K122878 · Silicon Vally Medical Instruments, Inc. · OBJ · Jan 11, 2013 · Cardiovascular
Device Facts
Record ID
K122878
Device Name
HD-IVUS ULTRASOUND IMAGING SYSTEM
Applicant
Silicon Vally Medical Instruments, Inc.
Product Code
OBJ · Cardiovascular
Decision Date
Jan 11, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The HD-IVUS Ultrasound Imaging System (Console) is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Device Story
HD-IVUS Ultrasound Imaging System is a minimally invasive diagnostic tool for coronary artery imaging. System components include a console with touch screen monitor, roll stand or bed-rail mount, Patient Interface Module (PIM), power supply, and a separately cleared intravascular ultrasound catheter. The catheter transducer emits sound energy into coronary arteries; reflected waves are processed by the console to generate real-time, high-resolution cross-sectional images. The Linear Translation System (LTS) accessory provides motorized, controlled-speed pullback of the catheter imaging element. Used in clinical settings by physicians or under their order. The system provides visualization of coronary lumen, wall morphology, and stents. Analysis tools include area/linear measurements and annotations. Data is stored in DICOM format and exportable via DVD or USB. The device aids clinical decision-making by providing in-vivo visualization of vascular pathology.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1), and acoustic safety (IEC 60601-2-37).
Technological Characteristics
System utilizes PZT crystal transducer technology for ultrasound imaging at 40 MHz or 60 MHz. Features a PC-based console with touch screen GUI. Connectivity includes DICOM export via DVD or USB. Powered by medical-grade AC to DC supply. Compliant with IEC 60601-1, IEC 60601-1-4, and IEC 60601-2-37 standards. Includes motorized Linear Translation System for catheter pullback.
Indications for Use
Indicated for patients who are candidates for transluminal coronary interventional procedures requiring ultrasound examination of coronary intravascular pathology.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
SVMI's HD-IVUS Ultrasound Imaging System (K112997)
Boston Scientific Corporation's iLab Ultrasound Imaging System (K051679)
{0}------------------------------------------------
K12 2878 pg 1 of 6
JAN 1 1 2013
Special 510(k) (accessory to device) Silicon Valley Medical Instruments, Inc. 510(k) Summary
Image /page/0/Picture/3 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo is in black and white and features the letters "SVMI" in a stylized font. The words "Silicon Valley Medical Instruments Inc." are written in a smaller font below the letters.
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| Submitter Information | |
|--------------------------------------|-----------------------------------------------------------------|
| Name: | Silicon Valley Medical Instruments, Inc. |
| Address | 47697 Westinghouse Drive, Suite 101, Fremont, CA 94539-7401 USA |
| Phone Number: | 510-897-4695 |
| Fax Number | 510-226-1230 |
| Establishment Registration<br>Number | N/A – Not yet registered. |
| Contact Person | Richard E. Anderson; VP, RA/QA (richarda@svmii.com) |
| Date Prepared | 07 September 2012 |
| Name of Device | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade or Proprietary Name | HD-IVUS Ultrasound Imaging System |
| Common or Usual Name | System, Imaging, System, Imaging, Pulsed Echo, Ultrasonic (IYO) |
| Classification Name | Class II |
| Classification Panel | Cardiovascular |
| Regulation | 21 CFR 892.1560 (ΙΥΟ) |
| Product Code(s) | IYO |
| Legally Marketed Device(s)<br>to which Equivalence is<br>Claimed | Boston Scientific Corporation's iLab Ultrasound Imaging System<br>(Cleared 14Jul05 via K051679) |
| Reason for Submission | New device |
| Device Description | The HD-IVUS Ultrasound Imaging System is a minimally invasive<br>intravascular ultrasound diagnostic imaging tool when used in<br>conjunction with an intravascular ultrasound catheter. The System is a<br>medical device for use by or on the order of a physician. The HD-IVUS<br>Ultrasound Imaging System is comprised of a System Console with a<br>touch screen monitor, a Roll Stand, or bed-rail swing arm mount, a<br>Patient Interface Module (PIM), a Power Supply, and an Intravascular<br>Ultrasound Catheter (510(k) cleared and sold separately). The Catheter<br>emits sound energy from a transducer at the distal tip of the catheter,<br>which is guided into the coronary arteries of the heart. Sound waves<br>that reflect from the inner vascular tissues are received by the<br>transducer and sent to the System Console where a high resolution,<br>cross-sectional image is displayed in real time. The technique provides<br>for in-vivo visualization of the coronary artery lumen, coronary artery<br>wall morphology, and devices (such as stents) at or near the surface of |
| the coronary artery wall. | |
| • System Console | |
| The HD-IVUS system features a point-of-care system console that is mounted to a roll stand, or bed-rail swing arm mount. The System Console is the central component that is responsible for: | |
| • managing the generation and display of IVUS images<br>• storing and exporting the image data<br>• providing the main user interface<br>• providing analysis tools (i.e., area measurements, linear measurements, and annotations) | |
| The System Console receives, processes, displays, and records ultrasound image data from the transducer in the Kodama HD-IVUS catheter. The IVUS images are displayed on a high-resolution, 19-inch, flat panel, touch screen monitor. The monitor serves as the graphical user interface (GUI) for operating the system. All system information and controls are located on the touch screen monitor. The HD-IVUS system has internal storage for patient studies. Each study can contain data in the form of loops or stills. Study data is saved in DICOM format and can be exported to DVD, using the integrated DVD drive, or to a USB storage device, using one of the four USB ports. | |
| • Roll Stand | |
| The variable height roll stand supports the System Console and enables the console to be attached to a mobile stand that can be rolled up to the patient bed. | |
| • Bed Rail Swing Arm | |
| The bed rail swing arm mount supports the System Console and enables the console to be attached to a patient bed rail. | |
| • Patient Interface Module (PIM) | |
| The Patient Interface Module (PIM) provides the electromechanical interface between the catheter and the System Console. The PIM provides the mechanical interface to secure the catheter, as well as the mechanical energy to rotate the catheter's imaging assembly. | |
{1}------------------------------------------------
K122878 pg 2016
Image /page/1/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo is composed of the letters SVMII in a stylized font. The words "Silicon Valley Medical Instruments Inc." are written below the letters in a smaller font.
{2}------------------------------------------------
K122878 pg 30P6
:
Image /page/2/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo is composed of the letters "SVMII" in a stylized font. The words "Silicon Valley Medical Instruments Inc." are written below the letters in a smaller font.
The PIM also provides the electrical interface that transmits the signal from the integrated ultrasound generator to the catheter and receives the return signals. The PIM is placed into a sterile bag for normal use and can be situated near the patient in the sterile field. The PIM features an imaging on/off button and a recording on/off button. These controls enable the sterile operator to start or stop imaging and recording from within the sterile field. A LED light on the PIM indicates active recording.
**Linear Translation System (LTS) (subject of this submission)**
The Linear Translation System (LTS) is an accessory (companion product) to the HD-IVUS system and provides for linear translation (pullback) of the catheter imaging element under motorized (controlled speed) linear translation when used in conjunction with the PIM. The LTS is an optional accessory and not required for basic IVUS imaging.
**Power Supply**
The system includes a universal, medical grade external AC to DC power supply that enables direct connection to the hospital power source.
**Intravascular Ultrasound Catheter (510(k) cleared under K113008 and sold separately)**
The Kodama Intravascular Ultrasound Catheter is a minimally invasive intravascular ultrasound coronary imaging catheter. The catheter emits acoustic energy from a transducer at its distal tip, which is guided into the coronary arteries of the heart. Sound waves that are reflected from vascular tissues are received by the transducer and sent through the PIM to the system console. The catheter can be operated at two different frequencies, 40 MHz or 60 MHz, depending on user preference. An integrated telescope allows the imaging of multiple regions of interest in a single procedure by advancing or retracting the imaging assembly without moving the catheter sheath. (*NOTE: Refer to the Directions for Use supplied with the Kodama Intravascular Ultrasound Catheter for additional information.*)
| Intended Use of the Device | The HD-IVUS Ultrasound Imaging System (Console) is intended for ultrasound examination of coronary intravascular pathology only. |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------|
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------|
{3}------------------------------------------------
K122878 pg 40R6
Image /page/3/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo is in black and white and features the letters "SVMI" in a stylized font. The words "Silicon Valley Medical Instruments Inc." are written in a smaller font below the letters.
| Indications for Use | The HD-IVUS Ultrasound Imaging System (Console) is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures. |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{4}------------------------------------------------
K12 2878 pg 50P6
Image /page/4/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters 'SVMI' in a stylized font. The words 'Silicon Valley Medical Instruments Inc.' are written in a smaller font below the letters 'SVMI'.
| Summary of the Technological Characteristics of the Device Compared to the Predicate Device | | |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Predicate Device<br>(SVMI's HD-IVUS Ultrasound<br>Imaging System - Cleared 17Dec11<br>via K112997) | Additional Predicate<br>(Boston Scientific Corporation's<br>iLab Ultrasound Imaging System -<br>Cleared 14Jul05 via K051679) |
| Signal Type | Echo | Doppler, Echo |
| Catheter Mechanical Drive | Patient Interface Module (PIM)<br>(cleared with ultrasound system) | Motor Drive Unit (MDU)<br>(cleared with ultrasound system) |
| Linear Translation System<br>(accessory to device(s)) | Subject of this submission - Linear<br>Translation System that allows for<br>motorized (controlled speed) linear<br>pullback (translation) of the<br>catheter when used in conjunction<br>with the PIM (above) | Pullback Sled that allows for linear<br>pullback (translation) of the<br>catheter when used in conjunction<br>with the MDU (above) |
| Signal Conductor | Coaxial wire traveling down<br>catheter | Coaxial wire traveling down<br>catheter |
| Signal Generator/ Receiver<br>Materials | PZT crystal transducer | PZT crystal transducer |
| Acoustic Output | Does not exceed Track 3 limits | Does not exceed Track 3 limits |
| Ultrasound Frequency | 40 MHz, 60 MHz ± 10% | 10-40 MHz ± 10% |
| Method of Use | Intravascular | Intravascular |
| User Interface | PC-based system with keyboard<br>and touch screen and mouse | PC-based system with keyboard<br>and touch screen and mouse |
| Software Interface | Custom GUI | Custom GUI |
| Intended Use | The HD-IVUS Ultrasound Imaging<br>System (Console) is intended for<br>ultrasound examination of<br>coronary intravascular pathology<br>only. Intravascular ultrasound<br>imaging is indicated in patients<br>who are candidates for<br>transluminal coronary<br>interventional procedures. | The iLab Ultrasound Imaging<br>System is intended for ultrasound<br>examinations of intravascular<br>pathology. Intravascular<br>ultrasound is indicated in patients<br>who are candidates for<br>transluminal interventional<br>procedures such as angioplasty<br>and atherectomy. |
{5}------------------------------------------------
K122878 pg 6aP6
Image /page/5/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo is in black and white and features the letters "SVMI" in a stylized font. The words "Silicon Valley Medical Instruments Inc." are written in a smaller font below the letters.
| | Performance Data | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------|
| Summary of Non-clinical tests Conducted for Determination of Substantial Equivalence | | |
| Performance Test Summary - New Device | | |
| Characteristic | Standard/Test/FDA Guidance | Results Summary |
| Electromagnetic<br>Compatibility | Compliant with IEC 60601-1-2 | Compliant with IEC 60601-1-2 |
| Electrical Safety and<br>Acoustic Safety | Compliant with IEC 60601-1, IEC<br>60601-1-4, and IEC 60601-2-37 | Compliant with IEC 60601-1, IEC<br>60601-1-4, and IEC 60601-2-37 |
. Conclusions Drawn from Design Verification/Validation and Non-clinical Data The HD-IVUS Ultrasound Imaging System (with Linear Translation System) is substantially equivalent in design and technology to the predicate device(s) with regard to intended use.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in black and is centered on the page. There is a line underneath the text. To the left of the text is a logo.
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
JAN 1 1 2013
Silicon Valley Medical Instruments, Inc. (SVMI) c/o Richard E. Anderson 47697 Westinghouse Drive, Suite 101 Fremont, CA 94539-7401
Re: K122878
Trade Name: HD-IVUS Ultrasound Imaging System (with Linear Translation System) Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: QBJ Dated: December 10, 2013 Received: December 13, 2013
Dear Mr. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{7}------------------------------------------------
Page 2 – Mr. Richard Anderson
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Owen P. Faris -S
Bram D. Zuckerman, M.D. for Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
K 122878 ps lofl
Special 510(k) (accessory to device) Silicon Valley Medical Instruments, Inc. Indications for Use
SVMI
Silicon Valley Medical Instruments Inc."
## INDICATIONS FOR USE
510(k) Number (if known):
Device Name: HD-IVUS Ultrasound Imaging System (with Linear Translation System)
Indications for Use: Silicon Valley Medical Instrument's HD-IVUS Ultrasound Imaging System (Console) is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
___ AND/OR Over-The-Counter Use _ Prescription Use _____________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K122878
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.