K130478 · Globus Medical, Inc. · MAX · Jul 26, 2013 · Orthopedic
Device Facts
Record ID
K130478
Device Name
SUSTAIN AND SUSTAIN RADIOLUCENT SPACERS
Applicant
Globus Medical, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jul 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as a lumbar intervertebral body fusion device, SUSTAIN® and SUSTAIN® Radiolucent (SUSTAIN®R) Spacers are intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The SUSTAIN® and SUSTAIN® R Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the REVERE®, REVOLVE™ or BEACON® Stabilization Systems. When used as a cervical intervertebral body fusion device, the SUSTAIN® and SUSTAIN®R Spacers are intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The SUSTAIN® and SUSTAIN®R Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the ASSURE®, PROVIDENCE® or XTEND® Anterior Cervical Plate System. When used as a vertebral body replacement device, SUSTAIN® and SUSTAIN® R Spacers are intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems and anterior screw and rod systems). The interior of the spacer can be packed with bone grafting material. SUSTAIN® and SUSTAIN® R Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period.
Device Story
SUSTAIN® and SUSTAIN® R Spacers are spinal implants used for intervertebral body fusion or vertebral body replacement. Devices feature axial holes for packing autogenous bone graft and superior/inferior surface protrusions to grip vertebral endplates and resist expulsion. Spacers are implanted by surgeons in clinical settings using various approaches (posterior, transforaminal, anterior, anterolateral, or lateral). Devices provide structural stability and anterior spinal column support; they must be used with supplemental spinal fixation systems. Benefits include restoration of spinal column support and facilitation of fusion in patients with DDD, tumor, or trauma.
Clinical Evidence
Bench testing only. Mechanical testing including static and dynamic compression, static and dynamic torsion, static and dynamic compression shear, and subsidence was performed in accordance with the FDA guidance document for Intervertebral Fusion Devices.
Technological Characteristics
Materials: Commercially pure titanium or titanium alloy (ASTM F67, F136, F1295) for SUSTAIN®; PEEK polymer with titanium alloy or tantalum markers (ASTM F136, F560, F1295, F2026) for SUSTAIN® R. Form factor: Spacers with axial holes and surface protrusions. Energy source: None (mechanical implant).
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at 1-2 contiguous lumbar levels (L2-S1) or 1 cervical level (C2-T1) requiring fusion, or thoracolumbar (T1-L5) vertebral body replacement due to tumor or trauma.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
BAK/Cervical (BAK-C®) Interbody Fusion System (P980048)
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 2 6 2013
## 510(k) SUMMARY: SUSTAIN® and SUSTAIN® Radiolucent Spacers
| Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Christina Kichula<br>Group Manager, Regulatory Affairs |
| Date Prepared: | February 22, 2013 |
| Device Name: | SUSTAIN® and SUSTAIN® Radiolucent (SUSTAIN® R)<br>Spacers |
| Classification: | Per 21 CFR as follow:<br>§888.3060 Implant, fixation, spinal intervertebral body<br>fixation orthosis devices<br>§888.3080 Intervertebral Body Fusion device<br>Product Codes MQP, MAX, and ODP.<br>Regulatory Class II, Panel Code 87. |
| Predicate(s): | PATRIOT® Spacers (Lumbar) K072970<br>PATRIOT® Spacers (Cervical) K072991<br>PATRIOT® Transcontinental™ Spacers (Lumbar) K093242<br>CALIBER™ Spacers (Lumbar) K102293<br>SUSTAIN® Spacers K031302<br>SUSTAIN® Radiolucent Spacers K040284<br>COALITION® Spacers K083389<br>BAK/Cervical (BAK-C®) Interbody Fusion System P980048 |
## Purpose:
The purpose of this submission is to add intervertebral indications for SUSTAIN® and SUSTAIN® R Spacers and to make sterile spacers available.
## Device Description:
SUSTAIN® and SUSTAIN® R Spacers are devices that can be used as intervertebral fusion devices or as vertebral body replacement devices. These spacers are available in different shapes and heights to accommodate various surgical approaches and anatomical needs. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to resist expulsion. Each spacer has an axial hole to allow grafting material to be packed inside the spacer.
These spacers are used to provide structural stability in skeletally mature individuals following discectomy, corpectomy, or vertebrectomy (including
{1}------------------------------------------------
partial). Lumbar spacers may be inserted using a posterior, transforaminal, anterior, anterolateral, or lateral lumbar approach. Cervical spacers are inserted using an anterior cervical approach.
The SUSTAIN® Spacers are made from commercially pure titanium or titanium alloy as specified in ASTM F67, F136, and F1295.
The SUSTAIN® R Spacers are made from radiolucent peek polymer with titanium alloy or tantalum markers as specified in ASTM F136, F560, F1295, and F2026.
#### Indications for Use
When used as lumbar intervertebral body fusion devices, SUSTAIN® and SUSTAIN® Radiolucent (SUSTAIN® R) Spacers are intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The SUSTAIN® and SUSTAIN® R Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the REVERE®, REVOLVE™ or BEACON® Stabilization Systems.
When used as cervical intervertebral body fusion devices, the SUSTAIN® and SUSTAIN®R Spacers are intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The SUSTAIN® and SUSTAIN® R Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the ASSURE®, PROVIDENCE® or XTEND® Anterior Cervical Plate System.
When used as vertebral body replacement devices. SUSTAIN® and SUSTAIN®R Spacers are intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems and anterior screw and rod systems). The interior of the spacer can be packed with bone grafting material. SUSTAIN® and SUSTAIN® R Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period.
#### Performance Data
Mechanical testing (static and dynamic compression, static and dynamic torsion. static and dynamic compression shear, and subsidence) was conducted in
{2}------------------------------------------------
accordance with the "Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Intervertebral Fusion Device," June 12, 2007. Performance and comparative analysis data demonstrate substantial equivalence to the predicate devices.
#### Basis for Substantial Equivalence:
SUSTAIN® and SUSTAIN® R Spacers are identical to the predicate SUSTAIN® and SUSTAIN® R Spacers with respect to design, technical characteristics, performance and vertebral body replacement indications for use. SUSTAIN® and SUSTAIN® R Spacers are equivalent to the predicate PATRIOT® Spacers with respect to intervertebral body fusion devices indications for use. The information provided within this premarket notification supports substantial equivalence to the predicate devices. SUSTAIN® and SUSTAIN® R Spacers are as safe, as effective, and perform as well as or better than predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image contains the text "Public Health Service". The text is in a bold, sans-serif font. The text is black and the background is white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 26, 2013
Globus Medical Incorporated % Ms. Christina Kichula Group Manager, Regulatory Affairs 2560 General Armistead Avenue Audubon, Pennsylvania 19403
Re: K130478
Trade/Device Name: SUSTAIN® and SUSTAIN® Radiolucent Spacers Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, ODP, MQP Dated: June 25, 2013 Received: June 26, 2013
Dear Ms. Kichula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Christina Kichula
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincercly vours,
# Erin I. Keith
#### For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
K130478 510(k) Number:
SUSTAIN® and SUSTAIN® Radiolucent Spacers Device Name:
#### INDICATIONS
When used as a lumbar intervertebral body fusion device, SUSTAIN® and SUSTAIN® Radiolucent (SUSTAIN®R) Spacers are intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The SUSTAIN® and SUSTAIN® R Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the REVERE®, REVOLVE™ or BEACON® Stabilization Systems.
When used as a cervical intervertebral body fusion device, the SUSTAIN® and SUSTAIN®R Spacers are intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The SUSTAIN® and SUSTAIN®R Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the ASSURE®, PROVIDENCE® or XTEND® Anterior Cervical Plate System.
When used as a vertebral body replacement device, SUSTAIN® and SUSTAIN® R Spacers are intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems and anterior screw and rod systems). The interior of the spacer can be packed with bone grafting material. SUSTAIN® and SUSTAIN® R Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period.
X x OR Prescription Use (Per 21 CFR §801.109)
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.