IB Neuro™ software allows the post-processing and display of dynamically acquired MR datasets to evaluate image intensity variations over time. IB Neuro™ v1.0 plug-in accepts data from existing MRI systems, performs quality control checks and generates parametric perfusion maps such as Relative Cerebral Blood Volume (rCBV), Cerebral Blood Flow (CBF), Mean Transit Time (MTT) and Time to Peak (TTP) and sends the maps to a PACS for subsequent viewing. These images when interpreted by a trained physician may yield information useful in clinical applications. Our advanced technology is designed to be compliant with healthcare standards such as DICOM and is easily and rapidly integrated into existing medical image visualization applications.
Device Story
IB Neuro v1.0 is a software plug-in for existing medical image visualization applications (e.g., OsiriX, PACS). It accepts dynamically acquired MR datasets as input. The software performs quality control, generates parametric perfusion maps (rCBV, CBF, MTT, TTP), performs automated contrast agent leakage correction for rCBV, generates automated brain masks, and allows parameter normalization to normal appearing white matter (NAWM). Output consists of DICOM-compliant perfusion maps and reports sent to PACS. Used in clinical settings by trained physicians to assist in image interpretation and clinical decision-making. Benefits include rapid integration into existing workflows and automated generation of perfusion metrics.
Clinical Evidence
No clinical data. Bench testing only, including software validation and verification per FDA software validation guidance.
Technological Characteristics
Software-based image processing plug-in; platform-independent base library; DICOM compliant; integrates with existing PACS/workstations; automated algorithms for perfusion mapping, brain mask generation, and contrast leakage correction.
Indications for Use
Indicated for use by trained physicians to process and display dynamically acquired MR datasets for the evaluation of image intensity variations over time, generating parametric perfusion maps (rCBV, CBF, MTT, TTP) for clinical interpretation.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
BrainStat Software Option for GE Signa MR Scanners (K073167)
Dynamic MRI Software Option for C-Scan, E-ScanXQ and E-Scan Opera (K061429)
Perfusion Package for Magnetom Vision and Symphony MR Systems (K984224)
Submission Summary (Full Text)
{0}------------------------------------------------
K080762
EXHIBIT #1
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:
### 1. Submitter's Identification:
MAY 1 5 2008
Imaging Biometrics, LLC 1035 Katherine Drive Elm Grove, WI 53122
Contact Person Michael Schmainda Phone: (262) 385-1796 Email: mike@imagingbiometrics.com
Date Summary Prepared: 3/4/2008
## 2. Name of the Device:
| Trade Name: | IB Neuro™ v1.0 |
|----------------------------------------------------------|--------------------------------------------------------------------|
| Classification Name:<br>Classification:<br>Product Code: | MR Diagnostic Device Accessory: 21 CFR 892.1000<br>Class II<br>LNH |
| Classification Name: | Picture Archiving and Communication Systems<br>21 CFR 892.2050 |
| Classification: | Class II |
| Product Code: | LLZ |
## 3. Common or Usual Name:
IB Neuro™ 1.0 / PACS device
#### 4. Predicate Device Information:
IB Neuro™ is substantially equivalent to the following legally marketed devices that are currently cleared by the FDA:
{1}------------------------------------------------
| 510(k) | Trade Name | Manufacturer |
|---------|-------------------------------------------------------------------------|----------------------------------------------------------------------------|
| K073167 | BrainStat Software Option for GE<br>Signa MR Scanners | GE Medical Systems<br>3200 N. Grandview Blvd. W-827<br>Waukesha, WI 53188 |
| K061429 | Dynamic MRI Software Option for<br>C-Scan, E-ScanXQ and E-Scan<br>Opera | ESAOTE S.p.A.<br>Via Siffredi, 58<br>16153 Genova - Italy |
| K984224 | Perfusion Package for Magnetom<br>Vision and Symphony MR Systems | Siemens Medical Systems, Inc.<br>186 Wood Avenue South<br>Iselin, NJ 08830 |
## 5. Device Description:
IB Neuro "" OsiriX Plugin is software designed to analvze dynamically acquired datasets. Using well-established algorithms, parametric perfusion maps can be generated such as Relative Cerebral Blood Volume (rCBV), Cerebral Blood Flow (CBF), Mean Transit Time (MTT) and Time to Peak (TTP). The strength of our software is its ability to extend the productivity of any existing viewer, CAD workstation or PACS via a platform-independent base library that allows for quick and seamless integration into existing server and workstation applications. It also includes other critical features such as:
- Enables rapid creation of a complete array of critical perfusion parameter o maps of rCBV, CBF, MTT, TTP
- o Automated correction of contrast agent leakage for rCBV maps
- o Automated brain mask generation
- Ability to normalize parameters to normal appearing white matter (NAWM) o
- o Automated report generation
- View dynamic signal time course on a per-voxel basis o
- Interactive Arterial Input Function (AIF) selection o
- o Automatic export of perfusion parameter maps to DICOM images within the same study
## 6. Intended Use:
IB Neuro™ software allows the processing and display of dynamically acquired MR datasets to evaluate image intensity variations over time. IB Neuro™ v1.0 plug-in accepts data from existing MRI systems, performs quality control checks
{2}------------------------------------------------
and generates parametric perfusion maps such as Relative Cerebral Blood Volume (rCBV), Cerebral Blood Flow (CBF), Mean Transit Time (MTT) and Time to Peak (TTP) and sends the maps to a PACS for subsequent viewing. These images when interpreted by a trained physician may vield information useful in clinical applications. Our advanced technology is designed to easily and rapidly integrate into existing medical image visualization applications.
## 7. Substantial Equivalence/ Comparison to Predicate Devices:
The intended use and performance characteristics for IB Neuro™ are substantially equivalent to the predicate devices listed in section 4 above for image analysis and processing and generation of parametric maps to provide additional information bevond standard imaging.
## 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Performance testing included software validation, verification and testing per FDA's software validation guidance.
## 9. Discussion of Clinical Tests Performed:
N/A
## 10. Conclusions:
The IB Neuro™ 1.0 device has a similar intended use and similar characteristics as the predicate devices. Moreover, documentation supplied in this submission demonstrates that any difference in technological characteristics do not raise any new questions of safety or effectiveness. Thus, the IB Neuro™ device is substantially equivalent to predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 1 5 2008
Mr. Michael Schmainda CEO Imaging Biometrics, LLC 1035 Katherine Dr. ELM GROVE WI 53122
Re: K080762
Trade/Device Name: IB Neuro™ v1.0 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 17, 2008 Received: March 18, 2008
Dear Mr. Schmainda:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and i adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA.'s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements act threact or any Federal statutes and regulations administered by other Federal agencies. You nust comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to I ogally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Crant of Crane of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device 2-so-zing (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Exhibit B
# Indications for Use
510(k) Number (if known):
IB Neuro™ v1.0
Indications For Use:
Device Name:
IB Neuro™ software allows the post-processing and display of dynamically acquired MR datasets to evaluate image intensity variations over time. IB Neuro™ v1.0 plug-in accepts data from existing MRI systems, performs quality control checks and generates parametric perfusion maps such as Relative Cerebral Blood Volume (rCBV), Cerebral Blood Flow (CBF), Mean Transit Time (MTT) and Time to Peak (TTP) and sends the maps to a PACS for subsequent viewing. These images when interpreted by a trained physician may yield information useful in clinical applications. Our advanced technology is designed to be compliant with healthcare standards such as DICOM and is easily and rapidly integrated into existing medical image visualization applications.
Prescription Use: X (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
**Concurrence of CDRH, Office of Device Evaluation (ODE)**
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K080762
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.