K123006 · Nimbic Systems, Inc. · ORC · Dec 20, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123006
Device Name
AIR BARRIER SYSTEM (ABS)
Applicant
Nimbic Systems, Inc.
Product Code
ORC · General, Plastic Surgery
Decision Date
Dec 20, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5070
Device Class
Class 2
Indications for Use
The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty and posterior vertebral fusion and laminoplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the nozzle and the surface. (2) the incision plane is parallel with the direction of air flow, and (3) the incision dimensions are within: 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.
Device Story
Portable air-handling apparatus; produces directed, non-turbulent HEPA-filtered airflow to surgical sites. Components: reusable Filter/Blower unit; sterile, single-use Air Supply Hose and Nozzle. Nozzle placed on anatomical surface adjacent to incision; airflow displaces airborne colony-forming units and particulates within 6" x 20" area. Used in OR during hip arthroplasty, posterior vertebral fusion, and laminoplasty. Benefits patient by reducing potential surgical site contamination. Operated by surgical staff; effectiveness requires specific nozzle placement and incision orientation.
Clinical Evidence
Two clinical studies provided. Study 1 (n=29, hip arthroplasty): randomized experiment vs. sham vs. control; ABS group showed significantly lower (p<0.001) airborne particulates and microorganisms. Study 2 (n=38, posterior vertebral fusion/laminoplasty): randomized experiment vs. control; ABS group showed significantly lower (p≤0.0001) CFU (1.55 vs 5.05 CFU/m³) and particulates ≥5μm (1325 vs 4837 particles/ft³).
Technological Characteristics
Stainless steel housing; 150 CFM blower output; HEPA filtration (99.97% efficiency at 0.3 micron) using continuous pleated glass microfiber; polyurethane foam pre-filter. Sterile, single-use ETO-sterilized hose/nozzle. 120VAC/60Hz power. No software/algorithm.
Indications for Use
Indicated for use in surgical operating rooms to reduce airborne particulate matter and microorganisms at the surgical site during hip arthroplasty, posterior vertebral fusion, and laminoplasty. Intended for clinical use by healthcare professionals.
Regulatory Classification
Identification
Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
# 510(k) Summary
Date Prepared: December 9, 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92 for the Air Barrier System (ABS).
# 1. Company making the submission:
| Owner/Submitter: | Nimbic Systems, Inc.<br>4910 Wright Road, Suite 170<br>Stafford, Texas 77477 |
|------------------|--------------------------------------------------------------------------------------|
| Contact Person: | Sean Self<br>President<br>281-565-5715<br>281-565-5712 fax<br>self@nimbicsystems.com |
# 2. Device Name and Classification:
| Common/Usual Name | : | Air-handing apparatus for a surgical<br>operation room |
|-------------------|---|--------------------------------------------------------|
| Proprietary Name | : | Air Barrier System (ABS) |
| Device Class | : | Class II |
| Regulation Number | : | 21 CFR 878.5070 |
| Product Code | : | ORC |
| 510(k) Number | : | K123006 |
#### 3. Predicate Device
This submission for the Air Barrier System (ABS) Model 5001 is for new claims only. The Air Barrier System (ABS) Model 5001 specifications are the same as the model 1001 released to market [K092801]. The Air Barrier System (ABS) Model 1001 is the primarily predicate device.
# 4. Intended Use Statement
The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty and posterior vertebral fusion and laminoplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the
{1}------------------------------------------------
nozzle and the surface. (2) the incision plane is parallel with the direction of air flow, and (3) the incision dimensions are within: 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.
# 5. Description of Device
The ABS device controls the airborne surgical environment at a specific and limited location adjacent to and over surgical incisions by emitting a non-turbulent flow of HEPA filtered air that displaces the presence of airborne colony-forming units and particulate matter within surgical site dimensions of 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable outside of these dimensions and effectiveness depreciates beyond this specified area.
Clinical studies at two sites by different investigators have been conducted to demonstrate the performance characteristics of the Air Barrier System (ABS) during surgical procedures:
In the first study, twenty-nine (29) patients were randomized into one of three groups: with the ABS active (experiment), with the ABS present but not active (sham), and with no ABS device present (control). Airborne particulate and microorganism samples were collected simultaneously at a discrete single location at the surgical incision within the ABS area of effect and from a discrete single location within the sterile field, but not within the ABS area of effect. The airborne particulate and microorganism counts observed in the experiment group were significantly lower (P<0.001) than that observed in the sham and control groups. In the experiment group, the mean microorganism density at the sampling location within the ABS area of effect was 1.60 colony-forming units per cubic meter compared to 10.73 at the location outside the ABS area of effect. For particulate of size ≥5um, the mean observed particulate density in the experiment group was 524 particles per cubic foot compared with 3853 and 4092 particles per cubic foot in the sham and control groups, respectively.
In the second study, thirty-eight (38) patients undergoing instrumented posterior vertebral fusion or laminoplasty were randomized into one of two groups: with the ABS device (experiment) and with no ABS device (control). Airborne particulate and microorganism samples were collected simultaneously at a discrete single location at the surgical incision in both groups at ten-minute intervals for the first 100 minutes of the procedures. The airborne particulate and microorganism counts observed in the experiment group were significantly lower (P<0.001) than that observed in the control group. The mean experiment group microorganism density at the sampling location within the ABS area of effect was 1.55 colony-forming units per cubic meter compared to 5.05 in the control group. The mean experiment group density of particles sized ≥5um at the sampling location within the ABS area of effect was 1325 particles per cubic foot compared to 4837 measured in the control group.
The Air Barrier System (ABS) has two components: a Filter/Blower and an Air Delivery System.
{2}------------------------------------------------
The ABS Filter/Blower filters ambient air found in a typical surgical operating room through a High Efficiency Particle Arresting (HEPA) filter. The HEPA filtered air exits the Filter/Blower from an exit port on the top of the unit. The Filter/Blower is nonsterile and reusable.
The Air Delivery System, consisting of an Air Supply Hose and Nozzle, is sterile, single-use. The Nozzle portion is applied to the surgery site, on top of the incision drape and adjacent to a surgical incision. The end of the Air Supply Hose is plugged into the Filter/Blower's air exit port. HEPA filtered air then flows through the Air Supply Hose through the Nozzle and directly to the incision site.
# 6. Summary of the technological characteristics of the device compared to predicate device
The difference from the predicate Air Barrier System (ABS) Model 1001 [K092801], is new and extended claims of application to posterior vertebral fusion and laminoplasty procedures that are in covenant with Indications for Use statement.
The technology, design and method of construction of the Air Barrier System (ABS) Model 5001 are identical to the Air Barrier System (ABS) Model 1001, predicate device.
| Parameter | This submission [K123006] Nimbic Systems' Air Barrier System with new claims | Predicate [K092801] Nimbic Systems Air Barrier System |
|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | ORC | ORC |
| Regulation No. | 878.5070 | 878.5070 |
| Review Panel | General Hospital | General Hospital |
| Indications for Use | The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non- turbulent flow of air to the surgical site. The air flow from the device is HEPA- filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty and posterior vertebral fusion and laminoplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical | The Air Barrier System (ABS) is indicated as a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow form the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the nozzle and |
| Parameter | This submission [K123006] Nimbic Systems' Air Barrier System with new claims | Predicate [K092801] Nimbic Systems Air Barrier System |
| | surface with no gap between the bottom of the nozzle and the surface, (2) the incision plane is parallel with the direction of air flow, and (3) the incision dimensions are within: 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area. | the surface, (2) the incision plane is parallel with the direction of air flow, and (3) the incision dimensions are within: 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area. Patient age restrictions: none. |
| Physical | | |
| Housing Dimensions | 11 X 11 X 25 inches | 11 X 11 X 25 inches |
| Weight in pounds | 48 | 48 |
| Material type | Stainless Steel | Stainless Steel |
| Filter/Blower Output (CFM) | 150 | 150 |
| Blower Recertification Interval | Annual | Annual |
| Electrical | | |
| Voltage (VAC) | 120 | 120 |
| Frequency (Hertz) | 60 | 60 |
| Max Current (Amps) | 5 | 5 |
| HEPA Filtration Specifications | | |
| Filtration Efficiency | 99.97% @ 0.3 micron | 99.97% @ 0.3 micron |
| HEPA Filter Media | Continuous pleated glass microfiber | Continuous pleated glass microfiber |
| Air Inlet Pre-Filter Media | Polyurethane Foam | Polyurethane Foam |
| Measured average particulate density @ surgical incision (0.5 μm particles per cubic foot) | 68,122 | 29,992 |
| Parameter | This submission [K123006] | Predicate [K092801] |
| | Nimbic Systems'<br>Air Barrier System<br>with new claims | Nimbic Systems<br>Air Barrier System |
| CFU's @ Surgical<br>Incision (Colony<br>Forming Units/m³) | 1.57 | 1.60 |
| Air Volume<br>Delivered to<br>incision in CFM | 41 | 41 |
| Hose and Nozzle Specifications | | |
| Sterility | Sterile | Sterile |
| Use | Single-Use | Single-Use |
| Sterility Assurance<br>Level (SAL) | 10-6 SAL | 10-6 SAL |
| Sterilization Method | ETO | ETO |
Comparison table of characteristics between current submission and predicate:
{3}------------------------------------------------
{4}------------------------------------------------
# 7. Testing
The Air Barrier System (ABS) Model 5001 physical and operational specifications are identical to Air Barrier System (ABS) Model 1001 [K092801]. No changes have been made to the design or method of construction between Models 5001 and 1001.
The Electrical Safety and Electromagnetic Compatibility testing results are identical to Air Barrier System (ABS) Model 1001 {K092801}. For completeness the testing documentation is presented in this submission in the appropriate Sections.
Bench Testing - The bench performance testing submitted in this submission is identical to that submitted in the predicate submission IK0928011. Specifically, it includes two feasibility studies (Ambient and Challenge Tests), and the final bench testing reports (Validation and Performance Qualification). The Validation and Performance Qualification test results demonstrate that the ABS successfully met the criteria for achieving reduction of airborne contamination at the simulated incision site and over the specified control area.
Clinical Testing - Clinical testing was provided in the previous submission [K092801] demonstrating clinical performance of the ABS during hip surgery. Additional clinical data is provided in this submission demonstrating clinical performance of the ABS during posterior cervical and lumbar fusion procedures at the Michael E. DeBakey VA Medical Center (MEDVAMC). The goal of this submission is to incorporate these additional procedures into the ABS' indications for use.
The objective of the MEDVAMC study was to determine the degree to which the Air Barrier System (ABS), which deploys directed HEPA air flow across the surgical field, reduces airborne particulate and airborne colony forming units (e.g., bacteria and fungal spores) at surgical incisions during posterior instrumented lumbar and cervical vertebral fusion procedures. This was a prospective randomized trial with one control
{5}------------------------------------------------
group and one experiment group with an enrollment of up to 40 patients (41 patients actually enrolled). Statistically valid conclusions were able to be made regarding the ABS intended use after analyzing 38 surgery cases.
The clinical trial results demonstrated a statistically significant reduction (p≤0.0001. a=0.05) of both CFU and particulate sized ≥5um in the ABS group versus control group. CFU were reduced from a mean of 5.05 to 1.55 CFU/m³. Particulate sized ≥5pm were reduced from a mean of 4837 to 1325 particles/m3.
#### 8. Conclusion
The results of all of the tests performed, including clinical and nonclinical tests, demonstrate that the ABS-5001 is as safe, and as effective as the predicate device.
#### 9. Rx or OTC
The Air Barrier System (ABS) Model 5001 is a Rx Prescription device per 21 CFR Subpart D. The indication for use is for clinical settings only as stated in device labeling.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2013
Nimbic Systems, Incorporated C/O Mr. Harvey Knauss Contract Consultant DELPHI Consulting Group 11874 South Evelyn Circle HOUSTON TX 77477
Re: K123006
Trade/Device Name: Air Barrier System Regulation Number: 21 CFR 878.5070 Regulation Name: Air-handling Apparatus For A Surgical Operating Room Regulatory Class: II Product Code: ORC Dated: November 26, 213 Received: December 17, 2013
Dear Mr. Knauss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Knauss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely vours.
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/7/Picture/7 description: The image shows a signature and some text. The text includes the name "Tejashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director". The acronym "DAGRID" is also present. The word "FOR" is at the bottom right.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
、1、1、1、1、1、2、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、
Enclosure
{8}------------------------------------------------
#### Indications for Use Statement
#### 510(k) Number: K123006
# Device Name: Air Barrier System (ABS)
The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty and posterior vertebral fusion and laminoplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the nozzle and the surface, (2) the incision plane is parallel with the direction of air flow. and (3) the incision dimensions are within: 6° (15.2 cm) in width and 20° (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.
Prescription Use YES (Parl 21 CFR 801 Subpart D)
AND/OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Elizabeth F. Claverie -S 2013.12.18 18:45:25 -05'00'
indications for use Statement Al response 4-18-13.doc
Delphi Consulting Group
Houston, TX 77071
Section 4 510(k) Submission Page #
48
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.