K102677 · Neurologica Corporation · JAK · Mar 24, 2011 · Radiology
Device Facts
Record ID
K102677
Device Name
BODYTOM NL 4000
Applicant
Neurologica Corporation
Product Code
JAK · Radiology
Decision Date
Mar 24, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The NL4000 BodyTom is intended to be used for x-ray Computed Tomography applications for anatomy that can be imaged in the 85cm aperture.
Device Story
BodyTom (NL4000) is a mobile, high-resolution, multi-row, 85cm bore x-ray computed tomography (CT) system. Input: x-ray attenuation data collected via rotating gantry containing solid-state x-ray generator and detector array. Transformation: system uses reconstruction computer to process raw data into tomographic images. Output: 2D, multi-planar reconstruction (MPR), and 3D images displayed on workstation and archived via DICOM/PACS. Used in clinical settings (radiology, ICU, ED, surgical, interventional) by qualified healthcare professionals. Features motorized transport wheels and battery-powered operation (LiPO batteries) for mobility. Includes radiation safeguards (dose reporting, lockout, 110% timer) and quality assurance software. Benefits include portable diagnostic imaging capability for patients requiring CT scans outside traditional radiology suites.
Clinical Evidence
Bench testing only. No clinical data provided. System verified against recognized safety standards (UL60601-1, IEC 60601-1, EN 60601-1, IEC 60601-2-44). Battery system validated using 50kW passive load simulation under severe duty cycles.
Indicated for x-ray computed tomography imaging of any anatomy fitting within the 85cm aperture. Intended for prescription use.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K102677 Page 1 of 13
Image /page/0/Picture/1 description: The image shows a logo with the word "NeuroLogica" written in a bold, sans-serif font. The "O" in "Logica" is stylized as a circle with a dot in the center. Behind the text is a stylized drawing of a brain, with an elliptical shape encircling it. The overall design appears to be a logo for a company or organization related to neurology or logic.
MAR 2 4 2011
## 510(k) SUMMARY For NeuroLogica Corporation NL4000 BodyTom™ Computed Tomography System
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h)
| (1) Submitter: | NeuroLogica Corporation 14 Electronics Avenue Danvers, MA, 01923 |
|----------------|--------------------------------------------------------------------|
|----------------|--------------------------------------------------------------------|
## Establishment Registration number: FDA #3004938766
Contact person: Mr. Donald D. Fickett Vice President of Regulatory Affairs and Quality Systems Telephone 978-564-8523 FAX 978-564-8677 e-mail: dfickett@neurologica.com
Date this summary was prepared: September 9, 2010
#### (2) Device Name:
Proprietary or Trade Name: BodyTom
Device Model: NL4000
Classification Name: Computed Tomography X-ray System
Product code: 90JAK
Device classification: Class II
Regulation number: 21 CFR 892.1750
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo for "NeuroLogica". The logo features a stylized drawing of a brain with an orbit around it. The word "NeuroLogica" is written in a simple font below the brain drawing. The "O" in "Logica" is stylized as a circle with a dot in the center.
### (3) Predicate device:
The legally marketed device to which substantial equivalence is being claimed is as follows:
- NeuroLogica Corporation, NL3000 CereTom™ CT per Pre-Market . Notification Submission K051765
- Hitachi Medical Systems America, Inc. ELCOS16™ per Pre-Market . Notification Submission K071806
- NeuroLogica Corporation, inSPira HD™ per Pre-Market Notification � Submission K090811
#### (4) Device Description:
The NL4000 BodyTom is essentially a larger bore version of our predicate NL3000 CereTom CT system. It is a high resolution, multi row, 85 cm bore, 60cm field of view, x-ray computed tomography system. The lightweight translating gantry consists of a rotating disk with a solid state x-ray generator, solid state detector array, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation), and stepper drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system software is based upon the CereTom software, however, does not contain application specific software for contrast imaging. In addition, the system has the necessary safety features such as emergency stop switch, x-ray indicators, interlocks, patient alignment laser, and 110 percent x-ray timer. The gantry has retractable rotating caster wheels and electrical drive system so the system can be moved easily to different locations.
#### (5) Intended Use:
The NL4000 BodyTom is intended to be used for non-contrast x-ray computed tomography applications for anatomy that can be imaged in the 85cm aperture.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo for a company called "NeuroLogica". The logo features a stylized drawing of a brain with a ring around it, suggesting a planet. The text "NeuroLogica" is written in a simple, sans-serif font, with the "O" in "Logica" replaced by a circle with a dot in the center, resembling an eye or a target.
### (6) Comparison of Technological Characteristics with the predicate device:
NeuroLogica Corporation's NL4000 Computed Tomography System, for its intended use, is of comparable type in design, material, functionality, technology and is substantially equivalent to the following cleared predicate dey, ces: NeuroLogica Corporation NL3000 CereTom™ (K051765), Hitachi ELCOS16TM (K071806) and NeuroLogica Corporation inSPira HD™ (K090811) includine
- Material: The BodyTom uses similar material to the above listed o scanners such as solid state detectors, x-ray generator, slip ring, data acquisition ICs, rotational bearing, and motion control systems,
- Design: The BodyTom is similar in general design principle to the o above listed scanners except it does not contain application specific software for contrast imaging.
## 7) General Safety and Effectiveness Concerns:
All components of the NL4000 BodyTom system subject to Federal Diagnostic Equipment Performance Standard and applicable regulations of 21 CFR Part 1020.30 and 1020.33 are certified to meet those requirements.
An initial report as per 21CFR Part 1002.10 will be filed with the Center for Device and Radiological Health (CDRH).
To minimize electrical, mechanical and radiation hazards, Neurologica adheres to recognized and established industry practices. The NL4000 BodyTom system is designed to meet UL60601-1, IEC 60601-1 and EN 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60101-2-28, and IEC 60101-2-32 and IEC 60601-2-44.
Relating to concerns per unnecessary radiation exposure, the BodyTom, like the CereTom, has software safeguards such as: security scanner ID/password log-in, dose display/reporting, lockout to prevent excessive dose, protocol protection, and quality assurance.
{3}------------------------------------------------
# Substantial Equivalence Comparison
The NeuroLogica NL4000 BodyTom™ Computed tomography system has similarities/differences to the NeuroLogica NL 3000 CereTom® Computed Tomography system (K051765) as follows:
Similarities
- 1. Diagnostic imaging for Computed Tomography applications
- 2. Same Detector fabrication
- 3. Same Detector measurement electronics
- 4. Same mechanism to translate system
- 5. Similar caster wheels for transport
- 6. Same Laser alignment
- 7. Similar workstation
- 8. Computers and algorithms to compute tomographic images
- 9. Applications: Scout, Axial, Helical, Dynamic, contrast enhanced
- 10. Display(DICOM 3.0) and connectivity(PACS) capability
- 11. Emergency Stop switch
- 12. Slipring to transmit electrical power to rotating disk
- 13. Mobility via transport wheels
- 14. Battery system
- 15. Wireless transfer of Image Data
- 16. 2D, MPR, 3D Viewing & archiving
- 17. Quality assurance phantom and test software
- 18. Radiation Safeguards Hardware (110% timer, warning light, etc)
- 19. Radiation Safeguards Software (login, excessive dose lockout, dose reporting,etc)
- 20. Same Wireless communication with Workstation
- 21. Same Manufacturing Quality System
- 22. Same Design Control System
- 23. Same Software Development methodology
- 24. Same Validation and Verification methods
## Differences:
- Size of Bore/Field of View ( CereTom = 32/25 cm, BodyTom = 85/60 cm) 1.
- Applicable software modifications to accommodate larger Bore 2.
- Motorized transport system on BodyTom 3.
- 4. X-ray tube type (Rotating Anode(BodyTom) vs. fixed Anode (CereTom))
- ഗ് Type of Batteries (SLA vs. LiPO)
{4}------------------------------------------------
The NeuroLogica NL4000 BodyTom™ Computed tomography system has similarities/differences to the Hitachi ECLOS16 Computed Tomography system (K071806) as follows:
K102677 Page 5 of 13
### Similarities
- 1. Diagnostic imaging for Computed Tomography applications
- 2. Similar Multi slice Detector fabrication
- 3. Similar Detector measurement electronics
- 4. Laser alignment
- 5. Similar workstation with Windows Operating System
- 6. Computers and algorithms to compute tomographic images
- 7. Applications: Scout, Axial, Helical, Dynamic, contrast enhanced
- Display(DICOM 3.0) and connectivity(PACS) capability 8.
- 9. Emergency Stop switch
- 10. Slip ring to transmit electrical power to rotating disk
- 11. 2D, MPR, 3D Viewing & archiving
- 12. Quality assurance phantom and test software
- 13. Radiation Safeguards Hardware (110% timer, warning light, etc)
- 14. Rotating Anode X-ray Tube
- 15. Exactly the same X-ray Power (42 kilowatt)
- 16. Whole Body sized bore {BodyTom = 85, ECLOS = 70cm}
### Differences:
- 1. Motorized transport system on BodyTom
- ECLOS has a moving patient table, BodyTom has translate mechanism to move on floor 2.
- BodyTom is battery Powered 3.
- 4. Caster Wheels for mobility on BodyTom, ECLOS is fixed mount
- 5. Number of Slices (ECLOS = 16, BodyTom = 32)
- 6. BodyTom uses wireless link to communicate with workstation
{5}------------------------------------------------
The NeuroLogica NL4000 BodyTom™ Computed tomography system has K102677 Page 6 of 13 similarities/differences to the NeuroLogica NL1000 inSPira® SPECT system (K090811) as follows:
## Similarities
- 1. Same mechanism to translate system
- Similar caster wheels for transport 2.
- 3. Same Laser alignment
- 4. Similar workstation
- 5. Display(DICOM 3.0) and connectivity(PACS) capability
- 6. Emergency Stop switch
- 7. Slipring to transmit electrical power to rotating disk
- 8. Mobility via transport wheels
- 9. Battery system
- 10. Wireless transfer of Image Data
- 11. 2D, MPR, 3D Viewing & archiving
- 12. Same Manufacturing Quality System
- 13. Same Design Control System
- 14. Same Software Development methodology
- 15. Same Validation and Verification methods
## Differences:
- 1. InSPira is a SPECT system, BodyTom is a CT system
- 2. Motorized transport system on BodyTom
- 3. Type of Batteries (SLA vs. LiPO)
{6}------------------------------------------------
| Company | NeuroLogica | NeuroLogica | Hitachi | NeuroLogica |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model/Name | NL4000 BodyTom | NL 3000 CereTom | ECLOS16 | NL1000 inSPira |
| Predicate/Submission | Submission | Predicate | Predicate | Predicate |
| 510(k) number | | K051765 | K071806 | K090811 |
| Type of System | Computed<br>tomography | Computed<br>tomography | Computed<br>tomography | Single Photon<br>Emission CT |
| Aperture (cm) | 85 | 32 | 70 | 29 |
| Image Field of View (cm) | 60 | 25 | 50 | 20 |
| Detector material | Solid State CdWO4 | Solid State CdWO4 | Solid State Ceramic | Nal, PMT |
| Detector configuration | 32 x 1.25mm | 8 x 1.25mm | 16 x 1.25 | 3 x 24 |
| MTF at 0% (lp/cm) | 17 | 17 | 17 | N/A |
| Xray Tube type | Rotating Anode | Fixed Anode | Rotating Anode | N/A |
| Heat storage (MHU) | 3.5 and 5.0 | 0.3 | 5 | N/A |
| Cooling | Oil | Air | Oil | N/A |
| Xray fan angle (deg) | 54 | 62 | Unknown | N/A |
| Max X-ray Power (kW) | 42 | 1.2 | 42 | N/A |
| otate Speed (seconds) | 1,2 | 1,2,4,6 | 0.8 | 1,2 |
| Jantry weight, kg | 1200 | 400 | 1320 | 850 |
| Wireless | Yes | Yes | No | Yes |
| Mobile | Yes (Motorized) | Yes (manual) | No | Yes (manual) |
| Battery system | Yes (LiPO) | Yes (SLA) | No | Yes (SLA) |
| Wheels | Wheels (6 inch) | Wheels (3 inch) | No | Wheels (4 inch) |
| Input voltage | 1 phase 110-<br>240Volt | 1 phase 110-<br>240Volt | 208VAC, 3- phase | 1phase 110-<br>240Volt |
| Input power max | 3.6kw | 1.5kw | 75kVA | 1.5kw |
| PACS / DICOM 3.0 | Yes | Yes | Yes | Yes |
| 2D scout | Yes | Yes | Yes | NA |
| bolus tracking | Yes | Yes | Yes | NA |
| Dynamic scan | Yes | Yes | Yes | NA |
| Axial/Helical | Both | Both | Both | NA |
| MPR | Yes | Yes | Yes | NA |
| 3D Viewing | Yes | Yes | Yes | NA |
| Patient Table | Not Required<br>(optional) | Not Required<br>(optional) | Yes | Not Required<br>(optional) |
| Scan motion | Scanner Moves | Scanner Moves | NA | Scanner Moves |
| Company | NeuroLogica | NeuroLogica | Hitachi | NeuroLogica |
| odel/Name | NL4000 BodyTom | NL 3000 CereTom | ECLOS16 | NL1000 inSPira |
| Predicate/Submission | Submission | Predicate | Predicate | Predicate |
| 510(k) number | | K051765 | K071806 | K090811 |
| Radiation Safeguards | | | | |
| Hardware | | | | |
| - X-ray warning light | Yes | Yes | Yes | NA |
| -110% X-ray timer | Yes | Yes | Yes | NA |
| -E-Stop | Yes | Yes | Yes | Yes |
| - Internal lead shield | Yes | Yes | No | NA |
| -external Lead curtains | Yes | Yes | No | NA |
| -Operator x-ray on switch | Yes | Yes | Yes | NA |
| -Quality Test Phantom | Yes | Yes | Yes | NA |
| | | | | |
| Radiation Safeguards | | | | |
| Software | | | | |
| - login ID/password | Yes | Yes | Unknown | Yes |
| -Administrator privileges | Yes | Yes | Unknown | Yes |
| -Dose display | Yes | Yes | Unknown | NA |
| - Dose report/audit | Yes | Yes | Unknown | NA |
| -Protocol override | Yes | Yes | Unknown | NA |
| protection | | | | |
| - Protocols by age/weight | Yes | Yes | Unknown | NA |
| -cessive dose lockout | Yes | Yes | Unknown | NA |
| -QA test report | Yes | Yes | Unknown | NA |
| | | | | |
| Quality Test Phantom | Included | Included | Included | Included |
| Biocompatibility | N/A | N/A | N/A | N/A |
| EM emissions | ETL testing | ETL testing | ETL testing | ETL testing |
| Sterility | N/A | N/A | N/A | N/A |
| Chemical Safety | N/A | N/A | N/A | N/A |
| Thermal Safety | ETL testing | ETL testing | ETL testing | ETL testing |
| IEC EN 60601 Electrical | ETL testing | ETL testing | ETL testing | ETL testing |
| Safety Testing | | | | |
| IEC EN 60601 | ETL testing | ETL testing | ETL testing | ETL testing |
| Mechanical Safety |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.