K111108 · Topcon Medical Laser Systems, Inc. · HQF · Aug 11, 2011 · Ophthalmic
Device Facts
Record ID
K111108
Device Name
PASCAL STREAMLINE 577 (WITH ACCESSORIES)
Applicant
Topcon Medical Laser Systems, Inc.
Product Code
HQF · Ophthalmic
Decision Date
Aug 11, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PASCAL® Streamline 577 (with Accessories) is intended for use in the treatment of ocular pathology in the posterior segment; Retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including: - proliferative and nonproliferative diabetic retinopathy . - macular edema . - choroidal neovascularization . - branch and central retinal vein occlusion . - the treatment of choroidal neovascularization associated with wet age-related macular . degeneration - . lattice degeneration - retinal tears and detachments . - retinopathy of prematurity . Intended for use in the treatment of ocular pathology in the anterior segment including: - . iridotomy - trabeculoplasty .
Device Story
PASCAL Streamline 577 is an Optically Pumped Semiconductor Laser (OPSL) system for ophthalmic photocoagulation. Input: 577 nm yellow laser light; 635 nm diode aiming beam. Operation: Physician uses integrated slit lamp or optional Laser Indirect Ophthalmoscope (LIO) to deliver laser energy to ocular tissue; energy delivery controlled via footswitch; system includes scanning pattern generator for precise application. Output: Laser pulses (100-2000 mW; 5-1000 ms duration) for thermal treatment of retinal/choroidal vascular/structural abnormalities or anterior segment procedures. Used in clinical settings by ophthalmologists. Output allows physician to perform targeted photocoagulation, potentially stabilizing vision or preventing further retinal damage in patients with specified ocular pathologies.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and comparison of technological characteristics to predicate devices.
Technological Characteristics
OPSL (Optically Pumped Semiconductor Laser) system; 577 nm treatment wavelength; 635 nm aiming beam. Integrated slit lamp with scanning pattern generator. Power output 0-2000 mW; pulse duration 5-1000 ms. Connectivity includes USB for printer. System includes control electronics, LCD touchscreen, footswitch, and emergency stop. Class II device.
Indications for Use
Indicated for patients requiring retinal or anterior segment photocoagulation for conditions including diabetic retinopathy, macular edema, choroidal neovascularization, retinal vein occlusions, wet age-related macular degeneration, lattice degeneration, retinal tears/detachments, retinopathy of prematurity, iridotomy, and trabeculoplasty.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
# Section 5 - 510(k) Summary
#### I. General Information
K111108
Submitter:
Topcon Medical Laser Systems, Inc. 3130 Coronado Drive Santa Clara, CA 95054, USA
Contact Person:
John Jossy Director of Quality Assurance and Regulatory Affairs 408-235-8202
Summary Preparation Date: June 7, 2011
#### II. Names
Device Name(s):
PASCAL® Streamline 577 (with Accessories)
Primary Classification Name(s): Ophthalmic laser; Laser powered surgical device
#### III. Predicate Devices
- PASCAL Streamline Photocoagulator, now owned/manufactured by Topcon Medical . Laser Systems (TMLS; previously owned by OptiMedica) (K100019)
- SUPRA 577.Y Photocoagulator, manufactured by Quantel Medical (K091581) .
- IO 577 Photocoagulator, manufactured by IRIDEX Corp. (K071687) .
#### IV. Product Description
The PASCAL® Streamline 577 (with Accessories) is an Optically Pumped Semiconductor Laser (OPSL: solid state) with integrated slit lamp that is intended for use in ophthalmic applications. The PASCAL® Streamline 577 system delivers 577 nm laser light (yellow) at a spot size range of 60 - 400 micrometers (um) to the focal plane of the slit lamp. The integrated slit lamp allows the physician (using a physician selected lens) to deliver the laser light to the treatment site. The treatment laser delivers single pulses or a scanned pattern of 100 to 2000 milliwatts (mW) per pulse duration of approximately 5 - 1000 milliseconds (ms) per pulse. A diode laser (635 nm) provides a visible aiming beam.
Laser light is delivered to the treatment site via the integrated slit lamp or via an optional Laser Indirect Ophthalmoscope (LIO) 577 with eye safety filter for ophthalmic indications.
The physician is able to control the delivery of laser energy using a footswitch.
The PASCAL® Streamline 577 (with Accessories) is comprised of the following functional components:
- Slit Lamp Table with integrated 577 nm laser system: .
- > 577 nm Laser system control electronics and power supply;
- > Control Panel with LCD (liquid crystal display) Touchscreen;
- > Keyswitch:
- Emergency Laser Stop Button;
{1}------------------------------------------------
- > Footswitch;
- 公 Door Interlock Plug;
- Power cable [4.6 m (15 ft.) long] A
- Connector Ports for the table lift power connector, the table lift up/down control 公 connector, the door interlock connector, the footswitch connector, and a USB connector for an optional printer.
- a Rolling base with lockable wheels.
- Printer (optional). a
- Slit Lamp with integrated scanning pattern generator (and laser eye safety filter). .
# Accessories (Optional)
- Laser Indirect Ophthalmoscope (LIO) 577 with eye safety filter. .
- Video Teaching System (VTS). .
- Physician Elbow Rest. .
### Intended Use & Indications for Use V.
The PASCAL® Streamline 577 (with Accessories) is intended for use in the treatment of ocular pathology in the posterior segment; Retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
- proliferative and nonproliferative diabetic retinopathy .
- macular edema .
- choroidal neovascularization .
- branch and central retinal vein occlusion .
- the treatment of choroidal neovascularization associated with wet age-related macular . degeneration
- . lattice degeneration
- retinal tears and detachments .
- retinopathy of prematurity .
Intended for use in the treatment of ocular pathology in the anterior segment including:
- . iridotomy
- trabeculoplasty .
{2}------------------------------------------------
## VI. Summary of Technological Characteristics
The technological characteristics of the PASCAL® Streamline 577 (with Accessories) are substantially equivalent to those of the predicate devices.
| | K 1 1 | K100019 | K091581 | K071687 |
|----------------------|-------------------------------|-------------------------------|-------------------------|-------------------------|
| | PASCAL Streamline 577 | PASCAL Streamline | SUPRA 577.Y | IQ 577 |
| | (with Accessories): Topcon | Photocoagulator; OptiMedica | Photocoagulator; | Photocoagulator; |
| | Medical Laser Systems | (now Topcon Medical Laser | Quantel Medical | IRIDEX Corp |
| Characteristics | | Systems) | | |
| Treatment Laser | | | | |
| Treatment 2 | 577 nm | 532 nm | 577 nm | 577 nm |
| Laser Type | OPSL - Optically Pumped | OPSL - Optically Pumped | Diode Pumped; Solid | Diode Pumped; Solid |
| | Semiconductor Laser | Semiconductor Laser | State | State |
| Power output | 0 - 2000 mW | 100 mW - 2000 mW | Up to 2000 mW | Up to 5000 mW |
| Duty cycle | 100% | 100% | Not reported | Variable |
| Pulse Duration | 5 ms - 1000 ms | 10 ms - 1000 ms | 10 ms to continuous | 10 us - 60 min |
| Repetition Rate | Off, 1.0, 1.5, 2.0, 3.0, 4.0, | Off. 1.0, 1.5, 2.0, 3.0, 4.0. | Not reported | < kHz |
| | 5.0, 6.0, 7.0, and 8.0 Hz | 5.0, 6.0, 7.0, and 8.0 Hz | | |
| | (available w/ single spot | (available w/ single spot | | |
| | pattern only) | pattern only) | | |
| Pulse counter | 0 - 99.9999 | 0 - 99 999 | Not reported | Not reported |
| | | | (pulse counter present) | (pulse counter present) |
| Aiming Beam | | | | |
| Aiming Beam 2 | 635 nm; Direct diode | 635 nm; Direct diode | 635 nm; diode | 630 - 670 nm (diode), |
| (power output) | (Adjustable to <1 mW) | (Adjustable to <1 mW) | | <1.0 mW |
| Other Specifications | | | | |
| Spot diameter | Slit Lamp Microscope | Slit Lamp Microscope | Microscope Adapter | Microscope Adapters |
| | (integrated) | (integrated) | | - 577 nm |
| | · 60 - 400 um delivered to | · 60 - 400 um delivered to | · 50 - 500 um | SLA/OMA Spot Size |
| | the focal plane of the slit | the focal plane of the slit | | Ranges: 50 - 5000 um |
| | lamp in air. | lamp in air. | | to the retina |
| | LIO 577 | LIO - 532 | LIO - 577 nm | LIO - 577 nm |
| | · 360 um (at fundus w/ 20D | • 360 um (at fundus w/ 20D | Unknown spot | 360 um & 1.3 mm (at |
| | lens) | lens) | diameter | retina w/ 20D lens) |
### Rationale for Substantial Equivalence VII.
The PASCAL® Streamline 577 (with Accessories) shares the same or similar indications for use, device operation, overall technical and functional capabilities, and therefore is substantially equivalent to the predicate devices.
# VIII. Safety and Effectiveness Information
The review of the indications for use and technical characteristics demonstrates that the PASCAL® Streamline 577 (with Accessories) is substantially equivalent to the predicate devices.
## IX. Conclusion
The PASCAL® Streamline 577 (with Accessories) was found to be substantially equivalent to the predicate devices.
The PASCAL® Streamline 577 (with Accessories) shares the same or similar indications for use, similar design features, and functional features with, and thus is substantially equivalent to, the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine. The caduceus has a staff with two snakes winding around it, and a pair of wings at the top.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Topcon Medical Laser Systems, Inc. % Mr. John Jossy Director, Quality Assurance and Regulatory Affairs 3130 Coronado Drive Santa Clara, California 95054
AUG 1 1 2011
Re: K111108
Trade/Device Name: PASCAL Streamline 577 (with Accessories) Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: Class II Product Code: HQF Dated: June 07, 2011 Received: June 09, 2011
Dear Mr. Jossy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Mr. John Jossy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Pari 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Malkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K11108
Device Name: PASCAL® Streamline 577 (with Accessories)
Indications for Use:
PASCAL® Streamline 577 (with Accessories) is intended for use in the treatment of ocular pathology in the posterior segment; Retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
- proliferative and nonproliferative diabetic retinopathy .
- . macular edema
- choroidal neovascularization ●
- branch and central retinal vein occlusion ●
- the treatment of choroidal neovascularization associated with wet age-related . macular degeneration
- lattice degeneration
- retinal tears and detachments .
- . retinopathy of prematurity
Intended for use in the treatment of ocular pathology in the anterior segment including:
- . iridotomy
- trabeculoplasty .
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evalu
NeLRPOgilem for mxn,
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
. 510(k) Number K111108
Page 1 of
TMLS Premarket Notification, 510(k) Submission for: PASCAL Streamline 577 (with Accessories)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.