TANITA INNERSCAN BODY COMPOSITION MONITOR, MODEL BC-53X (WHERE X=0 TO 9)
K040778 · Tanita Corp. of America · MNW · Dec 23, 2004 · Cardiovascular
Device Facts
Record ID
K040778
Device Name
TANITA INNERSCAN BODY COMPOSITION MONITOR, MODEL BC-53X (WHERE X=0 TO 9)
Applicant
Tanita Corp. of America
Product Code
MNW · Cardiovascular
Decision Date
Dec 23, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Body fat percentage
—
—
—
—
—
—
—
Total body water percentage
—
—
—
—
—
—
—
Visceral fat rating
—
—
—
—
—
—
—
Bone mass
—
—
—
—
—
—
—
Muscle mass
—
—
—
—
—
—
—
Physique rating
—
—
—
—
—
—
—
Daily calorie intake
—
—
—
—
—
—
—
Metabolic age
—
—
—
—
—
—
—
Indications for Use
The TANITA family of InnerScan Body Composition Monitors measure body weight and impedance and estimate percentage of body fat and body water, visceral fat rating, bone mass, muscle mass, physique rating, daily calorie intake (DCI) and metabolic age using BIA (bioelectrical impedance analysis). They are intended for use by healthy children 7-17 years old and healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment.
Device Story
Device is a computer-operated body composition monitor/scale; utilizes foot-to-foot bioelectrical impedance analysis (BIA) to measure body weight and impedance. Inputs include user weight and electrical impedance; device transforms these via proprietary algorithms to estimate body fat %, total body water %, visceral fat rating, bone mass, muscle mass, physique rating, metabolic age, and daily calorie intake. Used in home environment by consumers for self-assessment. Output displayed on device screen; provides users with body composition metrics to monitor health and lifestyle. Benefits include non-invasive tracking of body composition parameters.
Clinical Evidence
No clinical data provided. Substantial equivalence based on bench testing and validation of BIA methodology against reference methods including DEXA, deuterium dilution, and MRI.
Technological Characteristics
Foot-to-foot bioelectrical impedance analysis (BIA) scale. Measures weight and impedance. Models BC-533, BC-534, BC-550. Standalone consumer device. No specific material standards or software architecture details provided.
Indications for Use
Indicated for healthy children 7-17 years old and healthy adults with active, moderately active, or inactive lifestyles for body composition assessment in home settings.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
TANITA Body Fat Analyzer Professional and Consumer Models (K014009)
Submission Summary (Full Text)
{0}------------------------------------------------
2
TANITA InnerScan Body Composition Monitor 510(k) Submission
# DEC 2 3 2004
# 510(k) SUMMARY
This summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92
| Trade Name: | TANITA InnerScan Body Composition Monitor<br>Models BC-533, BC-534 and BC-550 |
|-------------------------|-------------------------------------------------------------------------------|
| Common Name: | Body Composition Analyzer / Body Fat Analyzer / Body Fat Monitor |
| Classification<br>Name: | ANALYZER, BODY COMPOSITION<br>21 CFR \$ 870.2770 |
## Description of Applicant Device:
The TANITA InnerScan Body Composition Monitor is a computer-operated body composition analyzer that utilizes BIA (bioelectrical impedance analysis) to determine body fat percent, body water percent, muscle mass, bone mass, visceral fat rating, metabolic age, physique rating and daily calorie intake.
| | Model Number | | |
|--------------------------------------------|--------------|--------|--------|
| Body Composition Capabilities | BC-533 | BC-534 | BC-550 |
| Total Body Water % (TBW %) | ✓ | ✓ | ✓ |
| Visceral Fat Rating | ✓ | ✓ | ✓ |
| Bone Mass | ✓ | ✓ | ✓ |
| Muscle Mass / Physique Rating | ✓ | ✓ | |
| Daily Calorie Intake (DCI) / Metabolic Age | ✓ | ✓ | |
| Daily Calorie Intake (DCI) | | | ✓ |
#### Intended Uses of Applicant Device:
Intended to be used as a body fat analyzer that determines body weight and estimates body fat with the use of BIA (bioelectrical impedance analysis). Other outputs include total body water, DCI (daily calorie intake), bone mass, visceral fat rating, metabolic age, physique rating and muscle mass.
## Predicate Devices:
TANITA Body Fat Analyzer Professional and Consumer Models K014009
{1}------------------------------------------------
### 510(k) SUMMARY, continued
This summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92
## Scientific Concepts and Significant Performance Characteristics:
| | Tanita Body Composition<br>Analyzer<br>Professional Models<br>K014009 | Tanita Body Composition<br>Analyzer<br>Consumer Models<br>K014009 | Tanita InnerScan Body<br>Composition Monitor<br>Models BC-533, BC-534 &<br>BC-550 |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INTENDED USE: | A combination non-invasive<br>device, which determines<br>weight and estimates body<br>fat and<br>total body water using BIA<br>(bioelectrical impedance<br>analysis). | A combination non-<br>invasive device, which<br>determines weight and<br>estimates body fat using<br>BIA (bioelectrical<br>impedance analysis). | A combination non-<br>invasive device, which<br>determines weight and<br>estimates total body fat,<br>visceral fat rating, bone<br>mass, total body water,<br>muscle mass, metabolic<br>age, physique rating and<br>daily calorie intake with<br>the use of BIA<br>(bioelectrical impedance<br>analysis). |
| PRODUCT<br>DESCRIPTION: | Body composition<br>analyzer/scale that utilizes<br>a "foot-to-foot" BIA<br>(bioelectrical impedance)<br>technology to determine<br>internal body composition. | Body composition<br>analyzer/scale that utilizes<br>a "foot-to-foot" BIA<br>(bioelectrical impedance)<br>technology to determine<br>internal body composition. | Body composition<br>monitor/scale that utilizes<br>a "foot-to-foot" BIA<br>(bioelectrical impedance)<br>technology to determine<br>internal body composition. |
| ANALYTICAL<br>METHOD/<br>MEASUREMENT | - Foot-to-Foot BIA<br>- In-house BIA and<br>- DEXA reference<br>methods | - Foot-to-Foot BIA<br>- In-house BIA and<br>- DEXA reference<br>methods | - Foot-to-Foot BIA<br>- In-house BIA, DEXA,<br>- Deuterium Dilution<br>and MRI reference<br>methods |
Side by side comparison of the TANITA InnerScan Body Composition Monitor to the predicate devices clearly demonstrates that the applicant device is substantially equivalent to the legally marketed devices.
Based on the results of using the previously approved "Foot-to-Foot" BlA methodology with our patented inhouse BIA, it was concluded that the TANITA InnerScan Body Composition Monitor performs as well as the predicate devices and therefore have proven its safety and efficacy.
Rhoda Lynn Valera, RAC TANITA Corporation of America Regulatory Affairs Specialist
2625 S. Clearbrook Dr. Arlington Heights, IL 60005 Phone: (847) 434-3966 (847) 640-7978 Fax:
December 20, 2004
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling birds in flight, stacked vertically. The emblem is rendered in black, contrasting with the white background.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 3 2004
Ms. Rhoda Lynn N. Valera Regulatory Affairs Specialist TANITA Corporation of America, Inc. 2625 South Clearbrook Drive ARLINGTON HEIGHTS IL 60005
Re: K040778
Trade/Device Name: TANITA InnerScan Body Composition Monitor BC-533, BC-534 and BC-550 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: 74 MNW Dated: November 10, 2004 Received: November 12, 2004
Dear Ms. Valera:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Scetion 510(k) I ms letter will anow you to organism
premarket notification. The FDA finding of substantial equivalence of your device to a legally premaired predicated on: "The PDF maxing sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you dosire specific act 100 of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Anot, production other general information on your responsibilities under the Act from the 601:27). " Ou may octurers, International and Consumer Assistance at its toll-free number (800) DIVISION of Unian Hallacturers address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon
Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
TANITA InnerScan Body Composition Monitor 510(k) Submission
## INDICATIONS FOR USE
| 510(k) Number: | K040778 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | TANITA InnerScan BC-533, BC-534 and BC-550 Body Composition Monitor |
| Indications for Use: | The TANITA family of InnerScan Body Composition Monitors measure body weight and impedance and estimate percentage of body fat and body water, visceral fat rating, bone mass, muscle mass, physique rating, daily calorie intake (DCI) and metabolic age using BIA (bioelectrical impedance analysis). They are intended for use by healthy children 7-17 years old and healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment. |
### Model Number
| Body Composition Capabilities | BC-533 | BC-534 | BC-550 |
|--------------------------------------------|--------|--------|--------|
| Total Body Water % (TBW %) | ✓ | ✓ | ✓ |
| Visceral Fat Rating | ✓ | ✓ | ✓ |
| Bone Mass | ✓ | ✓ | ✓ |
| Muscle Mass / Physique Rating | ✓ | | |
| Daily Calorie Intake (DCI) / Metabolic Age | ✓ | ✓ | |
| Daily Calorie Intake (DCI) | | | ✓ |
Prescription Use_ (Part 21 CFR 801 Subpart D) AND/OR
> Over-the-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David h. Lyman
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number _
Page 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.