TUBE-CLEAR, CONTROL BOX (COMPONENT OF TUBECLEAR), CLEARING STEM (COMPONENT OF TUBECLEAR)
K123659 · Actuated Medical, Inc. · KNT · Dec 20, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K123659
Device Name
TUBE-CLEAR, CONTROL BOX (COMPONENT OF TUBECLEAR), CLEARING STEM (COMPONENT OF TUBECLEAR)
Applicant
Actuated Medical, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Dec 20, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TubeClear™ is indicated for use only and solely in clearing occlusions / clogs in nasogastric (NG) tubes (i.e., tubes that are placed through the nose and reside in the stomach) of adult patients. Clearing Stem Model NG-1036 is indicated for use in nasogastric tubes that are of size 10 - 14 French and have a length of 36 - 42 inches (91 - 108 cm). Clearing Stem Model NG-1043 is indicated for use in nasogastric tubes that are of size 10 - 18 French and have a length of 43 - 50 inches (109 - 127 cm).
Device Story
TubeClear is an electromechanical system for clearing occlusions in nasogastric (NG) tubes. System consists of a reusable Control Box (Model 101) and a single-use Clearing Stem. Operator manually inserts Clearing Stem into the NG tube; Control Box provides electromechanical linear reciprocating motion to a wire within the Clearing Stem. Reciprocating wire tip mechanically breaks up occlusions to restore tube patency. Control Box remains outside the patient. Used in clinical settings by healthcare providers. Benefits include restoration of tube function without requiring tube replacement.
Clinical Evidence
Bench testing only. Usability and retainability testing by end users performed to support design validation; all acceptance criteria passed.
Technological Characteristics
Electromechanical system comprising a reusable Control Box and single-use Clearing Stem. Actuation via motor-driven linear reciprocating wire. No software/firmware involved in device function. Design is identical to predicate except for the removal of centimeter markings on the Clearing Stem.
Indications for Use
Indicated for clearing occlusions in nasogastric (NG) tubes in adult patients. Model NG-1036 for 10-14 French tubes (36-42 inches); Model NG-1043 for 10-18 French tubes (43-50 inches).
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
K123654 pg 1 of 4
Image /page/0/Picture/1 description: The image shows the logo for Actuated Medical. The logo consists of a stylized symbol to the left of the company name, "ACTUATEDMEDICAL". The symbol appears to be a stylized representation of a medical device or technology.
Innovative motion + Positive outcomes
310 Rolling Ridge Drive Bellefonte, PA 16823 + p (814) 355.0003 + f (814) 355.1532 +w ActuatedMedical.com
# DEC 2 0 2012
- + ISO 13485 & 14971
- + 9001 Compliant
- 1 WEF
Certified
#### ATTACHMENT 4
## 510(K) SUMMARY
summary of the Special 510(k) information is submitted in This accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR807.92.
510(k) Number:
#### 4.1 APPLICANT INFORMATION
Date Prepared: November 5, 2012
Name and Address: Actuated Medical, Inc. 310 Rolling Ridge Drive Bellefonte, PA 16823 Ph: 814-355-0003 Fax: 814-355-1532
Contact Person:
Debora L. Demers, Ph.D., Director of QA/RA Ph: 814-355-0003 x112 Fax: 814-355-1532 Email: debora.demers@actuatedmedical.com
#### 4.2 DEVICE INFORMATION
Classification: KNT TubeClear™ Trade Name: In Patient Nasogastric Tube Clearing System Common Name: Tubes, Gastrointestinal and Accessories, Classification Name: 21 C.F.R. §876.5980
Attachment 4, Pg. 1 of 4
{1}------------------------------------------------
#### 4.3 PREDICATE DEVICE
TubeClear, manufactured by Actuated Medical, Inc. (AMI) and cleared under 510(k) K121571. is the Predicate Device. TubeClear is composed of a reusable Control Box and single use Clearing Stem. One Control Box model, Model 101 is used to actuate all Clearing Stem models. TubeClear has two Clearing Stem models, Models NG-1036 and NG-1043.
#### 4.4 PROPOSED DEVICE DESCRIPTION
The Proposed Device is TubeClear. TubeClear is composed of a reusable Control Box and single use Clearing Stem. One Control Box model. Model 101 is used to actuate all Clearing Stem models. TubeClear has two Clearing Stem models, Models NG-1036 and NG-1043.
The Operator manually inserts the Clearing Stem into the feeding or decompression tube (i.e., Tube) and directs the Clearing Stem's advancement along the inside of the Tube. The Control Box Motor via electromechanical actuation creates a linear reciprocating motion. The linear reciprocating motion is transferred to the proximal end of the Clearing Stem which contains a Wire that also reciprocates. Because the Wire is continuous throughout the Clearing Stem, the reciprocating motion is further transferred to the distal Tip of the Wire. The motion at the Wire Tip mechanically acts on the occlusion, breaks up the occlusion and restores Tube patency.
Because the Control Box remains outside of the patient and it functions only to provide actuation to the Clearing Stem when the two are attached, the Clearing Stem is considered to be the primary element of TubeClear.
Attachment 4, Pg. 2 of 4
{2}------------------------------------------------
#### 4.5 INDICATIONS FOR USE
tuatedMedical.com
The Proposed Device and the Predicate Device, have the exact same indications for use. TubeClear is indicated for use only and solely in clearing occlusions / clogs in nasogastric (NG) tubes (i.e., tubes that are placed through the nose and reside in the stomach) of adult patients. Clearing Stem Model NG-1036 is indicated for use in nasogastric tubes that are of size 10 - 14 French and have a length of 36 - 42 inches (91 -Clearing Stem Model NG-1043 is indicated for use in 108 cm). nasogastric tubes that are of size 10 - 18 French and have a length of 43 - 50 inches (109 - 127 cm).
## 4.6 TECHNOLOGICAL CHARACTERISTICS
The Proposed Device and the Predicate Device, have the exact same design features with one exception. The Predicate Device has pad printed centimeter markings on a 45 cm (18 in) segment of the Clearing Stem and the Proposed Device has NO pad printed centimeter markings on the 45 cm (18 in) segment of the Clearing Stem.
#### 4.7 RISK ANALYSIS
The risk analysis methods used to assess the safety and efficacy impact of the design modification were Preliminary Hazard Analysis (PHA) and Failure Modes and Effects Analysis (FMEA). The Design Requirements Document, Design Input Output Matrix, and Finished Product Specification were also reviewed.
It was determined that the removal of centimeter marks did not add any new safety risk to TubeClear. The presence of centimeter markings is not a design requirement.
Attachment 4, Pg. 3 of 4
{3}------------------------------------------------
It was determined that the previously conducted design verification activities were still applicable. The proposed design modification will not affect the following:
- 1. Operation and performance of the Control Box.
- 2. Performance of the Clearing Stem.
- 3. Shelf Life.
- 4. Biocompatibility.
- 5. Packaging and Transportation.
- 6. Software / Firmware.
- 7. Electromagnetic Compatibility / Electrical Safety.
# 4.8 NON-CLINICAL PERFORMANCE DATA
Bench Testing focused on Usability Testing to support Design Validation. Usability and Retainability testing by end users passed all acceptance criteria.
## 4.9 CONCLUSIONS
The Proposed Device, TubeClear is substantially equivalent to the Predicate Device, TubeClear.
The Devices have the SAME:
- 1. Indications for Use.
- 2. Operating Principle.
- 3. Materials.
- 4. Construction.
- 5. Shelf Life.
- 6. Packaging.
Attachment 4, Pg. 4 of 4
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three lines representing its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2012
Actuated Medical, Inc. % Debora L. Demers, Ph.D. Director, Quality Assurance and Regulatory Affairs 310 Rolling Ridge Drive BELLEFONTE PA 16823
Re: K123659
Trade/Device Name: TubeClear™ Model 101 Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: November 26, 2012 Received: November 30, 2012
Dear Dr. Demers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use based in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of
{5}------------------------------------------------
Page 2 - Debora L. Demers, Ph.D.
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you desire speethly advisor out FDA/Centers Offices/CDRH/CDRHOffices/ucm115809.htm for go to mtp.//www.radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regulation untilied CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### ATTACHMENT 2
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _K123659
Device Name: TubeClear™ Model 101
# Indications for Use Statement
TubeClear™ is indicated for use only and solely in clearing occlusions / clogs in nasogastric (NG) tubes (i.e., tubes that are placed through the nose and reside in the stomach) of adult patients. Clearing Stem Model NG-1036 is indicated for use in nasogastric tubes that are of size 10 - 14 No-1050 is Indication 187 acc 18 - 42 inches (91 - 108 cm). Clearing Stem Preficit and have a longth of Sor use in nasogastric tubes that are of size Model NO-1040 10 have a length of 43 - 50 inches (109 - 127 cm).
Over-The-Counter Use AND/OR Prescription Use x (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fisher -S 2012.12.20 10:14:32 -05'00'
(Division Sign-Off) i. sion of Reproductive, Gastro-Renal, and rological Devices 510(k) Number 1123659
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.