K130858 · Reverse Medical Corporation · DQY · Oct 11, 2013 · Cardiovascular
Device Facts
Record ID
K130858
Device Name
REVERSE MEDICAL MICROCATHETER
Applicant
Reverse Medical Corporation
Product Code
DQY · Cardiovascular
Decision Date
Oct 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reverse Medical Microcatheter-027 is intended for use in neuro, peripheral and coronary vasculature for the infusion of diagnostic agents such as contrast media, and therapeutic agents such as occlusion coils.
Device Story
The Reverse Medical Microcatheter-027 is a single-lumen, flexible, wire-reinforced composite catheter designed for accessing distal tortuous vasculature. It features a hydrophilic coating to reduce friction and dual radiopaque markers at the distal tip for fluoroscopic visualization. The device includes a shaping mandrel for physician-performed steam-shaping of the tip and a peel-away introducer. It is operated by physicians in clinical settings (neuro, peripheral, coronary). The catheter connects to standard luer-compatible accessories. By facilitating the delivery of diagnostic and therapeutic agents, the device assists in interventional procedures, potentially improving patient outcomes in vascular treatments.
Clinical Evidence
Bench testing only. No clinical data. Verification testing included biocompatibility (ISO 10993), sterilization validation (ISO 11135), and mechanical performance (dimensional, tip buckling, lubricity, flow rate, pressure, kink resistance, tensile strength, and navigation/pushability). All tests met acceptance criteria.
Technological Characteristics
Single-lumen, wire-reinforced composite catheter; variable stiffness; hydrophilic coating; dual radiopaque markers; luer adapter. Sterilized via EtO (SAL 10^-6). Complies with ISO 10993 for biocompatibility and ISO 11135 for sterilization.
Indications for Use
Indicated for use in neuro, peripheral, and coronary vasculature for infusion of diagnostic agents (contrast media) and therapeutic agents (occlusion coils).
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K130858
Image /page/0/Picture/1 description: The image shows the logo for Reverse Medical. The text "REVERSE MEDICAL" is in bold, black font. There is a curved line underneath the text that ends in an arrow pointing upwards.
# 510(k) Summary according to 807.92
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Reverse Medical Corporation is providing the summary of Substantial Equivalence for the Reverse Medical Microcatheter-027.
Sponsor /Applicant Name and Address 5.1
Reverse Medical Corporation 13700 Alton Parkway, Suite 167 Irvine, CA 92618
> 5.2 Sponsor Contact Information
Linda D'Abate, Vice President RA/CA/QA Phone: 949-215-0660 x202 Mobile 714-235-6608 Fax: 949-215-0661 Email: Idabate@reversemed.com
> 5.3 Date of Preparation of 510(k) Summary
October 4, 2013
5.4 Device Trade or Proprietary Name
Reverse Medical Microcatheter-027
ર .5 Device Common/Usual or Classification Name
Catheter, Percutaneous (Product Code: DQY) (21 CFR.870. 1250)and Catheter, Infusion (Product Code KRA) (21 CFR 870.1210) and Diagnostic Intravascular Catheter (DQO, 21CFR870.1200)
Identification of the Legally Marketed Devices to which Equivalence is Being 5.6 Claimed:
| Name of Predicate Devices | Name of Manufacturer<br>(City, State) | 510(k)<br>Number |
|-----------------------------------|---------------------------------------|------------------|
| Headway™ 27 Microcatheter | MicroVention, Inc.<br>Tustin, CA | K110813 |
| Excelsior® XT-27™ Microcatheter | Stryker Neurovascular<br>Fremont, CA | K113778 |
| Reverse Medical Microcatheter-021 | Reverse Medical Corp.<br>Irvine, CA | K122684 |
#### 5.7 Device Description
The Reverse Medical Microcatheter-027 is a single lumen, flexible stiffness composite catheter. The catheter shaft has a hydrophilic coating to reduce friction during use. The Reverse Medical Microcatheter-027 dimensions are included on the individual device labels. The Reverse Medical Microcatheter-027 inner lumen can accommodate guidewires up to 0.025 inches inner diameter to
{1}------------------------------------------------
access distal tortuous vasculature. Dual radiopaque markers at the distal portion of the catheter facilitate fluoroscopic visualization.
Each Reverse Medical Microcatheter-027 is provided with accessories, which include a shaping mandrel and peel away introducer within a Tyvek™ pouch.
The shaping mandrel allows the catheter tip to be steam shaped by the physician for proper adjustment to the anatomy prior to use. Test data is presented in this submission for both the 027 and 021 Reverse Medical Microcatheters to support the inclusion of the shaping mandrel and the peel away introducer.
The Reverse Medical Microcatheter-027 incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter is provided sterile, non-pyrogenic, and is intended for single use only.
#### 5.8 Intended Use
The Reverse Medical Microcatheter-027 is intended for use in neuro, peripheral, and coronary vasculature for the infusion of diagnostic agents such as contrast media, and therapeutic agents such as inclusion coils.
{2}------------------------------------------------
| | MicroVention<br>Headway™ 27<br>Microcatheter | Excelsior® XT-27™<br>Microcatheter | Reverse Medical<br>Microcatheter-021 | Reverse Medical<br>Microcatheter-027 |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K110813 | K113778 | K122684 | TBD |
| Classification | Class II, DQY | Class II, DQY | Class II, DQY and KRA | Class II, DQY and KRA |
| Indication | Intended for use in<br>the peripheral, coro-<br>nary and neurovascu-<br>lature for the infusion<br>of diagnostic agents,<br>such as contrast<br>media and therapeutic<br>agents such as<br>occlusion coils. | Intended to assist in<br>delivery of diagnostic<br>agents (such as con-<br>trast media), thera-<br>peutic agents, and<br>non-liquid intervene-<br>tional devices (such as<br>stents) intended for<br>use in the neurovascu-<br>lature and with a<br>catheter of 0.027" I.D. | Intended for use in<br>neuro, peripheral and<br>coronary vasculature<br>for the infusion of<br>diagnostic agents such<br>as contrast media, and<br>therapeutic agents<br>such as occlusion coils. | Intended for use in<br>neuro, peripheral and<br>coronary vasculature<br>for the infusion of<br>diagnostic agents such<br>as contrast media, and<br>therapeutic agents<br>such as occlusion coils. |
| Shaft<br>Materials | Coaxial lumen braided<br>shaft variable stiffness<br>catheter with<br>radiopaque marker on<br>distal end. | Semi-rigid proximal<br>shaft that transitions<br>into the flexible distal<br>shaft with single or<br>dual radiopaque mark-<br>ers at the distal end. | Single lumen, wire<br>reinforced shaft,<br>variable stiffness<br>catheter with dual<br>radiopaque markers<br>on distal end. | Single lumen, wire<br>reinforced shaft,<br>variable stiffness<br>composite catheter<br>with dual radiopaque<br>markers on distal end. |
| Proximal End<br>Configuration | Luer Hub | Luer Hub | Luer Hub | Luer Hub |
| Radiographic<br>markers/radi<br>opacity | Dual radiopaque<br>markers at distal tip. | Single or dual<br>radiopaque markers at<br>distal end of shaft. | Dual radiopaque<br>markers at distal tip. | Dual radiopaque<br>markers at distal tip. |
| Packaging | Polyethylene hoop<br>and PET/PE/Tyvek<br>pouch inside SBS<br>carton. | Polyethylene hoop<br>and PET/PE/Tyvek<br>pouch inside SBS<br>carton. | Polyethylene hoop<br>and PET/PE/Tyvek<br>pouch inside SBS<br>carton. | Polyethylene hoop<br>and PET/PE/Tyvek<br>pouch inside SBS<br>carton. |
| Sterilization | EtO | EtO | EtO | EtO |
| Peel Away<br>introducer | | Yes | Yes (data presented in<br>this submission) | Yes |
| Catheter Tip<br>Shaping<br>Mandrel | Yes | Yes | Yes (data presented in<br>this submission) | Yes |
#### 5.9 Comparison to Predicate Devices
{3}------------------------------------------------
#### Summary of Non-Clinical Data 5.10
## 5.10.1 Biocompatibility and Sterilization
The Reverse Medical Microcatheter-027 is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (s24 hours). Results of the testing demonstrate that the blood-contacting materials are biocompatible.
Blood-contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993-1 guidelines "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." The Reverse Medical Microcatheter-027 successfully passed all of the following biocompatibility tests:
| Test | Results | Conclusion |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Cytotoxicity<br>L929 MEM Elution Test | MEM Elution test scored a grade 0 (No cell<br>lysis) per ISO 10993-5. | Non-Cytotoxic |
| Sensitization<br>Kligman Maximization | Sensitization test scored a grade 0 (no visible<br>change) per ISO 10993-10 | Non-Sensitizing |
| Systemic Toxicity (Acute)<br>ISO Acute Systemic<br>Injection Test | Acute Systemic Injection Test Articles scored<br>0 with no toxicity or animal weight loss for<br>both the cottonseed oil and saline extracts<br>per ISO 10993-11 | Non-Toxic |
| Hemocompatibility:<br>Complement Activation | Under the conditions of the C3a assay, the<br>test article exhibited activation at 9421<br>ng/ml. This was 5.8%of the normalized C3a<br>concentration produced by CVF. Under the<br>conditions of the SC5b-9 assay, the test<br>article exhibited activation at 8836 ng/ml.<br>This was 0.1%of the normalized SC5b-9<br>concentration produced by CVF. All<br>biomaterials have the potential to affect the<br>make-up of the complement activation<br>components of the blood. | No greater biological<br>response than<br>corresponding control |
| | | |
| Hemolysis | The test article with negative control exhibited a hemolytic grade score of zero and is considered non-hemolytic. | Non- Hemolytic |
| Inactivated Partial<br>Thromboplastin Time | The negative and positive controls met the criteria for a valid assay. The variance between duplicate readings was less than 15%. The clotting times were as follow:<br>Positive control: 107 seconds<br>Negative control: 300 seconds<br>Reference material: 300 seconds<br>Test Article: 290 seconds<br><br>The test article is considered to be a minimal activator of the intrinsic coagulation pathway. The test article was considered to pass the test. | Non Activator |
| <i>In vivo</i> Thrombogenicity | Implantation of the test and control devices resulted in no adverse effects or clinical signs. | Non-Thrombogenic |
| Pyrogenicity<br>USP Material Mediated<br>Rabbit Pyrogen Test | Material-mediated pyrogenicity test was non-pyrogenic with no individual rabbit at any time having a temperature rise of greater than or equal to 0.5 deg. C per ISO 10993-11 | Non- Pyrogenic |
| EtO Residuals<br>Ethylene Oxide and<br>Ethylene Chlorohydrins<br>Residuals | ETO residuals met the criteria in accordance with ISO 10993-7 Part 7 | Passed, acceptable limits |
| Genotoxicity/Bacterial<br>Reverse Mutation | A statistically significant increase in the number of revertant colonies was not observed with the test article extracts as compared to the negative controls in both non-activated and S9 activated conditions. | Non-Mutagenic |
| Genotoxicity - Mouse<br>lymphoma | The test article extracts (saline and DMSO)<br>did not produce significantly more revertant<br>colonies than the negative controls. | Non-Mutagenic |
| Genotoxicity - In vivo<br>micronucleus | There was no statistically significant increase<br>in the number of micronucleus in the test<br>article extract (saline and sesame oil) vs.<br>negative control group. | Non-Mutagenic |
{4}------------------------------------------------
{5}------------------------------------------------
Sterilization conditions have been validated according to ANSI / AAMI / ISO 11135, Sterilization of Health Care Products-Part 1: Requirement, for development, validation and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level (SAL) of 10 °.
## 5.10.2 Design Verification (Bench-Top Testing)
The physical, mechanical, and performance testing of the Reverse Medical Microcatheter-027 demonstrate that the product is substantially equivalent to the currently marketed predicate devices. Design verification testing was conducted to evaluate the physical and mechanical properties of the Reverse Medical Microcatheter-027 with and without steam-shaping capabilities. All studies were conducted in accordance with Reverse Medical Design Control procedures. All testing was performed on units that were sterilized and met all inspection criteria. Tests on the Reverse Medical Microcatheter-027 included:
| Test Descriptions | Acceptance Criteria | Requirements | Pass/Fail |
|-------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------|-----------|
| Dimensional/Visual<br>Inspection | Per test protocol rev B | Pass/Fail | Pass |
| Tip Buckling Test | Peak Force Perform the same or better than<br>the predicate devices | 95/90 | Pass |
| Coating Lubricity Test | Frictional force<br>Coating Length | 95/90 | Pass |
| Flexibility/Shaft Stiffness<br>Test | The distal tip and proximal shaft flexibility<br>should perform comparably or better than the<br>predicate | Pass/Fail | Pass |
| Priming Volume Test | For comparison to predicate | Equal to or better than<br>predicate | Pass |
| Flow Rate Test | For comparison to predicate | Equal to or better than<br>predicate | Pass |
| Guidewire Compatibility<br>Test | Free movement of appropriately sized<br>guidewires<br>Frictional force | 95/90 | Pass |
{6}------------------------------------------------
| Guide Catheter<br>Compatibility Test | Free movement inside an appropriately sized<br>Guide Catheter. Frictional force measurement | 95/90 | Pass |
|-----------------------------------------------|---------------------------------------------------------------------------------------------|-----------|------|
| Catheter Air Leakage Test<br>-per ISO 10551-1 | Per ISO 10551-1 | Pass/Fail | Pass |
| Catheter Liquid Leakage<br>Test | No leaks in accordance with protocol | Pass/Fail | Pass |
| Dynamic Pressure Test | No leaks due to dynamic pressure test.<br>Peak Pressure: | Pass/Fail | Pass |
| Static Pressure Test | Burst pressure in accordance with protocol | 95/90 | Pass |
| Kink Resistance Test | Distal kink resistance<br>Medial/proximal: for characterization<br>purposes. | 95/90 | Pass |
| Torque Strength Test | X revolutions without failure. | Pass/Fail | Pass |
| Corrosion Resistance | No corrosion on metallic components. | Pass/Fail | Pass |
| USP Particulate Testing | Report total of particles. Compare to<br>predicate devices. | USP 788 | Pass |
| Tensile Strength | Distal/medial ; medial/proximal<br>Proximal/hub<br>Proximal/hub | 95/90 | Pass |
{7}------------------------------------------------
The physical, mechanical, and performance testing of the Reverse Medical Microcatheter-027 demonstrate that the product is safe and effective for its labeled indications and is Substantially Equivalent to the currently marketed predicate devices
| Test Descriptions | Acceptance Criteria | Requirement | Pass/Fail |
|-------------------------------------------|---------------------------------|-------------|-----------|
| Navigation/ Accessibility/<br>Pushability | Comparable to predicate devices | Pass/Fail | Pass |
| Micro-devices<br>Compatibility | Comparable to predicate devices | Pass/Fail | Pass |
In vitro Test Results for Reverse Microcatheter-027 Without Steam-Shaping Capabilities
## Performance/Functional Verification Test Results for Reverse Microcatheter-027 and 021 With Steam-Shaping Capabilities
| Test Descriptions | Acceptance Criteria | Requirements | Pass/Fail |
|-----------------------------------------------------|--------------------------------------------------|--------------|-----------|
| Dimensional and Visual<br>Inspection | Per test protocol rev B. | Pass/Fail | Pass |
| Steam Shaping Capabilities | No Damage<br>Shape Inspection: at various angles | Pass/Fail | Pass |
| Coating Lubricity Test - Post<br>Steam Shaping Test | Frictional force | Pass/Fail | Pass |
| Static Pressure Test - Post<br>Steam Shaping Test | Burst pressure | 95/90 | Pass |
| Tensile Strength - Post Steam<br>Shaping Test | Distal/Medial<br>Medial/Proximal<br>Proximal/Hub | 95/90 | Pass |
## In vitro Bench Test Results for Reverse Microcatheter-027 and 021 Post Steam-Shaping
| Navigation/ Accessibility/<br>Pushability | Comparable to predicate | Pass/Fail | Pass |
|-------------------------------------------|--------------------------------------------------------------------|-----------|------|
| Micro-devices Compatibility | Can allow uses of other micro-devices, in<br>simulated flow model. | | |
{8}------------------------------------------------
#### 5.11 Substantial Equivalence
The performance of the Reverse Medical Microcatheter-027 with and without steam-shaping capabilities in this submission demonstrates that the product is substantially equivalent to the performance of the predicate devices. The equivalence was shown through comparison of component materials and specifications, performance, biocompatibility testing, and sterilization validation.
The Reverse Medical Microcatheter-027 is substantially equivalent in intended use, design, technology/principles of operation, materials, and performance to the predicate devices. Differences between the devices do not raise any significant issues of safety or effectiveness.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure with three flowing lines extending from the figure's head, representing health, services, and the human aspect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6(19 Silver Spring, MI) 20993-0002
October 11, 2013
Reverse Medical Corporation Mr. Jeffrey Valko President/Chief Executive Officer 13700 Alton Parkway, Suite 167 Irvine. CA 92618
Re: K130858
Trade/Device Name: Reverse Medical Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Perculancous Regulatory Class: Class II Product Code: DQY. KRA and DQO Dated: September 6, 2013 Received: September 6, 2013
Dear Mr. Valko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical
{10}------------------------------------------------
Page 2 - Jeffrey Valko
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
## Indications for Use
510(k) Number: K130858
Device Name: Reverse Medical
Indications For Use:
The Reverse Medical Microcatheter-027 is intended for use in neuro, peripheral and coronary vasculature for the infusion of diagnostic agents such as contrast media, and therapeutic agents such as occlusion coils.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Joyce M. Whang -S
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.