The AIRO™ is intended to be used for non-contrast X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients weighing over 400 lbs (182 kg). Airo is contraindicated as the principal means of guidance during surgical procedures. The physician must verify navigation accuracy using known anatomical landmarks or an equivalent verification method when used in surgical procedures.
Device Story
Mobile high-resolution CT system for intra-operative and clinical use. Inputs: X-ray irradiation of patient anatomy. Mechanism: Rotating gantry with solid-state X-ray generator and CdWO4/photodiode detector array; helical scanning; exact filtered-back projection reconstruction algorithm. Outputs: CT images for clinical review. Operation: Motorized mobile base; battery-powered; operated by clinical staff in OR, ICU, ED, or radiology settings. Healthcare providers use images for diagnostic purposes; surgical navigation requires independent verification of accuracy. Benefits: Provides high-resolution imaging in mobile, intra-operative environments.
Clinical Evidence
Non-Significant Risk (NSR) study conducted to evaluate diagnostic image quality. Images were independently and subjectively reviewed by clinicians and determined to be of diagnostic quality. No quantitative clinical metrics (sensitivity/specificity) reported. Bench testing included image performance, radiation safety, electrical/mechanical safety, and EMI/EMC testing.
Indicated for non-contrast X-ray CT imaging of anatomy fitting within 107cm aperture. Excludes pediatric patients and patients >400 lbs (182 kg). Contraindicated as primary surgical guidance tool; requires physician verification of navigation accuracy.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K131431
Page 1 of 6
# MOBIUS
# 510(k) Summary
This 510(k) Summary of Safety and Effectiveness information is prepared in accordance with the requirements of 21 CFR Part 807.92.
#### 1. Submitter:
Mobius Imaging, LLC 323 West Main Street Ayer, MA 01432 USA
Contact: Norma LeMay Director of Regulatory Affairs
Telephone: 978-391-9644 Fax: 978-772-3441 Email: nlemay@mobiusimaging.com
Date of Preparation: August 26, 2013
#### Device Name & Regulatory Classification: 2.
Proprietary or Trade Name: AIRO™ CT System Classification Name: Computed Tomography X-ray System Product Code: 90 JAK Device Classification: Class II Regulation Number: 21 CFR 892.1750
### 3. Predicate Device(s):
The legally marketed device to which substantial equivalence is being claimed is as follows:
- Neurologica NL4000 BodyTom™ CT System Premarket Notification K102677 (cleared � on March 24, 2011, product code 90 JAK)
### 4. Device Description:
The Mobius Airo is a mobile high resolution Computed Tomography (CT) System similar to the NL4000 BodyTom CT System. The Airo has a large-diameter bore designed for intra-operative use; the main features include a 107cm bore, with a 51.2cm field of view (FOV). The Airo has two modes of operation; transport and scanning mode, translation along the longitudinal axis is achieved through movement of the gantry along the length of the system base (rather than through movement of the patient support table).
SEP 26 2013
{1}------------------------------------------------
The lightweight translating gantry consists of a rotating disk with a solid state x-ray generator, solid state detector array (that includes detector modules that consist of a layered Cadmium Tungstate (CdWO4) and Photodiode Array). Each detector module includes a 32 x 16 pixel scintillator array that produces scintillation events responsive to irradiation by X-rays. The Airo also includes a collimator, control computer, communications link, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation), and a DC brushless servo drive system (translation).
The power system consists of batteries which provide system power while unplugged from a standard power outlet (e.g., during transport of the System and also during scanning). The base has retractable rotating caster wheels and electrical drive system can be easily moved to different locations.
In addition, the System has the necessary safety features such as emergency stop button, X-ray indicators, interlocks, patient alignment lasers, and 110 percent X-ray timer. The software helical reconstruction is based on an exact filtered-back projection algorithm.
### 5. Indications for Use:
The AIRO™ is intended to be used for non-contrast X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients weighing over 400 lbs (182 kg).
Airo is contraindicated as the principal means of guidance during surgical procedures. The physician must verify navigation accuracy using known anatomical landmarks or an equivalent verification method when used in surgical procedures.
### 6. Comparison of Technological Characteristics with the Predicate Device:
As detailed in Section 12 of this 510(k) Premarket Notification, the Airo CT System, for its intended use, is of comparable type in design, material, functionality, technology and is considered substantially equivalent to the NeuroLogica BodyTom™ CT System (K102677) based on the following comparison. Most importantly, the differences noted below raise no new issues of safety or effectiveness based on all testing performed and presented in this 510(k) submission.
| Model Name | Airo™ CT System | NL 4000 BodyTom™ (K102677) |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Characteristics | | |
| Indications for Use | The AIRO™ is intended to be used for non-<br>contrast X-ray computed tomography<br>applications for anatomy that can be<br>imaged in the 107cm aperture excluding<br>pediatric patients and patients weighing<br>over 400 lbs (182 kg). Airo is<br>contraindicated as the principal means of<br>guidance during surgical procedures. The<br>physician must verify navigation accuracy<br>using known anatomical landmarks or an | The NL4000 BodyTom™ is intended to be<br>used for x-ray computed tomography<br>applications for anatomy that can be imaged<br>in the 85cm aperture |
| Model Name | Airo™ CT System | NL 4000 BodyTom™ (K102677) |
| Technological Characteristics | equivalent verification method when used<br>in surgical procedures. | |
| Aperture (cm) | 107 | 85 |
| Image Field of View (cm) | 51.2 | 60 |
| Detector Material | Solid State CdWO4 | Solid State CdWO4 |
| Detector Configuration | 32 x 2.0mm | 32 x 1.25mm |
| Spatial Resolution for Sharpest<br>Clinical Algorithm (lp/cm at 2%) | 6.9 | 12.0 |
| X-ray Tube Type | Rotating Anode | Rotating Anode |
| Heat Storage (MHU) | 2.3 | 3.5 and 5.0 |
| X-ray Tube Cooling | Liquid (50% Water, 50% Propylene Glycol) | Oil |
| X-ray Fan Angle (deg) | 45 | 54 |
| Max X-ray Power (kW) | 32 | 42 |
| Rotating Speed (seconds) | 2 | 1, 2 |
| Gantry Weight (kg) | 1000 (approx) | 1200 |
| Transfer of electric current | Data Dock system | Slip ring |
| Mechanism to translate Gantry | Rails on Mobile Base System | On floor treads |
| Mobile | Yes (motorized) | Yes (motorized) |
| Wheels (casters) | Wheels (3 inch) | Wheels (6 inch) |
| Input Voltage | 1 phase 100-240 volt | 1 phase 110-240 volt |
| PACS/DICOM 3.0 | Yes | Yes |
| 2D Scout | Yes | Yes |
| Bolus tracking | No | Yes |
| Dynamic Scan | No | Yes |
| Axial/Helical | Helical | Both |
| MPR | Yes | Yes |
| 3D Viewing | No | Yes |
| Patient Table | Yes (Trumpf table column integrated with<br>base) | Optional, not required |
| Scan Motion | Scanner Moves | Scanner Moves |
| Model Name | Airo™ CT System | NL 4000 BodyTom™ (K102677) |
| *Technological Characteristics | | |
| Laser Alignment | Patient Alignment | Patient Alignment |
| X-ray warning light | Yes | Yes |
| 110% X-ray Timer | Yes | Yes |
| Emergency Stop | Yes | Yes |
| Operator X-ray On Switch | Yes | Yes |
| Quality Test Phantom | Yes | Yes |
| Login ID/password | Yes | Yes |
| Administrator Privileges | Yes | Yes |
| Dose Display | Yes | Yes |
| Dose Report/Audit | Yes | Yes |
| Protocol Override Protection | Yes | Yes |
| Protocols by weight/body region | Yes | Yes (includes Age) |
| QA Test Report | Yes/Third Party Software | Yes |
| Operating System | Microsoft Windows | Microsoft Windows |
| Where Used | Mobile or Fixed Radiology, ICU, ED,<br>Surgical, Clinic, Office | Mobile or Fixed Radiology, ICU, ED, Surgical,<br>Interventional, Clinic, Office |
#### Page 2 of 6
{2}------------------------------------------------
10
.
{3}------------------------------------------------
### 7. General Safety and Effectiveness Concerns:
Identical to its predicate device, all components of the Airo CT System are subject to Federal Diagnostic Equipment Performance Standard and applicable regulations of 21 CFR Part 1020.30 and 1020.33 and will be certified to meet those requirements. An Initial Report will be filed with the Center for Devices and Radiological Health (CDRH) according to 21 CFR 1002.10 prior to commercialization of the Airo CT System, respectively.
To minimize electrical, mechanical and radiation hazards, Mobius adheres to recognized and established industry practices. Additionally, the Airo CT System was designed and tested to the following FDA recognized International harmonized standards:
- . 1EC 60601-1 Issued: 2005/01/01 Ed:3 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2 Issued: 2007/03/01 Ed:3 Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Perfórmance – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
{4}------------------------------------------------
- . IEC 60601-1-3 Issued: 2008/01/22 Ed:2 Medical Elec. Equipment - P. 1: General Req. for Safety 3. Collateral Standard: General Req. for Radiation Protection in Diagnostic X-Ray Equipment
- IEC 62366 Issued: 2007/10 Ed:1, Medical devices Application of usability engineering . to medical devices
- . IEC 60601-2-44 Issued: 2009/02/25 Ed:3 Medical Electrical Equipment - Part 2-44: Particular Requirements for the basic safety and essential performance of X-ray Equipment for computed Tomography
- . IEC 61223:2004 – Evaluation and routine testing in medical imaging departments - Part 3-5 Acceptance Tests - Imaging Performance of Computed Tomography X-ray equipment
- IEC 60825-1:2007 Safety of laser products .
- IEC 62133:2002: Safety requirements for portable sealed secondary cells, and for . batteries made from them, for use in portable applications
- . UL 1642:1995: Standard for Lithium Batteries
Relating to concerns regarding unintentional radiation exposure, the Airo CT System, like the NL4000 BodyTom System, has software safeguards such as: ID password/login, dose display/reporting, safety warning to prevent excessive dose, protocol protection and required quality assurance testing.
- Determination of Substantial Equivalence: 8.
### Summary of Non-clinical tests:
The Airo CT System complies with the voluntary harmonized standards as detailed above and in Section 9 and 17 of this 510(k) Premarket Notification. In addition, the following quality assurance measures were applied to the development of the System:
- . Risk Analysis
- . Design Reviews
- . Design Verification Testing
- . SW Unit Integration Testing
- . System Software Verification & Validation Testing
- . Image Performance & Radiation Safety Testing
- . Electrical Safety, Mechanical & Stability Testing
- . EMI/EMC Testing
{5}------------------------------------------------
K131431
Page book 6
the
#### Summary of Clinical Tests:
A Non-Significant Risk (NSR) study was performed to determine the diagnostic image quality of the Airo CT System for typical intended use anatomy. Once collected, all images were independently and subjectively reviewed and found to be of diagnostic image quality. All images, as well as the final report, are provided in Section 20 of this 510(k) submission.
The results of all testing performed indicate that the Airo CT System meets the acceptance criteria and is substantially equivalent to the currently cleared predicate device (Neurologica BodyTom CT System).
### 9. Conclusion:
Based upon the above considerations, including all testing presented in this 510(k) submission, Mobius considers the Airo CT System to be as safe, as effective, and performance is substantially equivalent to its predicate device. We also believe the Airo CT System raises no new issues of safety and/or efficacy and the device performs as intended.
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 26. 2013
Mobius Imaging. LLC % Ms. Norma J. LeMay Director of Regulatory Affairs 323 West Main Street AYER MA 01432
Re: K131431
Trade/Device Name: AIRO" Computed Tomography (CT) X-ray System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: August 26, 2013 Received: August 27, 2013
Dear Ms. LeMay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2-Ms. LeMay
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### Indications for Use
510(k) Number (if known): K131431
AIROTM Computed Tomography X-ray System Device Name:
Indications for Use:
The AIRO™ is intended to be used for non-contrast X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients weighing over 400 lbs (182 kg).
Airo is contraindicated as the principal means of guidance during surgical procedures. The physician must verify navigation accuracy using known anatomical landmarks or an equivalent verification method when used in surgical procedures.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Archibald D. Hixon
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k) K131431
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.