ACE ALBUMIN REAGENT, ACE TOTAL PROTEIN REAGENT, ACE CALCIUM-ARSENAZO REAGENT AND ACE PHOSPHORUS REAGENT
Applicant
Alfa Wassermann Diagnostics Technologies, LLC
Product Code
CIX · Clinical Chemistry
Decision Date
Aug 19, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1035
Device Class
Class 2
Indications for Use
The ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum and lithium heparin plasma using the ACE, ACE Alere and ACE Axcel Chemistry Systems. Albumin measurements are used in the diagnosis and treatment of numerous diseased involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Device Story
Reagents for quantitative photometric/colorimetric analysis of serum and lithium heparin plasma samples; used on ACE, ACE Alera, and ACE Axcel clinical chemistry systems. Reagents react with analytes (albumin, total protein, calcium, phosphorus) to form colored complexes; intensity measured bichromatically. Systems operate in clinical laboratories or physician office laboratories; operated by lab personnel. Output is quantitative concentration value; used by clinicians for diagnosis and treatment monitoring of liver, kidney, bone, and metabolic diseases.
Clinical Evidence
Bench testing only. Precision studies (20-day, n=80; POL studies, n=20) showed total CVs generally <5%. Linearity confirmed across reportable ranges (r² >0.99). Method comparison (n=50-51 per analyte) against predicate showed high correlation (r² >0.99). Matrix comparison (serum vs. lithium heparin plasma) confirmed equivalence. Interference testing performed per CLSI EP07-A2.
Technological Characteristics
Photometric/colorimetric assays. Reagents: Albumin (Bromcresol green), Total Protein (Copper sulfate/tartrate), Calcium (Arsenazo III), Phosphorus (Ammonium molybdate/sulfuric acid). Bichromatic measurement at specific wavelengths (e.g., 629/692 nm, 544/692 nm, 647/692 nm, 340/378 nm). Standalone reagents for use on automated clinical chemistry analyzers. Traceability to NIST SRM 927, 909, 915b, 200, 3139a.
Indications for Use
Indicated for quantitative determination of albumin, total protein, calcium, and inorganic phosphorus in serum and lithium heparin plasma. Used in clinical laboratories or physician office laboratories for diagnosis/treatment of liver, kidney, bone marrow, parathyroid, and metabolic disorders.
Regulatory Classification
Identification
An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Predicate Devices
Alfa Wassermann ACE and ACE Axcel Clinical Chemistry Systems and ACE Reagents (K930104, K113253, K113374)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k131488
B. Purpose for Submission:
New reagents (Albumin, Total Protein, Calcium-Arsenazo and Inorganic Phosphorus U.V. Reagent added onto ACE Alera instrument
Addition of lithium heparin plasma samples to already cleared reagents on the ACE (k930104) and ACE Axcel (k113389) instruments.
C. Measurand:
Albumin, Total Protein, Calcium, Phosphorus
D. Type of Test:
Quantitative, photometric/colorimetric methods
E. Applicant:
Alfa Wasserman Diagnostic Technologies, LLC
F. Proprietary and Established Names:
ACE Albumin Reagent
ACE Total Protein Reagent
ACE Calcium-Arsenazo Reagent
ACE Inorganic Phosphorus U.V. reagent
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CIX | Class II | 21 CFR § 862.1035
Albumin Test System | Clinical Chemistry (75) |
| CEK | Class II | 21 CFR § 862.1635
Total Protein Test System, meets limitations of | Clinical Chemistry (75) |
{1}
| | | exemptions per 21 CFR § 862.9 (c) (9) | |
| --- | --- | --- | --- |
| CJY | Class II | 21 CFR § 862.1145
Calcium Test System | Clinical Chemistry (75) |
| CEO | Class I, reserved | 21 CFR § 862.1580
Phosphorus (inorganic) test system | Clinical Chemistry (75) |
## H. Intended Use:
1. Intended use(s):
See indications for use below
2. Indication(s) for use:
The ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum and lithium heparin plasma using the ACE, ACE Alere and ACE Axcel Chemistry Systems. Albumin measurements are used in the diagnosis and treatment of numerous diseased involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
3. Special conditions for use statement(s):
For *in vitro* diagnostic use only
For prescription use and Point-of-Care settings
{2}
4. Special instrument requirements:
For use on the ACE, ACE Axcel and ACE Alera Clinical Chemistry Systems.
I. Device Description:
The ACE Albumin Reagent consists of a single reagent bottle containing Bromcresol green (0.39 mmol/L) Acetate buffer, Preservative and Surfactant.
The ACE Total Protein Reagent consists of a single reagent bottle which contains Copper sulfate (12 mmol/L), Sodium potassium tartrate (32 mmol/L), Potassium iodide (30 mmol/L), Sodium hydroxide (600 mmol/L) and non-reactive ingredients.
The ACE Calcium-Arsenazo Reagent consists of a single reagent bottle containing Arsenazo III (≥0.15 mmol/L) Buffer and Surfactant.
The ACE Inorganic Phosphorus U.V. Reagent consists of a single reagent bottle containing Ammonium molybdate (0.48 mmol/L), Sulfuric Acid (220 mmol/L), and Surfactant
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACE Albumin Reagent
ACE Total Protein Reagent
ACE Calcium-Arsenazo Reagent
ACE Inorganic Phosphorus U.V. Reagent
2. Predicate 510(k) number(s):
k930104
3. Comparison with predicate:
ACE Albumin reagent:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Intended Use | Same | The ACE Albumin Reagent is intended for the quantitative determination of albumin concentration |
| Measurand | Same | Albumin |
| Assay Method | Same | Photometric |
| Measuring Range | 0.3-7.0 g/dL | 0.1-7.6 g/dL |
| Matrix | Human serum and Li | Human Serum |
{3}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| | heparin plasma | |
| Instrument platform | ACE, ACE Alera, and ACE Axcel clinical chemistry system | ACE clinical chemistry system |
ACE Total Protein Reagent:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Intended Use | Same | The ACE Total Protein reagent is for the quantitative determination of Total Protein concentration. |
| Measurand | Same | Total Protein |
| Assay Method | Same | Colorimetric |
| Measuring Range | 0.4-14.0 g/dL | 0.2-15.1 g/dL |
| Matrix | Human serum and Li heparin plasma | Human serum |
| Instrument platform | ACE, ACE Alera, and ACR Axcel clinical chemistry system | ACE clinical chemistry system |
ACE Calcium Reagent:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Intended Use | Same | The ACE Calcium-Arsenazo Reagent is for the quantitative determination of Calcium concentration. |
| Measurand | Same | Calcium |
| Assay Method | Same | Photometric |
| Measuring Range | 0.4-15.0 mg/dL | 0.2-16.5 mg/dL |
| Matrix | Human urine, serum and Li heparin plasma | Human serum |
| Instrument platform | ACE, ACE Alera, and ACR Axcel clinical chemistry system | ACE clinical chemistry system |
ACE Inorganic Phosphorus U.V. Reagent:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Intended Use | Same | The ACE Inorganic Phosphorus U.V. reagent is for the quantitative determination of Phosphorus concentration. |
| Measurand | Same | Phosphorus |
{4}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Assay Method | Same | Colorimetric |
| Measuring Range | 0.4-20.0 mg/dL | 0.2-21 mg/dL |
| Matrix | Human urine, serum and Li heparin plasma | Human serum |
| Instrument platform | ACE, ACE Alera, and ACR Axcel clinical chemistry system | ACE clinical chemistry system |
K. Standard/Guidance Document Referenced (if applicable):
Evaluation of Precision Performance of Quantitative Measurement Methods; (CLSI EP05-A2)
Method Comparison and Bias Estimation using Patient Samples; (CLSI EP09-A2-IR)
Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; (CLSI EP-17-A2)
Evaluation of the linearity of Quantitative Measurement Procedures: A statistical approach: (CLSI EP6-A)
Interference Testing in Clinical Chemistry; (CLSI EP07-A2)
L. Test Principle:
The ACE Albumin Reagent is a photometric assay in which Bromcresol green (BCG) binds specifically to albumin to form a green colored complex, which is measured bichromatically at 629 nm/692 nm. The intensity of color produced is directly proportional to the albumin concentration in the sample.
The ACE Total Protein Reagent is a colorimetric assay in which Cupric ions react with the peptide bonds of proteins under alkaline conditions to form a violet colored complex, which is measured bichromatically at 544 nm/692 nm. The intensity of color produced is directly proportional to the total protein concentration in the sample.
The ACE Calcium-Arsenazo Reagent utilizes a dye binding procedure in which calcium forms a blue-purple complex with Arsenazo III under acidic conditions
{5}
Calcium reacts with Arsenazo III in an acidic solution to form a blue-purple colored complex, which is measured bichromatically at 647 nm/692 nm. The intensity of color produced is directly proportional to the calcium concentration in the sample.
The ACE Inorganic Phosphorus U.V. Reagent is a colorimetric assay in which under acidic conditions, inorganic phosphorus in serum reacts with ammonium molybdate to form an unreduced phosphomolybdate complex, which absorbs strongly at 340 nm. The increase in absorbance, measured bichromatically at 340 nm/378 nm, is directly proportional to the amount of phosphorus in the sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision of the ACE Albumin, Total Protein, Calcium and Phosphorus assays was evaluated in-house using an ACE Alera analyzer. Human serum samples at 3 concentrations of the respective measurand were evaluated using a 20 day precision study. The study was performed measuring each sample 2 times per run, 2 runs per day for 20 days for a total of 80 measurements per sample using 1 reagent lot. The results are provided in the table below.
| Measurand | Sample Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV |
| Albumin g/dL | 2.6 | 0.03 | 1.3 | 0.05 | 2.0 |
| | 3.4 | 0.07 | 1.9 | 0.09 | 2.5 |
| | 4.3 | 0.03 | 0.7 | 0.10 | 2.3 |
| Calcium mg/dL | 6.5 | 0.08 | 1.3 | 0.13 | 2.1 |
| | 9.8 | 0.12 | 1.2 | 0.22 | 2.3 |
| | 12.6 | 0.23 | 1.8 | 0.29 | 2.3 |
| Total Protein g/dL | 4.2 | 0.10 | 2.3 | 0.11 | 2.6 |
| | 6.8 | 0.09 | 1.3 | 0.14 | 2.1 |
| | 10.1 | 0.23 | 2.3 | 0.32 | 3.1 |
| Inorganic Phosphorus mg/dL | 2.0 | 0.04 | 2.3 | 0.11 | 5.7 |
| | 3.8 | 0.12 | 3.2 | 0.16 | 4.2 |
| | 6.5 | 0.17 | 2.5 | 0.24 | 3.6 |
Plasma and Serum in house precision study:
The precision of the ACE Albumin, Calcium, Total Protein and Phosphorus assays for matched serum and Lithium heparin plasma samples at 3 concentrations of the respective measurand was evaluated in house over a period of 4 to 6 days. In this
{6}
study, matched serum and plasma with a low concentration of measurand from a single donor was spiked to a high concentration of measurand with commercially-available analyte and then the remaining sample prepared by intermixing. ACE Albumin was measured 2 times per run for 2 runs per day for 5 days (n=20). ACE Calcium was measured 2 times per run for 2 runs per day for 5 days (n=20). ACE Total Protein was measured 2 times per run for 2 runs per day for 6 days (n=24). ACE Inorganic Phosphorus was measured 2 times per run for 2 runs per day for 4 days (n=16)
All samples were tested on the ACE, ACE Alera and ACE Axcel analyzers.
# Albumin
| ACE analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (g/dL) | Sample Type | SD | %CV | SD | %CV |
| 4.1 | Serum | 0.05 | 1.3 | 0.07 | 1.6 |
| 3.8 | Plasma | 0.06 | 1.7 | 0.06 | 1.7 |
| 5.4 | Serum | 0.08 | 1.6 | 0.10 | 1.8 |
| 5.0 | Plasma | 0.05 | 1.0 | 0.07 | 1.4 |
| 6.5 | Serum | 0.07 | 1.1 | 0.11 | 1.6 |
| 6.2 | Plasma | 0.09 | 1.5 | 0.10 | 1.7 |
| ACE Alera Analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (g/dL) | Sample Type | SD | %CV | SD | %CV |
| 4.1 | Serum | 0.04 | 0.9 | 0.04 | 1.1 |
| 3.7 | Plasma | 0.03 | 0.8 | 0.05 | 1.4 |
| 5.3 | Serum | 0.05 | 1.0 | 0.06 | 1.1 |
| 5.0 | Plasma | 0.08 | 1.7 | 0.08 | 1.7 |
| 6.5 | Serum | 0.05 | 1.3 | 0.08 | 1.3 |
| 6.1 | Plasma | 0.08 | 1.3 | 0.10 | 1.6 |
| ACE Axcel Clinical Chemistry analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (g/dL) | Sample Type | SD | %CV | SD | %CV |
| 4.1 | Serum | 0.02 | 0.5 | 0.04 | 1.0 |
| 3.7 | Plasma | 0.06 | 1.6 | 0.06 | 1.6 |
| 5.3 | Serum | 0.03 | 0.6 | 0.03 | 0.6 |
{7}
8
| 4.9 | Plasma | 0.04 | 0.9 | 0.05 | 1.1 |
| --- | --- | --- | --- | --- | --- |
| 6.4 | Serum | 0.06 | 1.0 | 0.09 | 1.3 |
| 6.1 | Plasma | 0.05 | 0.9 | 0.08 | 1.3 |
Total Protein
| ACE analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (g/dL) | Sample Type | SD | %CV | SD | %CV |
| 6.7 | Serum | 0.06 | 1.0 | 0.07 | 1.0 |
| 7.2 | Plasma | 0.06 | 0.9 | 0.06 | 0.9 |
| 8.4 | Serum | 0.11 | 1.3 | 0.11 | 1.3 |
| 8.8 | Plasma | 0.04 | 0.5 | 0.06 | 0.7 |
| 10.1 | Serum | 0.07 | 0.7 | 0.08 | 0.8 |
| 10.3 | Plasma | 0.13 | 1.3 | 0.14 | 1.4 |
| ACE Alera analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (g/dL) | Sample Type | SD | %CV | SD | %CV |
| 6.7 | Serum | 0.05 | 0.7 | 0.05 | 0.8 |
| 7.1 | Plasma | 0.08 | 1.1 | 0.09 | 1.2 |
| 8.4 | Serum | 0.08 | 1.0 | 0.08 | 1.0 |
| 8.7 | Plasma | 0.06 | 0.7 | 0.10 | 1.2 |
| 10.0 | Serum | 0.07 | 0.7 | 0.09 | 0.9 |
| 10.2 | Plasma | 0.11 | 1.1 | 0.14 | 1.3 |
| ACE Axcel analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (g/dL) | Sample Type | SD | %CV | SD | %CV |
| 6.8 | Serum | 0.08 | 1.1 | 0.09 | 1.3 |
| 7.2 | Plasma | 0.05 | 0.7 | 0.08 | 0.9 |
| 8.4 | Serum | 0.07 | 0.8 | 0.11 | 1.4 |
| 8.8 | Plasma | 0.07 | 0.8 | 0.11 | 1.4 |
| 10.1 | Serum | 0.07 | 0.7 | 0.09 | 0.9 |
| 10.4 | Plasma | 0.08 | 0.8 | 0.10 | 1.0 |
{8}
# Calcium
| ACE analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (mg/dL) | Sample Type | SD | %CV | SD | %CV |
| 9.3 | Serum | 0.12 | 1.3 | 0.25 | 2.7 |
| 8.4 | Plasma | 0.04 | 0.5 | 0.20 | 2.4 |
| 11.7 | Serum | 0.18 | 1.6 | 0.20 | 1.7 |
| 10.7 | Plasma | 0.19 | 1.7 | 0.20 | 1.9 |
| 13.9 | Serum | 0.20 | 1.4 | 0.20 | 1.4 |
| 13.0 | Plasma | 0.25 | 1.9 | 0.26 | 2.0 |
| ACE Alera analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (mg/dL) | Sample Type | SD | %CV | SD | %CV |
| 9.3 | Serum | 0.09 | 0.9 | 0.22 | 2.4 |
| 8.3 | Plasma | 0.10 | 1.2 | 0.17 | 2.0 |
| 11.6 | Serum | 0.14 | 1.2 | 0.14 | 1.2 |
| 10.7 | Plasma | 0.13 | 1.2 | 0.15 | 1.4 |
| 13.8 | Serum | 0.19 | 0.14 | 0.19 | 1.4 |
| 12.9 | Plasma | 0.13 | 1.0 | 0.14 | 1.1 |
| ACE Axcel analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (mg/dL) | Sample Type | SD | %CV | SD | %CV |
| 9.3 | Serum | 0.08 | 0.8 | 0.17 | 1.8 |
| 8.3 | Plasma | 0.08 | 0.9 | 0.11 | 1.4 |
| 11.6 | Serum | 0.10 | 0.9 | 0.11 | 0.9 |
| 10.7 | Plasma | 0.12 | 1.2 | 0.13 | 1.2 |
| 13.8 | Serum | 0.09 | 0.7 | 0.11 | 1.2 |
| 13.1 | Plasma | 0.15 | 1.2 | 0.18 | 1.4 |
{9}
# Phosphorus
| ACE analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (mg/dL) | Sample Type | SD | %CV | SD | %CV |
| 3.5 | Serum | 0.15 | 4.4 | 0.17 | 5.0 |
| 3.1 | Plasma | 0.16 | 5.1 | 0.18 | 5.9 |
| 10.2 | Serum | 0.04 | 0.3 | 0.05 | 0.5 |
| 9.8 | Plasma | 0.09 | 0.9 | 0.09 | 0.9 |
| 17.0 | Serum | 0.26 | 1.5 | 0.26 | 1.6 |
| 16.7 | Plasma | 0.23 | 1.4 | 0.24 | 1.4 |
| ACE Alera analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (mg/dL) | Sample Type | SD | %CV | SD | %CV |
| 3.4 | Serum | 0.11 | 3.1 | 0.14 | 4.0 |
| 3.0 | Plasma | 0.11 | 3.7 | 0.15 | 5.0 |
| 9.9 | Serum | 0.08 | 0.8 | 0.08 | 0.8 |
| 9.6 | Plasma | 0.07 | 0.8 | 0.08 | 0.8 |
| 16.6 | Serum | 0.22 | 1.3 | 0.22 | 1.3 |
| 16.3 | Plasma | 0.24 | 1.5 | 0.29 | 1.8 |
| ACE Axcel analyzer | | Within run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Sample Mean (mg/dL) | Sample Type | SD | %CV | SD | %CV |
| 3.5 | Serum | 0.11 | 3.1 | 0.14 | 4.1 |
| 3.1 | Plasma | 0.15 | 5.0 | 0.19 | 6.1 |
| 10.2 | Serum | 0.04 | 0.4 | 0.12 | 1.2 |
| 9.9 | Plasma | 0.06 | 0.6 | 0.12 | 1.2 |
| 17.3 | Serum | 0.28 | 1.6 | 0.30 | 1.7 |
| 16.9 | Plasma | 0.30 | 1.8 | 0.32 | 1.9 |
# Serum POL Precision study:
The precision of the ACE Albumin, Calcium, Total Protein and Phosphorus assays at in-house and 3 physician office laboratories (POLs) for matched serum samples at 3 concentrations of the respective measurand was evaluated using a multiple day precision study. In this precision study, serum with a low concentration of measurand from a single
{10}
donor was spiked to a high concentration of measurand with commercially-available analyte and then the remaining sample prepared by intermixing. The study was performed measuring each sample 2 times per run, 2 runs per day for 5-6 days using 1 reagent lot for a total of 20 measurements per sample on the ACE Alera analyzer. The results of the precision study are presented in the table below.
# Albumin
| | Experimental Result | | | |
| --- | --- | --- | --- | --- |
| | | | SD (g/dL) or %CV | |
| Lab | Sample | Mean | Within-Run | Total |
| In-House | 1 | 3.5 | 0.02 | 0.04 |
| | | | 0.6% | 1.1% |
| POL 1 | 1 | 3.5 | 0.05 | 0.06 |
| | | | 1.4% | 1.7% |
| POL 2 | 1 | 3.6 | 0.05 | 0.05 |
| | | | 1.4% | 1.5% |
| POL 3 | 1 | 3.5 | 0.05 | 0.05 |
| | | | 1.6% | 1.6% |
| | | | | |
| In-House | 2 | 5.0 | 0.05 | 0.05 |
| | | | 1.0% | 1.1% |
| POL 1 | 2 | 5.0 | 0.08 | 0.09 |
| | | | 1.7% | 1.9% |
| POL 2 | 2 | 5.0 | 0.06 | 0.08 |
| | | | 1.2% | 1.6% |
| POL 3 | 2 | 4.9 | 0.03 | 0.03 |
| | | | 0.6% | 0.7% |
| | | | | |
| In-House | 3 | 6.2 | 0.06 | 0.07 |
| | | | 1.0% | 1.1% |
| POL 1 | 3 | 6.2 | 0.07 | 0.10 |
| | | | 1.1% | 1.6% |
| POL 2 | 3 | 6.2 | 0.06 | 0.07 |
| | | | 1.0% | 1.1% |
| POL 3 | 3 | 6.1 | 0.08 | 0.08 |
| | | | 1.3% | 1.4% |
{11}
Total Protein
| | Experimental Result | | | |
| --- | --- | --- | --- | --- |
| | | | SD (g/dL) or %CV | |
| Lab | Sample | Mean | Within-Run | Total |
| In-House | 1 | 5.3 | 0.08 | 0.10 |
| | | | 1.5% | 1.8% |
| POL 1 | 1 | 5.5 | 0.07 | 0.10 |
| | | | 1.4% | 1.8% |
| POL 2 | 1 | 5.2 | 0.07 | 0.15 |
| | | | 1.3% | 2.8% |
| POL 3 | 1 | 5.3 | 0.07 | 0.12 |
| | | | 1.4% | 2.2% |
| | | | | |
| In-House | 2 | 8.3 | 0.10 | 0.11 |
| | | | 1.2% | 1.4% |
| POL 1 | 2 | 8.4 | 0.09 | 0.10 |
| | | | 1.1% | 1.2% |
| POL 2 | 2 | 8.4 | 0.10 | 0.11 |
| | | | 1.2% | 1.4% |
| POL 3 | 2 | 8.2 | 0.09 | 0.14 |
| | | | 1.1% | 1.7% |
| | | | | |
| In-House | 3 | 11.3 | 0.14 | 0.15 |
| | | | 1.3% | 1.4% |
| POL 1 | 3 | 11.3 | 0.14 | 0.14 |
| | | | 1.2% | 1.2% |
| POL 2 | 3 | 11.5 | 0.09 | 0.16 |
| | | | 0.8% | 1.4% |
| POL 3 | 3 | 11.1 | 0.26 | 0.31 |
| | | | 2.3% | 2.8% |
12
{12}
13
# Calcium
| | Experimental Result | | | |
| --- | --- | --- | --- | --- |
| | | | SD (mg/dL) or %CV | |
| Lab | Sample | Mean | Within-Run | Total |
| In-House | 1 | 6.9 | 0.08 | 0.15 |
| | | | 1.2% | 2.1% |
| POL 1 | 1 | 6.9 | 0.07 | 0.19 |
| | | | 1.0% | 2.7% |
| POL 2 | 1 | 7.0 | 0.19 | 0.19 |
| | | | 2.7% | 2.7% |
| POL 3 | 1 | 7.0 | 0.14 | 0.14 |
| | | | 1.9% | 1.9% |
| | | | | |
| In-House | 2 | 10.5 | 0.05 | 0.06 |
| | | | 0.5% | 0.6% |
| POL 1 | 2 | 10.5 | 0.09 | 0.33 |
| | | | 0.9% | 3.2% |
| POL 2 | 2 | 10.6 | 0.21 | 0.22 |
| | | | 1.9% | 2.1% |
| POL 3 | 2 | 10.6 | 0.16 | 0.16 |
| | | | 1.5% | 1.5% |
| | | | | |
| In-House | 3 | 13.5 | 0.17 | 0.20 |
| | | | 1.3% | 1.5% |
| POL 1 | 3 | 13.4 | 0.14 | 0.34 |
| | | | 1.1% | 2.5% |
| POL 2 | 3 | 13.6 | 0.21 | 0.23 |
| | | | 1.5% | 1.7% |
| POL 3 | 3 | 13.6 | 0.14 | 0.18 |
| | | | 1.0% | 1.3% |
{13}
Phosphorus
| | Experimental Result | | | |
| --- | --- | --- | --- | --- |
| | | | SD (mg/dL) or %CV | |
| Lab | Sample | Mean | Within- | Total |
| In-House | 1 | 2.8 | 0.06 | 0.06 |
| | | | 2.1% | 2.1% |
| POL 1 | 1 | 2.7 | 0.04 | 0.10 |
| | | | 1.4% | 3.8% |
| POL 2 | 1 | 2.5 | 0.02 | 0.11 |
| | | | 0.9% | 4.4% |
| POL 3 | 1 | 2.9 | 0.05 | 0.07 |
| | | | 1.9% | 2.4% |
| | | | | |
| In-House | 2 | 7.1 | 0.07 | 0.09 |
| | | | 0.9% | 1.3% |
| POL 1 | 2 | 7.1 | 0.07 | 0.18 |
| | | | 0.9% | 2.5% |
| POL 2 | 2 | 6.7 | 0.07 | 0.22 |
| | | | 1.1% | 3.2% |
| POL 3 | 2 | 7.4 | 0.10 | 0.13 |
| | | | 1.4% | 1.7% |
| | | | | |
| In-House | 3 | 11.3 | 0.09 | 0.11 |
| | | | 0.8% | 0.9% |
| POL 1 | 3 | 11.3 | 0.16 | 0.27 |
| | | | 1.4% | 2.4% |
| POL 2 | 3 | 10.6 | 0.15 | 0.21 |
| | | | 1.4% | 1.9% |
| POL 3 | 3 | 11.7 | 0.11 | 0.14 |
| | | | 0.9% | 1.2% |
b. Linearity/assay reportable range:
The linearity of the ACE Alera analyzer was assessed in one run using one lot of reagents with each sample tested in triplicate. Six diluted serum samples with
{14}
measuring concentrations evenly distributed were prepared by diluting a high measured concentration serum pool. This yielded linearity samples with levels that spanned the measuring range of each of the four analytes (Albumin, Total Protein, Calcium and Phosphorus) measured.
The sponsor calculated linear and polynomial regressions from mean observed values versus expected values using an unweighted regression model. The linear regressions between the expected values and the measured values are found in the table below:
| Measurand | Range tested | Slope | Intercept | r² |
| --- | --- | --- | --- | --- |
| Albumin (g/dL) | 0.1-7.6 | 0.980 | 0.01 | 0.9982 |
| Total Protein (g/dL) | 0.2-15.1 | 0.991 | 0.04 | 0.9979 |
| Calcium (mg/dL) | 0.3-16.5 | 0.992 | 0.27 | 0.990 |
| Phosphorus (mg/dL) | 0.2-21 mg/dL | 1.001 | 0.03 | 0.9995 |
The linearity supports the sponsor's claimed reportable range of the Albumin assay is 0.3-7.0 g/dL, Total Protein assay is 0.4-14.0 g/dL, Calcium assay is 0.4-15.0 mg/dL and Phosphorus 0.4-20.0 mg/dL
Auto-dilution study:
Sponsor performed an auto-dilution study to confirm the auto-dilution function (1:3) on the ACE Alera analyzer for each measurand for plasma and serum samples by comparing the auto-diluted results to the manual dilution results. Sample recovery was within +/- 10% bias for all samples tested.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
| Analyte | Traceability |
| --- | --- |
| Albumin | NIST SRM 927 |
| Total Protein | NIST SRM 927 |
| Calcium | NIST SRM 909, NIST SRM 915b |
| Phosphorus | NIST SRM 200, NIST SRM 3139a |
The Albumin, Total Protein, Calcium and Phosphorus calibrators have been previously cleared in k930104.
d. Detection limit:
{15}
LoB, LoD and LoQ studies for each measurand were evaluated based upon CLSI EP-17A using the ACE Alera analyzer.
## LoB Test Protocol
Blank samples (n=5) were measured in replicates of 4 for 3 days for a total of n=60 measurements.
## LoD Protocol
Low serum-based samples (n=5) and blank samples (n=5) were measured in replicates of 4 per day for 3 days for a total of n=60 measurements
## LoQ Protocol
LoQ was determined by evaluating 5 low level samples with multiple replicates over multiple days for a total of 40 measurements per sample. LoQ values are based on inter-assay precision of <20% CV.
| | Albumin (g/dL) | Total Protein (g/dL) | Calcium (mg/dL) | Inorganic Phosphorus (mg/dL) |
| --- | --- | --- | --- | --- |
| LoB | 0.08 | 0.08 | 0.09 | 0.25 |
| LoD | 0.09 | 0.13 | 0.11 | 0.35 |
| LoQ | 0.09 | 0.20 | 0.23 | 0.35 |
The sponsor claimed the following measuring ranges:
| Measurand | Assay Range |
| --- | --- |
| Albumin (g/dL) | 0.3-7.0 |
| Total Protein (g/dL) | 0.4-14.0 |
| Calcium (mg/dL) | 0.4-15.0 |
| Phosphorus (mg/dL) | 0.4-20.0 |
## e. Analytical specificity:
Interference testing was performed according to CLSI-EP07 to determine whether the presence of hemoglobin, triglycerides (Intralipid), icterus (bilirubin), and ascorbic acid may interfere with results for each measurand using the ACE Alera analyzer.
Human serum samples were used with two different concentrations (normal and abnormal) of measurand (Albumin, Total Protein, Calcium and Phosphorus). Each sample level was spiked with increasing amounts of interferent for a total of 7 samples with interferent. Control samples at each level which were not spiked were
16
{16}
used to compare with spiked samples. The spiked and unspiked samples were tested and used to calculate % recovery (measured concentration compared to measured concentration with zero interferent). Measurand recovery of <10% of control value was defined as non-significant interference. The results of the highest concentration tested without significant interference are summarized in the table below.
| Measurand | Icterus (mg/dL) | Hemolysis (mg/dL) | Lipemia (mg/dL) | Ascorbic Acid (mg/dL) |
| --- | --- | --- | --- | --- |
| Albumin | 57 | 250 | 929 | 6 |
| Total Protein | 56.8 | 250 | 929 | 6 |
| Calcium | 58.8 | 1000 | 1000 | 6 |
| Phosphorus | 11.5 | 250 | 306 | 6 |
The following limitations are included in the labeling for Albumin and Total Protein "Do not use hemolyzed samples".
The following limitations are included in the labeling for Phosphorus "Use clear, unhemolyzed serum or lithium heparin plasma.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
**Serum Method Comparison:**
Method comparison studies were conducted at 3 POC sites and in house following CLSI EP9-A2. Samples were analyzed on the ACE Clinical Chemistry System at Alfa Wassermann and the results were compared against those gathered on the ACE Alera Clinical Chemistry Systems at 3 Physician Office Labs (POL). At least 50 determinations were made in singlicate for serum samples drawn from the same individuals on the same platform. To test across the reportable range of the assay, additional sets (<7 sets) of altered samples (spikes or diluted) were used to cover the full measuring range of each analyte. Results are summarized in the table below.
{17}
| Albumin | ACE (in-house) vs. ACE Alera (POL) | | |
| --- | --- | --- | --- |
| | POL 1 | POL 2 | POL 3 |
| N | 50 | 50 | 50 |
| Range (g/dL) | 1.0-6.4 | 1.0-6.4 | 1.0-6.4 |
| Regression Equation | y=1.004x-0.03 | y=1.005x-0.05 | y=0.982x+0.01 |
| R² | 0.9949 | 0.9960 | 0.9967 |
| Total Protein | ACE (in-house) vs. ACE Alera (POL) | | |
| --- | --- | --- | --- |
| | POL 1 | POL 2 | POL 3 |
| N | 51 | 51 | 51 |
| Range (g/dL) | 0.9-13.6 | 0.9-13.6 | 0.9-13.6 |
| Regression Equation | y=0.99x+0.16 | y=1.027x-0.06 | y=0.979x+0.24 |
| R² | 0.9969 | 0.9962 | 0.9964 |
| Calcium | ACE (in-house) vs. ACE Alera (POL) | | |
| --- | --- | --- | --- |
| | POL 1 | POL 2 | POL 3 |
| N | 50 | 50 | 50 |
| Range (mg/dL) | 1.9-13.7 | 1.9-13.7 | 1.9-13.7 |
| Regression Equation | y=0.992x-0.09 | y=1.007x-0.11 | y=1.008x-0.08 |
| R² | 0.9904 | 0.9929 | 0.9929 |
| Phosphorus | ACE (in-house) vs. ACE Alera (POL) | | |
| --- | --- | --- | --- |
| | POL 1 | POL 2 | POL 3 |
| N | 50 | 50 | 50 |
| Range (mg/dL) | 1.0-18.4 | 1.0-18.4 | 1.0-18.4 |
| Regression Equation | y=1.015x+0.14 | y=0.960x+0.12 | y=0.984x+0.05 |
| R² | 0.9992 | 0.9986 | 0.9991 |
# b. Matrix comparison
A matrix comparison was performed using paired serum/lithium heparin plasma samples. The matrix comparison study was performed for each measurand using paired samples on the ACE, ACE Alera and ACE Axcel analyzers.
| Measurand | | ACE | ACE Alera | ACE Axcel |
| --- | --- | --- | --- | --- |
| Albumin (g/dL) | N | 55 | 56 | 56 |
| | Range | 0.3-6.8 | 0.3-6.8 | 0.7-6.7 |
| | Slope | 0.991 | 1.002 | 0.956 |
| | Intercept | 0.03 | -0.01 | 0.20 |
| | r² | 0.987 | 0.991 | 0.985 |
{18}
| Total Protein (g/dL) | N | 56 | 56 | 81 |
| --- | --- | --- | --- | --- |
| | Range | 0.5-12.3 | 0.5-12.0 | 0.5-13.9 |
| | Slope | 1.001 | 0.999 | 0.994 |
| | Intercept | 0.12 | 0.14 | 0.34 |
| | r² | 0.979 | 0.984 | 0.989 |
| Calcium (mg/dL) | N | 56 | 56 | 81 |
| | Range | 1.0-13.7 | 1.0-13.7 | 0.7-15.0 |
| | Slope | 1.006 | 1.008 | 0.978 |
| | Intercept | -0.01 | -0.06 | 0.33 |
| | r² | 0.982 | 0.979 | 0.991 |
| Phosphorus (mg/dL) | N | 100 | 102 | 56 |
| | Range | 1.3-19.3 | 1.3-19.3 | 0.9-19.8 |
| | Slope | 1.042 | 1.049 | 0.999 |
| | Intercept | -0.26 | -0.28 | 0.04 |
| | r² | 0.993 | 0.993 | 0.995 |
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Expected values are based on the literature reference:
$^{1}$ Wu. A.H.B. (Ed), Tietz Clinical Guide to Laboratory Tests, 4th edition, Saunders Elsevier, St. Louis, MO (2006).
$^{2}$ Burtis, C.A., Ashwood, E.R. (Eds.). Tietz Fundamentals of Clinical Chemistry, 4th Edition, W.B. Saunders Co., Philadelphia, PA (1996).
$^{3}$ Heil W. Koberstein R. Zawta B. Reference ranges for Adults and Children, Pre-Analytical Considerations, 6th ed. 1999.
{19}
20
1. Albumin: 5.5-5.2 g/dL
2. Total Protein: 6.4-8.3 g/dL (Ambulatory); 6.0-7.8 g/dL (Recumbent)
3. Calcium: 8.5-10.2 mg/dL
4. Phosphorus: 2.7-4.5 mg/dL
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.