IB Clinic v1.0 (Clinic) is a post-processing software toolkit designed to be integrated into existing medical image visualization applications running on standard computer hardware. Clinic accepts relevant DICOM image sets, such as dynamic perfusion and diffusion image sets. Clinic generates various perfusion- and diffusion-related parameters, standardized image sets, and image intensity differences. The results are saved to a DICOM image file and may be further visualized on an imaging workstation. Clinic is designed to aid trained physicians in advanced image assessment, treatment consideration, and monitoring of therapeutic response. The information provided by Clinic should not be used in isolation when making patient management decisions. Dynamic Perfusion Analysis - Generates parametric perfusion maps used for visualization of temporal variations in dynamic datasets, showing changes in image intensity over time. These maps may aid in the assessment of the extent and type of perfusion, blood volume and vascular permeability changes. Dynamic Diffusion Analysis - Generates apparent diffusion coefficient maps used for the visualization of apparent water movement in soft tissue throughout the body on both voxel-by-voxel and sub-voxel bases. These images may aid in the assessment of the extent of diffusion in tissue. Image Comparison - Generates co-registered image sets. Generates standardized image sets calibrated to an arbitrary scale to facilitate comparisons between independent image sets. Generates voxel maps of the image intensity differences between image sets acquired at different times. Facilitates selection and DICOM export of user-selected regions of interest (ROIs). These processes may enable easier identification of image intensity differences between images and easier selection and processing of ROIs.
Device Story
IB Clinic is a software library integrated into medical image visualization workstations; processes MR and CT DICOM images. Inputs: dynamic perfusion and diffusion image sets. Operations: computes parametric perfusion maps (blood volume, vascular permeability), apparent diffusion coefficient (ADC) maps, image registration, intensity standardization, and voxel-wise intensity difference mapping. Outputs: DICOM-formatted parameter maps and ROI datasets for PACS/workstation export. Used by physicians in clinical settings to assist in therapeutic monitoring and treatment planning. Does not provide diagnostic conclusions; intended as an adjunct to clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only; verification and validation performed via software design process, requirements traceability, and risk analysis.
Technological Characteristics
Software-only library; platform-independent; runs on standard computer hardware. Functions: DICOM image retrieval, DSC perfusion computation, ADC diffusion computation, image registration, intensity scaling, and ROI generation. Connectivity: DICOM-compliant export to PACS/OS-based storage. Software-based processing.
Indications for Use
Indicated for trained physicians to aid in advanced image assessment, treatment consideration, and monitoring of therapeutic response using MR and CT DICOM image sets. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Imaging Biometrics. The words "Imaging Biometrics" are stacked on top of each other, with the "I" in Imaging and the "B" in Biometrics being larger than the other letters. Below the words "Imaging Biometrics" is the phrase "Creating the Standard."
# 510(k) SUMMARY
: . : . :
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.....
JAN 1 1 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92
The assigned 510(k) number is: k123302
# Submitter's Identification:
Imaging Biometrics, LLC 13416 Watertown Plank Road, Suite 260 Elm Grove, WI 53122
Contact Person Michael Schmainda Submitter, Holder, and Owner Phone: (262) 439-8252 Email: mike@imagingbiometrics.com
- Steve Krueger Consultant Phone: (262) 894-6427 Email: steve@makcompliance.com
Date Summary Prepared: October 19th, 2012
# Name of the Device:
IB Clinic v1.0 (Clinic) Trade Name:
Classification Name: Picture Archiving and Communication Systems 21 CFR 892 2050
Classification: Class II Product Code: LLZ
Performance Standards: None established under Food Drug and Cosmetic Act
# Common or Usual Name:
0
Radiological Image Processing Systems
13416 Watertown Plan Road, Suite 260 |Elm Grove, WI 53122| (262) 439-82521
{1}------------------------------------------------
## 4. Predicate Device Information:
Clinic is substantially equivalent to the following legally marketed devices that are currently cleared by the FDA:
| 510(k) | Trade Name | Manufacturer |
|---------|---------------|--------------------|
| K080762 | IB Neuro v1.0 | Imaging Biometrics |
| K090546 | nordiclCE | Nordic Imaging Lab |
## 5. Device Description:
Clinic is a platform independent image processing library which consists of a set of code modules which run on standard computer hardware that computes a variety of numerical analyses, image parameter maps, and other image manipulations based on DICOM images captured via MR and CT modalities. These actions include:
- Retrieval of MR and CT DICOM image studies from PACS and/or OS-. based file storage.
- Computation of parameter maps for: .
- DSC perfusion (based on MR and CT studies) O
- o ADC diffusion (based on MR studies)
- Image manipulations (of MR and CT studies): .
- Registration of images generated at different time points o
- Standardized scaling of image intensities O
- o Comparison of registered and/or standardized images
- Region of Interest (ROI) selection o
- Generation of ROI datasets in DICOM formats O
- Output of the above maps in DICOM format for export to PACS and/or OS . file storage
- . Generation of reports summarizing the computations performed
The IB Clinic code library can be used within standalone FDA cleared applications or can be "plugged in" and launched from within other FDA cleared applications such as Aycan's OsiriX PRO workstation. They are intended for distribution both in combination and in modular form, with functional subsets geared toward specific types of image analysis and marketed with corresponding names, including IB Neuro, IB Diffusion, and IB Delta Suite.
#### 6. Intended Use:
IB Clinic v1.0 (Clinic) is a post-processing software toolkit designed to be integrated into existing medical image visualization applications running on standard computer hardware. Clinic accepts relevant DICOM image sets, such as dynamic perfusion and diffusion image sets. Clinic generates various perfusion- and diffusion-related parameters, standardized image sets, and image
{2}------------------------------------------------
intensity differences. The results are saved to a DICOM image file and may be further visualized on an imaging workstation.
Clinic is designed to aid trained physicians in advanced image assessment. treatment consideration, and monitoring of therapeutic response. The information provided by Clinic should not be used in isolation when making patient management decisions.
Dynamic Perfusion Analysis - Generates parametric perfusion maps used for visualization of temporal variations in dynamic datasets, showing changes in image intensity over time. These maps may aid in the assessment of the extent and type of perfusion, blood volume and vascular permeability changes.
Dynamic Diffusion Analysis - Generates apparent diffusion coefficient maps used for the visualization of apparent water movement in soft tissue throughout the body on both voxel-by-voxel and sub-voxel bases. These images may aid in the assessment of the extent of diffusion in tissue.
Image Comparison - Generates co-registered image sets. Generates standardized image sets calibrated to an arbitrary scale to facilitate comparisons between independent image sets. Generates voxel maps of the image intensity differences between image sets acquired at different times. Facilitates selection and DICOM export of user-selected regions of interest (ROIs). These processes may enable easier identification of image intensity differences between images and easier selection and processing of ROIs.
## 7. Substantial Equivalence / Comparison to Predicate Devices:
The intended use and performance characteristics for Clinic are substantially equivalent to the predicate devices listed in section 4 above for image analysis. image processing and generation of parametric maps to provide additional information bevond standard imaging.
## 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
lmaging Biometrics used the following quality assurance measures for the development of Clinic:
- Product requirements and requirements review .
- Product requirements traceability analysis .
- . Risk analvsis
- . Design reviews
- . Software design process
- . Performance testing (verification)
- . Safety testing (verification)
- Product software validation .
{3}------------------------------------------------
- Product readiness review and approval .
#### Discussion of Clinical Tests Performed: 9.
N/A
## 10. Conclusions:
Clinic has the same intended use and same characteristics as the predicate devices. Moreover, documentation supplied in this submission demonstrates that any difference in technological characteristics do not raise any new questions of safety or effectiveness. Thus, Clinic is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized depiction of an eagle or bird-like figure, with flowing lines representing its wings or feathers. The logo is presented in black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
January 11, 2013
Mr. Mike Schmainda IMAGING BIOMETRICS, LLC 13416 Watertown Plank Rd Suite 260 Elm Grove, WI 53122
Re: K123302
Trade/Device Name: IB CLINIC V1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: II Product Code: LLZ Dated: October 19, 2012 Received: October 23, 2012
Dear Mr. Schmainda:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
## Page 2 - Mr. Schmainda
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Sean M. Boyd -S
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123302
IB Clinic v1.0 Device Name:
Indications for Use:
IB Clinic v1.0 (Clinic) is a post-processing software toolkit designed to be integrated into existing medical image visualization applications running on standard computer hardware. Clinic accepts relevant DICOM image sets, such as dynamic perfusion and diffusion image sets. Clinic generates vanous perfusion- and diffusion-related parameters, standardized image sets, and image intensity differences. The results are saved to a DICOM image file and may be further visualized on an imaging workstation.
Clinic is designed to aid trained physicians in advanced image assessment, treatment consideration, and monitoring of therapeutic response. The information provided by Clinic should not be used in isolation when making patient management decisions.
Dynamic Perfusion Analysis - Generates parametric perfusion maps used for visualization of temporal variations in dynamic datasets, showing changes in image intensity over time. These maps may aid in the assessment of the extent and type of perfusion, blood volume and vascular permeability changes.
Dynamic Diffusion Analysis - Generates apparent diffusion coefficient maps used for the visualization of apparent water movement in soft tissue throughout the body on both voxel-by-voxel and sub-voxel bases. These images may aid in the assessment of the extent of diffusion in tissue.
Image Comparison - Generates co-registered image sets. Generates standardized image sets calibrated to an arbitrary scale to facilitate comparisons between independent image sets. Generates voxel-by-voxel maps of the image intensity differences between image sets acquired at different times. Facilitates selection and DICOM export of user-selected regions of interest (ROIs) These processes may enable easier identification of image intensity differences between images and easier selection and processing of ROIs.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
# Sean M. Boyd -S
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) K123302
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.