K130212 · Wuxi Yushou Medical Applicances Co., Ltd. · FMI · Nov 20, 2013 · General Hospital
Device Facts
Record ID
K130212
Device Name
SAFETY HYPODERMIC NEEDLES WITH/WITHOUT SYRINGE
Applicant
Wuxi Yushou Medical Applicances Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Nov 20, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Safety Hypodermic Needle (With/Without Syringe) device is intended for use in the aspiration and injection of fluids for medical purposes. The Safety Hypodermic Needle (With/Without Syringe) is compatible for use with standard luer slip and luer lock syringes.
Device Story
Device consists of hypodermic needle with hinged safety sheath attached to connector hub, optionally including hypodermic syringe. Used by medical practitioners for fluid aspiration or injection. After needle withdrawal, practitioner manually activates hinged safety sheath via finger, thumb, or surface contact to cover needle, reducing sharps injury risk. Device features locking mechanism in sheath center/proximal end; hinge allows adjustment to desired position. Compatible with standard luer slip/lock syringes. Benefits include enhanced protection against accidental needle-sticks for healthcare personnel.
Clinical Evidence
Bench testing only. Device performance verified through conformance to ISO 7864 for hypodermic needles, ISO 7886-1 for piston syringes, and ISO 10993 for biocompatibility. No clinical data provided.
Technological Characteristics
Materials: Stainless steel needle, polypropylene (PP) hub/barrel, polyethylene (PE) sheath/plunger, rubber piston. Lubricant: Silicone oil. Features: Hinged safety sheath for sharps injury prevention. Connectivity: None. Sterilization: Not specified. Standards: ISO 7864, ISO 7886-1, ISO 10993. Form factor: Various gauges (16G-30G) and lengths (0.5-1.5 inch).
Indications for Use
Indicated for aspiration and injection of fluids for medical purposes in patients requiring hypodermic delivery or withdrawal. Compatible with standard luer slip and luer lock syringes. Includes a manual safety shield to minimize risk of accidental needle-stick after withdrawal from the body.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
## 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92
Date Prepared: Jan. 01.2013
## 1. Submitter Name and Address:
| Name: | WUXI YUSHOU MEDICAL APPLICANCES CO., LTD | |
|-----------------|-------------------------------------------------------------------------------|--------------|
| Address: | 215NO. XIGANG ROAD, DONGBEITANG WUXI CITY,<br>JIANGSU, JIANGSU, 214191, CHINA | |
| Contactor Name: | Garfield Wang | |
| TEL: | +86-510-83777555 | |
| E-mail: | Wangxuebo_11@hotmail.com | NOV 2 0 2013 |
| US Agent: | | |
| Contact Name: | QIAN SHEN | |
| Business Name: | P&L SCIENTIFIC INC. | |
| Address: | 6840 SW 45TH LN, UNIT 5 Miami, Florida, 33155, UNITED<br>STATES | |
| Phone: | 305-6094701 | |
| Fax: | 305 -3970289 | |
| E-mail Address: | info@plscientificinc.com | |
# 2. Submission Devices Information:
Trade/Proprietary Name: Safty Hypodermic Needles With/Without Syringe.
Common Name: Hypodermic Needle and Piston Syringe
Classification name: Hypodermic Single Lumen Needle and Piston Syringe.
Class: II and II.
Panel: 80 and 80.
Procodes: FMI - Hypodermic Single Lumen Needle and FMF - Piston Syringe and FMI - Hypodermic Single Lumen Needle
# 3. Predicate Devices Information:
Safety Hypodermic Needle With/Without Syringe. TERUMO® SurGuard®3 Safety Needle Trade Name: 510(K) Number: K113422 Trade Name: BD Single Use, Hypodermic Syringe
{1}------------------------------------------------
510(K) Number: K110771
### 4. Devices Description:
## Safety Hypodermic Needle With/Without Syringe
The Safety Hypodermic Needle With/Without Syringe consists of a hypodermic needle with a hinged safety sheath attached to the connector hub with or without an attached hypodermic syringe. This device features a hinged safety sheath attached to the needle hub. The safety sheath which are simultaneously activated when manually pressed over the needle immediately after use and just prior to disposal to minimize the possibility of sharps injury. The safety sheath is activated with a one-handed operation by pressing the sheath either by finger or thumb, or by surface activation.
The locking mechanisms are positioned within the center and proximal end of the Sheath. The hinge feature allows the medical practitioner to flexibly adjust the sheath to its desired position for use.
| Ref<br>Number | Model<br>Number | Description | Length | Gauge | Syringe |
|---------------|-----------------|--------------------------|-------------|-------|-------------|
| YSSN001 | YSSN | Safety Hypodermic Needle | 1/2 to 1" | 30G | 0.5 to 10cc |
| YSSN002 | YSSN | Safety Hypodermic Needle | 1/2 to 1" | 29G | 0.5 to 10cc |
| YSSN003 | YSSN | Safety Hypodermic Needle | 1/2 to 1" | 28G | 0.5 to 10cc |
| YSSN004 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 27G | 0.5 to 10cc |
| YSSN005 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 26G | 0.5 to 10cc |
| YSSN006 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 25G | 0.5 to 10cc |
| YSSN007 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 24G | 0.5 to 10cc |
| YSSN008 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 23G | 0.5 to 10cc |
| YSSN009 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 22G | 0.5 to 10cc |
| YSSN010 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 21G | 0.5 to 10cc |
| YSSN011 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 20G | 0.5 to 10cc |
| YSSN012 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 19G | 0.5 to 10cc |
| YSSN013 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 18G | 0.5 to 10cc |
| YSSN014 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 17G | 0.5 to 10cc |
| YSSN015 | YSSN | Safety Hypodermic Needle | 1 to 1 1/2" | 16G | 0.5 to 10cc |
### 5. Intended Use:
### Safety Hypodermic Needle With/Without Syringe
The Safety Hypodermic Needle (With/Without Syringe) device is intended for use in the aspiration and injection of fluids for medical purposes. The Safety Hypodermic Needle (With/Without Syringe) is compatible for use with standard luer slip and luer lock syringes.
{2}------------------------------------------------
Additionally, after withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle-stick.
# 6. Technological Characteristics:
Through comparisons between the applicant devices with the predicate devices as follows tables. We believe the applicant devices are substantially equivalent with the predicate devices.
| Element of Comparison | Submission Device | Predicate Device<br>K113422 |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The hypodermic needle is a device<br>intended to inject fluids into, or<br>withdraw fluids from, parts of the<br>body below the surface of the skin.<br>The device consists of a metal<br>tube that is sharpened at one end<br>and at the other end joined to a<br>female connector (hub) designed<br>to mate with a male connector<br>(nozzle) of a piston syringe. | The TERUMOO SurGuard®3<br>Safety Needle device is<br>intended for use in the<br>aspiration and injection of fluids<br>for medical purposes. The<br>TERUMO SurGuard®3 Safety<br>Needle is compatible for use<br>with standard luer slip and luer<br>lock syringes.<br>Additionally, after withdrawal of<br>the needle from the body, the<br>attached needle safety<br>shield can be manually<br>activated to cover the needle<br>immediately after use to<br>minimize risk of accidental<br>needlestick. |
| Principle of Operation | Normal | Normal |
| Needle Gauge and Length | Various Sizes | Various Sizes |
| Lubricant for Needle | Silicone Oil | Silicone Oil |
| Materials | | |
| Needle Hub | PP | PP |
| Needle | Stainless Steel | Stainless Steel |
| Needle Sheath | PE | PE |
| Sharps Injury Prevention | Needle safety shield | Needle safety shield |
| Features | | |
| Performances | Conforms to ISO7864 | Conforms to ISO7864 |
| Biocompatibility | Conforms to ISO10993 | Conforms to ISO10993 |
| Labeling | Meet the requirements of 21 CFR<br>Part 801 | Meet the requirements of 21<br>CFR Part 801 |
# Safety Hypodermic Needle With Syringe Comparison Table
| Element of Comparison | Submission Device | Predicate Device<br>K110771 |
|-----------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Intended Use | Sterile Piston Hypodermic<br>Syringes (With/Without Needle) is<br>a device intended for medical | The BD Single Use,<br>Hypodermic Syringe is intended<br>for use by health care |
{3}------------------------------------------------
| | purposes that consists of a<br>calibrated hollow barrel and a<br>movable plunger. At one end of the<br>barrel there is a male connector<br>(nozzle) for fitting the female | professionals for general<br>purpose fluid aspiration/<br>injection. |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| | connector (hub) of a hypodermic<br>single lumen needle. The device is | |
| | used to inject fluids into, or | |
| | withdraw fluids from, the body. | |
| Principle of Operation | Normal | Normal |
| Syringe Capacity | Various Sizes | Various Sizes |
| Nozzle Type | Luer | Luer |
| Lubricant for Barrel | Silicone Oil | Silicone Oil |
| Barrel Transparency | Transparent and Clear | Transparent and Clear |
| Gradations Legibility | Legible | Legible |
| Needle Gauge and Length | Various Sizes | Various Sizes |
| Lubricant for Needle | Silicone Oil | Silicone Oil |
| Materials | | |
| Barrel | PP | PP |
| Plunger | PE | PE |
| Piston | Rubber | Rubber |
| Nozzle Cap | PE | PE |
| Needle Hub | PP | PP |
| Needle | Stainless Steel | Stainless Steel |
| Performances | Conforms to ISO7886-1 | Conforms to ISO7886-1 |
| Biocompatibility | Conforms to ISO10993 | Conforms to ISO10993 |
| Labeling | Meet the requirements of 21 CFR<br>Part 801 | Meet the requirements of 21<br>CFR Part 801 |
7. Conclusion:
The materials, performance, and operational features of both the submitted device and the predicate device are substantially equivalent and are safe and effective for their intended use. The new device is safety.
Our device and the predicate device are same in intended use, Essential Component, Material, Sterile, Function etc.
END
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that resembles a stylized caduceus, a symbol often associated with medicine and healthcare.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## November 20, 2013
Wuxi Yushou Medical Appliances Company, Limited C/O Garfield Wang Regulatory Associate 215 No. Xigang Road Dongbeitang Wuxi City Jiangsu, China 21491
Re: K130212
Trade/Device Name: Safety Hypodermic Needles With/Without Syringe Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodernic Single Lumen Needle Regulatory Class: II Product Code: FMI/FMF Dated: October 18, 2013 Received: October 21, 2013
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), piease go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Image /page/5/Picture/6 description: The image shows the name "Kwame Ulmer" in a stylized font. The letters "EDA" are in between the first and last name. The letters "EDA" are in a blocky, geometric font, while the rest of the name is in a more rounded font.
for
Erin Keith M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device and Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K130212
#### Device Name
Safety Hypodermic Needles With/Without Syringe
#### Indications for Use (Describe)
The Safety Hypodermic Needle With/Without Syringe device is intended for use in the aspiration of fluids for medical purposes. The Safety Hypodernic Needle With/Without Syringe is compatible for use with standard her lock syringes.
Additionally, after withdraval of the needle from the body, the attached can be manually activated to cover the needle immediately after use to minimize risk of accidental needle-stick.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: December 31, 2013
See PRA Statement on last page.
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
2017 YOU WE BE BE ME FOR FDATUSEIONLY FOR FDATUSEIONEY ME BEAT BE BEAR BEAR BEAR BEAR BEAR CONSERVED Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Image /page/6/Picture/13 description: The image shows the FDA logo. The letters "FDA" are written in a stylized font, with each letter composed of multiple parallel lines. The logo is black and white and appears to be a vectorized image.
Digitally signed by Richard C. Chapman Date: 2013.11.19 12:50:38 -05:00,
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.