Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
Product Code
MHX · Cardiovascular
Decision Date
Jun 14, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment. The monitors are additionally intended for use in transport situations within hospital environments. The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI ECI1). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
Device Story
IntelliVue MX400/MX450 are multiparameter patient monitors (9" and 12" displays) designed for surgical, cardiac, medical, and neonatal care. Devices lack built-in sensors; they connect to external Philips measurement modules (e.g., M3001A, M3002A, gas modules) to acquire physiological signals (ECG, SpO2, NBP, CO2, etc.). Monitors process these inputs to display waveforms, trends, and values, and generate physiological/technical alarms. Used at bedside or during hospital transport by trained clinicians. Touch-screen interface, remote control, or mouse/keyboard input. Data management and connectivity to Central Station via LAN. Benefits include continuous patient status visibility and clinical decision support (e.g., ProtocolWatch, IPI) to aid provider assessment and intervention.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including regression testing of physiological measurement parameters, alarm systems, battery functionality, and human interface. Environmental (temperature, humidity) and mechanical (shock, vibration, free fall) testing performed. Compliance with consensus standards (IEC 60601-1, 60601-1-2, 60601-1-8, 62304) verified.
Technological Characteristics
Plastic housing; TFT LCD widescreen display; CPU/memory/interface upgrades. Connectivity via LAN to Central Station. Powered by internal battery. Reuses existing Philips measurement modules (M3001A, M3002A, etc.) and accessories. Software life cycle per IEC 62304. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-8 (alarms).
Indications for Use
Indicated for monitoring physiological parameters in adult, pediatric, and neonatal patients by healthcare professionals in hospital environments, including transport. Contraindications: Not for home use; not therapeutic. Specific limitations: ST segment monitoring (adults only); SSC Sepsis Protocol (adults only); IPI (adults and pediatric 1-12 years); PPV (sedated adults on controlled mechanical ventilation, free from arrhythmia).
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
## 510(k) Summary
# JUN 1 4 2013
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- l . Submitter of this premarket notification
Markus Stacha Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: +49 7031 463-2840 Fax: +49 7031 463-2442 e-mail: markus.stacha@philips.com
This summary was prepared on May 13, 2013.
- 2. The name and classification of the device
Trade name: IntelliVue Patient Monitor MX400 and MX450 Common name: Multiparameter Patient Monitor
Classification:
| Device Panel | Classification | ProCode | Description |
|---------------------------|----------------|---------|--------------------------------------------------------------------------|
| Cardiovascular<br>Devices | §870.2300, II | DRT | Monitor, Cardiac (incl. Cardiotachometer<br>& Rate Alarm) |
| | §870.1025, II | DSI | Detector and alarm, arrhythmia |
| | §870.1025, II | MLD | Monitor, ST Segment with Alarm |
| | §870.1025, II | MHX | Monitor, Physiological, Patient (with<br>arrhythmia detection or alarms) |
| | §870.1100, II | DSJ | Alarm, Blood Pressure |
| | §870.1110, II | DSK | Computer, Blood Pressure |
| | §870.1130, II | DXN | System, Measurement, Blood-Pressure,<br>Non-Invasive |
| | §870.1435, II | DXG | Computer, Diagnostic, Pre-Programmed,<br>Single-Function |
| | §870.1915, II | KRB | Probe, Thermodilution |
| | §870.2060, II | DRQ | Amplifier and Signal Conditioner,<br>Transducer Signal |
| | §870.2340, II | DPS | Electrocardiograph |
| | §870.2340, II | MLC | Monitor, ST Segment |
| | §870.2350, II | DRW | Electrocardiograph, Lead Switching<br>Adapter |
| | §870.2370, II | KRC | Tester, Electrode, Surface,<br>Electrocardiograph |
| | §870.2600, I | DRJ | System, Signal Isolation |
| | §870.2700, II | DQA | Oximeter |
| | §870.2770, II | DSB | Plethysmograph, Impedance |
| | §870.2800, II | DSH | Recorder, Magnetic tape, Medical |
| | §870.2810, I | DSF | Recorder, Paper Chart |
Page 106 of 110
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are large and black, contrasting with the white background. The font appears to be slightly distressed, giving it a textured look. The word is horizontally oriented and centered.
| - Device Panel | Classification | ProCode· | Description<br>: . |
|----------------------------------------------|----------------|----------|----------------------------------------------------------------------------|
| | 8870.2850, II | DRS | Extravascular Blood Pressure Transducer |
| | 8870.2900, I | DSA | Cable, Transducer and Electrode, incl.<br>Patient Connector |
| | | MSX | System, Network and Communication,<br>Physiological Monitors |
| | | | |
| Anesthesiology<br>Devices | \$868.1400, II | ССК | Analyzer, Gas, Carbon Dioxide, Gaseous-<br>Phase |
| | \$868.1500, II | сво | Analyzer, Gas, Enflurane, Gaseous-Phase<br>(Anesthetic Concentration) |
| | \$868.1500, II | NHO | Analyzer, Gas, Desflurane, Gaseous-<br>Phase (Anesthetic Concentration) |
| | \$868.1500, II | NHP | Analyzer, Gas, Sevoflurane, Gaseous-<br>Phase (Anesthetic Concentration) |
| | 8868.1500, II | NHQ | Analyzer, Gas, Isoflurane, Gaseous-Phase<br>(Anesthetic Concentration) |
| | 8868.1620, II | CBS | Analyzer, Gas, Halothane, Gaseous-Phase<br>(Anesthetic Concentration) |
| | 8868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide, Gaseous-<br>Phase (Anesthetic Concentration) |
| | \$868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-Phase |
| | \$868.1880, II | BZC | Data calculator Pulmonary-function |
| | \$868.2375, II | BZQ | Monitor, Breathing Frequency |
| | \$868.2480, II | LKD | Monitor, Carbon Dioxide, Cutaneous |
| | 8868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for Infant<br>not under Gas Anesthesia |
| General Hospital and<br>Personal Use Devices | 8880.2910, II | FLL | Thermometer, Electronic, Clinical |
- 3. The subject devices Philips IntelliVue Patient Monitor MX400 and MX450, release K,00, are substantially equivalent to the previously cleared Philips Intelli Vue Patient Monitor MP30 marketed pursuant to K122439, K120366, K113657, K113441, K110622, K102562, . K093268, K083517, K082633, K071426, K063315, K062392, K062283, K061610, K060221, K060541, K061052, K052961, K052801, K050762, K050141, K042845, K041235.
- Description of the device 4.
The IntelliVue Patient Monitor MX400 and MX450 is a modification of the legally marketed IntelliVue Patient Monitor MP30.
The MX400 and MX450 are display units with a TFT LCD flat panel display and built-in CPU. The models MX400 and MX450 differ only in the size of their flat panel displays: MX400 has a 9" and MX450 has a 12" display.
The MX400 and MX450 do not have any built-in measurements. They can be connected to one Philips M3001A Multi-Measurement Server, M3002A (X2) Multi-Measurement Module and any one of the M3012A, M3014A, M3015A/B or M3016A Measurement Server Extensions, as well as to the M1013A (G1) or M1019A (G5) Gas Modules. These external
Page 107 of 110
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are black and appear to have a slightly distressed or textured effect, as if they have been printed on a rough surface. The word is presented in all capital letters and is horizontally oriented.
devices provide non-invasive and invasive measurements such as ECG with arrhythmia and ST, respiration, SpO2, NBP, invasive blood pressure, temperature, CO2, C.O., CCO.
The MX400 and MX450 acquire multiple physiological patient signals (via connected external measurement modules), display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management.
The MX400 and MX450 offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located in the patient vicinity either at the bedside or used mobile, during patient transport in the hospital.
The MX400 and MX450 have a color display with touch-screen as primary input device. They also support a specialized remote control, keyboard and pointing devices such as a mouse.
The MX400 and MX450 interact with the connected external measurement devices locally at the bedside or in transport situations and with the Central Station via LAN.
- న. Intended Use
The subject devices IntelliVue Patient Monitor MX400 and MX450 have the same intended use as the legally marketed predicate IntelliVue Patient Monitor MP30.
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The monitors are additionally intended for use in transport situations within hospital environments.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMIECII).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
{3}------------------------------------------------
### 6. Technological Characteristics
The subject devices IntelliVue Patient Monitors MX400 and MX450 have the same range of functions as the legally marketed predicate IntelliVue Patient Monitor MP30. MX400 and MX450 reuse unchanged external measurement modules and measurement accessories as already used by the predicate MP30.
MX400 and MX450 have the same relevant-technological characteristics as the predicate MP30 with regard to energy sources, portability, user interface, and robustness.
Compared to the predicate MP30, the MX400 and MX450 have modified technological characteristics with regard to the usage of state of the art hardware components such as widescreen display, CPU, memories and interfaces, regarding a more compact form factor, resulting in slightly different dimensions and weight, regarding environmental conditions and operating time with internal battery.
The housing of the MX400 and MX450 is made of plastic whereas the housing of the predicate IntelliVue Patient Monitor MP30 is made of plastic and partly of magnesium alloy. Biocompatibility aspects are not affected because the devices do not have contact with patients. The MX400 and MX450 reuse plastic materials of the predicate MP30.
The fundamental scientific technology of the new IntelliVue Patient Monitors MX400 and MX450 is the same as that of the predicate IntelliVue Patient Monitor MP30. The differences in technological characteristics between the subject devices MX400 and MX450 and the predicate device MP30 do not diminish safety or effectiveness.
- 7. Summary of V&V activities
- Testing according to the recognized consensus standards:
- IEC 60601-1-2:2001 +A1:2004 (EMC) .
- IEC 60601-1:1988 +A1:1991+A2:1995 (General Safety) .
- . IEC 60601-1-8:2003 (Alarms)
- IEC 62304:2006 (Software life cycle processes) .
All applicable requirements have been met.
- -Environmental testing (temperature, humidity), mechanical testing (shock, vibration, and free fall).
All specified test requirements have been met. The tests confirmed that the Intelli Vue Patient Monitor MX400 and MX450 worked safely and according to their specifications and indicated claims during tests simulating general hospital conditions, handling and transport in hospital environments, and storage.
- -Regression testing on the IntelliVue Patient Monitor MX400 and MX450. The performed regression testing comprised function tests of all physiological measurement parameters (which remain unchanged in this submission), integrated recorder, display, batterv functionality, battery operating and charging time, human interface, external interfaces, alarming system, and white box tests as identified by the hazard analysis. All specified criteria have been met. The tests demonstrated that the IntelliVue Patient Monitor MX400 and MX450 worked safely, effectively, and correctly in accordance with specifications and labeling claims in the intended environment of use.
{4}------------------------------------------------
# PHILIPS
## 8. Conclusion
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified devices with respect to the predicate.
Testing comprised regression tests, white box tests, electrical and mechanical safety tests, EMC tests, and environmental tests.
Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence.
The results demonstrate that the Philips IntelliVue Patient Monitor MX400 and MX450 meet all defined reliability requirements and performance claims.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, with three wavy lines intersecting a staff-like element.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
June 14, 2013
Philips Medizinsysteme Boeblingen Gmbh c/o Markus Stacha Sr. Regulatory Affairs Engineer Hewlett-packard Str.2 71034 Boeblingen Germany
Re: K130849
> Trade/Device Name: IntelliVue patient monitor mx400; IntelliVue patient monitor mx450
Regulation Number: 21 CFR 870.1250
Regulation Name: Multiparameter Patient Monitor
Regulatory Class: Class II
Product Code: MHX, DRT, DSI, MLD, DSJ, DSK, DXN, DXG, KRB, DRQ, DPS, MLC, DRW, KRC, DRJ, DQA, DSB, DSH, DSF, DRS, DSA, MSX, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZC, BZO, LKD, KLK, FLL
Dated: May 13, 2013 Received: May 16, 2013
Dear Markus Stacha:
-We-have-reviewed-your-Section-510(k)-premarket-notification of intent-to-market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2 - Markus Stacha
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Owen-P.Faris-S
for Bram D. Zuckerman, M.D. Director
> Division-of-Cardiovascular-Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: IntelliVue Patient Monitor MX400 and MX450
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The monitors are additionally intended for use in transport situations within hospital environments.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI ECI1).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
Prescription Use Yes Over-The-Counter Use No AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER: PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen P. Faris -S
2013.06.13
15:50:16 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.