K113008 · Silicon Vally Medical Instruments, Inc. · OBJ · Jun 14, 2012 · Cardiovascular
Device Facts
Record ID
K113008
Device Name
KODAMA INTRAVASCULAR ULRASOUND CATHETER
Applicant
Silicon Vally Medical Instruments, Inc.
Product Code
OBJ · Cardiovascular
Decision Date
Jun 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The Kodama Intravascular Ultrasound Catheter is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Device Story
The Kodama Intravascular Ultrasound Catheter is a minimally invasive device used by physicians in clinical settings to visualize coronary artery lumen, wall morphology, and stents. The catheter features a distal PZT transducer that emits ultrasonic pressure waves into coronary arteries. Reflected acoustic signals are captured by the transducer and transmitted via a coaxial cable to the SVMI HD-IVUS Ultrasound Imaging System console. The console rotates the transducer to generate real-time, cross-sectional images. The device includes a telescope mechanism for adjusting the imaging assembly position without moving the sheath, and a guide wire lumen for navigation. By providing high-resolution, in-vivo visualization, the device assists clinicians in assessing vascular pathology and guiding interventional procedures, potentially improving procedural outcomes.
Clinical Evidence
Bench testing only. Performance testing included tensile strength, torque strength, tip flexibility, crossability, pushability, and traceability. Biocompatibility testing was performed per ISO 10993-1 (cytotoxicity, hemolysis, pyrogenicity). Sterilization and shelf-life validation were conducted per ISO 11135-1. All tests met specifications.
Technological Characteristics
PZT-based intravascular ultrasound catheter. Features include a distal transducer, drive cable, coaxial transmission line, and sheath assembly with lubricious coating. Operates at 40MHz or 60MHz. Connectivity via electromechanical rotating interface to system console. Sterilization via EtO. Compliant with ISO 10993 (biocompatibility) and ISO 60601-2-37 (ultrasound).
Indications for Use
Indicated for patients who are candidates for transluminal coronary interventional procedures requiring ultrasound examination of coronary intravascular pathology.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
Boston Scientific Corporation's Atlantis SR Pro/Pro 2
Submission Summary (Full Text)
{0}------------------------------------------------
# K113008
Traditional 510(k) (new device) Silicon Valley Medical Instruments, Inc. 510(k) Summary CONFIDENTIAL
Image /page/0/Picture/3 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters "SVMI" in a bold, sans-serif font. The words "Silicon Valley Medical Instruments Inc." are printed in a smaller font below the letters.
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Name: | Silicon Valley Medical Instruments, Inc. |
|-----------------------------------|-----------------------------------------------------------------|
| Address | 47697 Westinghouse Drive, Suite 101, Fremont, CA 94539-7401 USA |
| Phone Number: | 510-897-4695 |
| Fax Number | 510-226-1230 |
| Establishment Registration Number | N/A - Not yet registered. |
| Contact Person | Richard E. Anderson; VP, RA/QA (richarda@svmii.com) |
| Date Prepared | 7 October 2011 |
| Name of Device | |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade or Proprietary Name | Kodama Intravascular Ultrasound Catheter |
| Common or Usual Name | Catheter, Ultrasound, Intravascular |
| Classification Name | Class II |
| Classification Panel | Cardiovascular |
| Regulation | 21 CFR 870.1200 |
| Product Code(s) | OBJ |
| Legally Marketed Device(s) to which Equivalence is Claimed | Boston Scientific Corporation's Atlantis SR Pro/Pro 2 |
| Reason for Submission | New device |
| Device Description | The Kodama Intravascular Ultrasound Catheter is a minimally invasive<br>intravascular ultrasound coronary imaging catheter. The Catheter is a<br>medical device for use by or on the order of a physician. The Catheter<br>emits sound energy from a transducer at the distal tip of the catheter,<br>which is guided into the coronary arteries of the heart. Sound waves<br>that reflect from the inner vascular tissues are received by the<br>transducer and sent to the Console (SVMI's HD-IVUS Ultrasound Imaging<br>System) where a high resolution, cross-sectional image is displayed in<br>real time. The technique provides for in-vivo visualization of the<br>coronary artery lumen, coronary artery wall morphology, and devices<br>(such as stents) at or near the surface of the coronary artery wall. |
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters "SVMI" in a bold, sans-serif font. The words "Silicon Valley Medical Instruments Inc." are printed in a smaller font below the letters.
The Catheter can be operated at two different frequencies (40MHz or 60MHz) depending on user preference. A telescope allows the imaging of multiple regions of interest in a single procedure by advancing or retracting the imaging assembly without moving the catheter sheath. The Catheter is comprised of the following key components:
- · Imaging Assembly
The imaging assembly consists of an ultrasonic transducer, drive cable, transmission line (coaxial cable) and electromechanical rotating catheter to system interface (rotor):
- Transducer
- Drive Cable
- · Coaxial Cable
- · Rotor
The transducer is attached to the distal end of the drive cable via the distal housing. The drive cable contains the coaxial cable which is electrically terminated at the distal end to the transducer and at the proximal end to an electromechanical rotating interface (rotor). The rotor contains an electrical contact board (rotor board) and provides an electrical interface between the system console and the transducer. In addition, the rotor provides a rotating mechanical interface between the console and drive cable to rotate the drive cable and the transducer. The transducer emits ultrasonic pressure waves (acoustic signals) orthogonally from its face which are reflected back from the region of interest to the transducer surface. The console continuously rotates the drive cable in order to rotationally sweep the transducer beam to image a region of interest.
- · Sheath Assembly
The sheath assembly is the outer sleeve of the catheter that houses the imaging core and enters the coronary anatomy. The Sheath Assembly consists of:
- Distal Sheath
- Proximal sheath
- Femoral Marker
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo features the letters "SVMII" in a bold, sans-serif font. The words "Silicon Valley Medical Instruments Inc." are printed in a smaller font below the letters. The letters are large and black, and the words below are smaller and also black.
| · Lubricious Coating | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Distal Sheath | | | |
| The Distal Sheath consists of: | | | |
| Distal Tip<br>● | | | |
| Imaging Window | | | |
| Distal Tip | | | |
| The Distal Tip consists of: | | | |
| Soft Tip | | | |
| Guide Wire Lumen Entry Port | | | |
| Guide Wire Lumen Exit Port | | | |
| Guide Wire Lumen | | | |
| Radio Opaque Marker | | | |
| Flushing Vent | | | |
| The soft tip comprises the distal most end of the<br>catheter and is the softest member of the sheath<br>assembly. The smaller profile of the soft tip compared<br>to that of the rest of the sheath and its softness ensures<br>good catheter crossability and prevents damage to the<br>vasculature. The Soft Tip includes a guide wire lumen<br>(accessed by a guide wire entry port at the distal end of<br>the soft tip by the proximal end of the guide wire) to<br>engage a guide wire for advancing and retracting the<br>catheter. A radio opaque marker that is visible under<br>fluoroscopy is integrated circumferentially around the<br>guide wire lumen to provide a visual indication of the<br>location of the distal tip. A flushing vent just proximal to<br>the guide wire lumen exit port (located proximal to the<br>distal tip) allows the flushing media to vent out of the<br>sheath. | | | |
| Imaging Window | | | |
| Just proximal to the distal tip is the imaging window, the<br>section of the sheath that is transparent to acoustic<br>signals and through which the ultrasound beam is | | | |
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters "SVMI" in a bold, sans-serif font. The letters are arranged horizontally, with "Silicon Valley Medical Instruments Inc." written in a smaller font size below the letters. The letters are black, and the background is white.
emitted and reflected back from the anatomy and regions of interest to be imaged. The flexibility of the imaging window is graded such that it is more flexible at the distal tip and gradually becomes stiffer proximally. The imaging window is bonded to the proximal sheath at the midshaft (proximal most section of the imaging window). The midshaft is of an intermediate stiffness between that of the flexible imaging window and the stiff proximal sheath. The flexibility of the imaging window ensures good catheter trackability.
### Proximal Sheath
The proximal sheath is the structural member of the catheter between the distal sheath and the telescope. The proximal sheath material is selected based on mechanical properties that provide sufficient longitudinal stiffness. Proximal sheath stiffness ensures good catheter pushability, allowing the user to access and cross regions of interest. The proximal sheath is distally bonded to the imaging window at the midshaft and proximally bonded to the telescope. The proximal sheath also contains an inner lumen that contains the imaging assembly. The inner lumen is sized to provide sufficient clearance between the drive cable and the inner wall of the lumen, allowing flushing fluid to be delivered to the transducer at the distal end of the imaging assembly to flush air bubbles from the inner lumen and transducer surface that might otherwise compromise image quality. ·
#### Femoral Marker
The Femoral Marker is a visual indicator located at a fixed distance from the distal tip on the proximal end of the proximal sheath to alert the user that a pre-determined length of the catheter has been inserted into the patient anatomy. The femoral marker is intended to alert the user that the catheter is about to exit the guide catheter.
#### Lubricious Coating
The Sheath Assembly is coated with a lubricious coating on the surface of the working length of the catheter sheath to enhance
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo features the letters 'SVMII' in a bold, sans-serif font. Below the letters is the full name of the company, 'Silicon Valley Medical Instruments Inc.,' in a smaller font.
{5}------------------------------------------------
r
Image /page/5/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo is black and white and features the letters "SVMI" in a stylized font. The words "Silicon Valley Medical Instruments Inc." are printed in a smaller font below the letters.
| Intended Use of the Device | The Kodama Intravascular Ultrasound Catheter is intended for<br>ultrasound examination of coronary intravascular pathology only. |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Kodama Intravascular Ultrasound Catheter is intended for<br>ultrasound examination of coronary intravascular pathology only.<br>Intravascular ultrasound imaging is indicated in patients who are<br>candidates for transluminal coronary interventional procedures. |
| Summary of the Technological Characteristics of the Device Compared to the Predicate Device | | |
|---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Predicate | New Device |
| Materials, Design | Intravascular catheter with PZT transducer at the distal tip and a rotational drive line to enable ultrasound imaging. | Minor changes in materials/design. |
| Technology | PZT-based intravascular ultrasound. | Same. |
| Intended Use | The Kodama Ultrasound Imaging Catheter is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures. | The Atlantis SR Pro/Pro2 intravascular ultrasound catheter is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures. |
| | Performance Data | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------|
| Summary of Non-clinical tests Conducted for Determination of Substantial Equivalence | | |
| Performance Test Summary - New Device | | |
| Characteristic | Standard/Test/FDA Guidance | Results Summary |
| Performance Testing | Tensile Strength, torque<br>strength, tip flexibility,<br>crossability, pushability,<br>traceability, etc. | Met all specifications. |
| Biocompatibility Testing | ISO 10993 inc. Cytotoxicity,<br>Hemolysis, and Pyrogenicity. | Compliant with ISO 10993-1. |
| Sterilization and Shelf Life | EtO Sterilization validation inc.<br>EtO residuals. Shelf life validated<br>to 6 months (for launch). | Compliant with ISO 11135-1. |
Comparative Performance Information Summation Summary
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters 'SVMI' in a bold, sans-serif font. The words 'Silicon Valley Medical Instruments Inc.' are printed in a smaller font below the letters.
| Characteristic | Requirement | New Device | Predicate Device(s) |
|------------------|---------------------|---------------------|---------------------|
| Sterility | ISO 11135 (Sterile) | Compliant (Sterile) | Compliant (Sterile) |
| Biocompatibility | ISO 10993 | Compliant | Compliant |
| Ultrasound | ISO 60601-2-37 | Compliant | Compliant |
్ డిగ్రీ రాజ్ Conclusions Drawn from Design Verification and Non-clinical Data 1999 11 The Kodama Intravascular Ultrasound Catheter is substantially equivalent in design, materials, and technology to the predicate device(s) with regard to intended use.
{7}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 4 2012
Silicon Valley Medical Instruments, Inc. (SVMI) c/o Richard E. Anderson 47697 Westinghouse Drive, Suite 101 Fremont, CA 94539-7401
Re: K113008
Trade Name: Kodama Intravascular Ultrasound Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II (two) Product Codes: OBJ Dated: June 8, 2012 Received: June 11, 2012
Dear Mr. Richard Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Richard Anderson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
fo Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
#### K113008
Traditional 510(k) {new device} Silicon Valley Medical Instruments, Inc. Indications for Use CONFIDENTIAL .
Image /page/9/Picture/2 description: The image shows the logo for Silicon Valley Medical Instruments Inc. The logo consists of the letters 'SVMII' in a bold, sans-serif font. Below the letters is the full name of the company, 'Silicon Valley Medical Instruments Inc.', in a smaller font. The letters are all black, and the background is white.
#### INDICATIONS FOR USE
510(k) Number:
Device Name: Kodama Intravascular Ultrasound Catheter
Indications for Use: Silicon Valley Medical Instrument's Kodama Intravascular Ultrasound Catheter is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Prescription Use ______________________ AND/OR Over-The-Counter Use _
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
diovascular De
**510(k) Number** K113008
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.