The U-Scope device is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include: abnormal bleeding; infertility and pregnancy wastage; evaluation of abnormal hysterosalpingogram; intrauterine foreign body; amenorrhea; and pelvic pain.
Device Story
Handheld, battery-operated portable hysteroscope; consists of reusable handle and sterile, single-use disposable hysteroscopic cannula. Cannula tip contains miniature CMOS camera and LED illumination module; includes channel for irrigating fluid infusion. Handle contains video processor, LCD display, rechargeable battery, and firmware. Operated by trained medical professionals in hospitals or physician offices. Device captures real-time video of cervical canal and uterine cavity; displays images on integrated LCD for direct visualization during diagnostic procedures. Enables physician to perform diagnostic hysteroscopy; provides immediate visual feedback for clinical decision-making. Benefits include portability and disposable cannula to reduce reprocessing requirements.
Clinical Evidence
No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-1), electrical/EMC safety (ISO 60601-1, ISO 60601-1-2, IEC 60601-2-18), thermal safety, mechanical characteristics, optics performance, image quality, and software verification/validation.
Technological Characteristics
Handheld, battery-operated hysteroscope. Components: reusable handle, sterile single-use disposable cannula. CMOS sensor (approx. 160,000 pixels), LED illumination at distal tip. Oval shaft (3.8mm x 4.6mm). Materials compliant with ISO 10993. Integrated LCD display. Firmware-based video processing. Sterilization: Ethylene oxide (cannula).
Indications for Use
Indicated for diagnostic hysteroscopy in patients presenting with abnormal uterine bleeding, infertility, pregnancy wastage, abnormal hysterosalpingogram, intrauterine foreign body, amenorrhea, or pelvic pain.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
KR23151
Page 1 of 5
## MAR 0 8 2013
#### 510(k) Summary Prepared February 15, 2013
| Sponsor: | EndoSee Corp. |
|----------------------------|-----------------------------------------------------------------------------------|
| Contact Person: | Xiaolong (Bruce) Ou Yang<br>4546 El Camino Real, Suite 215<br>Los Altos, CA 94022 |
| Telephone: | 650 397 5174 |
| Fax: | 888 388 1420 |
| Device Name: | U-Scope Model 8000 |
| Common Name: | Hysteroscope |
| Regulatory Class: | II |
| Classification Regulation: | 21 CFR 884.1690 |
| Classification Panel: | Obstetrics and Gynecology |
## A. Legally Marketed Predicate Devices
The EndoSee device is substantially equivalent to Microspan Hysteroscope manufactured by Imagyn Medical (K961688) and Gyrus ACMI Digital Hysteroscope manufactured by Gyrus ACMI (K092278).
#### B. Device Description:
The EndoSee U-Scope Model 8000 is a handheld, battery-operated portable hysteroscope. It includes a sterile single use disposable hysteroscopic cannula and a reusable handle. The disposable cannula contains a miniature CMOS camera and a light-emitting diode (LED) illumination module at the tip and a channel for infusion of irrigating fluid. The disposable cannula is sterilized and packaged in a sealed bag. The handle is light weight and ergonomically designed. It has a connector and locking mechanism for attaching and detaching the cannula. The handle contains the remaining electronics including a power on/off button, a video processor, a display unit (LCD display), a rechargeable battery, management electronics, microcontrollers and firmware.
{1}------------------------------------------------
K123151
Page 2 of 5
## C. Intended Use
The U-Scope device is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include:
Abnormal uterine bleeding, infertility and pregnancy wastage; evaluation of abnormal hysterosalpingogram; intrauterine foreign body; amenorrhea; pelvic pain
## D. Substantial Equivalence
The U-Scope 8000 is substantially equivalent to the Microspan Hysteroscope manufactured by Imagyn (K961688 ) and Gyrus ACMI Digital Hysteroscope manufactured by Gyrus ACMI (K092278) with regard to both intended use and technological characteristics as described in the tables below:
| Feature | Subject Device<br>U-Scope 8000 | Predicate 1<br>Imagyn Medical<br>Microspan Hysteroscope<br>(K961688) | Predicate 2<br>Gyrus ACMI Invisio Digital<br>Hysteroscope System<br>(K092278) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Viewing of cervical canal<br>and uterine cavity | Same | Same |
| Indication for<br>use | The U-Scope device is<br>used to permit viewing of<br>the cervical canal and<br>uterine cavity for the<br>purpose of performing<br>diagnostic procedures.<br>Generally recognized<br>indications for diagnostic<br>hysteroscopy include:<br>Abnormal uterine<br>bleeding, infertility and<br>pregnancy wastage;<br>evaluation of abnormal<br>hysterosalpingogram;<br>intrauterine foreign body;<br>amenorrhea; pelvic pain | The Microspan<br>hysteroscope device is<br>used to permit viewing of<br>the cervical canal and<br>uterine cavity for the<br>purpose of performing<br>diagnostic and surgical<br>procedures | *The Gyrus ACMI Invisio<br>Digital Hysteroscope is<br>intended to be used to permit<br>viewing of the<br>cervical canal and uterine<br>cavity for the purpose of<br>performing diagnostic and<br>surgical procedures.<br>The Gyrus ACMI Invisio<br>Digital Hysteroscope System:<br>is intended to be used to<br>process the video signal from<br>the Invisio Digital<br>Hysteroscope and ensure<br>brightness, image clarity and<br>color.<br>Diagnostic Hysteroscopy:<br>Abnormal uterine bleeding<br>Infertility & pregnancy<br>wastage<br>'Evaluation of abnormal<br>hysterosalpingogrm |
Table 1 Comparison Table – Intended Use and Indications for Use
{2}------------------------------------------------
K123151
Page 3 of 5
| | | | Intrauterine foreign body |
|-----------------------------------------------|------------------------------------|------------------------------------------|--------------------------------------------|
| | | | Amenorrhea |
| | | | Pelvic pain |
| | | | Operative Hysteroscopy: |
| | | | • Directed biopsy |
| | | | • Removal of fibroids<br>and polyps |
| | | | • Transection of<br>intrauterine adhesions |
| | | | • Transection<br>intrauterine septa |
| Intended<br>Users | Trained Medical<br>Professionals | Trained Medical<br>Professionals | Trained Medical<br>Professionals |
| Clinical<br>application | Diagnostic hysteroscopy | Diagnostic and operative<br>hysteroscopy | Diagnostic and operative<br>hysteroscopy |
| Site of Use | Hospitals and Physician<br>Offices | Hospitals and Physician<br>Offices | Hospitals and Physician<br>Offices |
| Class | II | II | II |
| Classification<br>Regulation/<br>Product Code | 21 CFR 884.1690<br>HIH | 21 CFR 884.1690<br>HIH | 21 CFR 884.1690<br>HIH |
# Table 2 Comparison Table – Technological Characteristics
| Feature | Subject Device<br>U-Scope 8000 | Predicate 1<br>Imagyn Microspan<br>Hysteroscope<br>(K961688) | Predicate 2<br>Gyrus ACMI Invisio<br>Digital Hysteroscope<br>System (K092278) |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Components | Design incorporates two<br>separate components<br>(cannula and handle) and<br>incorporates a LCD<br>monitor as an integral part<br>of the handle. | Design incorporates<br>stainless steel shaft,<br>imaging fiber bundle, lens,<br>eyepiece, light source and<br>light cables | Design incorporates<br>handle, semi rigid<br>hysteroscope, digital<br>CMOS sensor, PCBs, LED<br>light source, electrical<br>cords, power supply and<br>cables |
| Battery<br>Operated | Yes | No | No |
| Optical<br>Image | Digital CMOS technology | Not specified in product<br>labeling | Digital CMOS technology |
| Image | The CMOS sensor consists | Not specified in product<br>labeling | The CMOS sensor consists |
{3}------------------------------------------------
K123151
Page 4 of 5
| Resolution | of approximately 160,000<br>pixels. | labeling | of approximately 100,000<br>pixels |
|----------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Illumination<br>Light source | LEDs at the distal tip of the<br>cannula | Fiber optic bundle /<br>external OES Xenon light<br>source | Fiber optic bundle<br>/integrated LEDs in the<br>handle |
| Image<br>transmission | Transmit images from a<br>video camera to a video<br>monitor | Eyepiece can be connected<br>through an optical coupler<br>to a camera | Transmit images from a<br>video camera to a video<br>monitor |
| Inflow and<br>outflow<br>channel for<br>saline<br>instillation | Single channel for inflow | None | In-flow and option for<br>continuous flow with<br>outside sheath |
| Sterile | Cannula is provided as<br>sterile single use;<br>Handle is provided non<br>sterile | Provided non-sterile and<br>requires sterilization by the<br>end user | Provided non-sterile and<br>requires sterilization by the<br>end user |
| Disposable/<br>Reusable | Cannula is provided sterile<br>for single use and the<br>handle is reusable. | Reusable device, provided<br>non-sterile which requires<br>sterilization by the end user | Reusable device, provided<br>non-sterile which requires<br>sterilization by the end user |
| Shaft<br>diameter and<br>working<br>length | Oval shape (3.8 mm x 4.6<br>mm )<br>Working length 287 mm<br>(11.3 in) | Not specified in product<br>labeling | 5.5mm with continuous<br>flow sheath<br>Working length not<br>available |
| Contact<br>Materials | Compliant with ISO 10993 | Not specified in product<br>labeling | Compliant with ISO 10993 |
## E. Performance Data
Bench testing was conducted on the device to verify that it is compliant with biocompatibility requirements for a short duration indwelling device as specified in ISO 10993 - Part 1. It was also tested by a certified test laboratory and met the compliance requirements for electrical safety as specified in ISO 60601-1, including provisions for EMC safety in ISO 60601-1-2 and IEC 60601-2-18 Medical electrical equipment – Part 2-18: "Particular requirements for the basic safety and essential performance of endoscopic equipment", including thermal safety at several points along the cannula. Mechanical characteristics were also tested with successful results. The software test data, including verification and validation results, were submitted as specified in FDA guidance documents related to requirements for software contained in medical devices. Other test data, as specified in FDA guidance documents related to 510(k) hysteroscopic devices included data on optics performance and results of image quality testing.
{4}------------------------------------------------
K123151
Page 5 of 5
Sterilization validation and shelf life testing were conducted to confirm the label shelf life and are in compliance with the following:
- ISO 11135-1 Sterilization of health care products -- Ethylene oxide -- Part 1: . Requirements for development, validation and routine control of a sterilization process for medical device
- ISO 11607 Packaging for Terminally Sterilized Medical Devices .
- AAMI TIR12:2010 -- Designing, testing, and labeling reusable medical devices . for reprocessing in health care facilities: A guide for medical device manufacturers
Handle reprocessing issues were addressed by test methodology described in AAMI TIR30:2011 - A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices.
In addition, the following battery and mechanical characteristics bench tests were performed as follows:
- testing of battery specifications .
- testing of durability of handle .
- testing of effects of repeated bending, pulling, and torque .
- testing of fluid delivery and controls .
All of these performance verification and validation test results were provided in the submission according to requirements specified in FDA guidance documents.
## F. Conclusion
In conclusion, the bench testing performed addressed materials, mechanical and electrical tests related to safety and effectiveness, and successful results demonstrated that the subject device is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with its wings spread, rendered in black. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle, also in black. The logo is simple and conveys a sense of national authority and public service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2013
EndoSee Corporation % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW BUFFALO MN 55313
Re: K123151
Trade/Device Name: U-Scope Model 8000 Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: February 26, 2013 Received: February 27, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{6}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin Fisher-S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K123151
Device Name: U-Scope Model 8000
Indications For Use:
The U-Scope device is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include: abnormal bleeding; infertility and pregnancy wastage; evaluation of abnormal hysterosalpingogram; intrauterine foreign body; amenorrhea; and pelvic pain.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R际库ish 2013.03.08 13.37.06 1
Page 1 of _ 1 _ _
(Division Sign-Off) ivision of Reproductive, Gastro-Renal, and
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.