K122567 · Sciton, Inc · GEX · Jun 19, 2013 · General, Plastic Surgery
Device Facts
Record ID
K122567
Device Name
JOULE 810/940/980 MULTI-PLATFORM SYSTEM
Applicant
Sciton, Inc
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 19, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The JOULE 810/940/980 Multi-Platform System (and delivery accessories used to deliver optical energy) are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue.
Device Story
JOULE 810/940/980 Multi-Platform System is a laser-powered surgical device consisting of a console, power supply, cooling system, and delivery accessories (fiber optic arm/handpiece). Device delivers thermal energy to soft tissue via 810nm, 940nm, or 980nm laser wavelengths. Operated by clinicians in clinical settings via footswitch activation. System uses focusing optics to perform vaporization, incision, excision, ablation, and coagulation. Output is controlled via LCD touchscreen interface. Clinical benefits include precise tissue management, vascular lesion treatment, hair reduction, and endovenous occlusion. Device supports various surgical specialties including dermatology, plastic surgery, and vascular surgery.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics and indications for use with legally marketed predicate devices.
Technological Characteristics
Diode laser system; wavelengths 810, 940, 980 nm; max power ≤ 100 W; pulse duration ≤ 2500 msec; repetition rate ≤ 200 Hz; fluence ≤ 120 J/cm2; spot size 0.6mm to 7.7 cm2; CW and pulsed output modes; fiber optic delivery with handpiece; LCD touchscreen interface; 230 VAC power; cooling system included.
Indications for Use
Indicated for patients requiring soft tissue surgery (incision, excision, ablation, coagulation, hemostasis) across dermatology, plastic surgery, vascular surgery, and podiatry. Specific uses include: vascular/pigmented lesion treatment, hair removal (skin types I-V), permanent hair reduction, endovenous occlusion of the greater saphenous vein, laser-assisted lipolysis, and temporary relief of minor muscle/joint pain/stiffness/spasm via elevated tissue temperature.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# Attachment IV
# 510(k) Summary
| Submitter: | Sciton, Inc. |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 925 Commercial Street, Palo Alto, CA 94303 |
| Phone: | (650) 493-9155 |
| Fax: | (650) 493-9146 |
| Contact Person: | Jay M. Patel, VP of Regulatory Affairs |
| Date Prepared: | June 17, 2013 |
| Device Trade Name: | JOULE 810/940/980 Multi-Platform System |
| Common Name: | Laser Powered Surgical Device (and Accessories) |
| Classification Name: | Laser Surgical Instrument, 21 CFR 878.4810. |
| Legally Marketed<br>Predicate Device: | K112031: Alma Lasers Modified Diode Laser Module<br>K102036: LAMBDA S.p.A. Doctor Diode Laser Family<br>K100558: Quanta System QUANTA Diode Laser Family<br>K100143: CAO Group Pilot Diode Laser<br>K053628: Lumenis LightSheer Duet |
| Description of<br>JOULE 810/940/980<br>Multi-Platform System: | The JOULE 810/940/980 Multi-Platform System consist of a console and laser deliver<br>accessories. It uses focusing optics to deliver thermal energy to the treatment site.<br>The control console houses the power supply, cooling system, articulated arm<br>delivery system and/or fiber optic arm delivery system with a handpiece. The<br>user activates laser emission by means of a footswitch. |
| Intended Use: | The JOULE 810/940/980 Multi-Platform System (and delivery accessories used<br>to deliver optical energy) are indicated for use in surgical applications requiring<br>the vaporization, incision, excision, ablation, cutting and hemostasis, or<br>coagulation of soft tissue. |
| | The device is specifically indicated for use as follows: |
| | 810nm and 980nm |
| | Dermatology/Aesthetics: |
| | - Photocoagulation of vascular & dermatological lesions of the face and<br>extremities<br>- Photocoagulation of telangiectasia, veinulectasia of the legs and face<br>- Treatment of reticular veins and branch varicosities<br>- Pyrogenic granuloma, lymphangioma and lymphangiomatosis disease,<br>angiofibromas<br>- Superficial benign vascular lesions including Telangiectasias, Rosacea,<br>Angioma, venous lakes Couperosis, Cherry angioma, hemangioma, Port<br>wine stains, angiokeratoma, and benign epidermal pigment lesions as<br>lentigines. Epidermal nevi, spider nevi.<br>- Dermatological surgery: Condyloma acuminate, warts, small non<br>malignant skin tumors, small semi-malignant tumors as basalomas,<br>Bowe, Kaposi sarcom. Warty leucoplasty and ulcers debridment. |
{1}------------------------------------------------
- Seborrheic keratosis
- Mixoid cyst
ﺮ ﺰﺭ ﺍﻟ
- Papillary varix
- Acne treatment
- Hair removal of unwanted hair from skin type I-V
#### Plastic Surgery:
- Cut, coagulation & vaporization
- Resurfacing
- Blepharoplasty
### Vascular Surgery:
- Endoluminal or endovenous laser surgery for saphenous incompetent veins
### 810nm
Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions.
The JOULE Multi-Platform System is indicated for the treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia and other benign vascular lesions, and the treatment for pseudofolliculitis barbae. It is also indicated for hair removal, permanent hair reduction, and the treatment of benign pigmented lesions and leg veins. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
#### 940nm
The JOULE Multi-Platform System (and their delivery accessories used to deliver optical energy) are intended for use in cutting, vaporization, ablation and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopies), in incision/excision, vaporization, ablation and coagulation of soft tissue in contact and non-contact open surgery (with or without a handpiece), in the treatment and/or removal of vascular lesions (fumors) and removal of unwanted hair, and for endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux.
### 980nm
The JOULE Multi-Platform System (and their delivery accessories used to deliver optical energy) is indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue endovenous occlusion of the greater saphenous vein. The JOULE 980 System is further indicated for laser assisted lipolysis.
The device is specifically indicated for use as follows:
#### Dermatology, Plastic Surgery and Podiatry:
Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, skin, fat or muscle tissue and dermabrasion. Examples include:
- Matrixectomy
- Excision of neuromas
- Excision of periungual and subungual warts
- Excision of plantar warts
- Excision of keloids
- Excision of cutaneous lesions
- Dermabrasion
- Vaporization and hemostasis of capillary hemangioma
- Debridement of wounds
- Photocoagulation of teleangectasia of the legs and face
- Photocoagulation of vascular lesions of the face and extremities
{2}------------------------------------------------
- Endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux.
- Treatment of reticular veins and branch varicosities
## Endovenous Occlusion of the Greater Saphenous Vein in Patients with Superficial Vein Reflux:
Indicated for use with their delivery accessories in the endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
Technological Characteristics: The JOULE 810/940/980 Multi-Platform System shares the same indications for use, and as noted below, shares similar design features (including wavelength, laser medium and delivery systems, power supply, cooling and control system), functional features (including power output, repetition rate, energy, spot size and fluence), and is therefore substantially equivalent to the above legally marketed predicate devices.
{3}------------------------------------------------
OFFICE
BUILDING
Param System Site
6040 Earle Brown Dr
Brooklyn Center, MN 55430
| Specification | This Application<br>JOULE 810/940/980 Multi-Platform System | Alma Lasers<br>Modified Diode<br>Laser Module | LAMBDA S.p.A.<br>Doctor Diode Laser<br>Family | Predicate Devices<br>Quanta System<br>QUANTA Diode<br>Laser Family | CAO Group Pilot<br>Diode Laser | Lumenis<br>LightSheer Duet | Substantially<br>Equivalent |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Indications for Use | General surgical applications.<br>Intended for hair removal and<br>treatment for benign<br>pigmented and vascular<br>lesions.<br>Temporary relief of muscle &<br>joint pain due to elevated<br>tissue temperature | General surgical<br>applications.<br>Intended for hair<br>removal and<br>treatment for benign<br>pigmented and<br>vascular lesions. | General surgical<br>applications.<br>Intended for hair<br>removal and<br>treatment for benign<br>pigmented and<br>vascular lesions. | General surgical<br>applications.<br>Intended for hair<br>removal and<br>treatment for benign<br>pigmented and<br>vascular lesions. | General surgical<br>applications.<br>Temporary relief of<br>muscle & joint pain<br>due to elevated<br>tissue temperature. | General surgical<br>applications.<br>Intended for hair<br>removal and<br>treatment for benign<br>pigmented and<br>vascular lesions. | Yes |
| Ref. 510(k) | N/A | K112031 | K102036 | K100558 | K100143 | K053628 | -- |
| Wavelength | 810, 940 & 980 nm | 810 nm | 810, 940, 980 &<br>1064 nm | 532, 810, 940, 980,<br>1064, 1320, 1470,<br>1950 nm | 810 nm | 790 - 950 nm<br>(800 nm nominal) | Yes |
| Fluence | ≤ 120 J/cm2 | ≤ 120 J/cm2 | -- | -- | -- | 10 - 100 J/cm2 | Yes |
| Max Power | ≤ 100 W | ≤ 100 W | ≤ 15 W | ≤ 30 W | ≤ 15 W | 1600 W peak power | Yes |
| Pulse Duration | ≤ 2500 msec | ≤ 1350 msec | -- | ≤ 2500 msec | -- | 5 - 400 msec | Yes |
| Spot Size | 0.6mm Ø to 7.7 cm2 | 1.2 cm2 | -- | 0.6 - 2.4 mm Ø | -- | 22x35mm<br>(7.7 cm2) | Yes |
| Output Mode | CW & pulsed mode | CW & pulsed mode | CW & pulsed mode | CW & pulsed mode | CW & pulsed mode | CW & pulsed mode | Yes |
| Repetition Rate | ≤ 200 Hz | ≤ 10 Hz | -- | ≤ 200 Hz | 2.5 - 20,000 Hz | ≤ 3 Hz | Yes |
| Laser Type | Diode | Diode | Diode | Diode | Diode | Diode | Yes |
| Electrical<br>Requirements | 230 VAC, 50/60 Hz, 1Φ | 120/230 VAC, 50/60<br>Hz, 1Φ | -- | 100-200VAC,<br>50/60Hz, 1Φ | -- | 110 - 240VAC<br>50/60 Hz. 1 Ф | Yes |
| Console Dimensions | 16" x 30" x 43" high | 20" x 23" x 47" high | -- | 16" x 13" x 10" high | -- | 17" x 20" x 44" high | Yes |
| Delivery System | Fiber optic with handpiece | Fiber optic with<br>handpiece | Fiber optic with<br>handpiece | Fiber optic with<br>handpiece | Fiber optic with<br>handpiece | Fiber optic with<br>handpiece | Yes |
| User Interface | LCD touchscreen | LCD touchscreen | LCD touchscreen | LCD touchscreen | LCD touchscreen | LCD touchscreen | Yes |
Page 4 of 5
.
{4}------------------------------------------------
| Safety and<br>Effectiveness: | The indications for use are based upon the indications for use for predicate systems.<br>Technologically, the JOULE 810/940/980 Multi-Platform System is substantially<br>equivalent to the listed predicate devices. Therefore, the risks and benefits for the<br>JOULE 810/940/980 Multi-Platform System are comparable to the predicate devices. |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | JOULE 810/940/980 Multi-Platform System shares similar indications for use, design<br>features, and similar functional features as, and therefore is substantially equivalent<br>to, the currently marketed predicate devices. |
:
. .
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
June 19, 2013
Sciton, Inc. % Mr. Jay M. Patel 925 Commercial Street Palo Alto, California 94303
Re: K122567
Trade/Device Name: Joule 810/940/980 Multi-Platform System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX
Dated: May 31, 2013 Received: June 04, 2013
Dear Mr. Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Mr. Jay M. Patel
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ycm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely yours, . FOR Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Attachment II
### Statement of Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K422567
JOULE 810/940/980 Multi-Platform System Device Name:
Indications for Use:
The JOULE 810/940/980 Multi-Platform System (and delivery accessories used to deliver optical energy) is indicated for use in surgical applications requiring the vaporization, incision, ablation, cutting and hemostasis, or coagulation of soft tissue.
The device is specifically indicated for use as follows:
#### 810nm and 980nm
Dermatology/Aesthetics:
- Photocoagulation of vascular & dermatological lesions of the face and ﮯ extremities
- Photocoagulation of telangiectasia, veinulectasia of the legs and face
- Treatment of reticular veins and branch varicosities
- Pyrogenic granuloma, lymphangioma and lymphangiomatosis disease, angiofibromas
- Superficial benign vascular lesions including Telangiectasias, Rosacea, Angioma, venous lakes Couperosis, Cherry angioma, hemangioma, Port wine stains, angiokeratoma, and benign epidermel pigment lesions as lentigines. Epidermal nevi,
- spider nevi. - Dermatological surgery: Condyloma acuminate, warts, small non malignant skon
- tumors, small semi-malignant tumors as basalomas, Bowe, Kaposi sarcom. Warty leucoplasty and ulcers debridment.
- Seborheic keratosis
- Mixoid cyst
- Papillary varix
- Acne treatment
- Hair removal of unwanted hair from skin type I-V .
#### Plastic Surgery:
- Cut. coaculation & vaporization .
- Resurfacing
- Blepharopiasty
#### Vascular Surgery:
- Endoluminal or endovenous laser surgery for saphenous incompetent veins
Prescription Use × (Per 21CFR801)
Over-The-Counter Use
21CFR801)
OR
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2013.06.18 16:11:45 -04'00"
(Division Sign-Off) for MXM
Oivision of Surgical Devices
510(k) Number_K122567
Page 1 of 3
{8}------------------------------------------------
#### 810nm
Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions
The JOULE Multi-Platform System is indicated for the treatment of vascular lesions, including angiomas, hemangiomas, telanglectasia and other benign vascular lesions, and the treatment for pseudofollicultis barbae. It is also indicated for hair removal, permanent hair reduction, and the treatment of benign pigmented lesions and leg veins. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
#### 940nm
The JOULE Multi-Plation System (and their delivery accessories used to deliver optical energy) are intended for use in cutting, vaporization, ablation and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopies), in incision, vaporization, ablation and coagulation of soft tissue in contact and non-contact open surgery (with or without a handpiece), in the treatment and/or removal of vascular lesions (tumors) and removal of unwanted hair, and for endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux.
Prescription Use × Over-The-Counter Use
(Per 21CFR801)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Neil R Ogden 2013.06.18 16:12:16 -04'00'
(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number _K122567
{9}------------------------------------------------
#### 980am
The JOULE Multi-Platform System (and their delivery accessories used to deliver optical energy) is indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue endovenous occlusion of the greater saphenous vein. The JOULE 980 System is further indicated for laser assisted fipolysis.
The device is specifically indicated for use as follows:
#### Dermatology, Plastic Surgery and Podiatry:
Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, skin, fat or muscle tissue and dermabrasion. Examples include:
- Matrixectomy .
- Excision of neuromas
- Excision of periungual and subungual warts -
- Excision of plantar warts
- Excision of keloids
- Excision of cutaneous lesions
- Dermabrasion
- Vaporization and hemostasis of capillary hemangioma
- Debridement of wounds
- Photocoagulation of teleangectasia of the legs and face
- Photocoagulation of vascular lesions of the face and extremities .
- Endovascular coagulation of the greater saphenous vein of the thigh in patients with . superficial vein reflux.
- Treatment of reticular veins and branch varicosities
### Endovenous Occlusion of the Greater Saphenous Vein in Patients with Superficial Vein Reflux:
Indicated for use with their delivery accessories in the endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
Over-The-Counter Use Prescription Use OR × (Per 21CFR801)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Neil R Ogden 2013.06.18 16:12:47 -04'00'
(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number_ K122567
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.