The Zoe Medical Nightingale Monitoring System is indicated for use in adult & pediatric patient populations. The Zoe Medical Nightingale Monitoring System facilitates the monitoring of: - ECG - - Impedance respiration - - Non-Invasive blood pressure - - Invasive blood pressure - - Body temperature - -Functional arterial oxygen saturation (SpO2) - -End-tidal & inspired CO2 The Zoe Medical Nightingale Monitoring System is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
Device Story
Nightingale Monitoring System (NMS) provides bedside (PPM3 monitor) and remote (MPC central station) physiological monitoring. Inputs: ECG, impedance respiration, NIBP, IBP, body temperature, SpO2, and CO2 (via Oridion Microstream). System processes these signals to display real-time waveforms and vital signs; provides 72-hour trending; supports up to 64 monitors at central station. Operated by healthcare professionals in clinical settings. Output used for patient status assessment and clinical decision-making. Benefits include continuous monitoring of critical physiological parameters to facilitate timely clinical intervention.
Clinical Evidence
No clinical data. Bench testing only, including performance, functional, environmental, and shipping/transportation testing. Electrical safety and EMC testing performed per AAMI ES 60601-1:2005, IEC 60601-1-2:2007, and related standards.
Technological Characteristics
Physiological patient monitor; portable; indoor use. Parameters: ECG, impedance respiration, NIBP, IBP, temp, SpO2, CO2. 8.4" display; push-knob/key interface. Power: Medical grade supply with internal Li-ion battery. Connectivity: Central station support for 64 monitors. Standards: IEC 60601-1-2:2007 (EMC), AAMI ES 60601-1:2005 (Safety). Ingress protection: IPX1. Software: Verified and validated per IEC 62304.
Indications for Use
Indicated for adult and pediatric patients requiring physiological monitoring of ECG, impedance respiration, NIBP, IBP, body temperature, SpO2, and end-tidal/inspired CO2. Prescription use only by healthcare professionals in healthcare facilities.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
Spacelabs Healthcare Vital Signs Monitoring System (K090556)
Submission Summary (Full Text)
{0}------------------------------------------------
| Company: | Zoe Medical, Inc.<br>460 Boston Street<br>Topsfield, MA 01983<br>Phone: (978) 887-1410<br>FAX: (978) 887-1406 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration: | 3003294644 |
| Primary Contact: | James Chickering<br>Regulatory Affairs Manager |
| Date Prepared: | March 13, 2013 |
| Trade Name: | Zoe Medical Nightingale Monitoring System |
| Common Name: | Patient Physiological Monitor |
| Classification Name: | Monitor, Physiological Patient |
| Product Code: | MWI |
| Classification<br>Regulation: | 21 CFR 870.2300 |
#### Device Description
The Zoe Medical Nightingale Monitoring System (NMS) facilitates the monitoring of patient physiological parameters both at the bedside on the Nightingale PPM3 monitor and remotely on the Nightingale MPC central station. The physiological parameters monitored by the NMS include: ECG. impedance respiration, non-invasive blood pressure, invasive blood pressure, body temperature, functional arterial oxygen saturation (SpO2), and end-tidal & inspired CO2 (capnography). The Nightingale Monitoring System as described in this submission is a new device, and will be marketed as the next generation product superseding the original and previously cleared Zoe Medical NMS (K001775).
The part numbers for the four Nightingale PPM3 variants and their associated configurations are given below.
179-0003 NIGHTINGALE PPM3 (with NIBP, SPO2, ECG, and Temp)
179-1000 NIGHTINGALE PPM3 WITH IBP (with NIBP, SPO2, ECG, Temp, & IBP)
179-1001 NIGHTINGALE PPM3 WITH ETCO2 (with NIBP, SPO2, ECG, Temp, & Microstream ETCO2)
179-1002 NIGHTINGALE PPM3 (with NIBP, SPO2, ECG, Temp, IBP, & Microstream ETCO2)
The Nightingale MPC central station (171-0001) will be sold as an option available to all customers based on their needs.
{1}------------------------------------------------
#### Intended Use
The Zoe Medical Nightingale Monitoring System is indicated for use in adult & pediatric patient populations.
The Zoe Medical Nightingale Monitoring System facilitates the monitoring of:
- ECG -
- Impedance respiration -
- Non-Invasive blood pressure -
- Invasive blood pressure -
- Body temperature -
- -Functional arterial oxygen saturation (SpO2)
- -End-tidal & inspired CO2
The Zoe Medical Nightingale Monitoring System is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
## Substantial Equivalence
The Nightingale Monitoring System utilizes the same Zoe Medical and Oridion Medical components utilized in the primary predicate device, the Spacelabs Healthcare Vital Signs Monitoring System (K090556).
| Characteristic | Zoe Medical Nightingale<br>Monitoring System<br>(this submission) | Spacelabs Healthcare élance<br>Vital Signs Monitoring<br>System<br>(K090556) | Discussion of<br>Differences |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Intended Use | The Zoe Medical Nightingale<br>Monitoring System is indicated for<br>use in adult & pediatric patient<br>populations.<br><br>The Zoe Medical Nightingale<br>Monitoring System facilitates the<br>monitoring of:<br>- ECG<br>- Impedance respiration<br>- Non-Invasive blood pressure<br>- Invasive blood pressure<br>- Body temperature<br>- Functional arterial oxygen<br>saturation (SpO2)<br>- End-tidal & inspired CO2<br><br>The Zoe Medical Nightingale<br>Monitoring System is a<br>prescription device intended to be<br>used by healthcare professionals<br>in all areas of a healthcare facility. | The Spacelabs élance Vital<br>Signs Monitor is indicated for<br>use in patient populations for:<br>- Adult<br>- Pediatric<br><br>The Spacelabs élance Vital<br>Signs Monitor facilitates the<br>monitoring of:<br>- ECG with arrhythmia<br>detection<br>- Respiration<br>- Non-Invasive blood pressures<br>- Invasive blood pressures<br>- Body temperature<br>- Functional arterial oxygen<br>saturation<br>- Expired and/or inspired CO2<br><br>The Spacelabs élance Vital<br>Signs Monitor is a prescription<br>device intended to be used by<br>healthcare professionals in all<br>areas of a healthcare facility. | Nightingale not<br>intended for<br>arrhythmia<br>monitoring |
| Characteristic | Zoe Medical Nightingale<br>Monitoring System<br>(this submission) | Spacelabs Healthcare élance<br>Vital Signs Monitoring<br>System<br>(K090556) | Discussion of<br>Differences |
| Parameters | ECG, impedance respiration, non-<br>invasive blood pressure, invasive<br>blood pressure, body<br>temperature, functional arterial<br>oxygen saturation (SpO2), end-<br>tidal & inspired CO2 (Oridion<br>Microstream®). | Same | Optional<br>arrhythmia<br>components not<br>included in<br>Nightingale |
| Number of<br>Waveforms | 3 or 5 | Same | |
| User Interface | Push knob and dedicated keys | Touch screen | PPM3 meets<br>intended use |
| Alarm Levels | 3 per IEC 60601-1-8 | Same | |
| Operating Modes | Continuous | Same | |
| Trending | 72 hours | Same | |
| Size | 286 W x 182 H x 83 D (mm) w/<br>8.4" display | 252 W x 193 H x 70 D w/ 10.2"<br>display or 291 W x 221 H x 70<br>D (12.1" display) | PPM3<br>represents a<br>slimmed down<br>design |
| Electrical | Medical grade power supply with<br>internal lithium-ion battery | Same | |
| Equipment Type | Portable, Indoor Use Only | Same | |
| Ingress<br>Protection | IPX1 | Same | |
| Operating<br>Conditions | 0 to 40°C (32 to 104°F)<br>15 to 90% RH, non-condensing<br>0 to 4572 m (0 to 15,000') | Same | |
| Storage<br>Conditions | -20 to 60°C (-4 to 140°F)<br>15 to 95%, non-condensing<br>0 to 12192 m (0 to 40,000') | Same | |
| Electromagnetic<br>Compatibility | Meets IEC 60601-1-2:2007 | Same | |
| Central Station | Support for 64 monitors;<br>Parameters: ECG, Respiration,<br>NIBP, IBP, Temp, SpO2, EtCO2,<br>FiCO2; Alarming with Logging;<br>Printed Vital Sign, Summary,<br>Trend Reports | Same | |
## Table 1. Comparison of the Subject and Predicate Devices
{2}------------------------------------------------
{3}------------------------------------------------
| Category | Testing Summary |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization Validation | The Nightingale Monitoring System is not sterilized or sterilizable, and<br>therefore this section does not apply to the monitor itself. |
| Shelf Life Testing | The Nightingale Monitoring System does not have a shelf life, and<br>therefore this section does not apply to the monitor itself. |
| Biocompatibility Testing | The Nightingale Monitoring System has no patient contact materials,<br>and therefore this section does not apply to the monitor itself. |
| Software Testing | Software for the Nightingale Monitoring System was designed and<br>developed in accordance with Zoe Medical software development<br>processes, and was verified and validated. Test results indicated that<br>the Nightingale Monitoring System complies with its predetermined<br>specification. |
| Electrical Safety | The Nightingale Monitoring System was tested for patient safety in<br>accordance with the following applicable standards:<br>• AAMI ES 60601-1:2005<br>• ANSI/AAMI SP10:2002 / A1:2003<br>• IEC 60601-1-4:2000<br>• IEC 60601-1-8:2006<br>• IEC 60601-2-27:2011<br>• IEC 80601-2-30:2009<br>• IEC 60601-2-34:2011<br>• ISO 80601-2-55:2011<br>• ISO 80601-2-56:2009<br>• ISO 80601-2-61:2011<br>• IEC 62366:2007<br>• IEC 62304:2006<br>Test results indicated that the Nightingale Monitoring System complies<br>with its predetermined specification. |
| Electromagnetic<br>Compatibility Testing | The Nightingale Monitoring System was tested for EMC in accordance<br>with IEC 60601-1-2:2007. Test results indicated that the Nightingale<br>Monitoring System complies with its predetermined specification. |
| Performance Testing -<br>Bench | The Nightingale Monitoring System was tested in accordance with Zoe<br>Medical internal requirements and procedures, and test results<br>indicated that the device complies with the predetermined requirements.<br>This testing includes performance and functional, environmental, and<br>shipping and transportation testing. |
| Performance Testing -<br>Animal | Animal performance testing was not performed and is not necessary to<br>demonstrate safety and effectiveness of the Nightingale Monitoring<br>System. |
| Performance Testing -<br>Clinical | Clinical performance testing was not performed and is not necessary to<br>demonstrate safety and effectiveness of the Nightingale Monitoring<br>System. |
## Table 2. Performance Testing of the Nightingale Monitoring System
# Conclusion
Based upon a comparison of devices and performance testing results, the Zoe Medical Nightingale Monitoring System is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name written around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20995-00012
July 23, 2013
Zoe Medical, Inc. c/o Mr. James Chickering Regulatory Affairs Manager 460 Boston Street Topsfield, MA 01983
Re: K130740
Trade/Device Name: Nightingale Monitoring System Regulatory Number: 21 CFR 870.2300 Regulation Name: Physiological Patient Monitor (without arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MWI Dated: April 21, 2013 Received: April 24, 2013
Dear Mr. Chickering:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Mr. James Chickering
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
for
Sincerely yours,
Owen-Paris-S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K130740
# Indications for Use
510(k) Number: _
Device Name:
## Zoe Medical Nightingale Monitoring System
Indications for Use:
The Zoe Medical Nightingale Monitoring System is indicated for use in adult & pediatric patient populations.
The Zoe Medical Nightingale Monitoring System facilitates the monitoring of:
- ECG -
- Impedance respiration -
- Non-Invasive blood pressure -
- Invasive blood pressure -
- Body temperature "
- Functional arterial oxygen saturation (SpO2) -
- -End-tidal & inspired CO2
The Zoe Medical Nightingale Monitoring System is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by
Owen R. Faris -5
Date: 2013.07.23
16:23:12 -04'00'
Zoe Medical, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.