GYRUS ACMI INVISIO DIGITAL HYSTEROSCOPE, MODEL IDH-4
Applicant
Gyrus Acmi, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Sep 29, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Gyrus ACMI Invisio Digital Hysteroscope is intended to be used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures. The Gyrus ACMI Invisio Digital Hysteroscope System includes the Invisio Digital Hysteroscope and IDC Invision™ Controller. The IDC Controller is intended to be used to process the video signal from the Invisio Digital Hysteroscope and ensure brightness, image clarity and color.
Device Story
Semi-rigid (rigi-flex) hysteroscope system for visualization of cervical canal and uterine cavity. Input: optical images captured by distal-tip CMOS sensor. Processing: IDC-1500 Controller processes video signals (brightness, clarity, color) via internal PCBs and software. Output: displayed video feed for clinician. Used in clinical settings by physicians for diagnostic and operative procedures (biopsy, polypectomy, myomectomy, adhesiolysis, septal transection). Benefits: provides direct visualization to facilitate minimally invasive intrauterine surgery.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and bench testing comparisons to predicate devices.
Technological Characteristics
Semi-rigid hysteroscope with distal CMOS sensor, LED light source in handle, and bifurcated fiber optic illumination. Connects to IDC-1500 Controller via electrical cord. Features articulating distal section with deflection pulleys/wires. Patient-contacting materials consistent with predicate DUR-D. Software-driven image processing.
Indications for Use
Indicated for diagnostic and operative hysteroscopy in patients requiring visualization of the cervical canal and uterine cavity. Diagnostic indications: abnormal uterine bleeding, infertility, pregnancy wastage, abnormal hysterosalpingogram, intrauterine foreign body, amenorrhea, and pelvic pain. Operative indications: directed biopsy, removal of fibroids and polyps, and transection of intrauterine adhesions or septa.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
K092278
Gyrus ACMI® Invisio® Digital Hysteroscope System Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
#### Traditional 510(k) Notification Summary of S & E July 24, 2009
# 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc. Gyrus ACMI® Invisio® Digital Hysteroscope System
General Information
Manufacturer:
136 Turnpike Rd. Southborough, MA 01772-2104
Establishment Registration Number:
Contact Person:
3003790304
July 24, 2009
Gyrus ACMI Inc.
Graham A. L. Baillie, MS Sr Regulatory Affairs Specialist Tel. # 508 804-2738 Fax # 508 804-2624
Date Prepared:
Trade Name:
Device Description Classification Name:
Hysteroscope and accessories (21CFR 886.1690). Class II Surgical camera and accessories, (21 CFR 878.4160), Class I
Gyrus ACMI Invisio Digital Hysteroscope System
Endoscope, Video Camera and accessories Hysteroscope and accessories
## Predicate Devices
Generic/Common Name:
K060269 ACMI DUR®-Digital Ureteroscope & Choledochoscope, (DUR-D) Gyrus ACMI IPN-2505 Percutaneous Nephroscope K072594 K990411 Karl Storz MVM 7.5French Flexible Hysteroscope K022445 Olympus Visera HFY-V Hysterovideoscope
# Intended Use
The Gyrus ACMI Invisio Digital Hysteroscope is intended to be used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
The Gyrus ACMI Invisio Digital Hysteroscope System includes the Invisio Digital Hysteroscope and IDC Invision™ Controller. The IDC Controller is intended to be used to process the video signal from the Invisio Digital Hysteroscope and ensure brightness, image clarity and color.
# Diagnostic Hysteroscopy:
- Abnormal uterine bleeding o
- Infertility & pregnancy wastage o
{1}------------------------------------------------
Gyrus ACMI® Invisio® Digital Hysteroscope Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
Traditional 510(k) Notification Truthfuiness Statement July 24, 2009
- Evaluation of abnormal hysterosalpingogram �
- Intrauterine foreign body o
- Amenorrhea �
- Pelvic pain �
## Operative Hysteroscopy:
- Directed biopsy a
- Removal of fibroids and polyps �
- Transection of intrauterine adhesions .
- Transection intrauterine septa �
## Product Description
The Gyrus ACMI® Invisio® Digital Hysteroscope System is comprised of the Invisio Digital Hysteroscope and Invisio Digital Controller (IDC-1500). The Gyrus ACMI Invisio Digital Hysteroscope is a semi-rigid (rigi-flex) hysteroscope that incorporates CMOS (complimentary medical oxide semi-conductor) sensor technology to generate an image. There is a miniature CMOS sensor located in the distal tip, wiring running through the endoscope shaft, a printed circuit board (PCB), a light-emitting diode (LED) light source located in the handle, and an electrical cord that connects the endoscope to the Camera Control Unit (CCU).. The Invisio Digital Hysteroscope connects to a CCU, (the Invisio Digital Controller (IDC-1500)), that contains printed circuits boards (PCBs), software, power supply and power cables to process and display the images transmitted by the camera.
# Technological Characteristics and Substantial Equivalence
The Gyrus ACMI® Invisio® Digital Hysteroscope System, utilizes features included in the predicate Gyrus ACMI Invisio Digital, IPN-2505 Percutaneous Nephroscope K072594, and the ACMI DUR®-Digital Ureteroscope & Choledochoscope, DUR®-D K060269. The proposed device incorporates the same CMOS video imaging technology located in the endoscope as the aforementioned Invisio predicates. The subject Invisio Digital Hysteroscope System utilizes the same model Invisio Controller loaded with the same Invisio digital software. The internal shaft construction and flexible/articulating distal section contains similarly designed defection pulleys/wires, and electrical wiring as the predicate Invisio DUR®-D. The subject and two aforementioned Invisio predicates also incorporate similarly designed bifurcated fiber optic illumination bundles that carry/transmit light from the same design LEDs, located in the handles of each device, through the shaft to the distal tip. The subject hysteroscope and predicate DUR®-D also share the same patient contacting materials used in the distal tip, working channel and deflection cover.
Like other predicate hysteroscopes: the Karl Storz MVM 7.5 Fr Flexible Hysteroscope (K99041 1); and the Olympus Visera HFY-V Hysterovideoscope (K022445); the proposed hysteroscope has similar shaft diameter and working length and is intended for viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
In summary, the Gyrus ACMI® Invisio® Digital Hysteroscope System is substantially equivalent to the four predicate devices and presents no new questions of safety or efficacy.
10
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract, curved shapes that resemble human figures or flowing lines. The symbol is positioned in the center of the seal.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
SEP 2 9 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Graham A. L. Baillie, MS Sr. Regulatory Affairs Specialist Gyrus ACMI, Inc. 136 Turnpike Road SOUTHBOROUGH MA 01772-2104
Re: K092278
Trade/Device Name: Invisio Digital Hysteroscope System Regulation Number: 21 CFR 8884,1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: July 24, 2009 Received: July 28, 2009
Dear Mr. Baillie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Gyrus ACMI® Invisio® Digital Hysteroscope System Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
Traditional 510(k) Notification Intended Use Statement July 24, 2009
# Indications for Use
K092278 510(k) Number: Device Name: Invisio Digital Hysteroscope System
# Indications for Use
The Gyrus ACMI Invisio Digital Hysteroscope is intended to be used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
The Gyrus ACMI Invisio Digital Hysteroscope System includes the Invisio Digital Hysteroscope and IDC InvisionM Controller. The IDC Controller is intended to be used to process the video signal from the Invisio Digital Hysteroscope and ensure brightness, image clarity and color.
## Diagnostic Hysteroscopy:
- 0 Abnormal uterine bleeding
- Infertility & pregnancy wastage .
- Evaluation of abnormal hysterosalpingogram �
- Intrauterine foreign body 0
- � Amenorrhea
- � Pelvic pain
ð
Operative Hysteroscopy:
- � Directed biopsy
- Removal of fibroids and polyps ね
- Transection of intrauterine adhesions 0
- Transection intrauterine septa
Prescription Use: *X*
Over-the-Counter Use :________________________________________________________________________________________________________________________________________________________
7
(Per 21 CFR 801.109)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE),
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.