K982696 · Burdette Medical Systems, Inc. · MUJ · Apr 1, 1999 · Radiology
Device Facts
Record ID
K982696
Device Name
INTERPLANT
Applicant
Burdette Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Apr 1, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Burdette Medical Systems, Inc. Interplant™ system is used in planning and performing a prostate brachytherapy implant procedure to provide ultrasound probe alignment and radioactive seed implantation. The Interplant™ device will use standard 125-Iodine (125 I) or 103-Paladium (103 Pd) isotope permanent implant radiation seeds for the treatment of prostate cancer. The system will allow volume measurement and dose calculation, relevant prostate anatomy and dose visualization, and post-treatment seed localization.
Device Story
Interplant™ is a brachytherapy dose planning and implant system used in the operating room by physicians. It integrates delivery components (needles, template guide, probe stepper) with software for 125-Iodine or 103-Paladium seed implants. The system interfaces with existing ultrasound equipment to acquire real-time image planes; it automatically determines probe position to facilitate rapid image acquisition for dose planning. The system performs volume/dose calculations, image contouring, and generates dose volume histograms and isodose surface contours. It also supports post-implant seed localization. By providing immediate feedback and precision alignment, the system aims to improve quality assurance during prostate seed implant procedures, ultimately benefiting the patient through more accurate radiation delivery.
Clinical Evidence
No clinical data provided. The submission relies on bench testing and the integration of established hardware and software components for brachytherapy dose planning and seed implantation.
Technological Characteristics
System includes hardware (needles, template guide, probe stepper) and software for 2D/3D image visualization and dose calculation. Operates on a Pentium-based processor with graphics and digitizing hardware. Uses 2D ultrasound imaging input. Software features include image contouring, dose calculation, dose volume histograms, isodose surface contours, and post-implant seed localization. Designed for use with 125-Iodine or 103-Paladium seeds.
Indications for Use
Indicated for patients requiring prostate brachytherapy for the treatment of prostate cancer using 125-Iodine or 103-Paladium permanent implant radiation seeds.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Multimedia Medical Systems, Inc. B3DTUA TherapacPLUS dose planning
SSGI Prowess 500 and Prowess 2000 dose planning (K924240)
CIVCO Medical Instruments Ultra-Step stepper (K875241)
B&K Medical Systems HP-100 stepper/stand
Tayman Medical Accuseed stand/stabilizer (K963302)
COTAN stand/stabilizer
Northwest stand/stabilizer
Maroon Bels, Inc. Brachystand and Brachystepper (K972672)
Submission Summary (Full Text)
{0}------------------------------------------------
4/1/99
## Burdette Medical Systems, Inc. 206 N. Randolph Street, Suite 301 Champaign, IL 61820 Tel: (217) 239-0900 Fax: (217) 239-0905
K982694
## 510(k) Summary
Submitter: Burdette Medical Systems, Inc.
Address: 206 N. Randolph Suite 301 Champagne IL 61820
Phone number: (217) 239-0900
Fax number: (217) 239-0905
Contact person: Everette C. Burdette, Ph.D.
Date prepared: July 27, 1998
Trade name: Interplant™
Common name: Brachytherapy dose planning and implant system
Classification name: 892.5050 Medical charged-particle radiation therapy system Class II
Substantial equivalence claimed to:
- 1. Multimedia Medical Systems, Inc. B3DTUA TherapacPLUS dose planning
- 2. SSGI Prowess 500 and Prowess 2000 dose planning K924240
- 3. CIVCO Medical Instruments Ultra-Step stepper K875241
- 4. B&K Medical Systems HP-100 stepper/stand
- 5. Tayman Medical Accuseed stand/stabilizer K963302
- 6. COTAN stand/stabilizer
- 7. Northwest stand/stabilizer
- 8. Maroon Bels, Inc. Brachystand and Brachystepper K972672
## Description:
The Interplant™ system for brachytherapy was developed to address the need for a comprehensive system and procedure for prostate seed implants. The Interplant in system compresses the requirement for separate multiple procedures into one online setting,
This product integrates all of the delivery system components (needles, template guide, probe stepper) with the dose planning system for 125-Iodine (125 I) and 103-Paladium (103 Pd) implants. The system is designed for real-time dose planning in the Operating Room
{1}------------------------------------------------
(OR) and for performing post-implant seed localization assessment. Real time ultrasound image plane acquisition with automatic probe position determination makes it possible to complete the image acquisition needed for dose planning within about a minute. The interplant™ system provides the needed components for seed implant procedures except the seeds themselves and uses existing ultrasound imaging equipment. This will enable the physician to focus on the procedure, providing immediate feedback for improved quality assurance.
## Intended use:
The Burdette Medical Systems, Inc. interplant™ system is used to provide precision ultrasound probe alignment and radioactive seed implantation in the treatment of prostate cancer. The system will allow volume and dose calculation, image visualization, and posttreatment seed localization.
Summary of technological characteristics:
The Real-Time Brachytherapy Radiation Seed Implant Program and System includes all technology, computer software, and hardware, and methods associated with the implant of radioactive "seeds" or pellets in the prostate. System elements include the user interface, real-time image acquisition, image contouring, dose calculation and display software, dose volume histograms, isodose surface contours, post-implant seed localization, and the patient scheduling spreadsheet software.
The System consists of a two-dimensional and three-dimensional image visualization and brachytherapy dose system employing 2D ultrasound imaging for use in radioactive seed implants of the prostate. The software for the brachytherapy seed implant and dose calculation system was developed on a Pentium-based processor with supporting graphics and digitizing hardware.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 1999
Everette C. Burdette, Ph.D. President Burdette Medical Systems 206 N. Randolph, Suite 301 Champaign, IL 61820
Re:
K982696 Interplant™ (Brachytherapy Dose Planning and Implant System) Dated: February 14, 1999 Received: February 24, 1999 Regulatory class: II 21 CFR 892.5730/Procode: 90 MUJ
Dear Dr. Burdette:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requilations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal register , Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal. Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):
Device Name: Interplant™
Indications for Use:
The Burdette Medical Systems, Inc. Interplant™ system is used in planning and performing a prostate brachytherapy implant procedure to provide ultrasound probe alignment and radioactive seed implantation. The Interplant™ device will use standard 125-Iodine (125 I) or 103-Paladium (103 Pd) isotope permanent implant radiation seeds for the treatment of prostate cancer. The system will allow volume measurement and dose calculation, relevant prostate anatomy and dose visualization, and post-treatment seed localization.
n982696
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-the-Counter Use **__**
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K982696
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.