The Inton OTPS™ Biodegradable Mesh Plating System is intended to sustain the relative position of weak bony tissue, e.g. bone grafts, bone graft substitutes, or bone fragments position of weak bony chsode, eightsm Biodegradable Mesh Plating System is also indicated for cement restriction in total joint arthroplasty procedures. Only when used in conjunction with traditional rigid fixation, the Inion OTPSTM Biodegradable Mesh Plating System is intended to sustain the relative position of weak bony tissue in trauma and reconstructive orthopaedic procedures involving the following: - Long bones - Flat bones - Short bones - · Irregular bones - · Appendicular skeleton - Thorax When used alone (without traditional rigid fixation), the Inion OTPS™ Biodegradable when used arone (without and to maintain the relative position of bone grafts or bone graft substitutes in reconstructive orthopaedic procedures involving: · Tumor resections where bone strength has not been compromised. - · Iliac crest harvest sites.
Device Story
Inion OTPS™ Biodegradable Mesh Plating System consists of resorbable mesh and fixation screws; composed of polylactic acid copolymers [Poly (L-lactide-co-D,L-lactide) and poly (L-lactide-co-trimethylenecarbonate)]. Device maintains relative position of bone fragments, grafts, or substitutes; provides cement restriction in joint arthroplasty. Implants lose strength over 18-36 weeks; complete resorption occurs within 2-4 years. Used by orthopedic surgeons in trauma and reconstructive procedures. Provides temporary structural support to bony tissues; used either as an adjunct to traditional rigid fixation or alone for specific non-load-bearing applications like iliac crest harvest sites. Benefits include elimination of secondary removal surgery due to biodegradable nature.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Resorbable mesh and fixation screws. Materials: Poly(L-lactide-co-D,L-lactide) and poly(L-lactide-co-trimethylenecarbonate). Mechanical strength profile: 18-36 weeks in vivo; full resorption 2-4 years. Non-metallic, biodegradable fixation system.
Indications for Use
Indicated for patients requiring stabilization of weak bony tissue, bone grafts, or bone graft substitutes in orthopedic trauma and reconstructive procedures. Used for cement restriction in total joint arthroplasty. When used with rigid fixation, indicated for long, flat, short, and irregular bones, appendicular skeleton, and thorax. When used alone, indicated for tumor resections (where bone strength is not compromised) and iliac crest harvest sites.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K03196/ P42 '14
MAR - 1 2004
Image /page/0/Picture/2 description: The image shows a logo with the word "INION" in bold, white letters. Below "INION" is the phrase "ligent Solution" in a smaller, white font. The background is black, and there is a white circle to the left of the word "INION".
510(k) SUMMARY for the Inion OTPS™ Biodegradable Mesh Plating System
### MANUFACTURER
Inion Ltd. Lääkärinkatu 2 FIN-33520 Tampere
Contact Person: Hanna Marttila Regulatory Affairs Manager Lääkärinkatu 2 FIN-33520 Tampere Phone: +358 3 2306 600 Fax: +358 3 2306 601 Harma. Marttila@ Inion. fi
#### DEVICE NAME
Trade name: Inion OTPS™ Biodegradable Mesh Plating System Classification Name: Bone, Plate
ESTABLISHMENT REGISTRATION NUMBER 9710629
DEVICE CLASSIFICATION AND PRODUCT CODE
Classification panel: Orthopedic Regulatory Class: Class II 21 CFR 888.3030, 87 HRS (plate)
### PREDICATE DEVICES
(1) MacroPore OS Trauma System (K021164)
- (2) MacroPoreOS Protective Sheet (K994158)
(3) Synthes Resorbable Meshes and Sheets (K003788)
Date: 18.6.2003 Status: Final
{1}------------------------------------------------
Kos 1961 page 72
## DEVICE DESCRIPTION AND PRINCIPLES OF OPERATION
The Inion OTPS™ Biodegradable Mesh Plating System is intended to maintain the relative position of weak bony tissue such as bone grafts, bone graft substitutes, or bone fragments from comminuted fractures.
The Inion OTPS™ Biodegradable Mesh Plating System is composed of various size mesh Internation associated fixations screws made of resorbable polylactic acid / pianes and associate a copolymers [Poly (L-lactide-co-D,L-lactide) and poly (L-lactide-cotrimethylenecarbonate)]. The Inion OTPS™ Biodegradable Mesh Plating System implants armlually lose their strength during 18-36 weeks in vivo with complete strength loss and resorption within two to four years.
### EQUIVALENCE TO MARKETED PRODUCTS
The Inion OTPS™ Mesh Biodegradable Plating System is substantially equivalent to I in implants, indicated for the reinforcing of weak bony tissues in orthopaedic brodegradures and when used in conjunction with traditional rigid fixation., which have proodal 50 0(k) clearance, including MacroPoreOS Trauma System (K021164),
16ccroPoreOS Protective Sheet (K994158) and Synthes Resorbable Meshes and Sheets (KNO3788). These devices have the same intended use and principles of operation and very (1000700). Incol arrecteristics. Mechanical testing demonstrates that the device is substantially equivalent to the predicate ones. Differences between The Inion OTPS™ surstandary "Equirations on "The System and predicate devices do not raise any new questions of safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an abstract emblem resembling an eagle or bird-like figure with three horizontal lines above it, possibly representing feathers or wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 1 2004
Ms. Hanna Marttila Regulatory Affairs Manager Inion Ltd. Lääkärinkatu 2 FIN-33520 Tampere Finland
Re: K031961
Trade/Device Name: Inion OTPS™ Biodegradable Mesh Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: December 2, 2003 Received: December 4, 2003
Dear Ms. Marttila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been rcclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CHR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Hanna Marttila
This letter will allow you to begin marketing your device as described in your Section 510(k) I ins icter will anow you to ogin maing of substantial equivalence of your device to a legally premarket notineation: The Price in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrise for your as (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Oinci general mionnation on Journational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincercly yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K031961
Device Name: INION OTPS™ Biodegradable Mesh Plating System
### Indications For Use:
The Inton OTPS™ Biodegradable Mesh Plating System is intended to sustain the relative position of weak bony tissue, e.g. bone grafts, bone graft substitutes, or bone fragments position of weak bony chsode, eightsm Biodegradable Mesh Plating System is also indicated for cement restriction in total joint arthroplasty procedures.
Only when used in conjunction with traditional rigid fixation, the Inion OTPSTM Biodegradable Mesh Plating System is intended to sustain the relative position of weak bony tissue in trauma and reconstructive orthopaedic procedures involving the following:
- Long bones
- Flat bones
- Short bones
- · Irregular bones
- · Appendicular skeleton
- Thorax
When used alone (without traditional rigid fixation), the Inion OTPS™ Biodegradable when used arone (without and to maintain the relative position of bone grafts or bone graft substitutes in reconstructive orthopaedic procedures involving:
· Tumor resections where bone strength has not been compromised.
- · Iliac crest harvest sites.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use I (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
**(Division Sign-Off)**
Division of General, Restorative,
and Neurological Devices
Page 1 of *_*_
**510(k) Number** K03196/
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.