycardia/tachycardia/AF, and 30-day storage.
Indications for Use
Indicated for patients requiring ambulatory, multi-channel, continuous ECG monitoring to capture symptomatic and asymptomatic cardiac events. Contraindications: None stated.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo in black and white. The logo features a stylized heart shape at the top. Below the heart shape, there is a letter 't' in a bold, sans-serif font. Underneath the letter 't', the letters 'INC' are written in smaller font size.
**HeartCare**
Corporation of America
Cardiac Monitoring Services
Image /page/0/Picture/3 description: The image shows a logo for Universal Medical Incorporated. The logo features a heart shape on the left side, with the words "Universal Medical" stacked on top of each other to the right of the heart. Below the words "Universal Medical" is the word "Incorporated" in a smaller font. The logo is simple and professional, and it is likely used to represent a medical company.
# 5. 510(k) Summary or 510(k) Statement
K083535
# DEC 1 5 2008
Date: October 30, 2008
Universal Medical, Inc. Submitter/Owner:
Official Contact Person Authorized by the Submitter:
Mark Job, Third-Party Reviewer Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313 Telephone: 763 682 4139 FAX: 763 682 4420 Email: mark@markjob.com
cardiac event recorder; electrocardiograph transmitter Device Type:
Device Class: ==============================================================================================================================================================================
Regulation Number 870.2920
Review Panel: Cardiovascular
Basis for the
new design incorporating wireless technology. Submission:
Page 1 of 9
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Mednet Healthcare Technologies Inc. The word "mednet" is written in bold, black letters. Below the word "mednet" is the phrase "HEALTHCARE TECHNOLOGIES INC." in smaller, black letters. To the right of the word "mednet" is a stylized heart shape.
HeartCare
Corporation of America
Cardiac Monitoring Services
Image /page/1/Picture/2 description: The image shows the logo for Universal Medical Incorporated. The logo features a heart shape on the left side, with the words "Universal medical" stacked on top of each other. Below the words, it says "INCORPORATED" in smaller letters. Underneath that, it says "An ISO13485 Certified Company".
### rvice bevond the cal
# Predicate Devices
Predicate devices for which Universal Medical, Inc. is claiming equivalence:
| 510(k)<br>Number | Trade or Model<br>Name | Manufacturer | Classification<br>Name | Product<br>Code |
|------------------|---------------------------------------------|-----------------------------------------|--------------------------------------------------------------------------------|-----------------|
| K071130 | Heartrak Smart<br>AF | Universal Medical,<br>Inc | transmitters and<br>receivers,<br>electrocardiograph,<br>telephone 870.2920 | DXH |
| K060911 | Cg-6108<br>Arrhythmia ECG<br>Event Recorder | Card Guard Scientific<br>Survival, Ltd. | transmitters and<br>receivers,<br>electrocardiograph,<br>telephone<br>870.2920 | DXH |
## Device Description
Heartrak Smart ECAT is a cardiac event recorder that is used to continuously scan and capture ECG signals. Patients can use Heartrak Smart ECAT to capture ECG data both before and after they experience a cardiac symptom. Heartrak Smart ECAT can capture and automatically record asymptomatic, infrequent, or illusive arrhythmia events such as Bradycardia, and Atrial Fibrillation.
Heartrak Smart ECAT can store up to 30 days of ECG data in its memory. The physician can use a compatible wireless device to set event recording times and autotriggering parameters and then upload them to a patient's monitor.
Using wireless technology, Heartrak Smart ECAT, when placed within range (less than 10 meters) of an RF compatible receiver, uploads recorded ECG waveform and ECG parameter data to the receiver. When data upload is complete, data can be reviewed and analyzed at a physician's office, clinic, or monitoring center.
The physician is to instruct the patient on the proper use and care of the Heartrak Smart ECAT monitor. Patients should be told to contact their physician if they have any further questions.
## Intended Use
Heartrak ECAT is a hand-held, portable, externally applied, cardiac event recorder; electrocardiograph transmitter.
page 2 of 9
tel 800.606.5511 ・ fax 800.889.5415 ・ 275 Phillips Blvd., Ewing, NJ 08618 ・ www.mednethealth.net
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Mednet Healthcare Technologies Inc. The word "mednet" is written in bold, black letters. Below the word "mednet" is the phrase "HEALTHCARE TECHNOLOGIES INC." in smaller, black letters. To the right of the word "mednet" is a graphic that looks like two overlapping hearts.
Image /page/2/Picture/2 description: The image shows a logo for HeartCare Corporation of America Cardiac Monitoring Services. The logo is black and white and features the words "HeartCare" in a large, bold font. Below that, in a smaller font, are the words "Corporation of America" and "Cardiac Monitoring Services". To the left of the text is a graphic that resembles a heart rate monitor.
Image /page/2/Picture/3 description: The image shows a logo for Universal Medical Incorporated. The logo features the words "Universal Medical" in a bold, sans-serif font, with the word "Universal" stacked on top of "Medical". Below the words is a thin line, and below that is the word "Incorporated" in smaller letters. The text "An ISO13485 Certified Company" is at the bottom.
## Indications for Use
Heartrak Smart ECAT is a wireless ambulatory, multi-channel, continuous ECG event recorder with embedded arrhythmia detection algorithms. Heartrak Smart ECAT registers symptomatic and asymptomatic cardiac events triggered by a patient manually or auto-triggered by embedded arrhythmia detection algorithms. Using wireless technology, Heartrak Smart ECAT, when placed within range of an RF compatible receiver, uploads recorded ECG waveform and ECG parameter data to the receiver. When data upload is complete, data can be reviewed and analyzed at a physician's office, clinic, or monitoring center.
Heartrak Smart ECAT does not deliver any energy, administer any drugs, make any diagnosis, or control a patient's life. Heartrak Smart ECAT is for prescription use only.
Page 3 OF 9
tel 800.606.5511 · fax 800.889.5415 · 275 Phillips Blvd., Ewing, NJ 08618 · www.medneth.net
{3}------------------------------------------------
mednet
ervice beyond the call
HeartCarporation of Amer
Image /page/3/Picture/2 description: The image shows the logo for Universal Medical Incorporated. The logo has the words "Universal Medical" in a vertical orientation. To the right of the words is the phrase "Incorporated, An ROI Health Company" in a smaller font.
echnological Claracteristics (Subsatial Equivalers) Table)
In the belows the the technologics of Heartel Co. In art ECAT ae subsentially equivalent othe prodicated deviral, L
CG-6108 Arrhythmia ECG Event Recorder (K060911) manufactured by Card Guard Scientific Survival, Ltd. in wireless communication
# ble 1 Substantial Equivalency Tabl
| | Heartrak Smart ECAT | CG-6108 Arrhythmia ECG Event<br>Recorder (K060911) | Heartrak Smart AF<br>(K071130) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Universal Medical, Inc.<br>Heartrak Smart ECAT is a<br>hand-held, portable, externally<br>applied, cardiac event<br>recorder; electrocardiograph<br>transmitter. | Card Guard Scientific Survival,<br>Ltd.<br>Intended for use by patients who<br>experience transient symptoms that<br>may suggest cardiac arrhythmia. | Universal Medical, Inc.<br>Heartrak Smart AF is a hand-<br>held, portable, externally<br>applied, cardiac event recorder<br>that is intended for<br>transtelephonic use. |
| Indications for use | Heartrak Smart ECAT is a<br>wireless ambulatory, multi-<br>channel, continuous ECG<br>event recorder with embedded<br>arrhythmia detection<br>algorithms. Heartrak Smart<br>ECAT registers symptomatic<br>and asymptomatic cardiac<br>events triggered by a patient<br>manually or auto-triggered by<br>embedded arrhythmia<br>detection algorithms. Using | The CG-6108 system is an Arrhythmia<br>ECG Event Recorder designed for self-<br>testing by patients at home and for<br>analysis by medical professionals at a<br>remote monitoring center.<br><br>It comprises a chest-worn ECG sensor<br>and a handheld device with a<br>proprietary application, configured to<br>process and transmit the ECG<br>recordings. | Heartrak Smart AF is a hand-<br>held, portable, externally<br>applied, cardiac event<br>recorder that is intended for<br>transtelephonic use. Patient<br>calls a receiving center at the<br>hospital or physicians office<br>from the patient's home to<br>play back the recording.<br>Heartrak Smart AF converts<br>electrocardiogram (ECG)<br>signals into audio tones which |
el 800.606.5511 ・ fax 800.889.5415 ・ 275 Phillips Blvd., Ewing, NJ 08618 ・ www.mednethealth.ne
4 OF 9
Page
{4}------------------------------------------------
mednét
Image /page/4/Picture/7 description: The image shows the logo for HeartCare, a corporation of America. The text is vertically oriented and stacked on top of each other. Underneath the text is a graphic of a heart rate monitor.
Image /page/4/Picture/2 description: The image shows the logo for Universal Medical. The logo has the words "Universal Medical" stacked vertically. To the right of the words are some smaller words that are illegible.
| Service |
|----------|
| beyond |
| the call |
| Heartrak Smart ECAT | CG-6108 Arrhythmia ECG Event<br>Recorder (K060911)<br>Card Guard Scientific Survival, Ltd. | Heartrak Smart AF<br>(K071130) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Universal Medical, Inc.<br>wireless technology, Heartrak<br>Smart ECAT, when placed<br>within range of a compatible<br>RF receiver, uploads recorded<br>ECG waveform and ECG<br>parameter data to the receiver.<br>When data upload is complete,<br>data can be reviewed and<br>analyzed at a physician's<br>office, clinic, or monitoring<br>center. | The chest-worn unit includes 3<br>electrodes on a harness and it houses a<br>battery, an ASIC and a Bluetooth<br>transceiver for the acquisition,<br>recording, and transmission of the ECG<br>signal. The ECG signals are transmitted<br>via Bluetooth to the handheld device.<br>When an event is detected, it is<br>wirelessly transmitted to the CG<br>Monitoring Center for professional<br>analysis. The handheld device is<br>equipped with shared memory used to<br>record the signal received from the<br>sensor and to allow pre- and post-<br>processing options through the use of<br>this memory in a dual memory loop<br>configuration, both running in parallel.<br>One loop is auto-triggered, with<br>programmable thresholds, which starts<br>recording based on specific rhythms<br>and arrhythmias detected or manually<br>activated by the patient. The second,<br>and longer, recording loop is controlled<br>remotely to provide the physician with<br>more information, when requested by | Universal Medical, Inc.<br>are transmitted over the<br>telephone lines.<br><br>Heartrak Smart AF does not<br>deliver any energy, administer<br>any drugs, or control a<br>patient's life. Heartrak Smart<br>AF is not a diagnostic tool and<br>performs no diagnostic<br>functions. | |
| Heartrak Smart ECAT does<br>not deliver any energy,<br>administer any drugs, make<br>any diagnosis, or control a<br>patient's life. Heartrak Smart<br>ECAT is for prescription use<br>only. | | | |
| | Heartrak Smart ECAT | CG-6108 Arrhythmia ECG Event<br>Recorder (K060911)<br>Card Guard Scientific Survival,<br>Ltd. | Heartrak Smart AF<br>(K071130) |
| | Universal Medical, Inc. | | Universal Medical, Inc. |
| | | the CG Monitoring Center. | |
| | | The handheld device automatically<br>transmits the recorded ECG, via<br>cellular link, to the Monitoring Center.<br>When cellular service is unavailable the<br>patient can transmit via landline<br>telephone. | |
| Monitor Features | | | |
| Patient Cable | 3-lead patient cable | 3-lead patient cable | 2-lead patient cable |
| Lead off Detection | Yes | Unknown | No |
| Channel Recording | 3 | 3 | 1 |
| Monitoring Mode | Continuous | Continuous | Loop |
| Data Transmission | Radio Frequency (RF) | Radio Frequency (RF) | Transtelephonic FM |
| Recording Button | Yes | Yes | Yes |
| | No | No | Yes |
| Playback Button | (When User puts monitor within range<br>of a compatible RF receiver, the monitor<br>automatically uploads recorded<br>ECG data to the receiver.) | (When User puts monitor within range<br>of a compatible RF receiver, the<br>monitor automatically uploads recorded<br>ECG data to the receiver.) | Yes |
| Reset Button | No | No | Yes |
| Unintentional Erase Data<br>Protection | Yes | Unknown | Yes |
| | Heartrak Smart ECAT | CG-6108 Arrhythmia ECG Event<br>Recorder (K060911)<br>Card Guard Scientific Survival, Ltd. | Heartrak Smart AF<br>(K071130) |
| | Universal Medical, Inc. | Unknown | Universal Medical, Inc. |
| Power Loss Data Protection | Yes | | Yes |
| Programming/Configuration<br>Options | Available | | |
| Programmable pre/post<br>recording times | Multiple | Yes | Multiple (Total memory 10 minutes) |
| Number of events | Multiple | Multiple | Multiple |
| Patient Manual Activation | Yes | Yes | Yes |
| Silent Recording | Yes | Unknown | Yes |
| Auto-Triggering | Yes | Yes | Yes |
| Bradycardia | Yes (configuration option to<br>set range for rate) | Yes (configuration option to<br>set range for rate) | Yes (configuration option to<br>set range for rate) |
| Tachycardia | Yes (configuration option to<br>set range for rate) | Yes (configuration option to set<br>range for rate) | Yes (configuration option to<br>set range for rate) |
| Atrial Fibrillation | Yes | Yes | Yes |
| Auto-Trigger On/OFF<br>Capability | Yes | Unknown | Yes |
| Monitor Physical<br>Characteristics | | | |
| Dimensions | 7.4 cm length x 5.3 cm wide x<br>1.8 cm thick<br>Weight with batteries 90 gm | 75 x 58 x 23 mm (max.)<br>Net weight 54 gm | 7.4 cm length x 5.3 cm wide x<br>1.8 cm thick<br>Weight with batteries 90 gm |
| | Heartrak Smart ECAT | CG-6108 Arrhythmia ECG Event<br>Recorder (K060911)<br>Card Guard Scientific Survival, Ltd. | Heartrak Smart AF<br>(K071130)<br>Universal Medical, Inc. |
| Monitor Technical<br>Characteristics | Universal Medical, Inc. | | |
| Transmission Mode<br>(Bluetooth 2.0 SPP<br>Profile) | Yes | Yes | N/A |
| RF transmission range…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.