K122976 · Shenzhen Belter Health Measurement and Analysis TE · MNW · Jan 4, 2013 · Cardiovascular
Device Facts
Record ID
K122976
Device Name
BODY FAT SCALE MODEL SF-371
Applicant
Shenzhen Belter Health Measurement and Analysis TE
Product Code
MNW · Cardiovascular
Decision Date
Jan 4, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd. Body Fat Scale SF-371 measures body weight, and estimates body fat, body water, body muscle mass and body bone mass using BIA (bioelectrical impedance analysis). The device is intended to be used by generally healthy people (from age 16 to age 80) who are not ill, feverish, have a chronic or acute disease, or a condition that affects the level of hydration such as pregnancy.
Device Story
Body Fat Scale SF-371 uses bioelectrical impedance analysis (BIA) to estimate body composition. User inputs age, height, and gender; stands barefoot on four scale electrodes. Device applies 500 µA, 50 kHz current through body; measures impedance. Internal software calculates body fat, water, muscle, and bone mass based on impedance, weight, and user-provided variables. Used by consumers in home settings. Output displayed on scale screen. Helps users monitor body composition metrics.
Clinical Evidence
Clinical testing was conducted to validate the effectiveness and accuracy of the device. All test results were satisfactory.
Indicated for generally healthy individuals aged 16-80 years for measurement of body weight and estimation of body fat, water, muscle mass, and bone mass. Contraindicated for individuals who are ill, feverish, have chronic or acute disease, or conditions affecting hydration levels (e.g., pregnancy).
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
K122976
PAGE 1 OF 3
JAN 0 4 2013
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
## Date: September 20, 2012
1. Company and Correspondent making the submission:
Name - Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd.
Address - 702/704, Block C, Tsinghua Unis Science Park, No. 13 Langshan Rd, Hi-Tech Industrial Park(north), Nanshan District, Shenzhen City, Guangdong Province
Telephone - +86-755-61869839
Fax - +86-755-61869122
Contact - Mr. Li Peiyong
Email - charliemack@irc-us.com
2. Device :
Trade/proprietary name: Body Fat Scale, SF-371
: Analyzer, Body Fat Common Name
Classification Name : Impedance plethysmograph
3. Predicate Devices :
Tanita BC-533, (K040778)
### 4. Classifications Names & Citations :
21CFR 870.2770, MNW, Body Fat Analyzer, Class 2
{1}------------------------------------------------
# 5. Description :
- 5.1 General
The patient steps on the scale device, where four electrodes are located. The patient must step on the electrodes with bare feet, with normal moisture. Through harmless current stimulation of 500 µA, at 50 kHz, the Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd. Body Fat Scale calculates the body fat percentage. This calculation is done via the Bioelectrical Impedance Method. The current is passed through the body and the impedance of the body determines the body fat. The calculation is based upon electrical impedance, height, weight, age, and gender. The calculation is performed via internal software, which uses the variables programmed in by the user. There are elements of this process that can produce erroneous readings, such as dry feet or improper-programmed data. The User's Manual defines items which could cause erroneous readings.
## 5.2 Direction
As discussed in the General description, the Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd. Body Fat Scale is relatively simple to use. The user inputs the variable data of age, height, and gender. The user steps onto the scale and the devices measures the user weight and body impedance (via the Bioelectric Impedance through the four electrodes on the scale). The scale displays the user's body fat composition. Upon the user's selection, the device can also display the user's weight.
6. Indication for use :
The Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd. Body Fat Scale SF-371 measures body weight, and estimates body fat, body water, body muscle mass and body bone mass using BIA (bioelectrical impedance analysis). The device is intended to be used by generally healthy people (from age 16 to age 80) who are not ill, feverish, have a chronic or acute disease, or a condition that affects the level of hydration such as pregnancy.
{2}------------------------------------------------
- 7. Comparison with predicate device :
Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd. believes that the Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd. Body Fat Scale is substantially equivalent to the Tanita BC-533 (K040778).
- 8. Safety and Performance Data :
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1. Clinical testing was used to validate the effectiveness and accuracy of the device. All test results were satisfactory.
- 9. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification TH Leaguer. Inc. concludes that The Shenzhen Beiter Health Measurement and Analysis Technology Co., Ltd. Body Fat Scale is safe and effective and substantially equivalent to predicate devices as described herein.
END
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human form or a caduceus, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### January 4. 2013
Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd. % Mr. Charlie Mack Principal Engineer IRC 77325 Joyce Way ECHO OR 97826
Re: K122976
Trade/Device Name: Body Fat Scale, Model SF-371 Regulation Number: 21 CFR§ 870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: December 14, 2012 Received: December 20, 2012
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Charlie Mack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and
listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Benjamin** R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K122917
Device Name: Body Fat Scale, Model SF-371
Indications for Use:
The Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd. Body Fat Scale SF-371 measures body weight, and estimates body fat, body water, body muscle mass and body bone mass using BIA (bioelectrical impedance analysis). The device is intended to be used by generally healthy people (from age 16 to age 80) who are not ill, feverish, have a chronic or acute disease, or a condition that affects the level of hydration such as pregnancy.
| Prescription Use | Commens promote the first the first the consisted on the consisted |
|-----------------------------|--------------------------------------------------------------------|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
• Concurrence of CDRH, Office of Device Evaluation (ODE)
| Benjamin R. Fisher -S | |
|-----------------------------|--|
| 2013.01.04 14:08:35 -05'00' | |
| (Division Sign-Off) | Page __ of __ |
|-------------------------------------------------------------------|---------------|
| Division of Reproductive, Gastro-Renal, and<br>Urological Devices | |
| 510(k) Number | K122976 |
13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.