K131571 · Advanced Vascular Dynamics · KKX · Jan 31, 2014 · General, Plastic Surgery
Device Facts
Record ID
K131571
Device Name
AVD RIGHT RADIAL DRAPE
Applicant
Advanced Vascular Dynamics
Product Code
KKX · General, Plastic Surgery
Decision Date
Jan 31, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4370
Device Class
Class 2
Indications for Use
The AVD Left and Right Radial Drapes are single use sterile devices intended for use by medical professionals as a protective patient covering, such as to isolate a site of surgical incision or vascular puncture from microbial and other contamination. The drapes are attached to the patient in proximity to a vascular puncture site to isolate vascular access punctures from microbial and other contamination during procedures including angiographies and percutaneous interventions.
Device Story
AVD Radial Drapes are single-use, sterile surgical drapes designed for vascular access procedures. The device consists of a fluid-resistant film layer, an absorbent barrier layer, and an adhesive material. The drape features an oval opening surrounded by adhesive tape, allowing clinicians to secure the drape directly over the surgical or vascular puncture site. By isolating the site from microbial and other contamination, the drape provides a sterile field during angiographies and percutaneous interventions. The device is supplied in Left Radial and Right Radial configurations. It is intended for use by medical professionals in clinical settings.
Clinical Evidence
No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-5:2009, ISO 10993-10:2002), barrier protection (ANSI/AAMI PB70 Level 4), tensile strength (ASTM 5034-09), tear resistance (ASTM D5587-08, WSP 100.2), flammability (16 CFR 1610), and sterilization validation.
Technological Characteristics
Materials: Non-woven rayon absorbent layer, polypropylene adhesive laminating layer, polyester film barrier layer, medical grade double-sided adhesive tapes (PE, LDPE). Barrier Level 4 (ANSI/AAMI PB70). Sterilization: Ethylene Oxide (EO). Form factor: Approx. 38" x 29" with oval access opening.
Indications for Use
Indicated for use by medical professionals as a protective patient covering to isolate surgical incision or vascular puncture sites from microbial contamination during procedures such as angiographies and percutaneous interventions.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
HVO Various Disposable Non-Sterile Drapes (K050508)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K131571
## A. Submitter:
Advanced Vascular Dynamics
4252 SE International Way, Suite F .
Milwaukie, OR 97222 USA
Phone: (503) 223-2333
Fax: (503) 223-8585
Matthew Semler, President Contact:
> Semler Technologies, Inc., dba Advanced Vascular Dynamics 4252 SE International Way. Suite F Milwaukie, OR 97222 USA
| Phone: | (503) 223-2333 |
|--------|------------------------|
| Fax: | (503) 223-8585 |
| Email: | msemler@compressar.com |
Date Prepared: December 9, 2013
### B. Device Names:
| Trade Name: | AVD Radial Drapes |
|----------------------------|--------------------------------------|
| Common/Usual Name: | Surgical Drape |
| Classification Name: | Surgical Drape and Drape Accessories |
| Classification Regulation: | 21 CFR 878.4370, Class II |
| Product Code: | KKX |
### B. Predicate Device:
HVO Various Disposable Non-Sterile Drapes (K050508), currently legally distributed by Haywood Vocational Opportunities, Inc.
## D. Device Description:
The AVD Radial Drapes are composed of a fluid-resistant film layer, an absorbent barrier layer, and an adhesive material that bonds the two layers. The surgical site is accessed through an oval opening in the drape that is surrounded by an adhesive tape to keep the drape in position over the surgical or vascular puncture site. The AVD Drapes are supplied sterile and for single use only. The Drapes are available in both a Left Radial and a Right Radial configuration. They may be sold singly or as part of a procedure pack.
{1}------------------------------------------------
### E. Intended Use:
The AVD Left and Right Radial Drapes are single use sterile devices intended for use by medical professionals as a protective patient covering, such as to isolate a site of surgical incision or vascular puncture from microbial and other contamination. The drapes are attached to the patient in proximity to a vascular puncture site to isolate vascular access punctures from microbial and other contamination during procedures including angiographies and percutaneous interventions.
## F. Comparison with the Predicate Device:
The AVD Radial Drapes have the same or similar technological specifications and are the same or similar in design, function, performance characteristics, and intended use as the predicate device.
The AVD Radial Drapes and the HVO Drapes are similar in that:
- . They have the same intended use;
- They have the same or similar physical and mechanical specifications; .
- The materials of construction are the same or very similar; ●
- Thev have the same principle of operation and performance; �
- They both meet applicable performance requirements; .
- Thev are both biocompatible. o
The differences between the AVD Radial Drapes and the HVO Drapes are:
- The material used in the AVD Drapes are lighter weight than the HVO drape . material;
- The layout of the AVD Drapes and dimensions are slightly different from the . HVO Drapes to be specifically suited for vascular access procedures;
- The AVD Drapes are provided sterile, while the HVO Drapes are provided . non-sterile.
There are no significant differences between the AVD Drapes and the predicate device that would adversely affect the use of the device, and the differences do not raise any new issues of safety or effectiveness. The comparison table on the following page summarizes the similarities and differences.
{2}------------------------------------------------
·
| Parameter | AVD Radial Drapes,<br>K131571 (this submission) | Various Disposable Non-<br>Sterile Drapes, K050508 | Substantial<br>Equivalence |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | Single use sterile devices<br>intended for use by medical<br>professionals as a protective<br>patient covering, such as to<br>isolate a site of surgical incision<br>or vascular puncture from<br>microbial and other<br>contamination. The drapes are<br>attached to the patient in<br>proximity to a vascular puncture<br>site to isolate vascular access<br>punctures from microbial and<br>other contamination during<br>procedures including<br>angiographies and<br>percutaneous interventions. | Disposable, non-sterile<br>drape. Intended to be used<br>by medical professionals<br>as protective coverings,<br>such as a patient covering<br>to isolate a site for surgical<br>incision from<br>contamination. Designed<br>to be repackaged and/or<br>sterilized before use. | Substantially<br>equivalent; the AVD<br>Drapes are designed<br>for use in specific<br>procedures, while the<br>HVO Drapes are for<br>general surgical use.<br>The AVD Drapes are<br>provided sterile as a<br>convenience for the<br>user. |
| Sterility | Sterile (EO) | Non-Sterile<br>Must be sterilized before<br>use (EO) | No adverse effect on<br>S&E the sterile AVD<br>Drapes are a<br>convenience for the<br>user. |
| | | Physical Specifications | |
| Materials | Non-woven rayon absorbent<br>layer<br>Polypropylene adhesive<br>laminating layer<br>Polyester film barrier layer<br>Medical grade double-sided·<br>adhesive tapes (PE, LDPE) | SMS, Micro-embossed<br>LDPE, Clear LDPE,<br>Nonwoven (wetlaid<br>cellulose), Airtex, Sontara,<br>Krayton,<br>Medical Grade Single and<br>Double Coated Tapes,<br>Bridging, Polyfoam,<br>Velcro, Polyester Mesh,<br>Hot Melt, Cold Glue and<br>Coated Medical Grade<br>Liners. | Substantially<br>equivalent; the AVD<br>Drape materials are<br>included in K050508<br>materials list. |
| Size | Approx. 38" ± 1" long by 29"<br>±1" wide | Various | Not a significant<br>difference; the AVD<br>Drape is sized for use<br>in specific procedures. |
| Barrier<br>Protection | Barrier Level 4<br>(ANSI/AAMI PB70) | Not stated | Unknown |
| Resistance<br>to tears | Pass<br>(ASTM D5587-08, WSP 100.2) | Not stated | Unknown |
| Tensile<br>Strength | Pass<br>(ASTM 5034-09) | Not available | Unknown |
| Flammability | Class I (16 CFR 1610 | Not available | Unknown |
| Comparison Table, AVD Radial Drapes vs. HVO Drapes | | | |
|----------------------------------------------------|--|--|--|
| | | | |
.
.
.
.
・
.
{3}------------------------------------------------
| Parameter | AVD Radial Drapes,<br>K131571 (this submission) | Various Disposable Non-<br>Sterile Drapes, K050508 | Substantial<br>Equivalence |
|-----------------------|-------------------------------------------------|----------------------------------------------------|----------------------------|
| Biocompatibility | | | |
| Cytoxicity | Non-cytotoxic<br>(ISO 10993-5:2009) | Not available | Unknown |
| Skin<br>Sensitization | Non-sensitizing<br>(ISO 10993-10:2002) | Not available | Unknown |
| Skin<br>Irritation | Non-irritating<br>(ISO 10993-10:2002) | Not available | Unknown |
| EO<br>Residuals | Acceptable<br>(ISO 10993-10:2008) | Not available | Unknown |
### G. Non-clinical Testing
In accordance with applicable test standards and methods, non-clinical performance testing of the AVD Radial Drapes consisted of tensile strength, resistance to tearing, resistance to penetration by synthetic blood/liquid barrier protection, biocompatibility (cytotoxicity, irritation, sensitization, EO residuals), sterilization validation, and
- flammability testing. Testing specifics are listed in the Comparison Table in Section F above. The Drapes met the stated acceptance criteria for the non-clinical testing.
## H. Clinical Testing
Clinical performance is not applicable for this product. Therefore clinical testing is not included in this submission.
.
### 1. Conclusions Drawn from Testing
. .
The non-clinical testing results demonstrate that the AVD Drapes are substantially equivalent to the predicate device and do not raise new questions about safety and effectiveness. Testing included biocompatibility testing (cytotoxicity, sensitization, imitation) according to ISO 10993, barrier protection classification according to ANSI/AAMI PB:70, tensile strength, tear resistance, flammability, sterilization validation, and sterilization residuals.
Page 7
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized abstract symbol resembling a bird or a person with outstretched arms. The symbol is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 31, 2014
Advanced Vascular Dynamics, Incorporated Mr. Matthew Semler President 4252 SE International Way, Suite F Milwaukie, OR 97222
Re: K131571
Trade/Device Name: AVD Right Radial Drape Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape and Drape Accessories Regulatory Class: II Product Code: KKX Dated: December 9, 2013 Received: December 11, 2013
Dear Mr. Semler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Semler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/5/Picture/7 description: The image shows a signature and some text. The text includes "Tejashri Purohit-Sheth, M.D., Clinical Deputy Director, DAGRID, FOR". The signature is illegible. The image is in black and white.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Sincerely yours.
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known): _ K131571
Device Name:
AVD Radial Drapes
Indications for Use:
The AVD Left and Right Radial Drapes are single use sterile devices intended for use by medical professionals as a protective patient covering, such as to isolate a site of surgical incision or vascular puncture from microbial and other contamination. The drapes are attached to the patient in proximity to a vascular puncture site to isolate vascular access punctures from microbial and other contamination during procedures including angiographies and percutaneous interventions.
Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth Claverie -S
2014.01.30T17:27:33-05'00'
Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.