K123082 · Nephosity, Inc. · LLZ · May 16, 2013 · Radiology
Device Facts
Record ID
K123082
Device Name
MOBILECT VIEWER
Applicant
Nephosity, Inc.
Product Code
LLZ · Radiology
Decision Date
May 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The MobileCT Viewer software program provides for communication and display of CT, MRI, X-ray medical images on the Apple iPad (4th generation, late 2012) . It is intended for use as a diagnostic, review and analysis tool by trained professionals. MobileCT Viewer provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation. This device is not to be used for mammography.
Device Story
MobileCT Viewer is a software-based PACS component; retrieves DICOM-compliant CT, MRI, and X-ray images from a server via network connection. Images are losslessly compressed for transfer and displayed on an Apple iPad (4th generation). Operated by trained medical professionals in clinical environments; provides portable, wireless access to images when primary workstations are inaccessible. Features include window/level adjustment, zoom, pan, rotation, and distance measurement. Does not alter original image data. Assists clinicians in diagnostic interpretation and review; intended to support clinical decision-making by providing remote access to imaging data.
Clinical Evidence
No clinical or animal testing performed. Evidence consists of bench testing and performance evaluations. Qualified medical professionals tested image quality across supported modalities (CT, MRI, X-ray) under varying environmental conditions. Software passed all functional and performance requirements.
Technological Characteristics
Software-based PACS; operates on Apple iPad (4th generation). DICOM-compliant; supports CT, MRI, and X-ray. Features include window/level, zoom, pan, rotation, and distance measurement. Network-dependent; utilizes lossless compression for image transfer. User authentication included. No advanced image processing (e.g., 3D reconstruction, fusion).
Indications for Use
Indicated for trained professionals to view, review, and analyze CT, MRI, and X-ray medical images on an Apple iPad. Not for use in mammography. Not intended to replace full diagnostic workstations; for use only when workstation access is unavailable.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K123082
Page 1 of 5
Part V. 510(k) Summary of Safety and Effectiveness
(the following information is in conformance with 21 CFR 807.92)
MAY 1 6 2013
19
# Submitter
Nephosity, Inc. 615 Grant Avenue 3F San Francisco, CA 94108
Contact Person: Michael Pan, CEO 615 Grant Avenue 3F San Francisco, CA 94108 mjpan@nephosity.com +1 650 429 8917
Date Summary Prepared: April 24, 2013
# Device name
Trade Name: MobileCT Viewer
Common Name: Medical Imaging Software T
Classification Name: System, Image Processing, Radiological (21 CFR. Part 892.2050, Product Code: LLZ)
# Predicate device
K 103785 MobileMIM MIM Software Inc.
{1}------------------------------------------------
## Indications for Use
The MobileCT Viewer software program provides for communication and display of CT, MRI, X-ray medical images on the Apple iPad (4th generation, late 2012) . It is intended for use as a diagnostic, review and analysis tool by trained professionals.
MobileCT Viewer provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.
This device is not to be used for mammography.
## Device description
The MobileCT Viewer is a software-based Picture Archiving and Communication System (PACS) used with computing servers and specific mobile devices. DICOM-compliant medical images from CT : MRI, X-ray modalities are stored on the server component. MobileCT Viewer retrieves patent image data securely via a network connection with the server. DICOM files are losslessly compressed for network transfer and downloaded by MobileCT Viewer for display on the mobile device component. Communication and display on the mobile device assist trained professionals in the diagnostic interpretation, review and analysis of the medical images.
MobileCT Viewer includes the capability to perform to the displayed image:
- · adjust window width and level (i.e. contrast) values,
- · apply view transforms (e.g. zoom, pan, and rotation),
- · measure distances, and
- · display measurement lines and annotations.
MobileCT Viewer operates on off-the-shelf portable hardware devices and is therefore subject to factors not typical for reading room workstations (c.g. screen size, environmental variability,
{2}------------------------------------------------
network dependencies, etc.). It is therefore required that the user follows the operating instructions properly and utilizes the risk mitigation features in order to make decisions safely and effectively.
MobileCT Viewer provides wireless and portable access to medical innages. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.
## Substantial equivalence
Table 1 provides evidence to facilitate the substantial equivalence determination between MobileCT Viewer and our chosen predicate. Mobile MIM (K103785).
There is a direct correlation between the Indication Statement / Intended Use of Mobile T Viewer with Mobile MIM. Both devices are software applications used by medical professionals in the diagnosis of patients by means of medical images.
MobileCT Viewer and MobileMM run on the same software platform- Apple iOS, and hardware platform- the Apple iPad (4th generation, late 2012) . MobileCT Viewer's technological characteristics are more limited than that of MobileMM, as MobileCT Viewer provides only viewing and simple image manipulation (which do not alter the image data. such as window and level, pan and zoom, and image annotation) capabilities. MobileCT Viewer does not provide image processing functions which are intended to alter the image data (e.g. filtering, multiplanar reconstruction, and 3D reconstruction).
MobileCT Viewer's capabilities include support for viewing medical images from modalities not indicated in Mobile MIMs submission: X-ray. This difference does not alter the intended offect of the device (that is, the display of medical images) and does not raise any different types of safety and effectiveness questions. Information including performance data is provided in this submission to assess device performance in viewing images from these added modalities.
In addition, MobileCT Viewer does not support the following advanced image manipulations: image fusion, multiplanar reconstruction (MPR), maximum intensity projection (MIP), or standard update values (SUV). Consistently, the Indications for Use Statement for MobileCT Viewer excludes the advanced image manipulations.
{3}------------------------------------------------
.
| Item | MobileCT | MobileMIM |
|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use / Indication for Use | The MobileCT Viewer software program provides for communication and display of CT, MRI, X-ray medical images on the Apple iPad (4th generation, late 2012) . It is intended for use as a diagnostic, review and analysis tool by trained professionals.<br>MobileCT Viewer provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.<br>This device is not to be used for mammography. | The Mobile MIM software program is used for the registration, fusion, and/or display for diagnosis of medical images from the following modalities: SPECT, PET, CT and MRI.<br>Mobile MIM provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.<br>This device is not to be used for mammography. |
| Receive, Store, Retrieve, Display and Process Digital Medical Images | Yes | Yes |
| Display of Clinical Patient Data When No Access to a Workstation | Yes | Yes |
| Image Fusion | No | Yes |
| 3D reconstruction, e.g. Multi-Planar Reconstruction (MPR), Maximum Intensity Projection (MIP) | No | Yes |
| Standard Uptake Value (SUV) | No | Yes |
| Distance Measurements | Yes | Yes |
| Window/Level | Yes | Yes |
| Zoom/Pan | Yes | Yes |
| User Authentication | Yes | Yes |
| Modalities | CT, MRI. X-ray | SPECT, PET, CT, MRI |
| Remote Handheld Viewing Device | Yes | Yes |
| Operating Platform | Yes | Yes |
| Hardware Requirements | the Apple iPad (4th generation, late 2012) | Apple iOS handheld devices |
. . . . . .
TABLE 1: Device Comparison table between new device and predicate
{4}------------------------------------------------
23
#### Part V. 510(k) Summary of Safety and Effectiveness
For a complete discussion of how hazards related to the use of MobileCT Viewer as a diagnostic, review, and analysis tool by trained professionals should be addressed during device development as part of the risk management process, see the Device Use Safety discussion in Part XVI (Software) Chapter 3 (Hazard analysis) of this 510(k) premarket notification. Additionally, a summary of the results of the testion 9.6 and section 9.7) done during the Alpha and Beta development stages demonstrate that the device, when used according to operating instructions, can be used safely and effectively.
## Summary of testing
Nephosity, Inc. has performed multiple studies with qualified medical professionals. These medical professional tested MobileCT Viewer by evaluating the image quality of the medical inages of the supported modalities (i.e. CT, MRJ, X-ray ) under different environmental conditions. Results of these studies affirm the diagnostic viewing capabilities of MobileCT Viewer when used as indicated.
Additionally, Nephosity, Inc. has conducted performance and functional testing on the MobileCT Viewer software. In all cases, the software passed its performance requirements and met specifications. A summary of the results of the testion 9.6 and section 9.7) of Part XVI.
No animal or clinical testing was performed.
## Conclusion
Based on a comparison between the MobileCT Viewer and the Mobile MIM and on the performance data provided in this premarket notification submission, it is our belief that the new device is as safe and effective as the predicate device, and does not raise different questions of safety and effectiveness than the predicate device.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Silver Spring, MD 20993-0002
May 16, 2013
Nephosity, Inc. % Mr. Michael Pan CEO 615 Grant Avenue, 3F SAN FRANCISCO CA 94108
Re: K123082
Trade/Device Name: MobileCT Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II . Product Code: .LLZ Dated: March 30, 2013 Received: April 3. 2013
Dear Mr. Pan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh.p)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K123082
Device Name: MobileCT Viewer
Indications for Use:
The MobileCT Viewer software program provides for communication and display of CT, MRI, X-ray medical images on the Apple iPad (4th generation, late 2012) . It is intended for use as a diagnostic, review and analysis tool by trained professionals.
MobileCT Viewer provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.
This device is not to be used for mammography.
Prescription Use × (Part 21 CFR 801 Subpart D)
#### AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Sm.h.p).
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
K123082 510(k)
Page I of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.