K130529 · Raumedic AG · GWM · Oct 25, 2013 · Neurology
Device Facts
Record ID
K130529
Device Name
RAUMEDIC-PTO-MONITORING-SYSTEM
Applicant
Raumedic AG
Product Code
GWM · Neurology
Decision Date
Oct 25, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1620
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The RAUMEDIC® - PTO-Monitoring-System measures intracranial pressure, temperature and oxygen and is intended as an adjunct monitor of trends in these parameters, indicating the perfusion status of cerebral tissue local to sensor placement. RAUMEDIC® - PTO-Monitoring-System values are relative within an individual, and should not be used as the sole basis for decisions as to diagnosis or therapy. It is intended to provide data additional to that obtained by current clinical practice, in cases where hypoxia or ischemia is a concern. Use of the parenchymal intracranial pressure monitoring kit with bolt is indicated for children who are at least one year old.
Device Story
The RAUMEDIC PTO-Monitoring-System is a multi-parameter intracranial monitoring device used in clinical settings by neurosurgeons. It utilizes implanted catheters (NEUROVENT series) containing piezoresistive pressure sensors for ICP, thermistors for temperature, and luminescence optical fiber sensors for ptiO2. The system includes a user interface unit (EASY logO) that processes signals from the catheters via dedicated cables (PTO and LWL). The EASY logO displays real-time measurements of ICP (including pulsatile ICPA), temperature, and oxygen, and provides technical alarms. Data can be exported to external patient monitoring units via specialized cables. The system provides trend data to assist clinicians in assessing cerebral perfusion status; it is not intended as the sole basis for diagnosis or therapy. Benefits include continuous, localized monitoring of physiological parameters in patients at risk of cerebral hypoxia or ischemia.
Clinical Evidence
Bench testing only. Performance evaluated against ANSI/AAMI NS 28:1988/(R) 2006 for ICP monitoring and ASTM E1112-00 (2006) for temperature. Oxygen sensor performance compared to predicate (K980380) regarding drift, accuracy, and response time. Biocompatibility testing conducted per ISO 10993-1, -5, and -18.
Technological Characteristics
System comprises intracranial catheters (piezoresistive pressure sensor, thermistor, luminescence optical fiber), cranial bolt kits, and EASY logO interface. Materials comply with ISO 10993. Connectivity via dedicated cables to external patient monitors. Sterilization via Ethylene Oxide. ICP sensor: Wheatstone bridge/piezoresistive. Temperature sensor: thermistor. Oxygen sensor: fiber optic quenching.
Indications for Use
Indicated for children (≥1 year) and adults requiring adjunct monitoring of intracranial pressure, temperature, and interstitial fluid oxygen partial pressure (ptiO2) to assess cerebral tissue perfusion status in cases where hypoxia or ischemia is a concern.
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
Predicate Devices
DIAMETRIC MEDICAL LTD. NEUROTREND MULTIPARAMETER SENSOR (K980380)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten string of characters, which appears to be an identifier or code. The string reads 'K1305-29'. The characters are written in a cursive style with varying stroke thicknesses, and the overall impression is that of a quickly jotted down note.
Image /page/0/Picture/1 description: The image shows the logo for RAUMEDIC, a company that provides lifeline to health. The logo consists of a circular symbol on the left, followed by the company name in bold, sans-serif font. Below the name is the tagline "Lifeline to Health" in a smaller, lighter font. The overall design is clean and professional, conveying a sense of reliability and trustworthiness.
RAUMEDIC AG · Postfach 5 01 · D - 95205 Münchberg
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - W066-0609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
#### Regulatory Affairs
Telefon: (0 92 52) 359-0 Telefax: (0 92 52) 359-10 00 info@RAUMEDIC.com www.RAUMEDIC.de
Direktkontakt: Telefon: (0 92 52) 359-2782
Reiner. Thiem@RAUMEDIC.com
Tag 16th of October 2013
Ihre Nachricht vom Ihre Zeichen
H Thiem Sachbearbeiter Unsere Zeichen thi
510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
1. Name, address, phone and fax number of the applicant
RAUMEDIC AG Hermann-Staudinger-Straße 2 95233 Helmbrechts D - Germany Tel.: 0049/9252/359-0 Fax: 0049/9252/359-1000
#### 2. Contact person
Mr. Reiner Thiem Head of Regulatory Affairs Hermann-Staudinger-Straße 2 95233 Helmbrechts D - Germany Tel.: 0049/9252/359-2782
#### 3. Date of preparation of the summary
October, the 16th 2013
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left and the word "RAUMEDIC" in bold, sans-serif font on the right. Below the word "RAUMEDIC" is the tagline "Lifeline to Health" with a line on either side of the text. The circular graphic appears to be a stylized representation of a globe or sphere.
#### 4. Name of the device
The RAUMEDIC® -PTO-Monitoring-System is composed of the following components:
- NEUROVENT -PTO .
- NEUROVENT®-TO .
- NEUROVENT®-PTO 2L .
- EASY logO .
- BOLT KIT PTO .
- BOLT KIT PTO 2L .
- Cable LWL .
- Cable PTO .
- Cable DATALOGGER GE/MARQUETTE .
- Cable DATALOGGER Philips/HP .
- Cable DATALOGGER Siemens/Dräger Infinity .
- Cable DATALOGGER Datex-Ohmeda .
- Cable DATALOGGER Hellige .
- Cable DATALOGGER SpaceLabs �
- Cable DATALOGGER Nihon Kohden .
Device Classification Name:
Classification Panel: CFR Section: Device Class: Product Code:
Device, Monitoring, Intracranial Pressure, Temperature and Oxygen Neurology 21 CFR §882.1620 Class II GWM
#### 5. Device Description
The RAUMEDIC® -PTO-Monitoring-System determines the level and change in intracranial pressure (ICP) by using semi-conductor pressure sensors, in temperature using thermistors, and in partial oxygen pressure ptiO2 using a luminescence optical sensor fibre.
The PTO-Monitoring-System consists of intracranial catheters, a cranial bolt kit, the user interface unit (EASY logO), cables to connect the catheters to the EASY logO and cables for the EASY logO to send the measurements to specific patient monitoring units.
Three intracranial catheters are provided: NEUROVENT®-PTO, NEUROVENT® TO and NEUROVENT® PTO 2L. The NEUROVENT®-PTO measures three physiological parameters - ICP, temperature and partial oxygen pressure. The application has to be made with a single lumen Bolt (BOLT-KIT PTO). The NEUROVENT -PTO 2L also measures these three parameters, but has to be placed with a multi lumen Bolt (BOLT KIT PTO 2L) or via tunneling process. The NEUROVENT® TO measures two physiological parameters – temperature and partial oxygen pressure. The application has to be made with a single lumen Bolt (BOLT-KIT PTO).
The NEUROVENT®-PTO or the NEUROVENT®-TO is implanted in the parenchyma using a BOLT KIT PTO. The NEUROVENT®-PTO 2L is implanted in the parenchyma using a BOLT KIT PTO 2L or using a RAUMEDIC® spliceable tunnelling sleeve CH8 (already cleared for market under 510(k) K112017).
Device #:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word is the tagline "Lifeline to Health" in a smaller, italicized font, flanked by horizontal lines.
For connecting the catheters to the EASY logO two cables – cable PTO and cable LWL - are needed. The cable PTO transfers the electrical signal for ICP and temperature. The cable LWL transfers the optical signal for oxygen pressure.
The EASY logO is the user interface that displays the measurements and alerts the user to the status of the intracranial catheters. The EASY logO displays: oxygen partial pressure pO2, intracranial pressure ICP, pulse amplitude ICPA and temperature T. ICP is a pulsatile signal that consists of the ICP average value and a pulsatile ICPA component, which is superimposed over the ICP average value.
DATALOGGER cables allows the EASY logO to send the measurements to specific patient monitoring units.
#### 6. Indications of Use
The RAUMEDIC® - PTO-Monitoring-System measures intracranial pressure, temperature and oxygen and is intended as an adjunct monitor of trends in these parameters, indicating the perfusion status of cerebral tissue local to sensor placement. RAUMEDIC® - PTO-Monitoring-System values are relative within an individual, and should not be used as the sole basis for decisions as to diagnosis or therapy. It is intended to provide data additional to that obtained by current clinical practice, in cases where hypoxia or ischemia is a concern.
Use of the parenchymal intracranial pressure monitoring kit with bolt is indicated for children who are at least one year old.
#### 7. Substantial Equivalence Summary
The RAUMEDIC® - PTO-Monitoring-System is substantially equivalent to the DIAMETRIC MEDICAL LTD. NEUROTREND MULTIPARAMETER SENSOR which was cleared for market under 510(k) K980380 on 14" of July 1999, the LICOX Brain Oxygen Monitoring System which was cleared for market under 510(k) K002765 on 24th of November 2000, the RAUMEDIC® ICP-TEMP- MONITORING-SYSTEM which was cleared for market under 510(k) K120252 on 110 of April 2012 and the RAUMEDIC® ICP-MONITORING-SYSTEM which was cleared for market under 510(k) K103206 on 04* of March 2011.
For further information see device comparison tables below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for RAUMEDIC, a company that provides medical products. The logo consists of a circular symbol on the left and the word "RAUMEDIC" in bold letters on the right. Below the word "RAUMEDIC" is the tagline "Lifeline to Health".
Table 1: Comparison to Predicate RAUMEDIC® ICP-TEMP- MONITORING- SYSTEM (K120252)
| Feature | RAUMEDIC® PTO-<br>MONITORING- SYSTEM | RAUMEDIC® ICP-TEMP.<br>MONITORING- SYSTEM<br>(K120252) | SE? |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | NEUROVENT -PTO,<br>NEUROVENT -PTO 2L.<br>NEUROVENT®-TO | NEUROVENT P-TEMP.<br>NEUROVENT - TEMP-IFD-S.<br>NEUROVENT - TEMP-IFD-R | N/A |
| Indication For<br>Use | The RAUMEDIC® -PTO-<br>Monitoring-System is indicated for<br>use by a qualified neurosurgeon for<br>direct measurement of intracranial<br>pressure and temperature in the<br>parenchyma as well as to<br>determine the partial oxygen<br>pressure ptiO2 of the interstitial<br>fluid. | The RAUMEDIC® -ICP-TEMP-<br>Monitoring-System is indicated for<br>use by a qualified neurosurgeon for<br>direct measurement of intracranial<br>pressure and temperature in the<br>parenchyma and in the ventricle<br>and cerebrospinal fluid drainage<br>applications. | YES - for<br>measurement<br>of intracranial<br>pressure and<br>temperature; for<br>oxygen<br>measurement<br>see SE-List with<br>predicate<br>device:<br>K880380 |
| Anatomical Site | Brain parenchyma | Brain parenchyma / ventricle | YES |
| Catheter<br>Sensors Location | Catheter Tip | Catheter Tip | YES |
| Tip Diameter | SF | SF | YES |
| Single-use- | YES | YES | |
| Catheter | | | YES |
| Sterilization<br>Process | With Ethylene Oxide | With Ethylene Oxide | YES |
| Bolt | Bolt with compression cap | Bolt with compression cap | YES |
| Spliceable<br>Tunnelling Sleeve | Spliceable Tunnelling Sleeve with<br>Trocar | Splicable Tunnelling Sleeve with<br>Trocar | YES |
| Size Of Access<br>Hole | for Bolt:<br>4.2 mm >> NEUROVENT®-PTO,<br>NEUROVENT®-TO<br>6.3 mm → NEUROVENT®-PTO 2L<br>for Tunnelling:<br>> 6 mm → NEUROVENT®-PTO 2L | for Bolt:<br>4.2 mm > NEUROVENT®-P-<br>TEMP.<br>6.3 mm -> NEUROVENT -TEMP-<br>IFD-S,<br>NEUROVENT®-TEMP-<br>IFD-R<br>for Tunnelling:<br>≥ 6 mm → NEUROVENT®-TEMP-<br>IFD-S,<br>NEUROVENT®-TEMP-<br>IFD-R | YES |
| Fixation Of<br>Catheter | Fixation wing (sutured to skin) | Fixation wing (sutured to skin) | YES |
| Non-fluid<br>Coupling Catheter | YES | YES | YES |
| ICP - Sensor<br>Design Catheter | Piezoresistive pressure sensor<br>(Semi-conductor / Wheatstone<br>Bridge) | Piezoresistive pressure sensor<br>( Semi-conductor / Wheatstone<br>Bridge) | YES |
| ICP - Pressure<br>Range | -50 to +250 mmHg | -50 to +250 mmHg | YES |
| ICP - Maximum<br>Pressure | 1,250 mmHg | 1,250 mmHg | YES |
| ICP - Accuracy | ± 1 mmHg | ± 1 mmHg | YES |
| ICP - Resolution | 0.1 mmHg | 0.1 mmHg | |
| ICP - Zero Point<br>Stability | Less than 1 mmHg during first 24<br>ours<br>Less than 2 mmHg during the first<br>7 days | Less than 1mmHg during first 24<br>hours<br>Less than 2mmHg during the first 7<br>days | YES |
| ICP - Dynamic<br>Response Time | < 5 ms | < 5 ms | YES |
| ICP - Sensitivity<br>System | 5 µV/V/mmHg on the monitor side | 5 µV/V/mmHg on the monitor side | YES |
| Feature | RAUMEDIC® PTO-<br>MONITORING- SYSTEM | RAUMEDIC® ICP-TEMP-<br>MONITORING- SYSTEM<br>(K120252) | SE? |
| TEMP - Sensor<br>Design | Thermistor | Thermistor | YES |
| TEMP -<br>Temperature<br>Range | 25 °C - 45 °C | 25 °C - 45 °C | YES |
| TEMP - Accuracy | ± 0.1 °C | ± 0.1 °C | YES |
| TEMP -<br>Resolution | 0.01 °C | 0.01 °C | YES |
| TEMP - Zero<br>Point Stability | ± 0.2 °C | ± 0.2 °C | YES |
| TEMP - Dynamic<br>Response Time | < 150 s | < 150 s | YES |
| TEMP - Adapter | Cable PTO | ICP-TEMP-Adapter<br>ICP-TEMP-Adapter HP / Philips | YES |
| p₁O₂ - Sensor<br>Design Catheter | Fiber Optic / Quenching<br>→ measuring amount of reflected<br>light from sensor tip in comparison<br>to pre-calibrated values with<br>reference solutions (see section<br>"12. Substantial Equivalence<br>Discussion") | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K980380 |
| p₁O₂ - Oxygen<br>Pressure Range | 0 to 150 mmHg | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K980380 |
| p₁O₂ - Accuracy | ± 3 % or ± 2.5 mmHg of the<br>measured value → the higher<br>value is applicable for p₁O₂ < 120<br>mmHg:<br>< 10 % of the measured value for<br>p₁O₂ from 120 mmHg to 150 mmHg | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K980380 |
| p₁O₂ - Resolution | 0.1 mmHg | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K980380 |
| p₁O₂ - Zero Point<br>Stability | max. 1.5 mmHg in 5 days<br>(at a ptiO2 of ≤ 10 mmHg) | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K980380 |
| p₁O₂ - Dynamic<br>Response Time | < 200 s for p₁O₂-decrease from<br>150 ± 15 mmHg to 0.0 mmHg | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K980380 |
| p₁O₂ - Adapter | Cable LWL | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K980380 |
| Product Code | GWM | GWM | YES |
| Registration # | Pending | K120252 | N/A |
| Applicant | RAUMEDIC AG<br>Hermann-Staudinger-Str. 2<br>95233 Helmbrechts | RAUMEDIC AG<br>Hermann-Staudinger-Str. 2<br>95233 Helmbrechts | N/A |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "life.to health" in a smaller font size.
- Lifeline to Health -
* ICP measurement only with NEUROVENT®-PTO and NEUROVENT®-PTO 2L
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for RAUMEDIC, a company that provides a lifeline to health. The logo consists of a circular graphic on the left, followed by the company name in bold, sans-serif font. Below the name is the tagline "Lifeline to Health".
Table 2: Comparison to predicate DIAMETRIC MEDICAL LTD.NEUROTREND MULTIPARAMETER SENSOR (K980380)
| Feature | RAUMEDIC® PTO-<br>MONITORING- SYSTEM | DIAMETRIC MEDICAL LTD.<br>NEUROTREND<br>MULTIPARAMETER SENSOR<br>(K980380) | SE? |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | NEUROVENT®-PTO,<br>NEUROVENT®-PTO 2L,<br>NEUROVENT®-TO | Neurotrend™ Multiparameter<br>Sensor<br>(C7004S) | N/A |
| Indication For<br>Use | The RAUMEDIC® -PTO-Monitoring-<br>System is indicated for use by a<br>qualified neurosurgeon for direct<br>measurement of intracranial<br>pressure and temperature in the<br>parenchyma as well as to determine<br>the partial oxygen pressure ptiO2 of<br>the interstitial fluid. | The Neurotrend Multiparameter<br>Sensor is a disposable, sterile,<br>single-use, device for the<br>continuous measurement of<br>intracranial oxygen, carbon<br>dioxide, pH and temperature that<br>is used in conjunction with a<br>suitable intracranial access<br>device. | YES - for<br>measurement<br>of oxygen; for<br>intracranial<br>pressure and<br>temperature<br>measurement<br>see SE-List with<br>predicate<br>device:<br>K120252 |
| Anatomical Site<br>Catheter | Brain parenchyma | Brain ventricle | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| Sensors Location | Catheter Tip | Catheter Tip | YES |
| Tip Diameter | 5 F | 2 F | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| Single-use-<br>Catheter | YES | YES | YES |
| Sterilization<br>Process | With Ethylene Oxide | With Gamma - Radiation | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| Bolt | Bolt with compression cap | A suitable intracranial access<br>device | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| Spliceable<br>Tunnelling Sleeve | Spliceable Tunnelling Sleeve with<br>Trocar | A suitable intracranial access<br>device | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| Size Of Access<br>Hole | for BOLT usage:<br>4.2 mm → NEUROVENT®-PTO,<br>NEUROVENT®-TO<br>6.3 mm → NEUROVENT®-PTO 2L<br><br>for Tunnelling Sleeve usage:<br>≥ 6 mm → NEUROVENT®-PTO 2L | | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| Fixation Of<br>Catheter | Fixation wing (sutured to skin) | | YES |
| Non-fluid | YES | YES | YES |
| Feature | RAUMEDIC® PTO-<br>MONITORING- SYSTEM | DIAMETRIC MEDICAL LTD.<br>NEUROTREND<br>MULTIPARAMETER SENSOR<br>(Ka80380) | SE? |
| ICP - Sensor<br>Design Catheter | Piezoresistive pressure sensor<br>(Semi-conductor / Wheatstone<br>Bridge) | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| ICP - Pressure<br>Range | -50 to +250 mmHg | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| ICP - Maximum<br>Pressure | 1,250 mmHg | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| ICP - Accuracy | ± 1 mmHg | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| ICP - Resolution | 0.1 mmHg | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| ICP - Zero Point<br>Stability | Less than 1 mmHg during first 24<br>ours<br>Less than 2 mmHg during the first 7<br>days | ---- | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| ICP - Dynamic<br>Response Time | < 5 ms | --- | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| ICP - Sensitivity<br>System | 5 µV/V/mmHg on the monitor side | - | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| TEMP - Sensor<br>Design | Thermistor | Thermocouple | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| TEMP -<br>Temperature<br>Range | 25 °C - 45 °C | 10 °C - 42 °C | NO<br>- see SE-List<br>with predicate<br>device:<br>K120252 |
| TEMP - Accuracy | + 0.1 °C | +0.1 °C | YES |
| TEMP -<br>Resolution | 0.01 °C | 0.01 °C | YES |
| TEMP - Zero<br>Point Stability | + 0.2 °C | + 0.2 °C | YES |
| TEMP - Dynamic<br>Response Time | < 150 s | < 150 s | YES |
| TEMP - Adapter | Cable PTO | PDM Cable | YES |
| Feature | RAUMEDIC® PTO-<br>MONITORING-SYSTEM | DIAMETRIC MEDICAL LTD.<br>NEUROTREND<br>MULTIPARAMETER SENSOR<br>(K980380) | SE? |
| ptO2 - Sensor<br>Design Catheter | Fiber Optic / Quenching<br>→ measuring amount of reflected<br>light from sensor tip in comparison<br>to pre-calibrated values with<br>reference solutions (see section "12.<br>Substantial Equivalence<br>Discussion") | Fiber Optic / Quenching<br>→ measuring amount of reflected<br>light from sensor tip in comparison<br>to pre-calibrated values with<br>reference solutions | YES |
| ptO2 - Oxygen<br>Pressure Range | 0 to 150 mmHg | 10 to 430 mmHg | NO; the lower<br>limit is<br>demonstrated<br>by<br>performance<br>testing |
| ptO2 - Accuracy | ± 3 % or ± 2.5 mmHg of the<br>measured value → the higher value<br>is applicable for ptO2 < 120 mmHg;<br>< 10 % of the measured value for<br>ptO2 from 120 mmHg to 150 mmHg | ± 3 % or ± 2.5 mmHg of the<br>measured value → the higher<br>value is applicable for ptO2 < 120<br>mmHg:<br>< 10 % of the measured value for<br>ptO2 from 120 mmHg to 200<br>mmHg | YES |
| ptO2 - Resolution | 0.1 mmHg | 0.1 mmHg | YES |
| ptO2 - Zero Point<br>Stability | max. 1.5 mmHg in 5 days<br>(at a ptiO2 of ≤ 10 mmHg) | < 0.5 % per hour | YES<br>within the<br>stability range<br>of the<br>NEUROTREND |
| ptO2 - Dynamic<br>Response Time | < 200 s for ptO2-decrease from 150<br>± 15 mmHg to 0.0 mmHg | < 200 s for ptO2-decrease from<br>150±15 mmHg to 0.0 mmHg | YES |
| ptO2-Adapter | Cable LWL | PDM Cable | YES |
| Product Code | GWM | GWM | YES |
| Registration # | Pending | K980380 | N/A |
| Applicant | RAUMEDIC AG<br>Hermann-Staudinger-Str. 2<br>95233 Helmbrechts<br>Germany | Diametrics Medical, Ltd.<br>Short Street<br>High Wycombe<br>Buckinghamshire, UK HP11 2QH | N/A |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Underneath the word "RAUMEDIC" is the tagline "Lifeline to Health".
- Lifeline to Health ----
1
Device #:_____________________________________________________________________________________________________________________________________________________________________
,
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "life. united. health."
- Lifeline to Health ----
* ICP measurement only with NEUROVENT®-PTO and NEUROVENT®-PTO 2L
Device #:_
:
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Raumedic logo. The logo consists of a circular symbol on the left and the word "RAUMEDIC" in bold, sans-serif font on the right. Below the word is the tagline "Lifeline to Health" with a line above and below the text.
Table 3: Comparison to Predicate LICOX Brain Oxygen Monitoring System (K002765)
| Feature | RAUMEDIC® PTO-<br>MONITORING-<br>SYSTEM | RAUMEDIC® ICP-<br>MONITORING-<br>SYSTEM | LICOX Brain Oxygen<br>Monitoring System | SE? |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Trade Name | EASY logo | NPS3 | CMP® Monitor | N/A |
| Indication for Use | The RAUMEDIC® -<br>PTO-Monitoring-<br>System is indicated<br>for use by a qualified<br>neurosurgeon for<br>direct measurement<br>of intracranial<br>pressure and<br>temperature in the<br>parenchyma as wellto determine the<br>partial oxygen<br>pressure ptiO2 of the<br>interstitial fluid. | The RAUMEDIC® -<br>ICP-Monitoring-<br>System is indicated<br>for use by a qualified<br>neurosurgeon for<br>direct measurement<br>of intracranial<br>pressure in the<br>parenchyma. | The LICOX Brain<br>Oxygen Pressure<br>Monitoring System<br>directly measures<br>Partial Pressure of<br>Oxygen in the brain.<br>The LICOX system can<br>also utilize brain<br>temperature information<br>for temperature<br>compensation of the<br>pO2. | YES |
| ICP Measurement | YES | YES | NO | YES |
| Temperature<br>Measurement | YES | NO | YES | YES |
| pO2<br>Measurement | YES | NO | YES | YES |
| Pressure Range | -40 to +100 mmHg | -40 to +100 mmHg | --- | YES |
| Temperature<br>Range | 15 °C - 45 °C | --- | 0°C - 46 °C | NO; limits are<br>demonstrated<br>by<br>performance<br>testing in<br>accordance to<br>ASTM E1112 |
| pO2 Range | 0 to 150 mmHg | --- | 0 to 150 mmHg | YES; limits are<br>demonstrated<br>by<br>performance<br>testing |
| Screen | YES | YES | YES | YES |
| Monitoring | Continuous…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.