The intended use of Acoustic MedSystems, Inc. RadVision Dose Planning and Treatment Systems is to provide patient-specific planning, imaging, and implants or applicator device alignment for treating cancer using radioactive seed implants or high dose radiation (HDR) afterloader devices. In addition to planning and treatment, RadVision allows volume and dose calculations. 2D and 3D anatomy and dose visualizations, as well as post treatment seed localization. RadVision is a general purpose brachytherapy planning system used for prospective and confirmation dose calculations for patients undergoing a course of brachytherapy using either temporary or permanent implants of various radioisotopes.
Device Story
RadVision is a brachytherapy dose planning and treatment guidance system used in the OR. It ingests medical images (ultrasound, MR, CT, Fluoroscopy) via a video capture device. Software tools perform contouring, anatomical identification, and dose calculations for LDR and HDR radiation sources. The system facilitates pre-treatment planning, applicator alignment, and post-implant seed localization. Outputs include 2D/3D visualizations and dose distribution reports. Clinicians use these outputs to guide radiation delivery and verify implant placement. The system does not control radiation delivery; it serves as a planning and analysis tool to assist in cancer treatment.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed performance for image acquisition, contouring, anatomical identification, radiation source placement, and dose calculation (LDR/HDR).
Technological Characteristics
Software-based brachytherapy planning system. Compatible with LDR and HDR delivery systems. Supports 2D/3D image import (ultrasound, MR, CT, Fluoroscopy). Complies with IEC 60601-1-1, 1-2, 1-3, and 1-4. Operates on a standard computer with video capture hardware.
Indications for Use
Indicated for cancer patients undergoing brachytherapy using temporary or permanent radioactive seed implants or HDR afterloader devices. Used for pre-treatment planning, applicator alignment, and post-implant assessment.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Burdette Medical Systems, Inc. Interplant (K982696)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Acoustic Medsystems. The logo consists of a stylized eye-like symbol on the left, with the words "ACOUSTIC" above "MEDSYSTEMS" on the right. A horizontal line runs through the center of the eye symbol and extends to the right, underlining the words.
K131428
Page 1 of 3
208 Burwash Avenue Savoy, IL 61874
# JUL 1 6 2013
# Section 5 510(k) Summary
| (1) | Date Summary Prepared: | 04/29/2013<br>Traditional 510(k) Submission | | | |
|-----|------------------------|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|------|-----------------------------------|
| (2) | Submitter: | Acoustic MedSystems, Inc.<br>208 Burwash Avenue<br>Savoy, IL 61874 | | Tel: | 217-239-0900<br>Fax: 217-239-0905 |
| | | Establishment Registration No.: AMS will register following FDA clearance | | | |
| | | Contact Person: | Jennifer Williams<br>FDA Official Regulatory<br>Correspondent<br>e-mail: jwilliams@acousticmed.com | | |
| (3) | Device Name: | RadVision | | | |
| | Common Name: | Brachytherapy Dose Planning and Treatment System | | | |
| | Classification Name: | 21CFR 892.5050 Medical charged particle radiation therapy system, Class II. | | | |
| | Product Code: | MUJ | | | |
| | Proprietary Name: | RadVision | | | |
#### Legally Marketed Predicate Devices: (4)
Burdette Medical Systems, Inc. Interplant, 510(k) number K982696 dated 1 April 1999
Varian Medical Systems BrachyVision 6.0, 510(k) number K992762 dated 03 March 2000
Varian Medical Systems VariSeed 7.1, 510(k) number K030534 dated 21 May 2003
Telephone: 217-239-0900 Facsimile: 217-239-0905
{1}------------------------------------------------
K131428
Page 2 of 3
### Description of Acoustic MedSystems, Inc. RadVision Dose Planning and (5) Treatment System:
The RadVision system is a brachytherapy dose planning and treatment guidance system. The system consists of a computer, video capture device and software tools. The required software tools are installed in the computer. The system can be used for pre-treatment realtime dose planning in the Operating Room (OR), and for permanent seed implants performing post-implant seed localization assessment and post-implant dose distribution analysis.
#### (6) Intended Use:
The intended use of Acoustic MedSystems. Inc. RadVision Dose Planning and Treatment Systems is to provide patient-specific planning, imaging, and implants or applicator device alignment for treating cancer using radioactive seed implants or high dose radiation (HDR) afterloader devices. In addition to planning and treatment, RadVision allows volume and dose calculations. 2D and 3D anatomy and dose visualizations, as well as post treatment seed localization.
RadVision is a general purpose brachytherapy planning system used for prospective and confirmation dose calculations for patients undergoing a course of brachytherapy using either temporary or permanent implants of various radioisotopes.
#### (6) Performance Testing:
The computer systems that will be used to install RadVision for commercial distribution to customers are required to be tested and meet the following standards:
- IEC 60601-1-1 .
- IEC 60601-1-2 .
- IEC 60601-1-3 .
- IEC 60601-1-4 .
Verification and Validation test procedures were prepared considering the technological characteristics of the predicate devices. Non-clinical testing of RadVision was conducted for the following characteristics:
- Computer systems and software .
- Administration, Patient management and Help support .
- Calibration of template positioning and images to template registration �
- Equipment set identifications and specifications (associates all . equipment/instrumentation as configured for set up for implants)
- . Image acquisition, import, display (2D and 3D) from various modalities (ultrasound, MR, CT Fluoroscopy)
- . Contouring and anatomical identification
- Identification and loading of radiation source and placement with image quidance (low . dose radiation (LDR) and HDR)
- . Dose calculation (LDR and HDR)
- . Printing reports
Verification and validation testing was performed using RadVision Dose Planning and Treatment System. RadVision passed all the verification and validation tests successfully.
{2}------------------------------------------------
K131421
Page 3 of 3
Based on verification and validation test results and analysis of similarities and differences between the technological characteristics of the devices, RadVision is substantially equivalent to the stated predicate devices without raising new safety and/or efficacy issues.
#### Technological Characteristics: (7)
The intended use statement of RadVision is respectfully similar to the intended use statement of the predicate devices. RadVision is a Brachytherapy treatment planning software compatible with both LDR and HDR delivery systems, whereas Interplant and VariSeed 7.1 are for LDR and Varian BrachyVision 6.0 for HDR. There are no substantial differences between RadVision and Interplant and Variseed 7.1 for permanent seed implant functions and between RadVision and Varian BrachyVision for HDR functions.
The level of concern for RadVision software is major: FDA considers software contained in radiation therapy devices to be Major level of concern. RadVision is not intended for diagnostic purposes, nor will it control the delivery of the treatment plan to the patient.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol featuring three parallel lines that curve and merge into a single, wavy line at the bottom.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July, 16, 2013
Acoustic MedSystems, Inc. % Ms. Jennifer Williams Official Regulatory Correspondent 208 Burwash Avenue SAVOY IL 61874
Re: K131428
Trade/Device Name: RadVision Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class: II Product Code: KXK. MUJ Dated: April 29. 2013 Received: May 22, 2013
Dear Ms. Williams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind vou, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Williams
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
## 510(k) Number (if known): K131428
Device Name: RadVision
Indications for Use:
The intended use of Acoustic MedSystems, Inc. RadVision Dose Planning and Treatment Systems is to provide patient-specific planning, imaging, and implants or applicator device alignment for treating cancer using radioactive seed implants or HDR afterloader devices. In addition to planning and treatment, RadVision also allows volume and dose calculations, 2D and 3D analomy and dose visualizations, and post treatment seed localization.
RadVision is a general purpose brachytherapy planning system used for prospective and confirmation dose calculations for patients undergoing a course of brachytherapy using either temporary or permanent implants of various radioisotopes.
| Burdette Medical Systems, Inc<br>Interplant | Varian Medical Systems<br>BrachyVision 6.0 | Varian Medical Systems<br>VariSeed 7.1 |
|---------------------------------------------|--------------------------------------------|----------------------------------------|
| K982696 | K992762 | K030534 |
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k)_K131428
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.