Sterile Piston Hypodermic Syringes (With/Without Needle) is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Device Story
Device consists of calibrated hollow barrel, movable plunger, and male Luer connector (Slip/Lock). Used for fluid injection or withdrawal. Operated by healthcare professionals in clinical settings. Syringe attaches to hypodermic single lumen needle via female Luer hub. Device is sterile, single-use, and non-pyrogenic. Benefits include precise fluid delivery or extraction. No electronic or software components.
Clinical Evidence
Bench testing only. Device performance conforms to ISO 7886-1 for syringes and ISO 7864 for needles. Biocompatibility testing conforms to ISO 10993.
Technological Characteristics
Materials: PP (barrel, hub), PE (plunger, nozzle cap, sheath), rubber (piston), stainless steel (needle). Principle: Manual piston-driven fluid displacement. Connectivity: None. Sterilization: EtO gas. Standards: ISO 7886-1, ISO 7864, ISO 10993.
Indications for Use
Indicated for use by clinicians to inject fluids into or withdraw fluids from the body. Applicable to patients requiring fluid aspiration or injection via hypodermic single lumen needle.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
[WUXI YUSHOU MEDICAL APPLICANCES CO., LTD] (WUXI CITY, JIANGSU, CHINA] Iston Syringe With/Without Needle]
510(k) Submission [PMN2012001] Rev 1.00 17/10/13
HACHASZET 002
## 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92
Date Prepared: Oct. 17.2013
# 1. Submitter Name and Address:
| Name: | WUXI YUSHOU MEDICAL APPLICANCES CO., LTD |
|--------------------|-------------------------------------------------------------------------------|
| Address: | 215NO. XIGANG ROAD, DONGBEITANG WUXI CITY,<br>JIANGSU, JIANGSU, 214191, CHINA |
| Contactor Name: | Garfield Wang |
| TEL: | +86-510-83777555 |
| E-mail: | Wangxuebo_11@hotmail.com |
| US Agent: | |
| Contact Name: | QIAN SHEN |
| Business Name: | P&L SCIENTIFIC INC. |
| Address:<br>STATES | 6840 SW 45TH LN, UNIT 5 Miami, Florida, 33155, UNITED |
| Phone: | 305-6094701 |
| Fax: | 305 -3970289 |
| E-mail Address: | info@plscientificinc.com |
## 2. Submission Devices Information:
**NOV** 20 2013
Trade/Proprietary Name:
- 1. Sterile Piston Hypodermic Syringes (With/Without Needle).
- 2. Hypodermic Needles.
#### Common Name:
- 1. Piston Syringe
- 2. Hypodermic Needle.
#### Classification name:
- 1. Piston Syringe.
- 2. Hypodermic Single Lumen Needle.
#### Class:
- 1. II.
- 2. II.
## Panel:
1. 80.
2. 80.
[PMN-510(K) SUBMISSION]
{1}------------------------------------------------
510(k) Submission [PMN2012001] 17/10/13
## Procodes:
- 1. FMF Piston Syringe and FMI Hypodermic Single Lumen Needle
- 2. FMI Hypodermic Single Lumen Needle
### 3. Predicate Devices Information:
- 1. Piston Syringe: Trade Name: BD Single Use, Hypodermic Syringe 510(K) Number: K110771 2. Hypodermic Needle: BD Hypoint™ Trade Name: 510(K) Number: K070440
### 4. Devices Description:
### 4.1 Sterile Piston Hypodermic Syringes (With/Without Needle).
The piston syringe is a device intended for medical purposes, consisting of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male Luer Slip/Lock connector (nozzle) for attaching the female Luer connector (hub) of a hypodermic single lumen needle, or for attaching other devices with a female Luer. The syringe is sterilized by EtO gas. And it is a Non-Pyrogenic and single use device. The mainly raw materials are PP, PE and rubber.
| Ref<br>Number | Model<br>Number | Description | Size |
|---------------|-----------------|----------------------------------------|-------------------------------------|
| YSPS001 | YSPS | Piston syringe<br>with/without needles | 0.5cc/ml≤Capacity of Syringe<2cc/ml |
| YSPS002 | YSPS | Piston syringe<br>with/without needles | 2cc/ml≤Capacity of Syringe<5cc/ml |
| YSPS003 | YSPS | Piston syringe<br>with/without needles | 5cc/ml≤Capacity of Syringe<10cc/ml |
| YSPS004 | YSPS | Piston syringe<br>with/without needles | 10cc/ml≤Capacity of Syringe<20cc/ml |
| YSPS005 | YSPS | Piston syringe<br>with/without needles | 20cc/ml≤Capacity of Syringe<30cc/ml |
| YSPS006 | YSPS | Piston syringe<br>with/without needles | 30cc/ml≤Capacity of Syringe<50cc/ml |
| YSPS007 | YSPS | Piston syringe<br>with/without needles | 50cc/ml≤Capacity of Syringe |
#### 4.2 Hypodermic Needles.
The Hypodermic Needle is a single lumen needle, designed for use with syringes and injection devices for general purpose fluid injection. The needle is sterilized by EtO gas. and it is a Non-Pyrogenic and single use device. The mainly raw materials are PP and stainless steel.
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>"We do different and and and the different and the different of the dest<br>I<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>AND AND AND AND AND AND A<br>. .<br>.<br>But the count of control of concession<br>стания из да в доставления все, внего повед величает в при<br>I | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Mission of Shows of Shoung Property of Children Artists of Children<br>1<br>ALAL-ANDREA-ALAL-A<br>I |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| | | | | |
| | | | | |
[PMN-510(K) SUBMISSION]
{2}------------------------------------------------
510(k) Submission (PMN2012001) Rav 1.00 17/10/13
| Number | Number | | | |
|---------|--------|---------------------------------|-------------|-----|
| YSHN001 | YSHN | Hypodermic Needle Normal-walled | 1/2 to 1" | 30G |
| YSHN002 | YSHN | Hypodermic Needle Normal-walled | 1/2 to 1" | 29G |
| YSHN003 | YSHN | Hypodermic Needle Normal-walled | 1/2 to 1" | 28G |
| YSHN004 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 27G |
| YSHN005 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 26G |
| YSHN006 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 25G |
| YSHN007 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 24G |
| YSHN008 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 23G |
| YSHN009 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 22G |
| YSHN010 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 21G |
| YSHN011 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 20G |
| YSHN012 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 19G |
| YSHN013 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 18G |
| YSHN014 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 17G |
| YSHN015 | YSHN | Hypodermic Needle Normal-walled | 1 to 1 1/2" | 16G |
### 5. Intended Use:
### 5.1 Sterile Piston Hypodermic Syringes (With/Without Needle).
Sterile Piston Hypodermic Syringes (With/Without Needle) is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
#### 5.2 Hypodermic Needles.
The hypodermic needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe.
#### 6. Technological Characteristics:
Through comparisons between the applicant devices with the predicate devices as follows tables. We believe the applicant devices are substantially equivalent with the predicate devices.
## 6.1 Sterile Piston Hypodermic Syringes (With/Without Needle) Comparison Table
| Element of Comparison | Submission Device | Predicate Device<br>K110771 |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Sterile Piston Hypodermic<br>Syringes (With/Without Needle) is<br>a device intended for medical<br>purposes that consists of a<br>calibrated hollow barrel and a | The BD Single Use,<br>Hypodermic Syringe is intended<br>for use by health care<br>professionals for general<br>purpose fluid aspiration/ |
{3}------------------------------------------------
.. . . . . . .
| | movable plunger. At one end of the<br>barrel there is a male connector<br>(nozzle) for fitting the female<br>connector (hub) of a hypodermic<br>single lumen needle. The device is<br>used to inject fluids into, or<br>withdraw fluids from, the body. | injection. |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Principle of Operation | Normal | Normal |
| Syringe Capacity | Various Sizes | Various Sizes |
| Nozzle Type | Luer | Luer |
| Lubricant for Barrel | Silicone Oil | Silicone Oil |
| Barrel Transparency | Transparent and Clear | Transparent and Clear |
| Gradations Legibility | Legible | Legible |
| Needle Gauge and Length | Various Sizes | Various Sizes |
| Lubricant for Needle | Silicone Oil | Silicone Oil |
| Materials | | |
| Barrel | PP | PP |
| Plunger | PE | PE |
| Piston | Rubber | Rubber |
| Nozzle Cap | PE | PE |
| Needle Hub | PP | PP |
| Needle | Stainless Steel | Stainless Steel |
| Performances | Conforms to ISO7886-1 | Conforms to ISO7886-1 |
| Biocompatibility | Conforms to ISO10993 | Conforms to ISO10993 |
| Labeling | Meet the requirements of 21 CFR<br>Part 801 | Meet the requirements of 21<br>CFR Part 801 |
··············································································································································································
# 6.2 Hypodermic Needles Comparison Table
| Element of Comparison | Submission Device | Predicate Device<br>K070440 |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The hypodermic needle is a device<br>intended to inject fluids into, or<br>withdraw fluids from, parts of the<br>body below the surface of the skin.<br>The device consists of a metal<br>tube that is sharpened at one end<br>and at the other end joined to a<br>female connector (hub) designed<br>to mate with a male connector<br>(nozzle) of a piston syringe. | The BD Hypoint™ Hypodermic<br>Needle is intended for use with<br>Syringes and injection devices<br>for general purpose fluid<br>aspiration/ injection. |
| Principle of Operation | Normal | Normal |
| Needle Gauge and Length | Various Sizes | Various Sizes |
| Lubricant for Needle | Silicone Oil | Silicone Oil |
| Materials | | |
| Needle Hub | PP | PP |
| Needle | Stainless Steel | Stainless Steel |
| Needle Sheath | PE | PE |
| Performances | Conforms to ISO7864 | Conforms to ISO7864 |
| Biocompatibility | Conforms to ISO10993 | Conforms to ISO10993 |
| Labeling | Meet the requirements of 21 CFR<br>Part 801 | Meet the requirements of 21<br>CFR Part 801 |
{4}------------------------------------------------
510(k) SubmIsslon [PMN2012001] Rev 1.00 17/10/13
.. ...
-------
# 7. Conclusion:
The materials, performance, and operational features of both the submitted device and the predicate device are substantially equivalent. Our device and the predicate device are same in intended use, Essential Component, Material, Sterile, Function etc.
END
02 Page 5 of 5
{5}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 20, 2013
Wuxi Yushou Medical Appliances Company, Limited C/O Garfield Wang Regulatory Associate 215 No. Xigang Road, Dongbeitang Wuxi City Jiangsu, China 21491
Re: K130230
Trade/Device Name: Sterile Piston Hypodermic Syringes (With/Without Needle) and Hypodermic Needles Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF/FMI Dated: October 18, 2013 Received: October 21, 2013
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kwame
Ulmer
for
Erin Keith M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K130230
#### Device Name
Sterile Piston Hypodermic Syringes (With/Without Needle)
#### Indications for Use (Describe)
Sterile Piston Hypodermic Syringes (With/Without Needed for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the fennector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from the body.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PSC Institutios services ( 411) 44 1 1.71.
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
Contract Contract Concerner Concerner Concerner Concerner Comments of the Super Market Concern Comments of Concern Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/7/Picture/12 description: The image shows the FDA logo. The letters "FDA" are stylized with a triple-line effect, giving them a bold and somewhat retro appearance. The logo is simple and monochromatic.
Digitally signed by Richard C. Chapman Date: 2013.11.19 13:00:19 -05'00'
FORM FDA 3881 (9/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.