K111606 · Terumo Medical Corp. · DYB · Oct 3, 2011 · Cardiovascular
Device Facts
Record ID
K111606
Device Name
PINNACLE PRECISION ACCESS SYSTEM
Applicant
Terumo Medical Corp.
Product Code
DYB · Cardiovascular
Decision Date
Oct 3, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Pinnacle Precision Access System is used to facilitate placing a catheter through the skin into a vein or artery. The Entry Needle is an accessory device which is used to gain access to the vein or artery, for placement of the Mini Guide Wire. The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery.
Device Story
The Pinnacle Precision Access System is a manual introducer kit used to gain vascular access. It consists of an introducer sheath, dilator, metallic entry needle, mini guide wire, and guide wire inserter. The clinician uses the entry needle to access the vein or artery, inserts the mini guide wire through the needle, removes the needle, and then advances the sheath and dilator over the guide wire. Once the sheath is positioned, the dilator and guide wire are removed, allowing for subsequent catheter insertion. The sheath and dilator are radiopaque due to bismuth content. The device is used in clinical settings by healthcare professionals to facilitate interventional procedures. It benefits patients by providing a stable, controlled pathway for catheter delivery into the vasculature.
Clinical Evidence
Bench testing only. Performance testing included needle and guide wire mechanical integrity (joint strength, leakage, corrosion resistance, tip penetration), sheath/dilator functional testing (radiopacity, snap fit, wire passage), and simulated use in an anatomical model. Biocompatibility testing was performed per ISO 10993, including cytotoxicity, hemolysis, thrombogenicity, and systemic toxicity, confirming the device is non-hemolytic and non-cytotoxic.
Technological Characteristics
Materials: ETFE with bismuth/titanium dioxide (sheath), polypropylene (hubs/sleeves), stainless steel or Nitinol (guide wires), silicone rubber (valve). Principle: Manual percutaneous vascular access. Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Form factor: 4Fr-8Fr sheath sizes. Radiopacity: Bismuth-loaded sheath/dilator.
Indications for Use
Indicated for facilitating catheter placement into a vein or artery via percutaneous access. No specific patient population, age, or gender restrictions are stated.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
OCT - 3 2011
K1116.06
Terumo Medical Corporation Premarket Notification - Pinnacle Precision Access System™ Section II. 510(k) Summary
# SECTION II. 510(K) SUMMARY
#### A. Device Name
| Proprietary Name | Pinnacle Precision Access System |
|---------------------|---------------------------------------|
| Classification Name | Catheter Introducer (as per 870.1340) |
| Common Name | Introducer Sheath |
| Product Code | DYB |
#### B. Intended Use
The Pinnacle Precision Access System is used to facilitate placing a catheter through the skin into a vein or artery.
The Entry Needle is an accessory device which is used to gain access to the vein or artery, for placement of the Mini Guide Wire.
The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery.
## C. Device Description
The Pinnacle Precision Access System consists of an introducer sheath and a dilator which are packaged together with a metallic entry needle, a mini guide wire and a guide wire inserter, prior to sterilization. The Pinnacle Precision Access System is used to facilitate placing a catheter through the skin into a vein or artery. The sheath and dilator contain bismuth, making these devices visible under fluoroscopy
The entry needle (cannula) is an accessory device which is used to gain access to the vein or artery for placement of the mini guide wire.
The mini guide wire is an accessory device which is used for placement of the sheath and dilator into the vein or artery. The mini guide wire is offered in two versions, a stainless steel (spring coil) model and a Palladium tipped Nitinol model.
A guide wire inserter is also provided to assist in insertion of the mini guide wire into the cannula.
{1}------------------------------------------------
Terumo Medical Corporation Premarket Notification - Pinnacle Precision Access System™ Section II. 510(k) Summary
### D. Principle of Operation / Technology
The Pinnacle Precision Access System and its accessories are operated manually or by a manual process.
The mini guide wire is inserted through a cannula placed in the patient's blood vessel. A guide wire inserter is provided to assist in insertion of the mini guide wire into the cannula. Following guide wire insertion, the cannula is removed and the sheath and dilator are then inserted over the mini guide wire and into the blood vessel. The mini guide wire is then withdrawn from the vessel. The dilator maintains the integrity of the Sheath and dilates the blood vessel during insertion. Once the sheath is situated in the vessel, the dilator is removed and an appropriate catheter can then be inserted through the sheath.
#### E. Design / Materials and Comparison to Predicate Devices
The primary components of the Pinnacle Precision Access System in this submission (the Sheath and Dilator) are unchanged from those of the Radifocus Introducer II predicate device (K954234). The Pinnacle Precision Access System differs from the Radifocus Introducer II predicate device in the addition of the metallic entry needle and the optional Nitinol Mini Guide Wire. A metallic entry needle and Nitinol Mini Guide Wire are included in the Glidesheath predicate device (K102008).
The Pinnacle Precision Access System submitted in this 510(k), the Radifocus Introducer II (K954234) and the Glidesheath (K102008) have similar components which function in the same manner. Material differences between the devices, as shown in the following table, do not raise any new issues of safety and effectiveness.
| Component | Radifocus Introducer II (K954234) | Glidesheath (K102008) | Pinnacle Precision Access System |
|-------------------|--------------------------------------------|--------------------------------------------|-------------------------------------------------------|
| Wire | Stainless Steel | Nitinol w/Palladium<br>distal coil | Stainless Steel or Nitinol<br>w/Palladium distal coil |
| Needle | N/A | Stainless Steel | Stainless Steel |
| Needle Hub | N/A | Styrene-Butadiene<br>copolymer | Styrene-Butadiene<br>copolymer |
| Protective Sleeve | N/A | Polypropylene | Polypropylene |
| Dilator Tubing | Polypropylene<br>w/Bismuth<br>subcarbonate | Polypropylene<br>w/Bismuth<br>subcarbonate | Polypropylene<br>w/Bismuth subcarbonate |
| Dilator Stiffener | N/A | N/A | Stainless Steel (optional<br>on 4Fr and 5Fr) |
{2}------------------------------------------------
## Terumo Medical Corporation Premarket Notification - Pinnacle Precision Access System™ Section II. 510(k) Summary
| Dilator Caulking<br>Pin | Stainless Steel | Stainless Steel | Stainless Steel |
|-------------------------|--------------------------------------------------|--------------------------------------------------|--------------------------------------------------|
| Dilator Hub | Polypropylene | Polypropylene | Polypropylene |
| Sheath Tubing | ETFE w/15% BiO3,<br>w/2% TiO2 | ETFE w/15% BiO3,<br>w/2% TiO2 | ETFE w/15% BiO3,<br>w/2% TiO2 |
| Sheath Caulking<br>Pin | Stainless Steel | Stainless Steel | Stainless Steel |
| Sheath Housing | Polypropylene | Polypropylene | Polypropylene |
| Sheath Support | Styrene elastomer | Styrene elastomer | Styrene elastomer |
| Valve | Silicone rubber | Silicone rubber | Silicone rubber |
| Cap | Polypropylene | Polypropylene | Polypropylene |
| Sidetube | Polybutadiene | Polybutadiene | Polybutadiene |
| 3 WSC | Polycarbonate,<br>Polyethylene,<br>Polypropylene | Polycarbonate,<br>Polyethylene,<br>Polypropylene | Polycarbonate,<br>Polyethylene,<br>Polypropylene |
| Guide wire<br>inserter | Polypropylene | Polypropylene | Polypropylene |
### F. Specifications and Comparison to Predicate Devices
The Pinnacle Precision Access System submitted in this 510(k), the Radifocus Introducer II (K954234) and the Glidesheath (K102008) have similar device specifications. Differences in specifications between the devices do not raise any new issues of safety and effectiveness.
| | Radifocus Introducer<br>II (K954234) | Glidesheath<br>(K102008) | Pinnacle Precision<br>Access System |
|-----------------------|--------------------------------------|--------------------------|-------------------------------------|
| Sheath Sizes | 4Fr - 11Fr | 5Fr and 6Fr | 4Fr through 8Fr |
| Sheath Length | 5cm - 100cm | 10 cm | 10 cm |
| Sheath Coating | Silicone | Hydrophilic | Silicone |
| Dilator Length | 6cm - 110cm | 15.5 cm | 15.5cm or 16.0cm |
| Guide Wire OD | .021" - .038" | .021" | .021" |
| Guide Wire<br>Length | 40cm - 100cm | 45cm | 45cm |
| Entry Needle:<br>Size | N/A | 21G<br>1.5" | 21G/19G<br>2.75" |
{3}------------------------------------------------
Terumo Medical Corporation Premarket Notification - Pinnacle Precision Access System™ Section II. 510(k) Summary
# G. Performance
.
The Pinnacle Precision Access System successfully passed all of the following performance tests: .
| | Pinnacle Precision Access System Performance Testing |
|----------------|----------------------------------------------------------|
| Needle: | Needle surface free from defects |
| | Needle OD |
| | Needle length |
| | Needle ID |
| | Needle hub conical entry angle |
| | Bevel indicator visibility |
| | Bevel indicator position |
| | Needle to hub joint strength |
| | Gauge luer taper |
| | Liquid leakage from fitting assembly under pressure |
| | Air leakage into the fitting assembly during aspiration |
| | Separation force of fitting assembly |
| | Unscrewing torque of fitting assembly |
| | Ease of assembly |
| | Resistance to overriding |
| | Stress cracking |
| | Tip penetration through thin film |
| | Corrosion resistance |
| Guidewires: | Guidewire surface free from defects |
| | Tip buckling test |
| | Test for resistance of guidewires to damage by flexing |
| | Test for fracture of guidewires |
| | Test for distal tip retention and proximal end retention |
| | Guidewire OD |
| | Guidewire length |
| | Corrosion resistance |
| | Radiopacity |
| Dilator: | Dilator surface free from defects |
| | Dilator tip ID |
| | Dilator to hub joint strength |
| | Dilator length |
| | Dilator OD at sheath tip interface |
| | Dilator hub to sheath hub snap fit strength |
| | Hypotube length |
| | Hypotube to hub joint strength |
| | Hypotube fall-out |
| | Wire passage |
| | |
| Sheath: | Corrosion resistance (hypotube) |
| | Sheath surface free from defects |
| | Sheath tip ID |
| | Sheath length |
| | Sheath tip cracks |
| | Radiopacity |
| Simulated Use: | System use in anatomical model |
| | Dilator and sheath tip penetration |
{4}------------------------------------------------
Terumo Medical Corporation Premarket Notification - Pinnacle Precision Access System™ Section II. 510(k) Summary
# H. Biocompatibility and Sterilization
Biocompatibility testing was conducted in accordance with the FDA General Program memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing''.
The Pinnacle Precision Access System is classified as an Externally Communicating Device, Circulating Blood, Prolonged Contact (up to 30 days), the same as the Radifocus Introducer II Sheath predicate device. Biocompatibility of the Radifocus Introducer II Sheath, classified as Externally Communicating Device, Circulating Blood, Prolonged Contact (24 hours - 30 days) has been established under K954234. Differences in materials between the Pinnacle Precision Access System and the Radifocus Introducer II Sheath affect only those portions of the system that contact the patient for a short period of time (less than 24 hours). Therefore, the full battery of biocompatibility tests was performed on the whole device for Externally Communicating Devices, Circulating Blood, Limited Contact (up to 24 hours).
Testing was performed on the worse case configuration of the system: Introducer sheath with stiffener, stainless steel guide wire and echogenic step needle. Results are presented in the following table.
{5}------------------------------------------------
# Terumo Medical Corporation Premarket Notification -- Pinnacle Precision Access System™ Section II. 510(k) Summary
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
: :
| Biocompatibility Testing on non aged, 2x EO sterile Pinnacle Precision Access System | | |
|--------------------------------------------------------------------------------------|--------------|--------------------------------------------------------------|
| Test | Standard | Result |
| Physicochemical profile | USP | Meets requirements |
| Cytotoxicity | ISO 10993-5 | Not considered to have cytotoxic potential |
| Hemolysis | ASTM F756 | Non-hemolytic |
| In vitro<br>Hemocompatibility Assay | ISO 10993-4 | Pass |
| Thrombogenicity Study in<br>Dogs | ISO 10993-4 | Thrombosis was not considered significant |
| Complement Activation | ISO 10993-4 | Meets requirements |
| Unactivated Partial<br>Thromboplastin time | ISO 10993-4 | Meets requirements |
| Prothrombin Time | ISO 10993-4 | No adverse effect on the prothrombin time of<br>human plasma |
| Sensitization | ISO 10993-10 | Meets requirements |
| Intracutaneous Reactivity | ISO 10993-10 | Meets requirements |
| Acute Systemic Toxicity | ISO 10993-11 | Negative |
| Pyrogenicity | ISO 10993-11 | Meets requirements |
| Genotoxicity | ISO 10993-3 | Not considered to be mutagenic |
Alkalinity/acidity and extractables testing was performed per ISO 7864: Sterile Hypodermic Needles for Single Use on 2x EO sterile, non-aged needles.
| Pinnacle Precision Access System Echogenic Taper Needle (non-aged, 2xEO sterile) | | |
|----------------------------------------------------------------------------------------|--------------------------|--------------------|
| Test | Standard | Results |
| Analysis of Metals in<br>Extract by Inductively<br>Coupled Plasma Mass<br>Spectrometry | (ICP-MS) Per<br>ISO 7864 | Meets requirements |
{6}------------------------------------------------
Limited screening tests were conducted on accelerated aged, 2x EO sterile devices to demonstrate that aging does not affect the device's biocompatibility.
| Test | Standard | Results |
|-------------------------|-------------|--------------------------------------------|
| Physicochemical profile | USP | Meets requirements |
| Cytotoxicity | ISO 10993-5 | Not considered to have cytotoxic potential |
| Hemolysis | ASTM F756 | Non-hemolytic |
Sterilization conditions have been validated according to ANSI / AAMI / ISO 11135, Sterilization of Health Care Products-- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, to provide a Sterility Assurance Level (SAL) of 10-6.
Residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) residuals will meet requirements for limited exposure devices (contact up to 24 hours) prior to use, based on ISO 10993-7. Biological Evaluation of medical devices- Part 7: Ethylene Oxide Sterilization residuals. Residual EO will not exceed 4 mg per device and residual ECH will not exceed 9 mg per device.
The Pinnacle Precision Access System is certified to be non-pyrogenic in the unopened and undamaged package. Kinetic Turbidimetric Limulus Amebocyte Lysate (LAL) test is performed on each lot of production accordance to the United States Pharmacopoeia (USP) <85> Bacterial Endotoxins Test. Validation was performed in accordance with FDA published "Guideline on Validation of the Limulus Amebocyte Lysate Test as an End-Product Endotoxin Test for Human and Animal Parenteral Drugs, Biological Products, and Medical Devices"; 1987.
{7}------------------------------------------------
Terumo Medical Corporation Premarket Notification – Pinnacle Precision Access System™ Section II. 510(k) Summary
# I. Substantial Equivalence
The performance of the Pinnacle Precision Access System in this submission is substantially equivalent to the performance of the predicate devices. The equivalence was shown through comparison of component materials and specifications, performance and biocompatibility testing and sterilization validation.
The Pinnacle Precision Access System is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the predicate devices the Radifocus Introducer II (K954234) and the Glidesheath (K102008). Differences between the devices do not raise any significant issues of safety or effectiveness.
### J. Submitter Information
| Prepared By: | Mr. Daniel R. Plonski<br>Regulatory Affairs Specialist |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared For: | Terumo Medical Corporation<br>950 Elkton Blvd.<br>Elkton, MD 21921<br>Phone: (410) 392-7395<br>Fax: (410) 398-6079<br>Email: daniel.plonski@terumomedical.com |
| Date Prepared: | June 8, 2011 |
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the Department of Health and Human Services. The logo is a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the edge. Inside the circle is a symbol that looks like a stylized human figure with three lines representing the arms and legs. The figure is facing to the right.
Food and Drug Administration 10903 New Flampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Terumo Medical Corporation c/o Mr. Daniel R. Plonski, RAČ Regulatory Affairs Specialist 950 Elkton Blvd. Elkton. MD 21921
OCT - 3 2011
Re: K111606
Trade/Device Name: Pinnacle Precision Access System™ Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: September 23, 2011 Received: September 26, 2011
Dear Mr. Plonski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{9}------------------------------------------------
Page 2 – Mr. Daniel R. Plonski, RAC
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.him for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
sincerely yours,
Bram D. Zuckerman. M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use
510(k) Number (if known): _
K111606
Pinnacle Precision Access System™ Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
The Pinnacle Precision Access System is used to facilitate placing a catheter through the skin into a vein or artery.
The Entry Needle is an accessory device which is used to gain access to the vein or artery, for placement of the Mini Guide Wire.
The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sian-Off) Division of Cardiovascular Devices
510(k) Number K111606
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.