STRAUMANN NARROW NECK CROSSFIT (NNC) 03.3MM DENTAL IMPLANT SYSTEM
K111357 · Straumann USA · DZE · Oct 3, 2011 · Dental
Device Facts
Record ID
K111357
Device Name
STRAUMANN NARROW NECK CROSSFIT (NNC) 03.3MM DENTAL IMPLANT SYSTEM
Applicant
Straumann USA
Product Code
DZE · Dental
Decision Date
Oct 3, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Straumann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only large diameter implants. In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases. Specific indications for small diameter (Ø 3.3 mm) implants: Because of their reduced mechanical stability, small diameter implants are only used in cases with a low mechanical load. Placement in the molar region is not recommended. Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges and overdentures. The Straumann NNC Temporary Abutments are indicated for use in Straumann NNC Tissue Level Implants for temporary restorations of single crowns and bridges for up to six months. Straumann NNC Healing Caps and NNC Closure Screws are intended for use with the Straumann NNC Tissue Level Implant System to protect the inner configuration of the implant. Healing Caps have a secondary function to maintain, stabilize and form the soft tissue during the healing process.
Device Story
The Straumann Narrow Neck CrossFit (NNC) Ø3.3mm Dental Implant System is a tissue-level endosseous implant designed for smaller interdental spaces. The system includes the implant, gold abutments, temporary titanium abutments, closure screws, healing caps, impression posts, and analogs. The implant features a threaded body with an SLActive surface and an internal CrossFit connection. Used by dental clinicians in clinical settings to restore chewing function in edentulous or partially dentate patients. The device is placed surgically into the jawbone; prosthetic restorations (crowns, bridges, dentures) are attached via abutments. The system provides a stable foundation for dental prosthetics, enabling functional and esthetic oral rehabilitation. The device is intended for prescription use only.
Clinical Evidence
Bench testing only. Fatigue testing was conducted in accordance with the FDA guidance document 'Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments' to verify mechanical performance and ensure design output meets design input.
Indicated for edentulous and partially dentate patients requiring oral endosteal implantation for functional and esthetic rehabilitation. Small diameter (Ø 3.3 mm) implants are indicated only for low mechanical load cases; placement in molar region is not recommended. Temporary abutments indicated for single crowns and bridges for up to six months.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
K111357
OCT - 3 2011
#### Applicant's Name and Address 1.
Straumann US (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 Telephone Number: 800-448-8168, ext 2513 Fax Number: 978-747-0023 Elaine Alan Contact Person: Senior Regulatory Affairs Specialist
- Date of Submission: May 12, 2011 2.
#### 3. Name of the Device
| Trade Name: | Straumann Narrow Neck CrossFit® (NNC)<br>Ø3.3mm Dental Implant System |
|----------------------|-----------------------------------------------------------------------|
| Common Name: | NNC Ø3.3mm Dental Implant |
| Classification Name: | Implant, Endosseous, Root-form |
| Regulation Number: | §872.3640 |
### Legally Marketed Device to which Equivalence is Claimed (Predicate 1. Device)
K083550. Straumann Dental Implant System K060958, Straumann Narrow Neck Dental Implants K072497, P.004 NC Gold Abutment for Crowns K092814. Straumann NC Temporary Abutments K071585, NC Closure Screws K960634, Esthetic Healing Caps
#### Description of the Device 5.
The proposed Straumann Narrow Neck Connection CrossFit® (NNC) Ø3.3mm Dental Implant System is a new product to be added to our current Tissue Level implant product portfolio. Straumann currently has a Tissue Level dental implant with a Ø3.5mm prosthetic platform with an external hexagon abutment; the abutment and implant are machined as one piece.
{1}------------------------------------------------
The proposed system introduces a Ø3.5mm prosthetic platform with an internal CrossFit connection. The proposed system is an implant with no attached abutment. The abutment is sold separately.
The purpose of this 510(k) Premarket Notification is to introduce:
- Narrow Neck CrossFit® (NNC) Ø3.3mm dental implant in 4 lengths; a. 8.0mm, 10.0mm, 12.0mm and 14.0m,
- b. Gold abutment for use with the crown,
- c. NNC posts for temporary restorations made from Titanium. The posts are to be used for a single crown restoration or bridge,
- d. NNC closure screws (0mm and 1.5mm),
- e. Healing caps (3.0mm, 4.5mm and 2.0mm),
- f. NNC impression post, open tray, and
- g. NNC analog.
### Intended Use of the Device 6.
Narrow Neck CrossFit® (NNC) Dental Implants
Straumann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only
{2}------------------------------------------------
large diameter implants. In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.
Specific indications for small diameter (Ø 3.3 mm) implants: Because of their reduced mechanical stability, small diameter implants are only used in cases with a low mechanical load. Placement in the molar region is not recommended.
Straumann Narrow Neck CrossFit® (NNC) Gold Abutment for Crowns Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges and overdentures.
Straumann Narrow Neck CrossFit® (NNC) Temporary Abutments
The Straumann NNC Temporary Abutments are indicated for use in Straumann NNC Tissue Level Implants for temporary restorations of single crowns and bridges for up to six months.
Straumann Narrow Neck CrossFit® (NNC) Healing Caps and Closure Screws
Straumann NNC Healing Caps and NNC Closure Screws are intended for use with the Straumann NNC Tissue Level Implant System to protect the inner configuration of the implant. Healing Caps have a secondary function to maintain, stabilize and form the soft tissue during the healing process.
### 7. Technological Characteristics
The body of the proposed implant has a threaded Ø3.3mm implant body design with the SLActive surface of Straumann's TiZr Implant currently
{3}------------------------------------------------
cleared under K083550. The material of the proposed implant is Straumann's Titanium Zirconium (TiZr) currently cleared under K083550. This design proposes a narrow Tissue Level implant with an internal connection for smaller interdental spaces. There are no changes to the materials, indications for use, fundamental operating principles, or sterilization processes or procedures as a result of the new implant and its accessories and components. No new surgical instruments are being introduced as placement of the proposed implant will follow the established surgical protocols of the currently cleared Straumann Dental Implant Systems. The proposed devices are substantially equivalent to the currently marketed devices. .
### 8. Performance Testing
Verification and validation testing were performed to ensure that the devices subject to this 510(k) Premarket Notification function as intended and that design input matches design output. Testing included:
- 1. Performance Testing
- a. Fatigue Testing in accordance to FDA guidance document "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments."
### 9. Conclusion
The results from the testing conducted demonstrated that the Straumann Narrow Neck CrossFit® (NNC) Ø3.3mm Dental Implant System functions as intended and met the pre-determined acceptance criteria.
The Straumann Narrow Neck CrossFit® (NNC) Ø3.3mm Dental Implant System is a validated system. The results of the performance bench testing and risk analysis indicate that the Straumann Narrow Neck
{4}------------------------------------------------
CrossFit (NNC) Ø3.3mm Dental Implant System is substantially equivalent to the named predicate devices and is safe and effective for its intended use.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a sans-serif font. The text is black and is on a white background. There is a logo to the left of the text.
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a symbol. The symbol consists of three stylized human figures connected by flowing lines, representing the department's mission to protect the health of all Americans and provide essential human services.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 2093-0002
Ms. Elaine Alan Senior Regulatory Affairs Specialist Straumann USA 60 Minuteman Road Andover. Massachusetts 01810
OCT = 3 201
Re: K111357
Trade/Device Name: Straumann Narrow Neck CrossFit (NNC) 03.3 Dental Implant System, Straumann Narrow Neck CrossFit (NNC) Gold Abutment for Crowns. Straumann Narrow Neck CrossFit (NNC) Temporary Abutments, Straumann -Narrow Neck CrossFit (NNC) Healing Caps and Closure Screws Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: August 19, 2011 Received: August 22, 2011
Dear Ms. Alan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 – Ms. Alan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Fleath's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.hum.
Sincerely yours.
hn for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known):
Device Name: Narrow Neck CrossFit (NNC) Ø3.3mm Dental Implant System
Indications for Use:
Straumann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only large diameter implants. In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.
Specific indications for small diameter (Ø 3.3 mm) implants: Because of their reduced mechanical stability, small diameter implants are only used in cases with a low mechanical load. Placement in the molar region is not recommended.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
| Concurrence of DCRH, Office of Device Evaluation (ODE) | |
|-----------------------------------------------------------------------------------|----------------|
| (Division Sign-Off) | |
| Division of Anesthesiology, General Hospital<br>Infection Control, Dental Devices | Page 1 of 1 |
| RE: 510(k) Straumann NNC<br>Dental Implant System | |
| 510(k) Number: | K111357 |
| May 13, 2011 | Page 16 of 126 |
{8}------------------------------------------------
510(k) Number (if known):
Device Name: Straumann Narrow Neck CrossFit (NNC) Gold Abutment for Crowns
Indications for Use:
Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges and overdentures.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of DCRH, Office of Device Evaluation (ODE)
Division: State of Illinois
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111357
{9}------------------------------------------------
510(k) Number (if known):
Device Name: Straumann Narrow Neck CrossFit (NNC) Temporary Abutments
Indications for Use:
The Straumann NNC Temporary Abutments are indicated for use in Straumann NNC Tissue Level Implants for temporary restorations of single crowns and bridges for up to six months.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of DCRH, Office of Device Evaluation (ODE)
Zl
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k111357
{10}------------------------------------------------
510(k) Number (if known):
Device Name: Straumann Narrow Neck CrossFit (NNC) Healing Caps and Closure Screws
Indications for Use:
Straumann NNC Healing Caps and NNC Closure Screws are intended for use with the Straumann NNC Tissue Level Implant system to protect the inner configuration of the implant. Healing Caps have a secondary function to maintain, stabilize and form the soft tissue during the healing process.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of DCRH, Office of Device Evaluation (ODE)
h
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k 11357
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.