K123475 · Lepu Medical Technology (Beijing) Co., Ltd. · DYB · Feb 3, 2014 · Cardiovascular
Device Facts
Record ID
K123475
Device Name
SHOOCIN INTRODUCER KIT
Applicant
Lepu Medical Technology (Beijing) Co., Ltd.
Product Code
DYB · Cardiovascular
Decision Date
Feb 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The ShoocinTM Introducer Kits are intended for use to facilitate the introduction of guide wires, catheters and other accessory medical devices through the skin into a vein or artery and minimize blood loss associated with such introduction.
Device Story
Shoocin Introducer Kit is a vascular access device consisting of a puncture needle, guide wire with collimator, sheath introducer, and dilator. Puncture needle creates initial access; guide wire acts as a track for sheath insertion. Sheath introducer provides a conduit for interventional devices; includes hemostasis valve and side port with 3-way stopcock to minimize blood loss. Dilator provides stability during sheath deployment. Components are radiopaque for fluoroscopic visualization. Used in clinical settings by physicians to gain vascular access for interventional procedures. Benefits include facilitated device delivery and reduced hemorrhage during access.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included biocompatibility (ISO 10993-1, -4, -5, -10, -11), fracture strength, leakage performance, radiodetectability, and guide wire/dilator accessibility. All tests met specified requirements.
Technological Characteristics
Components: Stainless steel needle, stainless steel guide wire, polycarbonate/acrylobutyistyrene/polyethylene/polyurethane/silicone plastics. Sheath/dilator contain bismuth for radiopacity. Sterilization: Ethylene oxide per AAMI/ANSI/ISO 11135:2007. Shelf life: 2 years per ISO 11070. Connectivity: None. Software: None.
Indications for Use
Indicated for facilitating the percutaneous introduction of guide wires, catheters, and accessory medical devices into the venous or arterial system while minimizing blood loss.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
K.123475
FEB - 3 2014
## Section 5: 510k) Summary
ﻪ ﺍ
The Summary of Safety and Ifferintion on the Shocial" in the submitted hat or revience with the requirences in the requirences in the requirences in the regirents in the rece
| Applicant: | Lepu Medical Technology (Beijing) Co., Ltd.<br>No. 37 Chaoqian Road<br>Changping District, Beijing 102200<br>P.R. China |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone: | +86-10-80120641 |
| Contact: | Shan Zhao |
| Date: | November 6, 2012 |
| Name: | ShoocinTM Introducer Kit |
| Classification Name: | Catheter Introducer, 870.1340 |
| Product Code: | DYB |
| Predicate: | Pinnacle Precision Access System, Terumo Corporation, K111606 with market<br>clearance dates of October 3, 2011. |
| Description: | The ShoocinTM Introducer Kit consists of a puncture needle, a guide wire with<br>a guide wire collimator, a sheath introducer, and a dilator. The puncture<br>needle incorporates a lumen, which provides a conduit for the insertion of the<br>guide wire into the vascular system. The flexible stainless steel guide wire is<br>utilized as a guiding mechanism for the insertion of the introduction sheath<br>into the vascular system. The guide wire contains a wire collimator, which<br>assists in funneling the wire through the lumen of the puncture needle and<br>contains marking that are visible under fluoroscopy, which can determine in<br>length of the guide wire within the vascular system. The sheath introducer<br>provides a conduit for introducing other interventional devices, including<br>guide wires and interventional catheters, into the vascular system. The main<br>components of this assembly are a non-hydrophilic coated sheath introducer,<br>hemostasis valve housing, and a side port with tubing connected to a 3-way<br>stopcock/valve. The dilator is used to provide support and stability to the<br>sheath introducer during deployment into the vascular system. The proximal<br>end of the dilator includes a luer port and has a tapered, atraumatic distal tip.<br>Both the sheath and dilator contain bismuth, making these devices visible<br>under fluoroscopy. |
| Intended Use: | The ShoocinTM Introducer Kits are intended for use to facilitate the<br>introduction of guide wires, catheters and other accessory medical devices<br>through the skin into a vein or artery and minimize blood loss associated with<br>such introduction. |
{1}------------------------------------------------
## Section 5: 510k) Summary
ﺰ
:
·
| Specification<br>Comparison: | | | The Shoocin™ Introducer Kit specifications are similar to the Pinnacle Access<br>System (K111606) and the differences do not raise any new issues of safety or<br>effectiveness. |
|------------------------------|---------------------------|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Pinnacle | Shoocin |
| | Sheath Diameter: | 4F through 11F | 4F through 11F |
| | Sheath Length: | 10cm - 25cm | 7cm - 23cm |
| | Guide Wire Length: | 45cm | 45cm and 80 cm |
| | Guide Wire OD: | .035" - .038" | .018" - .038" |
| | Introducer Needle: | 19 – 21G | 18 – 20G |
| | Dilator: | 15.5cm or 16.0cm | 7cm, 11cm, 16cm, 23cm |
| | | | |
| Materials<br>Comparison: | | | The Shoocin ™ Introducer Kit materials have been subjected to<br>biocompatibility tests and the differences between the Pinnacle Access System<br>(K111606) do not raise any new issues of safety or effectiveness. |
| | | Pinnacle | Shoocin |
| | Needle: | Stainless Steel | Stainless Steel |
| | Needle Luer: | Styrene-Butadiene | Polycarbonate |
| | Guide Wire: | Stainless Steel or Nitinol<br>w/Palladium distal coil | Stainless Steel |
| | Guide Wire<br>Collimator: | Polypropylene | Polycarbonate |
| | Dilator Tube: | Polypropylene w/Bismuth<br>subcarbonate | Polypropylene |
| | Dilator Luer: | Polypropylene | Acrylobutyistyrene |
| | Outer Sheath: | ETFE w15% BiO3, w/2%<br>TiO2 | Fluorinated Ethylene<br>Propylene |
| | Sheath base: | Styrene elastomer | Acrylobutyistyrene |
| | Side Port Tubing | Polybutadiene | Polyurethane |
| | Hemostasis Valve: | Silicone Rubber | Silicone |
| | 3-Way Stop Cock: | Polypropylene | Acrylobutyistyrene |
| | Cap: | Polypropylene | Polyethylene |
| | Valve: | Polypropylene | Polyethylene |
| Sterilization: | | | The method used is based on practices recommended by AAMI / ANSI / ISO<br>11135:2007 and provides a Sterility Assurance Level (SAL) of 106 |
| Shelf Life: | | | In accordance with ISO 11070 the real time aging of Shoocin ™ Introducer Kit<br>demonstrated that the performance of the specific components met the<br>standard requirements without any significant difference to product<br>performance requirements before aging. So the product is stable and reliable<br>within the two-year useful life |
…… ち "
, . .. !
· 《中语·
*
. '.
ﺒﻴ
. '
{2}------------------------------------------------
| <b>Section 5: 510k) Summary</b> | | | |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility: | The ShoocinTM Introducer Kit produced by Lepu Medical was assessed<br>against the International Standard ISO 10993-1, "Biological evaluation of<br>medical devices. Part 1. Guidance on selection of tests." The ShoocinTM<br>Introducer Kit would be classified as an External Communicating Device in<br>contact with the Circulating Blood for a Limited Duration (<24 hours). The<br>following test would be required for any patient / user contacting material: | | |
| | Test | Standard | Results |
| | Haemolytsis | ISO 10993-4 | Both the test article in direct<br>contact with blood and the test<br>article extract were non-<br>hemolytic. |
| | Cytotoxicity | ISO 10993-5 | The test article extract showed<br>no evidence of causing cell<br>lysis or toxicity |
| | Maximum<br>Sensitization | ISO 10993-10 | The test article extracts showed<br>no evidence of causing delayed<br>dermal contact sensitization in<br>the guinea pig. |
| | Intracutaneous<br>Irritation | ISO 10993-10 | The test article met the<br>requirements for the SC and SO<br>test extracts. |
| | Systemic Toxicity | ISO 10993-11 | There was no mortality or<br>evidence of systemic toxicity<br>from the extracts injected into<br>mice. |
| | USP Pyrogen Study | ISO 10993-11 | The test article was judged as<br>nonpyrogenic. |
| | <i>In Vivo</i><br>Thromboresistance | ISO 10993-4 | The test article and control<br>were thromboresistant and<br>comparable. |
| | Partial Thromboplastin<br>Time | ISO 10993-4 | The test article would be<br>considered a minimal activator<br>and met the requirements of the<br>test. |
| Performance Testing: | The ShoocinTM Introducer Kit successfully passed all of the following<br>performance tests: | | |
| | ShoocinTM | Introducer Kit Performance Testing | |
| | Sheath Introducer | Sheath Introducer should be free of defects | |
| | | Sheath Introducer ID | |
| | | Sheath Introducer Effective Length | |
| | | Radiodetectability | |
| | | Dilator accessibility | |
| | | Fracture Strength Test | |
| | | Leaking Performance Test | |
| <b>Section 5: 510k) Summary</b> | | | |
| Performance Testing: | The ShoocinTM Introducer Kit successfully passed all of the following performance tests: | | |
| | | ShoocinTM Introducer Kit Performance Testing | |
| | Dilator | Dilator should be free of defects | |
| | | Dilator OD | |
| | | Dilator ID | |
| | | Dilator Effective Length | |
| | | Radiodetectability | |
| | | Guide wire accessibility | |
| | Guide wire | Fracture Strength Test | |
| | | Surface free from impurity | |
| | | Guide wire length | |
| | | Guide wire OD | |
| | | Radiodetectability | |
| | | Corrosion Resistance | |
| | | Fracture Strength Test | |
| | | Flexing Resistance Test | |
| | | Associative Guide Wire Strength Test | |
| | Puncture Needle | Surface free from defects | |
| | | Needle OD | |
| | | Needle ID | |
| | | Needle Effective Length | |
| | | Accessibility - Guide wire | |
| | | Corrosion Resistance | |
| | | Joint Strength - Needle to hub | |
| <b>Substantial<br/>Equivalency<br/>Information:</b> | The information provided in this submission, comparing intended use, principle of operation and performance, the ShoocinTM Introducer Kit device is substantially equivalent to existing legally marketed device. | | |
| <b>Conclusion:</b> | The information provided in this submission and comparing intended use, principle of operation and overall technological characteristics (i.e. puncture needle, guide wire, dilator, and sheath introducer to obtain access to the vascular system), the ShoocinTM Introducer Kit supports a determination of substantially equivalent to existing legally marketed predicate device Pinnacle Precision Access System. Any technological differences between the ShoocinTM Introducer Kit and the predicate Pinnacle Precision Access System device do not raise new questions of safety or effectiveness. | | |
:
{3}------------------------------------------------
4
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wavy lines underneath, representing the department's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The logo is printed in black and white.
## 11 February 3, 2014
Lepu Medical Technology (Beijing) Co., Ltd. c/o Arthur Goddard FDA Regulatory and Quality Systems Consultant 1531 Felton Road South Euclid, Ohio 44121-2722
Re: K123475
Trade Name: Shoocin™ Introducer Kit Dated: January 21, 2014 Received: January 27, 2014
Dear Mr. Goddard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
{5}------------------------------------------------
Page 2 - Mr. Goddard
" If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDeyices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M. Zillberman
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Section 4: Indication for Use Summary
| 510(k) Number (if known): | K123475 |
|---------------------------|--------------------------------------|
| Device Name: | Shoocin <sup>TM</sup> Introducer Kit |
Indications For Use:
The Shoocin<sup>TM</sup> Introducer Kits are intended for use to facilitate the introduction of guide wires, catheters and other accessory medical devices through the skin into a vein or artery and minimize blood loss associated with such introduction.
| Prescription Use | X | OR | Over-The-Counter-Use |
|----------------------|---|----|----------------------|
| (Per 21 CFR 801.109) | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. Z. Hillman
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.