The RAUMEDIC® -ICP-TEMP-Monitoring-System is indicated for use in parenchymal and ventricular pressure and temperature monitoring and can be used for the measurement of the intra-cranial pressure (ICP) as well as of the cerebral perfusion pressure (central arterial blood pressure minus ICP) which is the essential pre-requisite for an effective treatment of patients suspected of suffering from intra-cranial pressure increases (such as cranio-cerebral traumas, GCS ≤ 8; malignant medial cardiac infarctions; hepatic encephalopathy; SAB Hunt / tradinas, IV + V; cerebral edema; hydrocephalus) or of patients whose clinical picture may be linked to an increase of the ICP and cerebrospinal fluid drainage applications. Additional the measurement of the brain temperature allows the direct measurement of the cerebral the measurature. This temperature can deviate from the patient's overall body temperature. Measuring the brain's temperature permits episodes of cerebral hypothermia to temperated and also provides additional temperature information during hypothermia treatment.
Device Story
System monitors intracranial pressure (ICP) and brain temperature; utilizes semi-conductor pressure sensors and thermistors at catheter tip. Components include NEUROVENT®-TEMP IFD-S/IFD-R (ventricular) and NEUROVENT®-P-TEMP (parenchymal) catheters, ICP-TEMP-adapters, and cables. Used in clinical settings by neurosurgeons; catheters implanted via tunneling sleeves or bolt kits. System connects to hospital patient monitors to provide real-time pressure and temperature data. Enables calculation of cerebral perfusion pressure; assists in managing patients with elevated ICP or undergoing hypothermia treatment. Provides direct brain temperature measurement, which may differ from systemic body temperature, aiding clinical decision-making for neurological conditions.
Clinical Evidence
Bench testing only. Biocompatibility testing conducted per ISO 10993. Performance testing confirmed catheter design safety, pressure range (-50 to +250 mmHg), and zero-point stability (less than 1 mmHg during first 24 hours; less than 2 mmHg during first 7 days). No clinical data presented.
Technological Characteristics
System uses semi-conductor pressure transducers and thermistors. Catheters are single-use, non-fluid coupled. Pressure range: -50 to +250 mmHg; Temperature range: 25°C - 45°C. Sensitivity: 5 µV/V/mmHg. Sterilization: Ethylene Oxide. Connectivity: Wired via ICP-TEMP-Adapter to hospital patient monitors. Materials biocompatible per ISO 10993.
Indications for Use
Indicated for use by qualified neurosurgeons for direct measurement of intracranial pressure and temperature in parenchyma and ventricles, and for cerebrospinal fluid drainage. Patient population includes those with suspected ICP increases (e.g., trauma, GCS ≤ 8, cardiac infarction, hepatic encephalopathy, cerebral edema, hydrocephalus). Contraindicated in children under one year old for parenchymal and ventricular monitoring kits with bolts. Catheters are MR Unsafe.
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
Predicate Devices
Camino NeuroCare Intracranial Pressure-Temperature Monitoring Kit, Model 110-4BT (K962928)
Camino NeuroCare Intracranial Pressure-Temperature Monitoring Kit, Model 110-4HMT (K962928)
{0}------------------------------------------------
K120252
Image /page/0/Picture/1 description: The image shows a black and white symbol that appears to be a stylized representation of a letter "C" within a circle. The letter "C" is not fully closed, and there are two horizontal lines below the circle. The symbol is simple and graphic, with a bold, solid fill.
APR 1 1 2012
e to Health
| RAUMEDIC AG - Postfach 5 01 - D - 95205 Münchberg | Forschung & Entwicklung |
|---------------------------------------------------|-----------------------------------|
| Food and Drug Administration | Telefon: (0 92 52) 359-0 |
| Center for Devices and Radiological Health | Telefax: (0 92 52) 359-10 00 |
| Document Mail Center - W066-0609 | info@RAUMEDIC.com |
| 10903 New Hampshire Avenue | www.RAUMEDIC.de |
| Silver Spring, MD 20993-0002 | Direktkontakt: |
| | Telefon: (0 92 52) 359-2782 |
| Ihre Nachricht vom | |
| Ihre Zeichen | |
| | Reiner.Thiem@RAUMEDIC.com |
| | |
| | Tag |
| | 19 <sup>th</sup> of December 2011 |
| 510(k) Summary | |
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
1. Name, address, phone and fax number of the applicant
| RAUMEDIC AG | |
|-----------------------------|--|
| Hermann-Staudinger-Straße 2 | |
| 95233 Helmbrechts | |
| D - Germany | |
| Tel.: 0049/9252/359-0 | |
| Fax: 0049/9252/359-1000 | |
2. Contact person
| Mr. Reiner Thiem | |
|-----------------------------|--|
| Head of Regulatory Affairs | |
| Hermann-Staudinger-Straße 2 | |
| 95233 Helmbrechts | |
| D - Germany | |
| Tel.: 0049/9252/359-2782 | |
3. Date of preparation of the summary
December, the 19<sup>th</sup> 2011
| Device #: | Page 1 of 4 |
|-----------|-------------|
|-----------|-------------|
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "Lifeline to Health" in a smaller, italicized font. The circular graphic appears to be a stylized representation of a globe or planet.
Image /page/1/Figure/1 description: The image shows the logo for RAUMEDIC, a company that provides lifeline to health. The logo consists of a stylized globe-like icon on the left, followed by the company name in bold, sans-serif font. Below the name is the tagline "Lifeline to Health" in a smaller, italicized font, flanked by horizontal lines.
### 4. Name of the device
The RAUMEDIC® -ICP-TEMP-Monitoring-System is composed of the following elements:
- NEUROVENT®-TEMP IFD-S
- NEUROVENT®-TEMP IFD-R
- NEUROVENT®-P-TEMP
- ICP-TEMP-Adapter
- ICP-TEMP-Adapter HP / Philips
Device Classification Name:
Classification Panel: CFR Section: Device Class: Product Code:
Device, Monitoring, Intracranial Pressure and Temperature Neurology 21 CFR §882.1620 Class II GWM
#### 5. Device Description
The RAUMEDIC® -ICP-TEMP-Monitoring-System determines the level and change in intracranial pressure (ICP) by using semi-conductor pressure sensors and temperature monitoring using thermistors.
The NEUROVENT®-TEMP IFD-S and NEUROVENT®-TEMP IFD-R are indicated for use in ventricular pressure and temperature monitoring and cerebrospinal fluid drainage applications. The NEUROVENT®-P-TEMP is indicated for use in parenchymal pressure and temperature monitoring.
The NEUROVENT®-TEMP IFD-S is implanted in the ventricle by using a soft mandrel and the spliceable tunneling sleeve CH12 (already cleared to market under 510 (k) K112017) or via a RAUMEDIC® - BOLT KIT CH9 (already cleared to market under 510 (k) K112017). The NEUROVENT®-TEMP IFD-R is implanted in the ventricle only by using a rigid mandrel and a RAUMEDIC® - BOLT KIT CH9 (already cleared to market under 510 (k) K112017). The NEUROVENT® P-TEMP is implanted in parenchyma via a RAUMEDIC® - spliceable tunnelling sleeve CH8 (already cleared to market under 510 (k) K112017) or via a RAUMEDIC® - BOLT KIT CH5 (already cleared to market under 510 (k) K103206).
In addition to the catheters used for pressure and temperature monitoring an ICP-TEMPcable (already cleared to market under 510 (k) K103206) as well as an ICP-TEMP-Adapter or an ICP-TEMP-Adapter HP / Philips (depending on the type of monitor used) and a zero point module NPS2 x (already cleared to market under 510 (k) K103206) is needed to set up the measuring chain.
"x" depends on the type of patient monitor available in the hospital - there are 20 different references. To measure the pressure and temperature the NPS2 and the temperature plug have to be connected to the monitor.
Device #:
Page 2 of 4
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the text "RAUMEDIC" in bold, sans-serif font. Below the text is the tagline "Lifeline to Health" with a line on either side.
Image /page/2/Picture/1 description: The image shows the logo for RAUMEDIC. The logo consists of a circular graphic on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "Lifeline to Health" with a line extending from each side of the text.
The RAUMEDIC® -ICP-TEMP-Monitoring-System is composed of the following elements:
- NEUROVENT®-TEMP IFD-S
- neurovent®-temp IFD-R
- NEUROVENT®-P-TEMP
- ICP-TEMP-Adapter
- ICP-TEMP-Adapter HP / Philips
### 6. Device Intended Use
The RAUMEDIC® -ICP-TEMP-Monitoring-System is indicated for use in parenchymal and ventricular pressure and temperature monitoring and can be used for the measurement of the intra-cranial pressure (ICP) as well as of the cerebral perfusion pressure (central arterial blood pressure minus ICP) which is the essential pre-requisite for an effective treatment of patients suspected of suffering from intra-cranial pressure increases (such as cranio-cerebral traumas, GCS ≤ 8; malignant medial cardiac infarctions; hepatic encephalopathy; SAB Hunt / tradinas, IV + V; cerebral edema; hydrocephalus) or of patients whose clinical picture may be linked to an increase of the ICP and cerebrospinal fluid drainage applications.
Additional the measurement of the brain temperature allows the direct measurement of the cerebral the measurature. This temperature can deviate from the patient's overall body temperature. Measuring the brain's temperature permits episodes of cerebral hypothermia to temperated and also provides additional temperature information during hypothermia treatment.
### 7. Substantial Equivalence Summary
The RAUMEDIC® -ICP-TEMP-Monitoring-System is substantially equivalent to those of the legally marketed predicate devices, the Camino NeuroCare Intracranial Pressure-Temperature Monitoring Kit, Model 110-4BT and Camino NeuroCare Intracranial Pressure-Temperature Monitoring Kit, Model 110-4HMT, which were cleared to market under 510 (k) K962928 on December 20th 1996 and the RAUMEDIC® -ICP-Monitoring-System which was cleared to market under 510 (k) K103206 on 04th of March 2011 and the RAUMEDIC® -ICP-Monitoring-System ventricular which was cleared to market under 510 (k) K112017 on 11° of October 2011.
In details this means that
- ils this means that
the NEUROVENT® TEMP IFD-S and the NEUROVENT® -TEMP IFD-R are substantially equivalent to those of the Camino NeuroCare Intracranial Pressure-Temperature Monitoring Kit, Model 110-4HMT and RAUMEDIC® -ICP-Monitoring-System ventricular.
- the NEUROVENT®-P-TEMP is substantially equivalent to those of the Camino the NEGT Intracranial Pressure-Temperature Monitoring Kit, Model 110-4BT and the RAUMEDIC® -ICP-Monitoring-System.
Fur further information see device comparison tables attached.
Device #:
Page 3 of 4
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Raumedic logo. The logo consists of a circular symbol on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word "RAUMEDIC" is the tagline "Lifeline to Health" with a line above and below the text. The logo appears to be for a medical or healthcare-related company.
Image /page/3/Picture/1 description: The image shows the Raumedic logo. The logo consists of a stylized circle on the left, followed by the word "RAUMEDIC" in bold, sans-serif font. Below the word is the tagline "Lifeline to Health" in a smaller, italicized font, with a line on either side.
Based on performance testing and the available information concerning the referenced comparison device, the RAUMEDIC® -ICP-TEMP-Monitoring-System is equivalent in that:
The devices have the same intended use and indication for use.
- The devices are made of the same materials or substantially equivalent materials. .
- The devices have equivalent form, function, procedures and features. .
- Performance characteristics are suitable for designated indications for use
Based on this, the anticipated clinical performance of the RAUMEDIC® ICP-TEMP-Monitoring-System is equivalent to the referenced systems.
### 8. Device Testing
Biocompatibility studies were conducted per ISO 10993 standard and have demonstrated that the materials used to manufacture the RAUMEDIC® -ICP-TEMP-Monitoring-System are safe for its intended use.
In addition, the mentioned catheters were subjected to extensive performance testing. Results of the testing showed that the catheter designs are safe for their intended uses.
Finally, the manufacturing process of the RAUMEDIC® - ICP-TEMP- Monitoring - System complies with the United States Food and Drug Administration and European Standards for the manufacturing of medical devices.
Device #:
Page 4 of 4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Raumedic logo with the tagline "Lifeline to Health" underneath. The logo consists of a stylized circle with horizontal lines inside, followed by the word "RAUMEDIC" in bold, sans-serif font. Above the logo, there is a handwritten code "K120252".
See attachment 2: Device comparison tables (complete 5 pages).
Table 1
| Feature | RAUMEDIC® ICP-<br>TEMP-MONITORING-<br>SYSTEM | RAUMEDIC® ICP-<br>MONITORING-SYSTEM<br>(K103206) | RAUMEDIC® ICP-<br>MONITORING- SYSTEM<br>ventricular<br>(K112017) | SE? |
|---------------------------------------|-------------------------------------------------------------------------------|--------------------------------------------------|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Trade Name | NEUROVENT®-P-TEMP,<br>NEUROVENT®-TEMP-<br>IFD-S,<br>NEUROVENT®-TEMP-<br>IFD-R | NEUROVENT®-P,<br>NEUROVENT®-P-C | NEUROVENT®,<br>NEUROVENT®-IFD-S,<br>NEUROVENT®-IFD-R | N/A |
| Indication for Use | SAME | SAME | SAME | YES |
| Anatomical Site-Catheter | Brain parenchyma/<br>ventricle | Brain parenchyma | Brain ventricle | YES |
| ICP-Sensor Design<br>Catheter | Miniatur transducer | Miniatur transducer | Miniatur transducer | YES |
| Temperature-Sensor<br>Design Catheter | Thermistor | --- | --- | NO<br>-- see SE-List with<br>predicate devices:<br>Camino NeuroCare<br>Intracranial Pressure-<br>Temperature Monitoring Kit |
| Temperature Range | 25°C - 45°C | --- | --- | NO<br>-- see SE-List with<br>predicate devices:<br>Camino NeuroCare<br>Intracranial Pressure-<br>Temperature Monitoring Kit |
| Transducer Location | Catheter Tip | Catheter Tip | Catheter Tip | YES |
| Non-fluid coupling Catheter | YES | YES | YES | YES |
| Single-use-Catheter | YES | YES | YES | YES |
| Pressure Range | -50 to +250 mmHg | -50 to +250 mmHg | -50 to +250 mmHg | YES |
| Feature | RAUMEDIC® ICP-<br>TEMP- MONITORING-<br>SYSTEM | RAUMEDIC® ICP-<br>MONITORING- SYSTEM<br>(K103206) | RAUMEDIC® ICP-<br>MONITORING- SYSTEM<br>ventricular<br>(K112017) | SE? |
|----------------------------------|------------------------------------------------------|---------------------------------------------------|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Zero Point Stability<br>Catheter | Less than 1mmHg during<br>first 24 hours | Less than 1mmHg during<br>first 24 hours | Less than 1mmHg during<br>first 24 hours | YES |
| | Less than 2mmHg during<br>the first 7 days | Less than 2mmHg during<br>the first 7 days | Less than 2mmHg during<br>the first 7 days | |
| Sensitivity System | 5 µV/V/mmHg on the<br>monitor side | 5 µV/V/mmHg on the<br>monitor side | 5 µV/V/mmHg on the<br>monitor side | YES |
| Maximum pressure<br>Catheter | 1,250 mmHg | 1,250 mmHg | 1,250 mmHg | YES |
| Sterilization process | With Ethylene Oxide | With Ethylene Oxide | With Ethylene Oxide | YES |
| Bolt | Bolt with compression cap | Bolt with compression cap | Bolt with compression cap | YES |
| Spliceable tunnelling<br>sleeve | Spliceable Tunnelling<br>Sleeve with Trocar | --- | Spliceable Tunnelling<br>Sleeve with Trocar | YES |
| TEMP-Adapter | ICP-TEMP-Adapter<br>ICP-TEMP-Adapter HP /<br>Philips | --- | --- | NO<br>- see SE-List with<br>predicate devices:<br>Camino NeuroCare<br>Intracranial Pressure-<br>Temperature Monitoring Kit |
| Product Code | GWM | GWM | GWM | YES |
| Registration # | Pending | K103206 | K112017 | N/A |
| Applicant | RAUMEDIC AG<br>Hermann-Staudinger-Str.<br>2 | RAUMEDIC AG<br>Hermann-Staudinger-Str.<br>2 | RAUMEDIC AG<br>Hermann-Staudinger-Str.<br>2 | N/A |
| | 95233 Helmbrechts<br>Germany | 95233 Helmbrechts<br>Germany | 95233 Helmbrechts<br>Germany | |
{5}------------------------------------------------
# 11 6 - Lifeline to Health
# Table 2
.
ﺬ. ﻳﻌﺘ ﺮ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺮﺍﺟﻊ
| Feature | RAUMEDIC® ICP-TEMP-MONITORING-System | Camino® NeuroCare™ Intracranial Pressure and Temperature Monitoring Kit | Camino® NeuroCare™ Intracranial Pressure and Temperature Monitoring Kit | SE? |
|------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Trade Name | NEUROVENT®-P-TEMP, NEUROVENT®-TEMP-IFD-S, NEUROVENT®-TEMP-IFD-R | Micro-Ventricular Pressure-Temperature Monitoring Kit (110-4HMT) | Parenchymal Pressure-Temperature Monitoring Kit (110-4BT) | N/A |
| Indication for Use | SAME | SAME | SAME | YES |
| Anatomical Site-Catheter | Brain parenchyma/ventricle | Brain ventricle | Brain parenchyma | YES |
| ICP-Sensor Design Catheter | Miniatur transducer | Optical reflector | Optical reflector | NO - see SE-List with predicate devices: RAUMEDIC® ICP-MONITORING-SYSTEM and RAUMEDIC® ICP-MONITORING-SYSTEM ventricular |
| Temperature-Sensor Design Catheter | Thermistor | Thermistor | Thermistor | YES |
| Temperature Range | 25°C - 45°C | 30°C - 40°C | 30°C - 40°C | NO - see SE-discussion 1 below table |
| Transducer Location | Catheter Tip | Catheter Tip | Catheter Tip | YES |
| Non-fluid coupling Catheter | YES | YES | YES | YES |
| Single-use-Catheter | YES | YES | YES | YES |
- Page 1 of 3
| Feature | RAUMEDICO ICP-<br>TEMP- MONITORING-<br>System | Camino® NeuroCare™<br>Intracranial Pressure<br>and Temperature<br>Monitoring Kit | Camino® NeuroCare™<br>Intracranial Pressure<br>and Temperature<br>Monitoring Kit | SE? |
|----------------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pressure Range | -50 to +250 mmHg | -10 to +250 mmHg | -10 to +250 mmHg | NO<br>- see SE-List with<br>predicate devices:<br>RAUMEDICO® ICP-<br>MONITORING- SYSTEM<br>and<br>RAUMEDICO® ICP-<br>MONITORING- SYSTEM<br>ventricular |
| Zero Point Stability<br>Catheter | Less than 1 mmHg during<br>first 24 hours<br>Less than 2 mmHg during<br>the first 7 days | Less than 2 mmHg during<br>first 24 hours<br>Less than 1 mmHg per<br>day during the first 5 days | Less than 2 mmHg during<br>first 24 hours<br>Less than 1 mmHg per<br>day during the first 5 days | NO<br>- see SE-List with<br>predicate devices:<br>RAUMEDICO® ICP-<br>MONITORING- SYSTEM<br>and<br>RAUMEDICO® ICP-<br>MONITORING-SYSTEM<br>ventricular |
| Sensitivity System | 5 µV/V/mmHg on the<br>monitor side | 5 µV/V/mmHg on the<br>monitor side | 5 µV/V/mmHg on the<br>monitor side | YES |
| Maximum pressure<br>Catheter | 1,250 mmHg | 1,250 mmHg | 1,250 mmHg | YES |
| Sterilization process | With Ethylene Oxide | With Ethylene Oxide | With Ethylene Oxide | YES |
| Bolt | Bolt with compression cap | Bolt with compression cap | Bolt with compression cap | YES |
{6}------------------------------------------------
EDIC e to Health -Lifelin
.
| Feature | RAUMEDIC® ICP- TEMP-MONITORING- System | Camino® NeuroCare™ Intracranial Pressure and Temperature Monitoring Kit | Camino® NeuroCare™ Intracranial Pressure and Temperature Monitoring Kit | SE? |
|------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Spliceable tunnelling sleeve | Spliceable Tunnelling Sleeve with Trocar | - | - | NO - see SE-List with predicate devices: RAUMEDIC® ICP- MONITORING-SYSTEM and RAUMEDIC® ICP- MONITORING-SYSTEM ventricular |
| TEMP-Adapter | ICP-TEMP-Adapter ICP-TEMP-Adapter HP / Philips | ICT-XX (XX stands for specific Monitor-Type) | ICT-XX (XX stands for specific Monitor-Type) | YES |
| Product Code | GWM | GWM | GWM | YES |
| Registration # | Pending | K962928 | K962928 | N/A |
| Applicant | RAUMEDIC AG Hermann-Staudinger-Str. 2 95233 Helmbrechts Germany | RAUMEDIC AG Hermann-Staudinger-Str. 2 95233 Helmbrechts Germany | RAUMEDIC AG Hermann-Staudinger-Str. 2 95233 Helmbrechts Germany | N/A |
1) The beneature range the AAUKED(OP) (CATEM Street Met merature rager the Camines
News and Pressure and Temperation of the register vegilender register regisser regisser
:
Page 3 of 3
:
. '
.
-
.
{7}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR 1 1 2012
RAUMEDIC AG c/o Mr. Olaf Teichert TUV SUD America, Inc. 1775 Old Highway 8 NW New Brighton, MN 55112-1891
Re: K120252
Trade/Device Name: Raumedic ICP Monitoring System Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial Pressure Monitoring Device Regulatory Class: Class II Product Code: GWM Dated: January 24, 2012 Received: January 27, 2012
Dear Mr. Teichert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
### Page 2 - Mr. Olaf Teichert
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Earle R. Hunt Jr.
Malvina B. Eydelman, M Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Lifeline to Health
# Indications for Use
510(k) Number (if known): K
Device Name: Device, Monitoring, Intracranial Pressure and Temperature
Indications for Use:
The RAUMEDIC®-ICP-TEMP-Monitoring-System is indicated for use by a qualified
neurosurgeon for direct measurement of intracranial pressure and temperature in the
parenchyma and in the ventricle and cerebrospinal fluid drainage applications.
Use of the parenchymal and ventricular intracranial pressure monitoring kit with bolt is
contra-indicated in children under one year old.
The RAUMEDIC® catheters are MR Unsafe.
Prescription Use ✓
(Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Prescription Use X
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
510(k) Number K120252
K120252
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.