K131301 · Becton, Dickinson & CO · PER · Aug 2, 2013 · Clinical Chemistry
Device Facts
Record ID
K131301
Device Name
BD FACS SAMPLE PREP ASSISTANT III
Applicant
Becton, Dickinson & CO
Product Code
PER · Clinical Chemistry
Decision Date
Aug 2, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2750
Device Class
Class 1
Indications for Use
The BD FACS™ Sample Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto™ II and BD FACSCalibur™ flow cytometry systems.
Device Story
Microprocessor-controlled pipetting/diluting system; automates lyse/no-wash sample preparation for flow cytometry. Inputs: whole blood specimens in primary tubes (Vacutainer/Sarstedt), reagents, lysing solution. Operation: robotic pipetting module (X/Y/Z axes) pierces tube caps; syringe pumps volumetrically measure/transfer fluids to daughter tubes; mixes samples; adds lysing solution; automates probe cleaning. Output: prepared samples ready for acquisition on BD FACSCalibur or BD FACSCanto II flow cytometers. Used in clinical laboratories; operated by technicians/lab staff. Automates manual pipetting tasks; ensures consistent sample preparation; reduces manual error; improves workflow efficiency for flow cytometric analysis.
Clinical Evidence
Bench testing only. Accuracy evaluated via method comparison and bias estimation (CLSI EP9-A2); precision evaluated per CLSI EP5-A2; carryover evaluated per FDA guidance for automated differential cell counters. Results demonstrated equivalent performance to manual preparation and system specifications.
Technological Characteristics
Microprocessor-controlled robotic pipetting system; X/Y/Z axis movement; 1 mL sample/reagent syringes and 10 mL lyse syringe; fluidic, optic, robotic, and electronic components; barcode reader; automated probe rinse (3 pulses, ~1 sec each).
Indications for Use
Indicated for in vitro diagnostic use in preparing human whole blood samples for flow cytometric analysis using specific previously cleared BD Multitest and BD Tritest reagent kits on BD FACSCanto II and BD FACSCalibur flow cytometry systems.
Regulatory Classification
Identification
A pipetting and diluting system for clinical use is a device intended to provide an accurately measured volume of liquid at a specified temperature for use in certain test procedures. This generic type of device system includes serial, manual, automated, and semi-automated dilutors, pipettors, dispensers, and pipetting stations.
Predicate Devices
BD FACS™ Sample Prep Assistant III (K131301 - predicate for BD FACSCanto II system usage)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
k131301
B. Purpose for Submission:
Modify Intended Use to include FACSCalibur
C. Manufacturer and Instrument Name:
BD Biosciences
BD FACSTM Sample Prep Assistant III
D. Type of Test or Tests Performed:
Specimen Processor
E. System Descriptions:
1. Device Description:
The BD FACSTM Sample Prep Assistant III (SPA III) is a microprocessor-controlled pipetting and diluting system which automatically prepares whole blood samples using the lyse / no-wash sample preparation method for flow cytometry. Used as an accessory to the BD FACSCalibur™ and BD FACSCanto™ II flow cytometers, the SPA III combines fluidic, optic, robotic, and electronic components to automatically prepare samples for acquisition and analysis.
The SPA III pierces the sample tube cap to withdraw sample, aliquots blood and reagent into daughter tubes, and mixes the sample according to preprogrammed protocols. The device also adds lysing solution and automates cleaning procedures.
The unit consists of an enclosure, one robotic pipetting module moving in the X/Y/Z axes, a power supply, a central controller unit, fluid pumps, and a barcode reader.
2. Principles of Operation:
Two syringe pumps volumetrically measure fluids and move them through the system for aspiration and delivery.
3. Modes of Operation:
Automated
4. Specimen Identification:
Manual entry or bar code reader.
5. Specimen Sampling and Handling:
Specimens are whole blood specimens collected into approved tubes
6. Calibration:
{1}
Not applicable.
7. Quality Control:
The laboratory follows the manufacturer’s instructions for gravimetric verification of accuracy of pipetting and quality control instructions for BD cleared flow cytometric IVD reagents in the context of the BD flow cytometer systems.
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X or No ☐
F. Regulatory Information:
1. Regulation section:
21 CFR § 862.2750, Pipetting and Diluting System for Clinical Use
2. Classification:
Class I
3. Product code:
PER, Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis
4. Panel:
Clinical Chemistry (75)
G. Intended Use:
1. Intended Use:
The BD FACSTM Sample Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto™ II and BD FACSCalibur™ flow cytometry systems.
2. Indications for Use:
Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto™ II flow cytometry systems:
- BD Multitest 6-Color TBNK Reagent with or without BD Trucount Tubes
- BD Multitest IMK Kit with or without BD Trucount Tubes
- BD Multitest CD3 FITC/CD16+CD56 PE/CD45 PerCP/CD19 APC with or without BD Trucount Tubes
- BD Multitest CD3 FITC/CD8 PE/CD45 PerCP/CD4 APC with or without BD Trucount Tubes
Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCalibur™ flow cytometry systems:
- BD Multitest IMK Kit with or without BD Trucount Tubes
- BD Multitest CD3 FITC/CD16+CD56 PE/CD45 PerCP/CD19 APC with or without
2
{2}
BD Trucount Tubes
BD Multitest CD3 FITC/CD8 PE/CD45 PerCP/CD4 APC with or without BD
Trucount Tubes
BD Tritest CD3/CD16+56/CD45 with or without BD Trucount Tubes
BD Tritest CD3/CD19/CD45 with or without BD Trucount Tubes
BD Tritest CD3/CD4/CD45 with or without BD Trucount Tubes
BD Tritest CD3/CD8/CD45 with or without Trucount Tubes
BD Tritest CD4/CD8/CD3 with BD Trucount Tubes
For in vitro diagnostic use.
3. Special Conditions for Use Statement(s):
For prescription use only.
4. Special instrument requirements:
For use on the BD FACSCanto II (previously cleared under k041074) and BD
FACSCalibur (previously cleared under k923790) flow cytometry systems.
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
BD FACSTM Sample Prep Assistant III accessory to the BD FACSCanto™ system with BD
FACSCanto clinical software (k102064)
2. Comparison with Predicate Device:
| Similarities/Differences | | |
| --- | --- | --- |
| Item | New Device | Predicate |
| | The BD FACSTM Sample Prep Assistant III as an accessory to BD FACSCanto II and BD FACSCalibur flow cytometry systems. | BD FACSTM Sample Prep Assistant III accessory to the BD FACSCanto™ system with BD FACSCanto clinical software (K102064) |
| Intended Use | The BD FACS Sample Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II and BD FACSCalibur flow cytometry systems. | The BD FACS Sample Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II flow cytometry systems. |
| Controlling software | BD FACS SPA software version 4.0.2.1 (did not include any software changes. The software revision only added the cleared BD FACS Sample Prep Assistant III Instructions for Use with the software and also updated the Online Help Files) | BD FACS SPA software version 4.0.2 |
{3}
4
| Similarities/Differences | | |
| --- | --- | --- |
| Item | New Device | Predicate |
| | The BD FACSTM Sample Prep Assistant III as an accessory to BD FACSCanto II and BD FACSCalibur flow cytometry systems. | BD FACSTM Sample Prep Assistant III accessory to the BD FACSCanto™ system with BD FACSCanto clinical software (K102064) |
| Sample Type | Whole blood | Same |
| Preparation Method | Automated | Same |
| Pipetting Syringe | 1 mL sample/reagent syringe
10 mL lyse syringe | Same |
| Supported primary blood sample tubes | Vacutainer
Sarstedt | Same |
| Probe Rinse | 3 pulses of approximately 1 second | Same |
| Quality Control Techniques | Gravimetric calibration | Same |
I. Special Control/Guidance Document Referenced (if applicable):
CLSI EP5-A2: Evaluation of Precision Performance of Clinical Chemistry.
CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples. Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA, December 4, 2001.
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
The performance of the SPA III was compared to manual pipetting. Normal peripheral blood specimens (defined as specimens with normal CD4+ counts) and HIV-infected patient peripheral blood specimens were evaluated. Samples were stained with BD Multitest CD3 FITC/CD8 PE/CD45 PerCP/CD4 APC and BD Multitest CD3 FITC/CD16+CD56 PE/CD45 PerCP/CD19 APC with Trucount Tubes reagents using manual pipetting and using the SPAIII. The samples were acquired on BD FACSCalibur systems containing the BD FACSTM Loader (k953302). Both Sarstedt and BD Vacutainer EDTA tubes were used for blood collection. Three testing sites were used for the study.
Acceptance Criteria:
| Parameter | 95% CI on Mean Bias |
| --- | --- |
| CD3+ absolute counts | ≤ 10% |
| CD3+CD4+ absolute counts | ≤ 10% |
| CD3+CD8+ absolute counts | ≤ 10% |
| B cell absolute counts (CD19+, CD3-) | ≤ 20% |
{4}
| Parameter | 95% CI on Mean Bias |
| --- | --- |
| NK cell absolute counts (CD16+ + CD56+, CD3-) | ≤ 20% |
| % CD3, CD3+CD4+, CD3+CD8+, CD19+CD3-, CD16+CD56+CD3- | If the mean of % positive values of all specimens is ≤ 30% as determined by reference method, 95% CI of mean bias must be within ± 3. |
| NK cell (CD16+ + CD56+, CD3-) %Positive | If the mean of % positive values of all specimens is > 30% as determined by reference method, 95% CI of relative mean bias must be within ±10%. |
Summary of the Bland Altman (bias) Absolute Count Results
Vacutainer Tubes
| Absolute Count Subsets | Rep | N | Mean %Bias | SD | 95% CI | Criteria | Pass/Fail |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CD4 | 1 | 236 | -3.3 | 9.6 | (-4.3, -2.3) | 10% | Pass |
| CD8 | 1 | 236 | -3.2 | 8.8 | (-4.1, -2.2) | 10% | Pass |
| CD3 | 1 | 236 | -3.5 | 7 | (-4.2, -2.7) | 10% | Pass |
| CD19 | 1 | 236 | -3.7 | 17.5 | (-5.6, -1.8) | 20% | Pass |
| CD16+56 | 1 | 236 | -1.5 | 13.2 | (-2.9, -0.1) | 20% | Pass |
Sarstedt
| Absolute Count Subsets | Rep | N | Mean %Bias | SD | 95% CI | Criteria | Pass/Fail |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CD4 | 1 | 100 | 0.9 | 9.1 | (-0.6, 2.4) | 10% | Pass |
| CD8 | 1 | 100 | 0.0 | 8.4 | (-1.4, 1.4) | 10% | Pass |
| CD3 | 1 | 100 | -1.0 | 5 | (-1.9, -0.2) | 10% | Pass |
| CD19 | 1 | 100 | -1.5 | 10.8 | (-3.3, 0.3) | 10% | Pass |
| CD16+56 | 1 | 100 | 1.8 | 14.3 | (-0.6, 4.1) | 10% | Pass |
Summary of the Bland Altman (bias) Percent Positive Results
Vacutainer
| Percent Positive Subsets | Rep | N | Mean %Bias | SD | 95% CI | Criteria | Pass/Fail |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CD4 | 1 | 236 | -0.03 | 1.2 | (-0.2, 0.1) | 3 | Pass |
| CD8 | 1 | 236 | 0.3 | 3.6 | (0, 0.7) | 10% | Pass |
| CD3 | 1 | 236 | 0.2 | 1.8 | (0, 0.4) | 10% | Pass |
| CD19 | 1 | 236 | -0.1 | 0.9 | (-0.2, 0) | 3 | Pass |
| CD16+56 | 1 | 236 | 0.1 | 1 | (0, 0.2) | 3 | Pass |
Sarstedt
| Percent Positive Subsets | Rep | N | Mean %Bias | SD | 95% CI | Criteria | Pass/Fail |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CD4 | 1 | 100 | 0.1 | 1.2 | (-0.1,0.3) | 3 | Pass |
| CD8 | 1 | 100 | -0.5 | 3.6 | (-1.1,0.1) | 10% | Pass |
| CD3 | 1 | 100 | -0.3 | 1.4 | (-0.5, -0.1) | 10% | Pass |
| CD19 | 1 | 100 | 0 | 0.6 | (-0.1,0.1) | 3 | Pass |
| CD16+56 | 1 | 100 | 0.2 | 0.9 | (0, 0.4) | 3 | Pass |
{5}
# Summary of the Deming Regression Absolute Count results
Vacutainer Tubes
| Abs Counts | N | R² | Intercept | Confidence Limit (L/U) | Slope | Confidence Limit (L/U) |
| --- | --- | --- | --- | --- | --- | --- |
| CD4 | 236 | 0.96 | 4.64 | (-20.45, 29.78) | 0.96 | (0.93, 0.99) |
| CD8 | 236 | 0.98 | 1.67 | (-9.88, 13.23) | 0.96 | (0.92, 0.99) |
| CD3 | 236 | 0.93 | 11.27 | (-12.16, 34.71) | 0.95 | (0.90, 0.99) |
| CD19 | 236 | 0.96 | -0.65 | (-3.94, 2.64) | 0.96 | 0.93, 0.98) |
| CD16+56 | 236 | 0.97 | 0.23 | (-1.92, 2.4) | 0.97 | (0.95, 0.99) |
Sarstedt
| Abs Counts | N | R² | Intercept | Confidence Limit (L/U) | Slope | Confidence Limit (L/U) |
| --- | --- | --- | --- | --- | --- | --- |
| CD4 | 100 | 0.98 | -45.53 | (-77.4, -13.7) | 1.02 | (0.99, 1.04) |
| CD8 | 100 | 0.98 | 6.6 | (-17.6, 30.8) | 0.99 | (0.93, 1.04) |
| CD3 | 100 | 0.97 | -24.5 | (-53.3, 4.3) | 1.02 | (0.99,1.05) |
| CD19 | 100 | 0.97 | -1.08 | (-4.91, 2.76) | 0.99 | (0.96, 1.01) |
| CD16+56 | 100 | 0.96 | 4.22 | (-0.006, 8.44) | 0.98 | (0.94, 1.01) |
# Summary of the Deming Regression Percent Positive results
Vacutainer Tubes
| Percent Positive Subsets | N | R² | Intercept | Confidence Limit (L/U) | Slope | Confidence Limit (L/U) |
| --- | --- | --- | --- | --- | --- | --- |
| CD4 | 236 | 0.98 | -0.57 | (-1.94, 0.81) | 1.0 | (0.99,1.02) |
| CD8 | 236 | 0.99 | 0.22 | (-0.18,0.63) | 0.99 | (0.97, 1.0) |
| CD3 | 236 | 0.98 | -0.39 | (-1.31,0.52) | 1.01 | (0.99, 1.03) |
| CD19 | 236 | 0.98 | -0.008 | (-0.30, 0.28) | 0.99 | (0.96, 1.02) |
| CD16+56 | 236 | 0.96 | 0.19 | (-0.035, 0.42) | 0.99 | (0.97, 1.02) |
Sarstedt
| Percent Positive Subsets | N | R² | Intercept | Confidence Limit (L/U) | Slope | Confidence Limit (L/U) |
| --- | --- | --- | --- | --- | --- | --- |
| CD4 | 100 | 0.99 | -0.27 | (-1.69, 1.15) | 1.00 | (0.98, 1.01) |
| CD8 | 100 | 0.99 | 0.71 | (-1.06, 2.49) | 0.97 | (0.89, 1.04) |
| CD3 | 100 | 0.97 | 0.39 | (-1.22, 1.99) | 0.99 | (0.95, 1.02) |
| CD19 | 100 | 0.98 | 0.032 | (-0.21, 0.27) | 0.99 | (0.97, 1.02) |
| CD16+56 | 100 | 0.96 | 0.33 | (-0.06, 0.72) | 0.99 | (0.94, 1.03) |
The results showed that the $95\%$ confidence interval for the two collection tubes and for all parameters tested met the acceptance criteria. When used with the BD FACSCalibur system, the SPA III and manual pipetting yield equivalent performance.
# b. Precision/Reproducibility:
Precision for the SPA III was evaluated in three studies; two for Dispense Accuracy and Precision, and one for Assay Precision.
{6}
# Dispense Accuracy and Precision (1)
The pipetting accuracy and precision were assessed gravimetrically for the BD-defined Multitest protocol using normal donor blood, BD Multi-Check process Control and reagent buffer (to represent reagent). These studies were run on three instruments. The study design is shown in the following table.
Table 1: Study Design
| Testing Mode | Single Dispense | Multi Dispense |
| --- | --- | --- |
| Blood - “Multitest/Tritest/Absolute Count” | 50 μL blood (20 primary Tubes x1 daughter tube=20 daughter tubes) | 50 μL Multi-Check Control (5x4=20) 50 μL TruCount Control (3x7=21) |
| Reagent - “Multitest/Tritest/Absolute Count” | 20 μL blood (1x20=20) | 5 μL blood (20x1=20) Repeat three times (20x3=60) |
Table 2: Acceptance Criteria Table
| Sample Type | Volume Dispensed | N | Mean Value | %CV |
| --- | --- | --- | --- | --- |
| Blood | 50 μl | 20 | 50 μl ± 3% | ≤ 3% |
| Reagent | 20 μl | 20 | 20 μl ± 7% | ≤ 5% |
Conclusion: The SPA III demonstrated pipetting accuracy and precision within specification.
# Dispense Accuracy and Precision (2)
Testing was performed gravimetrically using reagent buffer.
The studies were run on three SPAIII instruments. Multitest/Tritest/Absolute Count mode were evaluated for accuracy and precision as well as reagent liquid detection, aspirate and dispense.
Table 3: Acceptance Criteria Table
| Sample Type | Volume Dispensed | N | Mean Value | %CV |
| --- | --- | --- | --- | --- |
| Reagent | 20 μl | 20 | 20 μl ± 7% | ≤ 5% |
Table 4: Accuracy and Precision
| Instrument | Mean Volume | % CV | 90% Upper Limit of % CV | Pass/Fail |
| --- | --- | --- | --- | --- |
| T0069 | 19.49 | 1.18 | 1.43 | Pass |
| T0412P | 21.12 | 1.90 | 2.30 | Pass |
| T0413P | 19.91 | 2.74 | 3.32 | Pass |
Conclusion: The SPA III demonstrated pipetting accuracy and precision within specification.
{7}
8
# Assay Precision:
Precision was assessed based on Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline, CLSI document EP5-A2.
Pipetting precision was assessed gravimetrically for the BD-defined Multitest protocol using normal donor blood, BD Multi-Check process control, reagent buffer (to represent reagent). The performance for pipetting blood was found to be within the manufacturer's specifications of $50~\mu \mathrm{L} \pm 3\%$ and the performance for pipetting reagent was found to be within the manufacturer's specifications of $20~\mu \mathrm{L} \pm 7\%$.
Precision of the staining and lysing of samples was also assessed. This was demonstrated by measuring the closeness of agreement between independent test/measurement results obtained under stipulated conditions. These conditions include multiple operators (3), multiple instruments (3 SPA III), and duration of testing (21 days) run twice daily in duplicate. The study samples were CD-Chex normal (CDN) and CD-Chex CD4 low (CDL) stained with BD Multitest 6-Color TBNK Reagent in BD Trucount Tubes. Measurements for this study were the absolute counts and percentages of the following lymphocyte subsets: CD3+CD4+, CD3+CD8+, CD3+, CD3-CD19+ and CD3-CD16+CD56+. The manufacturer's acceptance criteria were as follows:
- Absolute counts:
- $95\%$ upper CI on the CV: $\leq 10\%$ for CD3+, CD4+ and CD8+
- $95\%$ upper CI on the CV: $\leq 20\%$ for CD19+ and CD16+56+
- Percent positives:
- $95\%$ upper CI on the SD: $\leq 2.5$
Study results indicate that the device performed according to the manufacturer's acceptance criteria.
Conclusion: This study confirms that the SPA III system aspirates and dispenses specified volumes of blood and reagent accurately and precisely.
c. Linearity:
Not applicable
d. Carryover:
# Sample Carryover
Carryover was assessed based on recommendations contained in Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA, December 4, 2001.
Sample carryover was tested using three donors (in duplicate) on three SPA III instruments by running three consecutive high cell count samples followed by three consecutive low cell count samples. The carryover acceptance criterion was $\text{carryover} \leq 0.2\%$. High cell count samples ($\geq 50,000$ WBC/ $\mu$L) were prepared by addition of autologous RBC-lysed whole blood. Low cell counts samples ($400 \pm 200$ WBC/ $\mu$L) were prepared by dilution with buffer (PBS). Comparison between the gold standard of manual pipetting and SPA III was done with three instruments using
{8}
three normal donor samples run twice per day [as Replicates 1 and 2 (Rep1 and Rep2)]. Samples were stained with the pan-leucocyte marker CD45 in BD Trucount Tubes.
Sample carryover was calculated using the following formula:
% carryover = [(First Low – Third Low)/(Third High – Third Low)]*100
from “Guidelines for the Evaluation of Blood Cell Analysers including those use for Differential Leucocyte and Reticulocyte Counting and Cell Marker Applications,” Clinical Laboratory Haematology 16: 157-174 [1994].
% carryover = [(First Low – Third Low)/(Third High – First Low)]*100
from CLSI (formerly known as NCCLS) Guideline H52-A “Approved Guideline for Fetal Red Cell Detection.”
These studies, when analyzed, analyzed using both carryover formulae indicated that the device performed within the manufacturer’s specification of <0.2%.
## Reagent Carryover
Reagent carryover was assessed by determining carryover of reagent to buffer for reagent tube to reagent tube carryover and for secondary specimen tube to secondary specimen tube. The acceptance criteria was carryover < 0.01%. High cell count samples (≥ 50,000 WBC/μL) were used and were prepared by addition of autologous RBC-lysed blood. Comparison between manual pipetting and SPA III was done with three instruments using six donor samples run twice per day [as Replicates 1 and 2 (Rep1 and Rep2)]. The reagent dispensed was BD Multitest 6-Color TBNK Reagent (three donors) for CD45+ cells or BD Leucocount (three donors) for White Blood Cells (WBCs).
### e. Interfering Substances:
Not applicable
### 2. Other Supportive Instrument Performance Data Not Covered Above:
Not applicable
## K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.