SPACELABS ELANCE VITAL SIGNS MONITORING SYSTEM AND CENTRAL MONITOR SOFTWARE
K090556 · Spacelabs Healthcare Medical Equipment (Suzhou) CO · MHX · Apr 30, 2009 · Cardiovascular
Device Facts
Record ID
K090556
Device Name
SPACELABS ELANCE VITAL SIGNS MONITORING SYSTEM AND CENTRAL MONITOR SOFTWARE
Applicant
Spacelabs Healthcare Medical Equipment (Suzhou) CO
Product Code
MHX · Cardiovascular
Decision Date
Apr 30, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Spacelabs élance Vital Signs Monitor is indicated for use in patient populations for: - Adult - Pediatric The Spacelabs élance Vital Signs Monitor facilitates the monitoring of: - ECG with arrhythmia detection - Respiration - Non-invasive blood pressures - Invasive blood pressures - Body temperature - Functional arterial oxygen saturation, and - Expired and/or inspired CO2. The Spacelabs élance Vital Signs Monitor is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
Device Story
Portable patient monitoring system; monitors ECG, respiration, NIBP, IBP, temperature, SpO2, and EtCO2. Available in 10.2" and 12.1" display sizes. Used by clinicians in healthcare facilities. Includes optional central station software (93900) for remote monitoring via customer-provided computer. Provides real-time physiological data and arrhythmia detection to assist clinical decision-making. Benefits include continuous patient status tracking and alarm notification.
Clinical Evidence
No clinical data. Bench testing only, including performance, functional, environmental, and shipping/transportation testing. Software, electrical safety, and EMC testing performed per applicable standards.
Technological Characteristics
Portable patient monitor; 10.2" or 12.1" display. Parameters: ECG, respiration, NIBP, IBP, temperature, SpO2, EtCO2. Connectivity: Central station software package (93900) for networked monitoring. Non-sterile device. No patient contact materials.
Indications for Use
Indicated for adult and pediatric patient populations for monitoring ECG (with arrhythmia detection), respiration, non-invasive blood pressure, invasive blood pressure, body temperature, functional arterial oxygen saturation, and expired/inspired CO2. Prescription use only by healthcare professionals in healthcare facilities.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Zoe Medical Nightingale Monitoring System (K001775)
{0}------------------------------------------------
# K090556
"
## 510(k) Summary
### APR 3 0 2009
| Submission Date: | 23 February 2009 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Spacelabs Healthcare Medical Equipment (Suzhou) Co., Ltd.<br>Building 30, West Wing<br>Chuang Tou Industrial Square, Yang Xian Road |
| Submitter Contact: | Jennifer Cui<br>Phone: 011-086-0512-87178286<br>Fax: 011-086-0512-87178278<br>Email: Jennifer.Cui@spacelabs.com |
| Official Contact: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 |
| Trade Name: | Spacelabs élance Vital Signs Monitoring System |
| Common Name: | Patient Physiological Monitor (with arrhythmia detection or alarms) |
| Classification Name: | Monitor, physiological, patient (with arrhythmia detection or alarms) |
| Classification<br>Regulation: | 21 CFR §870.1025 |
| Product Code: | MHX |
Page 1 of 4
{1}------------------------------------------------
| Substantially<br>Equivalent Devices: | Spacelabs élance Vital Signs<br>Monitoring System | Predicate 510(k)<br>Number | Predicate<br>Manufacturer / Model |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|---------------------------------------------------------------|
| | ECG component | K001775 | Zoe Medical Nightingale<br>Monitoring System |
| | Respiration component | | |
| | Non-invasive blood pressures<br>component | | |
| | Invasive blood pressures<br>component | | |
| | Temperature component | | |
| | Expired and/or inspired CO2<br>component | K060065 | Oridion Capnostream 20 |
| | Functional arterial oxygen<br>saturation component | K033296 | Masimo Corporation SET<br>Rad 5 MS-11 board |
| | Arrhythmia detection and<br>invasive pressure monitoring<br>(IBP) | K062144 | Invivo Escort M8 Vital<br>Signs Patient Monitor<br>Model 3810 |
| | Functional arterial oxygen<br>saturation component | K012891 | Nellcor Oximax N-595<br>pulse oximeter |
| Device Description: | The Spacelabs Healthcare Medical Equipment (Suzhou) Co., Ltd.<br>(Spacelabs) élance Vital Signs Monitoring System (élance) is a family<br>of portable patient monitors intended to be used by clinicians and<br>medical qualified personnel for monitoring ECG, Respiration, NIBP,<br>Temperature, SPO2, Invasive Blood Pressure and EtCO2. Models<br>within the Spacelabs élance family come in two different sized viewing<br>areas (10.2" and 12.1"), two different housing colors (white and black)<br>and offer selected monitoring features. See the following Product<br>Configuration List for the identified models and Spacelabs defined<br>monitoring features. | | |
| Central Station: | A software package (93900) is available for use with customer acquired<br>computer based on specifications provided by Spacelabs Medical. This<br>package allows monitoring of the élance at a central workstation. | | |
| Technology<br>Comparison: | The élance utilizes the same or similar technology for each parameter<br>as utilized by the predicate devices. | | |
. . .
{2}------------------------------------------------
#### Intended Use:
The Spacelabs élance Vital Signs Monitor is indicated for use in patient populations for:
- Adult
- Pediatric
The Spacelabs élance Vital Signs Monitor facilitates the monitoring of:
- ECG with arrhythmia detection
- Respiration
- Non-invasive blood pressures
- Invasive blood pressures
- Body temperature
- Functional arterial oxygen saturation, and
- Expired and/or inspired CO2.
The Spacelabs élance Vital Signs Monitor is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
#### Performance Testing:
| Sterilization<br>Validation | The Spacelabs élance is not sterilized or sterilizable, and therefore this<br>section does not apply to the monitor itself. |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Shelf Life Testing | The Spacelabs élance is not sterilized or sterilizable, and therefore this<br>section does not apply to the monitor itself. |
| Biocompatibility<br>Testing | The Spacelabs élance has no patient contact materials, and therefore<br>this section does not apply to the monitor itself. |
| Software Testing | Software for the Spacelabs élance was designed and developed according<br>to a robust software development process, and was rigorously verified<br>and validated. Test results indicated that the Spacelabs élance complies<br>with its predetermined specification. |
| Electrical Safety | The Spacelabs élance was tested for patient safety in accordance with<br>applicable Standards. Test results indicated that the Spacelabs élance<br>complies with its predetermined specification. |
| Electromagnetic<br>Compatibility Testing | The Spacelabs élance was tested for EMC in accordance with<br>applicable Standards. Test results indicated that the Spacelabs élance<br>complies with its predetermined specification. |
| Performance Testing<br>– Bench | The Spacelabs élance was tested in accordance with internal requirements<br>and procedures, and test results indicated that the device complies with<br>the predetermined requirements. This testing includes performance and<br>functional, environmental, and shipping and transportation testing. |
| Performance Testing<br>— Animal: | Animal performance testing was not performed and is not necessary to<br>demonstrate safety and effectiveness of the Spacelabs élance. |
Page 3 of 4
{3}------------------------------------------------
| Performance Testing | Clinical performance testing was not performed and is not necessary to |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| – Clinical: | demonstrate safety and effectiveness of the Spacelabs élance. |
| Conclusion | Based upon a comparison of devices and performance testing results,<br>the Spacelabs élance is substantially equivalent to the predicate devices. |
. . . .
{4}------------------------------------------------
Public Health Service
Image /page/4/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design, resembling an eagle or a bird in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the design.
MAY 19 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Spacelabs Healthcare Medical Equipment (Suzhou) Co., Ltd. c/o Mr. Thomas Kroenke Speed To Market, Inc. P.O. Box 3018 Nederland, CO 80466
Re: K090556
Trade/Device Name: élance Vital Signs Monitoring System; and, élance Central Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: February 23, 2009 Received: March 2, 2009
Dear Mr. Kroenke:
This letter corrects our substantially equivalent letter of April 30, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Thomas Kroenke
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
signature
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use
| 510(k) Number (if known): | K |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Spacelabs Healthcare, Inc. (Spacelabs) élance Vital Signs<br>Monitoring System |
| Indications for Use: | The Spacelabs élance Vital Signs Monitor is indicated for use in<br>patient populations for:<br>- Adult<br>- Pediatric |
| | The Spacelabs élance Vital Signs Monitor facilitates the<br>monitoring of:<br>- ECG with arrhythmia detection<br>- Respiration<br>- Non-invasive blood pressures<br>- Invasive blood pressures<br>- Body temperature<br>- Functional arterial oxygen saturation, and<br>- End title CO2. |
| | The Spacelabs élance Vital Signs Monitor is a prescription<br>device intended to be used by healthcare professionals in all<br>areas of a healthcare facility. |
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(Division Sign-Off) Division of Cardiovascular Devices Division of Caraloversealar Bo
510(k) Number_Ko G0 53 G
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.