K050508 · Haywood Vocational Opportunities · KKX · Apr 27, 2005 · General, Plastic Surgery
Device Facts
Record ID
K050508
Device Name
HVO, INC., DISPOSABLE NON-STERILE SURGICAL DRAPES
Applicant
Haywood Vocational Opportunities
Product Code
KKX · General, Plastic Surgery
Decision Date
Apr 27, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4370
Device Class
Class 2
Indications for Use
HVO surgical drapes are made from natural and synthetic materials intended to be used by medical professionals as protective coverings, such as a patient covering to isolate a site for surgical incision from contamination. They are designed to be repackaged and/or sterilized before use. This single use product is a disposable non-sterile drape designed to be re-packaged and/or sterilized prior to use. This product may be sterilized using Ethylene Oxide (EO) following the Validation and Routine Control under ANSI/AMMI/ISO 11135. For more information about sterilization of this product, contact HVO, Inc.
Device Story
Disposable non-sterile surgical drapes; used as protective patient coverings to isolate surgical incision sites from contamination. Materials include SMS, LDPE, nonwoven cellulose, Airtex, Sontara, Krayton, medical-grade tapes, polyfoam, Velcro, polyester mesh, and adhesives. Supplied non-sterile; intended for customer-performed sterilization via Ethylene Oxide (EO) per ANSI/AAMI/ISO 11135. Used in clinical/surgical settings by medical professionals. Provides a physical barrier to reduce microbial transfer during surgical procedures.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Materials: SMS, micro-embossed/clear LDPE, nonwoven (wetlaid cellulose), Airtex, Sontara, Krayton, medical-grade single/double coated tapes, bridging, polyfoam, Velcro, polyester mesh, hot melt, cold glue, coated medical-grade liners. Form factor: Various surgical drapes (table covers, OB/GYN, general surgery, orthopedic, ENT/EENT, drape sheets, cystoscopy, craniotomy, angiography, fluid pouches, instrument covers). Sterilization: Non-sterile; validated for ETO sterilization per ANSI/AAMI/ISO 11135.
Indications for Use
Indicated for use by medical professionals as protective patient coverings to isolate surgical incision sites from microbial and other contamination. Intended for single use; must be sterilized by the customer prior to use in a sterile setting.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Submission Summary (Full Text)
{0}------------------------------------------------
# APR 2 7 2005
K050508
#### Haywood Vocational Opportunities, Inc. Various Disposable Non-Sterile Surgical Drapes
510(k) Summary
Summary of the Safety and Effectiveness Information Upon Which An Equivalence Could Be Based
#### SPONSOR INFORMATION
Haywood Vocational Opportunities, Inc. Name: P.O. Box 7 Address: Hazelwood, N.C. 28738 Telephone: (828) 456-4455 (828) 456-4401 Fax:
#### DEVICE NAMES
HVO, Inc., Various Disposable Non-Sterile Surgical Drapes Name: Surgical Drapes Common/Usual Name: Classifications Name (if known): Surgical and Drape Accessories Device Names: Table Covers, OB GYN, General Surgery, Orthopedic, ENT and EENT, Drape Sheets, Cystoscopy, Craniotomy, Angiography, Fluid Pouches, and Instrument Covers.
#### PREDICATE OR LEGALLY MARKETED DEVICES
There are several predicate devices currently on the market that have similar function and are made from same or similar materials.
#### DEVICE DESCRIPTION
The proposed HVO, Inc. Various Disposable Non-Sterile Surgical Drapes function in the same manner as predicate devices in that they are intended to be used as protective patient coverings, such as used to isolate a site of surgical incision from microbial and other contamination.
Device Design/ Materials Used/ Physical Properties: The HVO, Inc. Various Disposable Non-Sterile Surgical Drapes are made of materials commonly used for their purpose. The primary material components are SMS, Micro-embossed LDPE, Clear LDPE, Nonwoven (wetlaid cellulose), Airtex, Sontara, Krayton, Medical Grade Single and Double Coated Tapes, Bridging, Polyfoam, Velcro, Polyester Mesh, Hot Melt, Cold Glue and Coated Medical Grade Liners.
Primarily all of our drapes are offered in the color blue, which is the common color for medical device materials, but some materials are offered in white or with white backing. These materials include Nonwovens, SMS, Airtex, Sontara, Polyfoam, Back Table Covers, Bridging, and LDPE's. All other materials are clear plastics or adhesives, which are not colored, but natural.
#### DEVICE INTENDED USE
The HVO, Inc. Various Non-Sterile Surgical Drapes are intended to be used as protective patient covering, such as to isolate surgical incisions from microbial and other contamination. These proposed devices are intended to undergo sterilization by the customer prior to use in the sterile setting. These products can be sterilized by ETO. The products may be sterilized using Ethylene Oxide following the Validation and Routine Control under ANSVAAMI/ISO 11135, see section 12 for more details on sterilization.
{1}------------------------------------------------
### Haywood Vocational Opportunities, Inc. Various Disposable Non-Sterile Surgical Drapes
510(k) Summary
## TECHNOLOGICAL COMPARISON WITH PREDICATE OR LEGALLY MARKETED DEVICE(S)
| Characteristics | HVO, Inc. Devices | Other Devices |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Materials | SMS, Micro-Embossed LDPE, Clear<br>LDPE, Sontara, Nonwoven, Krayton,<br>Airtex, Polyfoam, Medical Grade Single<br>And Double Coated Tapes, Velcro, Cold<br>Glue, Hot Melt, Bridging, Polyester Mesh<br>Coated Medical Grade Liners | Same or similar materials |
| Absorbency | Absorbent and Non-Absorbent | Same |
| Packaging | Bulk and Single Use | Bulk and Single Use |
| Disposable | Yes | Yes |
| Sterility | Non-Sterile | Non-Sterile and Sterile |
Page 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.
APR 2 7 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Tim Kelley Director of Regulatory Affairs Haywood Vocational Opportunities, Incorporated 56 Scates Street Waynesville, North Carolina 28786
Re: K050508
Trade/Device Name: Various Disposable Non-Sterile Surgical Drapes (Table Covers, OB GYN, General Surgery, Orthopedic, ENT and ENNT, Drape Sheets, Cystoscopy, Craniotomy, Angiography, Fluid Pouches, and Instrument Covers) Regulation Number: 878.4370 Regulation Name: Surgical Drape and Drape Accessories Regulatory Class: II Product Code: KKX Dated: April 19, 2005 Received: April 20, 2005
Dear Mr. Kelley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Kelley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sultie Y. Michie Omd.
Chr. Livia Ph.D.
Chíu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication for Use
510(k) Number: K050508
Device Name: HVO, Inc. Various Disposable Non-Sterile Surgical Drapes (Table Covers, OB GYN, General Surgery, Orthopedic, ENT and ENNT, Drape Sheets, Cystoscopy, Craniotomy, Angiography, Fluid Pouches, and Instrument Covers).
Indications For Use:
HVO surgical drapes are made from natural and synthetic materials intended to be used by medical professionals as protective coverings, such as a patient covering to isolate a site for surgical incision from contamination. They are designed to be repackaged and/or sterilized before use.
This single use product is a disposable non-sterile drape designed to be re-packaged and/or sterilized prior to use. This product may be sterilized using Ethylene Oxide (EO) following the Validation and Routine Control under ANSI/AMMI/ISO 11135. For more information about sterilization of this product, contact HVO, Inc.
Prescription Use X (Part 21 CFR 801 Subpart D) and/or
Over-The-Counter Use (21 CFR 807 Subpart C)
Laird H. Mingay 13
on Sign-Off) sion Sign-Off)
sion of Anesthesiology, General Hospital, rection Control. Dental Devices
:(k) Number. 4656508
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.