THE PHILLIPS, MODEL MP20,MP30,MP40,MP50,MP60,MP70,MP80,AND MP90 INTELLIVUE PATIENT MONITORS
Applicant
Philips Medizin Systeme Boeblingen GmbH
Product Code
MHX · Cardiovascular
Decision Date
Feb 17, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare facilities. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within healthcare facilities. ST Segment monitoring is restricted to adult patients only. The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.
Device Story
IntelliVue Patient Monitors (MP20-MP90) collect physiological data via sensors/electrodes; process signals for real-time monitoring, recording, and alarming; display data for clinician review. Used in hospital environments (including transport for MP20-MP50) by healthcare professionals. Output informs clinical decision-making regarding patient status. Release D.02 software update provides enhanced monitoring capabilities. Benefits include continuous patient surveillance and timely alerts for physiological changes.
Clinical Evidence
Bench testing only. Performance, functionality, and reliability were established through system-level testing, pass/fail analysis, and safety testing compared against cleared predicate devices.
Technological Characteristics
Multi-parameter physiological monitoring system. Includes arrhythmia detection, ST segment monitoring, blood pressure measurement, gas analysis, and thermodilution. Connectivity via network/communication systems. Software-based signal processing and alarm management. Class II device.
Indications for Use
Indicated for monitoring, recording, and alarming of physiological parameters in adult, pediatric, and neonatal patients in healthcare facilities. ST segment monitoring restricted to adults. Transcutaneous gas measurement (tcpO2/tcpCO2) restricted to neonates.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Respironics Novametrix Rostion, Rossister and Accessories PN 3468-3475 (K053174)
Submission Summary (Full Text)
{0}------------------------------------------------
K060221
p1/3
## FEB 1 7 2006
## 510 (k) Summary 8.0
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe
is submitted in accordance with the requirements of the Safe 13 Submitted In accordinate of 1990 and 21 C.F.R. \$807.92.
1. The submitter of this premarket notification is: Herbert van Dyk Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany - 71054 Bocking - 463-1734 - Fax: ++49 7031 463-2442 e-mail: herbert.van_dyk@philips.com
This summary was prepared on January 27, 2006.
- 2. The names of the devices are the Philips MP20, MP30, MP40,
MP40, MP30, Lunco, Lunco, Intellilly, Pations Monitor The Mames Of the devices a 8990 IntelliVue Patient Monitors. Classification names are as follows:
| Device Panel | Classification | ProCode | Description |
|------------------------------------------|----------------|---------|-----------------------------------------------------------------------------|
| Circulatory<br>System Devices<br>(12625) | \$870.1025, II | DSI | Detector and alarm, arrhythmia |
| | \$870.1025, II | MLD | Monitor, ST Segment with Alarm |
| | \$870.1025, II | MHX | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) |
| | \$870.1100, II | DSJ | Alarm, Blood Pressure |
| | \$870.1110, II | DSK | Computer, Blood Pressure |
| | \$870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive |
| | \$870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function |
| | \$870.1915, II | KRB | Probe, Thermodilution |
| | \$870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer Signal |
| | \$870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) |
| | \$870.2300, II | MSX | System, Network and<br>Communication, Physiological<br>Monitors |
| | \$870.2340, II | DPS | Electrocardiograph |
| | \$870.2340, II | MLC | Monitor, ST Segment |
| | \$870.2350, II | DRW | Electrocardiograph, Lead<br>Switching Adapter |
| | \$870.2370, II | KRC | Tester, Electrode, Surface,<br>Electrocardiograph |
| | \$870.2450, II | DXJ | Display, Cathode-Ray Tube,<br>Medical |
| | \$870.2600, I | DRJ | System, Signal Isolation |
| | \$870.2700, II | DQA | Oximeter |
| | \$870.2770, II | DSB | Plethysmograph, Impedance |
| | \$870.2800, II | DSH | Recorder, Magnetic tape,<br>Medical |
| | \$870.2810, I | DSF | Recorder, Paper Chart |
| | \$870.2850, II | DRS | Extravascular Blood Pressure<br>Transducer |
COMPANY CONFIDENTIAL
{1}------------------------------------------------
| | \$870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector |
|------------------------------------------------------|-------------------------|----------------|------------------------------------------------------------------------------|
| | \$870.2910, II | DRG | Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency |
| Anesthesiology<br>and Respiratory<br>Therapy (12624) | \$868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase |
| | \$868.1500, II | CBQ | Analyzer, Gas, Enflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1500, II | NHO | Analyzer, Gas, Desflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1500, II | NHP | Analyzer, Gas, Sevoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1500, II | NHQ | Analyzer, Gas, Isoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1620, II | CBS | Analyzer, Gas, Halothane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-<br>Phase |
| | \$868.1880, II | BZC | Data calculator Pulmonary-<br>function |
| | \$868.2375, II | BZQ | Monitor, Breathing Frequency |
| | \$868.2480, II | LKD | Monitor, Carbon Dioxide,<br>Cutaneous |
| | \$868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia |
| General Hospital<br>and Personal Use<br>(12520) | \$880.2910, II | FLL | Thermometer, Electronic,<br>Clinical |
| | Neurological<br>(12513) | \$882.1400, II | GWR |
| | | \$882.1420, I | GWS |
- The modified devices are substantially equivalent to 3. me modified dored Philips devices marketed pursuant to K014159, K021778, K030038, K032858, K040304, K040183 and K040357, K041235, K41741, K042845, K050141, 050762 and K040337, K041259, K1273522, and to Respironics Novametrix Rostion, Ro32001 , Rossister and Accessories "PN 3468-3475" (K053174)
- The modification is the introduction of Release D.02 software 4 . The MOGILLEACION ID che Internitor devices, MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90
- The modified devices have the same intended use as the 5. legally marketed predicate devices. They are intended for the regarry marketa ding, and alarming of multiple physiological Monicoring, recording, and trics, and neonates in hospital
COMPANY CONFIDENTIAL
{2}------------------------------------------------
environment and during transport within hospital environments .
- The modified devices have the same technological 6. The modified devices have the band elected predicate devices.
- 7. Verification, validation, and testing activities establish verformance, functionality, and reliability the performance, functionally, mevice with respect to the characteristics of the modified as a level tests, performance
predicate. Testing involved system levely - Rang (Fail predicate. Testing involved bybeard analysis. Pass/Fail tests, and safety testing from hassisions cleared for the criteria were babed on storesults showed substantial predicate device and other demonstrate that the Philips equivalence. The rebuild acments and reliability requirements and performance claims.
## COMPANY CONFIDENTIAL
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 7 2006
Philips Medizinsysteme Boeblingen GmbH c/o Mr. Herbert van Dyk Sr. Regulatory Affairs Engineer Philips Medical Services 3000 Minuteman Road Andover, MA 01810
Re: K060221
文
K 060221
Trade Name: The Philips MP20, MP30, MP30, MP60, MP70, MP80 and MP90
Trade Name: The Philips MP30, Dalam Dalam D 02 Intellivue Patient Monitors, Release D.02 Regulation Number: 21 CFR 870.1025 Regulation Name: 21 CFR 870.1023
Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: Class II (two) Product Code: MHX Dated: January 27, 2006 Received: January 30, 2006
Dear Mr. Van Dyk:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 510(x) prematics in the manufated (for the indications referenced and nave decimined the device are devices marketed in interstate for use stated in the encrosule) to regally manced provice Americal Device Ameralments, or to commerce prior to May 20, 1970, the chaouthore with the provisions of the Federal Food, Drug, devices that have been reclassified in accordance what approval approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval applic and Cosmetic Act (Act) that do not require approvince approvisions of the Act. The Act. The You may, merelore, market the device, saloject to the greateration, listing of
general controls provisions of the Act include requirements for annual registration, listing of general controls provisions of the Are, Inc. Increase some in misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) me existing major regulations affecting your device can may be subject to such additional controlis. Entroling the 300 to 898. In addition, I'DA may be found in the Ood of Peachartergerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. Herbert van Dyk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that i Drederal statutes and regulations administered by other Federal agencies. You must or uny I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI K Part 607), laoomig (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control pro begin marketing your device as described in your Section 510(k) This letter will and w yourse FDA finding of substantial equivalence of your device to a legally marketed predicated. " caresults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific as not any at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bhimmar fa
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release D.02
Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarmine of multiple physiological parameters of adults, pediatrics and memates in healthcare facilities. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within healthcare facilities.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.
Prescription Use yes___________________________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ No (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummenstor
n Sign-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.